Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 11/2012

01-11-2012 | Refractive Surgery

Can we measure mesopic pupil size with the cobalt blue light slit-lamp biomicroscopy method?

Authors: Miguel J. Maldonado, Alberto López-Miguel, David P. Piñero, José R. Juberías, Juan C. Nieto, Jorge L. Alió

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 11/2012

Login to get access

Abstract

Background

The aim of this work is to assess a previously described slit-lamp biomicroscopy-based method (SLBM) for measuring pupil diameter and compare it to Colvard infrared pupillometry (CIP).

Methods

Two examiners performed three repeated measurements with each instrument in 40 healthy eyes. We determined the agreement of SLBM and CIP, intraobserver and interobserver repeatabilities, and interobserver concordance (kappa) and SLBM ability for detecting pupil sizes over 6.0 mm.

Results

The mean (±standard deviation [SD]) pupil diameter was 5.81 ± 0.70 mm with SLBM and 6.26 ± 0.68 mm with CIP (p = 0.01) averaging both examiner’s results. Mean differences between the SLBM and CIP were –0.60 mm and –0.30 mm for each examiner using the average of the three readings (p = 0.02), and they were very similar using the first reading. Intraobserver reproducibility: the width of the 95% LoA ranged from 1.79 to 2.30 mm. The ICCs were 0.97 and 0.92 for SLBM, and 0.96 and 0.90 for CIP. Interobserver reproducibility: the width of the LoA ranged from 1.82 to 2.09 mm. Kappa statistics were 0.39 and 0.49 for the first and mean SLBM readings, respectively, and 0.45 for both the first and mean CIP readings. Sensitivity and specificity of SLBM for detection of pupils larger than 6 mm ranged from 55.56% to 73.68% and from 76.19% to 95.45%, respectively. The best trade-off between sensitivity and specificity ranged from 5.4 mm to 6.2 mm.

