Published in: BMC Anesthesiology 1/2020

01-12-2020 | Care | Research article

Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study

Authors: Arpa Chutipongtanate, Churairat Yasaeng, Tanit Virankabutra, Somchai Chutipongtanate

Published in: BMC Anesthesiology | Issue 1/2020

Login to get access

Abstract

Background

Transfusion decision during the perioperative period mostly relies on the point-of-care testing for Hb measurement. This study aimed systematically compared four point-of-care methods with the central laboratory measurement of hemoglobin (LHb) regarding the accuracy, precision, and assay practicality to identify the preferred point-of-care method during the perioperative period.

Methods

This cross-sectional method comparison study was conducted in the surgical intensive care unit at Ramathibodi Hospital, Thailand, from September 2015 to July 2016. Four point-of-care methods, i.e., capillary hematocrit (HctCap), HemoCue Hb201+, iSTAT with CG8+ cartridge, and SpHb from Radical-7 pulse co-oximeter were carried out when LHb was ordered. Pearson correlation and Bland-Altman analyses were performed to assess the accuracy and precision, while the workload, turnaround time, and the unit cost were evaluated for the method practicality.

Results

Thirty-five patients were enrolled, corresponding to 48 blood specimens for analyses, resulting in the measured hemoglobin of 11.2 ± 1.9 g/dL by LHb. Ranking by correlation (r), mean difference (bias) and 95% limit of agreement (LOA) showed the point-of-care methods from the greater to the less performance as followed, iSTAT-LHb pair (r = 0.941; bias 0.15 (95% LOA; − 1.41, 1.12) g/dL), HemoCue-LHb pair (r = 0.922; bias − 0.18 (95% LOA; − 1.63, 1.28) g/dL), SpHb-LHb pair (r = 0.670; bias 0.13 (95% LOA; − 3.12, 3.39) g/dL) and HctCap-LHb pair (r = 0.905; bias 0.46 (95% LOA; − 1.16, 2.08) g/dL). Considering the practicality, all point-of-care methods had less workload and turnaround time than LHb, but only HemoCue and HctCap had lower unit cost.