Conclusions

Although the SLBM is quite repeatable, it underestimates mesopic pupil size and shows a too wide range of agreement with CIP. SLBM shows low sensitivity in detecting pupils larger than 6 mm, which may be misleading when planning anterior segment surgery. Previous grading-consensus training strategies may increase interrater reproducibility, and compensation for the systematic underestimation could improve accuracy of the SLBM.
Literature
1.
go back to reference Nixon WS (1997) Pupil size in refractive surgery. J Cataract Refract Surg 23:1435–1436PubMed Nixon WS (1997) Pupil size in refractive surgery. J Cataract Refract Surg 23:1435–1436PubMed
2.
go back to reference Colvard M (1999) Preoperative measurement of scotopic pupil dilation using an office pupillometer. J Cataract Refract Surg 24:1594–1597 Colvard M (1999) Preoperative measurement of scotopic pupil dilation using an office pupillometer. J Cataract Refract Surg 24:1594–1597
3.
go back to reference Fan-Paul NI, Li J, Miller JS, Florakis GJ (2002) Night vision disturbances after corneal refractive surgery. Surv Ophthalmol 47:533–546PubMedCrossRef Fan-Paul NI, Li J, Miller JS, Florakis GJ (2002) Night vision disturbances after corneal refractive surgery. Surv Ophthalmol 47:533–546PubMedCrossRef
4.
go back to reference Helgesen A, Hjortdal J, Ehlers N (2004) Pupil size and night vision disturbances after LASIK for myopia. Acta Ophthalmol Scand 82:454–460PubMedCrossRef Helgesen A, Hjortdal J, Ehlers N (2004) Pupil size and night vision disturbances after LASIK for myopia. Acta Ophthalmol Scand 82:454–460PubMedCrossRef
5.
go back to reference Alió JL, de la Hoz F, Pérez-Santonja JJ, Ruiz-Moreno JM, Quesada JA (1999) Phakic anterior chamber lenses for the correction of myopia. A 7-year cumulative analysis of complications in 263 cases. Ophthalmology 106:458–466PubMedCrossRef Alió JL, de la Hoz F, Pérez-Santonja JJ, Ruiz-Moreno JM, Quesada JA (1999) Phakic anterior chamber lenses for the correction of myopia. A 7-year cumulative analysis of complications in 263 cases. Ophthalmology 106:458–466PubMedCrossRef
6.
go back to reference Pieh S, Lackner B, Hanselmayer G, Zöhrer R, Sticker M, Weghaupt H, Fercher A, Skorpik C (2001) Halo size under distance and near conditions in refractive multifocal intraocular lenses. Br J Ophthalmol 85:816–821PubMedCrossRef Pieh S, Lackner B, Hanselmayer G, Zöhrer R, Sticker M, Weghaupt H, Fercher A, Skorpik C (2001) Halo size under distance and near conditions in refractive multifocal intraocular lenses. Br J Ophthalmol 85:816–821PubMedCrossRef
7.
go back to reference Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R (2007) Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens. J Cataract Refract Surg 33:430–438PubMedCrossRef Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R (2007) Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens. J Cataract Refract Surg 33:430–438PubMedCrossRef
8.
go back to reference Pop M, Payette Y, Santoriello E (2002) Comparison of the pupil card and pupillometer in measuring pupil size. J Cataract Refract Surg 28:283–288PubMedCrossRef Pop M, Payette Y, Santoriello E (2002) Comparison of the pupil card and pupillometer in measuring pupil size. J Cataract Refract Surg 28:283–288PubMedCrossRef
9.
go back to reference Wachler BS, Krueger RR (1999) Agreement and repeatability of infrared pupillometry and the comparison method. Ophthalmology 106:319–323PubMedCrossRef Wachler BS, Krueger RR (1999) Agreement and repeatability of infrared pupillometry and the comparison method. Ophthalmology 106:319–323PubMedCrossRef
10.
go back to reference Schnitzler EM, Baumeister M, Kohnen T (2000) Scotopic measurement of normal pupils: Colvard versus Video Vision Analyzer infrared pupillometer. J Cataract Refract Surg 26:859–866PubMedCrossRef Schnitzler EM, Baumeister M, Kohnen T (2000) Scotopic measurement of normal pupils: Colvard versus Video Vision Analyzer infrared pupillometer. J Cataract Refract Surg 26:859–866PubMedCrossRef
11.
go back to reference Rosen ES, Gore CL, Taylor D, Chitkara D, Howes F, Kowalewski E (2002) Use of a digital infrared pupillometer to assess patient suitability for refractive surgery. J Cataract Refract Surg 28:1433–1438PubMedCrossRef Rosen ES, Gore CL, Taylor D, Chitkara D, Howes F, Kowalewski E (2002) Use of a digital infrared pupillometer to assess patient suitability for refractive surgery. J Cataract Refract Surg 28:1433–1438PubMedCrossRef
12.
go back to reference Kohnen T, Terzi E, Buhren J, Kohnen EM (2003) Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameter. J Cataract Refract Surg 29:112–117PubMedCrossRef Kohnen T, Terzi E, Buhren J, Kohnen EM (2003) Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameter. J Cataract Refract Surg 29:112–117PubMedCrossRef
13.
go back to reference Michel AW, Kronberg BP, Narváez J, Zimmerman G (2006) Comparison of 2 multiple-measurement infrared pupillometers to determine scotopic pupil diameter. J Cataract Refract Surg 32:1926–1931PubMedCrossRef Michel AW, Kronberg BP, Narváez J, Zimmerman G (2006) Comparison of 2 multiple-measurement infrared pupillometers to determine scotopic pupil diameter. J Cataract Refract Surg 32:1926–1931PubMedCrossRef
14.
go back to reference Twa MD, Bailey MD, Hayes J, Bullimore M (2004) Estimation of pupil size by digital photography. J Cataract Refract Surg 30:381–389PubMedCrossRef Twa MD, Bailey MD, Hayes J, Bullimore M (2004) Estimation of pupil size by digital photography. J Cataract Refract Surg 30:381–389PubMedCrossRef
15.
go back to reference Yang H, Lee M, Kim JB, Ahn J (2006) Burst-shot infrared digital photography to determine scotopic pupil diameter. J Cataract Refract Surg 32:2113–2117PubMedCrossRef Yang H, Lee M, Kim JB, Ahn J (2006) Burst-shot infrared digital photography to determine scotopic pupil diameter. J Cataract Refract Surg 32:2113–2117PubMedCrossRef
16.
go back to reference Ho LY, Harvey TM, Scherer J, Balasubramaniam M, Dhaliwal DK, Mah FS (2009) Comparison of Rosenbaum Pupillometry Card Using Red and Blue Light to Colvard and Iowa Pupillometers. J Refract Surg 2:1–7 Ho LY, Harvey TM, Scherer J, Balasubramaniam M, Dhaliwal DK, Mah FS (2009) Comparison of Rosenbaum Pupillometry Card Using Red and Blue Light to Colvard and Iowa Pupillometers. J Refract Surg 2:1–7
17.
go back to reference Schallenberg M, Bangre V, Steuhl KP, Kremmer S, Selbach JM (2010) Comparison of the Colvard, Procyon, and Neuroptics pupillometers for measuring pupil diameter under low ambient illumination. J Refract Surg 26:134–143PubMedCrossRef Schallenberg M, Bangre V, Steuhl KP, Kremmer S, Selbach JM (2010) Comparison of the Colvard, Procyon, and Neuroptics pupillometers for measuring pupil diameter under low ambient illumination. J Refract Surg 26:134–143PubMedCrossRef
18.
go back to reference Bradley JC, Bentley KC, Mughal AI, Brown SM (2010) Clinical performance of a handheld digital infrared monocular pupillometer for measurement of the dark-adapted pupil diameter. J Cataract Refract Surg 36:277–281PubMedCrossRef Bradley JC, Bentley KC, Mughal AI, Brown SM (2010) Clinical performance of a handheld digital infrared monocular pupillometer for measurement of the dark-adapted pupil diameter. J Cataract Refract Surg 36:277–281PubMedCrossRef
19.
go back to reference Starck T, Liu Y, Prewett AL, Curup LG (2002) Comparison of scotopic pupil measurement with slit-lamp-based cobalt blue light and infrared video-based system. J Cataract Refract Surg 28:1952–1956PubMedCrossRef Starck T, Liu Y, Prewett AL, Curup LG (2002) Comparison of scotopic pupil measurement with slit-lamp-based cobalt blue light and infrared video-based system. J Cataract Refract Surg 28:1952–1956PubMedCrossRef
20.
go back to reference Chaglasian EL, Akbar S, Probst LE (2006) Pupil measurement using the Colvard pupillometer and a standard pupil card with a cobalt blue filter penlight. J Cataract Refract Surg 32:255–260PubMedCrossRef Chaglasian EL, Akbar S, Probst LE (2006) Pupil measurement using the Colvard pupillometer and a standard pupil card with a cobalt blue filter penlight. J Cataract Refract Surg 32:255–260PubMedCrossRef
21.
go back to reference Hsieh YT, Hu FR (2007) The correlation of pupil size measured by Colvard pupillometer and Orbscan II. J Refract Surg 23:789–795PubMed Hsieh YT, Hu FR (2007) The correlation of pupil size measured by Colvard pupillometer and Orbscan II. J Refract Surg 23:789–795PubMed
22.
go back to reference Bootsma S, Tahzib N, Eggink F, de Brabander J, Nuijts R (2007) Comparison of two pupillometers in determining pupil size for refractive surgery. Acta Ophthalmol Scand 85:324–328PubMedCrossRef Bootsma S, Tahzib N, Eggink F, de Brabander J, Nuijts R (2007) Comparison of two pupillometers in determining pupil size for refractive surgery. Acta Ophthalmol Scand 85:324–328PubMedCrossRef
23.
go back to reference Rocha KM, Soriano ES, Chamon W, Chalita MR, Nosé W (2007) Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses: a prospective randomized study. Ophthalmology 114:2050–2054PubMedCrossRef Rocha KM, Soriano ES, Chamon W, Chalita MR, Nosé W (2007) Spherical aberration and depth of focus in eyes implanted with aspheric and spherical intraocular lenses: a prospective randomized study. Ophthalmology 114:2050–2054PubMedCrossRef