Conclusion

This study identified HemoCue as the suitable point-of-care method for the sole purpose of Hb measurement in the surgical ICU setting, while iSTAT should be considered when additional data is needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference van KAMPEN E, ZIJLSTRA WG. Standardization of hemoglobinometry. II. The hemiglobincyanide method. Clin Chim Acta. 1961;6:538–44.CrossRef van KAMPEN E, ZIJLSTRA WG. Standardization of hemoglobinometry. II. The hemiglobincyanide method. Clin Chim Acta. 1961;6:538–44.CrossRef
2.
go back to reference Lienhart A, Auroy Y, Péquignot F, Benhamou D, Warszawski J, Bovet M, et al. Survey of anesthesia-related mortality in France. Anesthesiology. 2006;105(6):1087–97.CrossRef Lienhart A, Auroy Y, Péquignot F, Benhamou D, Warszawski J, Bovet M, et al. Survey of anesthesia-related mortality in France. Anesthesiology. 2006;105(6):1087–97.CrossRef
3.
go back to reference Gilliss BM, Looney MR, Gropper MA. Reducing noninfectious risks of blood transfusion. Anesthesiology. 2011;115(3):635–49.CrossRef Gilliss BM, Looney MR, Gropper MA. Reducing noninfectious risks of blood transfusion. Anesthesiology. 2011;115(3):635–49.CrossRef
4.
go back to reference Glance LG, Dick AW, Mukamel DB, Fleming FJ, Zollo RA, Wissler R, et al. Association between intraoperative blood transfusion and mortality and morbidity in patients undergoing noncardiac surgery. Anesthesiology. 2011;114(2):283–92.CrossRef Glance LG, Dick AW, Mukamel DB, Fleming FJ, Zollo RA, Wissler R, et al. Association between intraoperative blood transfusion and mortality and morbidity in patients undergoing noncardiac surgery. Anesthesiology. 2011;114(2):283–92.CrossRef
5.
go back to reference Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.CrossRef Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.CrossRef
6.
go back to reference Tsuei BJ, Hanseman DJ, Blakeman MJ, Blakeman TC, Yang SH, Branson RD, et al. Accuracy of non-invasive hemoglobin monitoring in patients at risk for hemorrhage. J Trauma Acute Care Surg. 2014;77(3):135–9. Tsuei BJ, Hanseman DJ, Blakeman MJ, Blakeman TC, Yang SH, Branson RD, et al. Accuracy of non-invasive hemoglobin monitoring in patients at risk for hemorrhage. J Trauma Acute Care Surg. 2014;77(3):135–9.
7.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.CrossRef Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.CrossRef
8.
go back to reference Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.CrossRef Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.CrossRef
9.
go back to reference Steinfelder-Visscher J, Teerenstra S, Gunnewiek JM, Weerwind PW. Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients. J Extra Corpor Technol. 2008;40(1):57–60.PubMedPubMedCentral Steinfelder-Visscher J, Teerenstra S, Gunnewiek JM, Weerwind PW. Evaluation of the i-STAT point-of-care analyzer in critically ill adult patients. J Extra Corpor Technol. 2008;40(1):57–60.PubMedPubMedCentral
10.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef
11.
go back to reference Montenij LJ, Buhre WF, Jansen JR, Kruitwagen CL, de Waal EE. Methodology of method comparison studies evaluating the validity of cardiac output monitors: a stepwise approach and checklist. Br J Anaesth. 2016;116(6):750–8.CrossRef Montenij LJ, Buhre WF, Jansen JR, Kruitwagen CL, de Waal EE. Methodology of method comparison studies evaluating the validity of cardiac output monitors: a stepwise approach and checklist. Br J Anaesth. 2016;116(6):750–8.CrossRef
12.
go back to reference U.S. Department of Health and Human Services: Centers for Medicare, Medicaid Services. 42 CFR part 493. Clinical laboratory improvement amendments of 1988 (CLIA) proficiency testing regulations related to Analytes and acceptable performance. Fed Regist. 2019;84(23):1536–67. U.S. Department of Health and Human Services: Centers for Medicare, Medicaid Services. 42 CFR part 493. Clinical laboratory improvement amendments of 1988 (CLIA) proficiency testing regulations related to Analytes and acceptable performance. Fed Regist. 2019;84(23):1536–67.
13.
go back to reference Kolotiniuk NV, Manecke GR, Pinsky MR, Banks D. Measures of blood hemoglobin and hematocrit during cardiac surgery: comparison of three point-of-care devices. J Cardiothorac Vasc Anesth. 2018;32(4):1638–41.CrossRef Kolotiniuk NV, Manecke GR, Pinsky MR, Banks D. Measures of blood hemoglobin and hematocrit during cardiac surgery: comparison of three point-of-care devices. J Cardiothorac Vasc Anesth. 2018;32(4):1638–41.CrossRef
14.
go back to reference Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-hemoglobin or 201+) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43(3):341–50.CrossRef Hiscock R, Kumar D, Simmons SW. Systematic review and meta-analysis of method comparison studies of Masimo pulse co-oximeters (Radical-7™ or Pronto-7™) and HemoCue® absorption spectrometers (B-hemoglobin or 201+) with laboratory haemoglobin estimation. Anaesth Intensive Care. 2015;43(3):341–50.CrossRef
15.
go back to reference Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue (®) and laboratory haemoglobin measurements in obstetric patients. Anaesthesia. 2013;68(1):40–5.CrossRef Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue (®) and laboratory haemoglobin measurements in obstetric patients. Anaesthesia. 2013;68(1):40–5.CrossRef
Metadata
Title
Systematic comparison of four point-of-care methods versus the reference laboratory measurement of hemoglobin in the surgical ICU setting: a cross-sectional method comparison study
Authors
Arpa Chutipongtanate
Churairat Yasaeng
Tanit Virankabutra
Somchai Chutipongtanate
Publication date
01-12-2020
Publisher
BioMed Central
Keyword
Care
Published in
BMC Anesthesiology / Issue 1/2020
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-020-01008-8

Other articles of this Issue 1/2020

BMC Anesthesiology 1/2020 Go to the issue