24.
go back to reference McDonnell C, Rolincova M, Venter J (2006) Comparison of measurement of pupil sizes among the Colvard pupillometer, Procyon pupillometer, and NIDEK OPD-scan. J Refract Surg 22:1027–1030 McDonnell C, Rolincova M, Venter J (2006) Comparison of measurement of pupil sizes among the Colvard pupillometer, Procyon pupillometer, and NIDEK OPD-scan. J Refract Surg 22:1027–1030
25.
go back to reference Yoon MK, Schmidt G, Lietman T, McLeod SD (2007) Inter- and intraobserver reliability of pupil diameter measurement during 24 hours using the Colvard pupillometer. J Refract Surg 23:266–271PubMed Yoon MK, Schmidt G, Lietman T, McLeod SD (2007) Inter- and intraobserver reliability of pupil diameter measurement during 24 hours using the Colvard pupillometer. J Refract Surg 23:266–271PubMed
26.
go back to reference Colvard M (1998) Preoperative measurement of scotopic pupil dilation using an office pupillometer. J Cataract Refract Surg 24:1594–1597PubMed Colvard M (1998) Preoperative measurement of scotopic pupil dilation using an office pupillometer. J Cataract Refract Surg 24:1594–1597PubMed
27.
go back to reference Ray WA, O'Day DM (1985) Statistical analysis of multi-eye data in ophthalmic research. Invest Ophthalmol Vis Sci 26:1186–1188PubMed Ray WA, O'Day DM (1985) Statistical analysis of multi-eye data in ophthalmic research. Invest Ophthalmol Vis Sci 26:1186–1188PubMed
28.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef
29.
30.
go back to reference Margo CE, Harman LE, Mulla ZD (2002) The reliability of clinical methods in ophthalmology. Surv Ophthalmol 47:375–386PubMedCrossRef Margo CE, Harman LE, Mulla ZD (2002) The reliability of clinical methods in ophthalmology. Surv Ophthalmol 47:375–386PubMedCrossRef
31.
go back to reference Fleiss J (1981) Statistical methods for rates and proportions. John Wiley & Sons Inc. New York, NY. pp 212-236. Fleiss J (1981) Statistical methods for rates and proportions. John Wiley & Sons Inc. New York, NY. pp 212-236.
32.
go back to reference Jin Y, Zabriskie N, Olson RJ (2009) Dysphotopsia outcomes analysis of two truncated acrylic 6.0-mm intraocular optic lenses. Ophthalmologica 223:47–51PubMedCrossRef Jin Y, Zabriskie N, Olson RJ (2009) Dysphotopsia outcomes analysis of two truncated acrylic 6.0-mm intraocular optic lenses. Ophthalmologica 223:47–51PubMedCrossRef
33.
go back to reference Kohnen T, Terzi E, Kasper T, Kohnen EM, Bühren J (2004) Correlation of infrared pupillometers and CCD-camera imaging from aberrometry and videokeratography for determining scotopic pupil size. J Cataract Refract Surg 30:2116–2123PubMedCrossRef Kohnen T, Terzi E, Kasper T, Kohnen EM, Bühren J (2004) Correlation of infrared pupillometers and CCD-camera imaging from aberrometry and videokeratography for determining scotopic pupil size. J Cataract Refract Surg 30:2116–2123PubMedCrossRef
34.
go back to reference Cheng AC, Lam DS (2004) Comparison of the Colvard pupillometer and the Zywave for measuring scotopic pupil diameter. J Refract Surg 20:248–252PubMed Cheng AC, Lam DS (2004) Comparison of the Colvard pupillometer and the Zywave for measuring scotopic pupil diameter. J Refract Surg 20:248–252PubMed
35.
go back to reference Griner PF, Mayewski RJ, Mushlin AI, Greenland P (1981) Selection and interpretation of diagnostic tests and procedures. Principles and applications. Ann Intern Med 94:557–592PubMed Griner PF, Mayewski RJ, Mushlin AI, Greenland P (1981) Selection and interpretation of diagnostic tests and procedures. Principles and applications. Ann Intern Med 94:557–592PubMed
37.
go back to reference Bouma H (1962) Size of the static pupil as a function of wavelength and luminosity of the light incident on the human eye. Nature 193:690–691PubMedCrossRef Bouma H (1962) Size of the static pupil as a function of wavelength and luminosity of the light incident on the human eye. Nature 193:690–691PubMedCrossRef
38.
39.
go back to reference Kardon R, Anderson SC, Damarjian TG, Grace EM, Stone E, Kawasaki A (2009) Chromatic pupil responses: preferential activation of the melanopsin-mediated versus outer photoreceptor-mediated pupil light reflex. Ophthalmology 116:1564–1573PubMedCrossRef Kardon R, Anderson SC, Damarjian TG, Grace EM, Stone E, Kawasaki A (2009) Chromatic pupil responses: preferential activation of the melanopsin-mediated versus outer photoreceptor-mediated pupil light reflex. Ophthalmology 116:1564–1573PubMedCrossRef
40.
go back to reference Howarth PA, Heron G, Whittaker L (2000) The measurement of pupil cycling time. Graefe’s Arch Clin Exp Ophthalmol 238:826–832CrossRef Howarth PA, Heron G, Whittaker L (2000) The measurement of pupil cycling time. Graefe’s Arch Clin Exp Ophthalmol 238:826–832CrossRef
41.
go back to reference Robl C, Sliesoraityte I, Hillenkamp J, Prahs P, Lohmann CP, Helbig H, Herrmann WA (2009) Repeated pupil size measurements in refractive surgery candidates. J Cataract Refract Surg 35:2099–2102PubMedCrossRef Robl C, Sliesoraityte I, Hillenkamp J, Prahs P, Lohmann CP, Helbig H, Herrmann WA (2009) Repeated pupil size measurements in refractive surgery candidates. J Cataract Refract Surg 35:2099–2102PubMedCrossRef
42.
go back to reference Schallhorn SC, Kaupp SE, Tanzer DJ, Tidwell J, Laurent J, Bourque LB (2003) Pupil size and quality of vision after LASIK. Ophthalmology 110:1606–1614PubMedCrossRef Schallhorn SC, Kaupp SE, Tanzer DJ, Tidwell J, Laurent J, Bourque LB (2003) Pupil size and quality of vision after LASIK. Ophthalmology 110:1606–1614PubMedCrossRef
43.
go back to reference Roberts CW, Koester CJ (1993) Optical zone diameters for photorefractive corneal surgery. Invest Ophthalmol Vis Sci 34:2275–2281PubMed Roberts CW, Koester CJ (1993) Optical zone diameters for photorefractive corneal surgery. Invest Ophthalmol Vis Sci 34:2275–2281PubMed
44.
go back to reference Kohnen T, Bühren J, Kühne C, Mirshahi A (2004) Wavefront-guided LASIK with the Zyoptix 3.1 system for the correction of myopia and compound myopic astigmatism with 1-year follow-up: clinical outcome and change in higher order aberrations. Ophthalmology 111:2175–2185PubMedCrossRef Kohnen T, Bühren J, Kühne C, Mirshahi A (2004) Wavefront-guided LASIK with the Zyoptix 3.1 system for the correction of myopia and compound myopic astigmatism with 1-year follow-up: clinical outcome and change in higher order aberrations. Ophthalmology 111:2175–2185PubMedCrossRef
45.
go back to reference Oshika T, Klyce SD, Applegate RA, Howland HC, El Danasoury MA (1999) Comparison of corneal wavefront aberrations after photorefractive keratectomy and laser in situ keratomileusis. Am J Ophthalmol 127:1–7PubMedCrossRef Oshika T, Klyce SD, Applegate RA, Howland HC, El Danasoury MA (1999) Comparison of corneal wavefront aberrations after photorefractive keratectomy and laser in situ keratomileusis. Am J Ophthalmol 127:1–7PubMedCrossRef
46.
go back to reference Martínez CE, Applegate RA, Klyce SD, McDonald MB, Medina JP, Howland HC (1998) Effect of pupillary dilation on corneal optical aberrations after photorefractive keratectomy. Arch Ophthalmol 116:1053–1062PubMed Martínez CE, Applegate RA, Klyce SD, McDonald MB, Medina JP, Howland HC (1998) Effect of pupillary dilation on corneal optical aberrations after photorefractive keratectomy. Arch Ophthalmol 116:1053–1062PubMed
47.
go back to reference Maguire LJ (1994) Keratorefractive surgery, success, and the public health. Am J Ophthalmol 117:394–398PubMed Maguire LJ (1994) Keratorefractive surgery, success, and the public health. Am J Ophthalmol 117:394–398PubMed
48.
go back to reference Bradley JC, Anderson JE, Xu KT, Brown SM (2005) Comparison of Colvard pupillometer and infrared digital photography for measurement of the dark-adapted pupil diameter. J Cataract Refract Surg 31:2129–2132PubMedCrossRef Bradley JC, Anderson JE, Xu KT, Brown SM (2005) Comparison of Colvard pupillometer and infrared digital photography for measurement of the dark-adapted pupil diameter. J Cataract Refract Surg 31:2129–2132PubMedCrossRef
49.
go back to reference Spadea L, Giammaria D, Ferrante R, Balestrazzi E (2005) Pre-excimer laser and post-excimer laser refractive surgery measurements of scotopic pupil diameter using 2 pupillometers. Ophthalmology 112:1003–1008PubMedCrossRef Spadea L, Giammaria D, Ferrante R, Balestrazzi E (2005) Pre-excimer laser and post-excimer laser refractive surgery measurements of scotopic pupil diameter using 2 pupillometers. Ophthalmology 112:1003–1008PubMedCrossRef
50.
go back to reference Chan A, Manche EE (2011) Effect of preoperative pupil size on quality of vision after wavefront-guided LASIK. Ophthalmology 118:736–741PubMedCrossRef Chan A, Manche EE (2011) Effect of preoperative pupil size on quality of vision after wavefront-guided LASIK. Ophthalmology 118:736–741PubMedCrossRef
51.
go back to reference Bühren J, Kühne C, Kohnen T (2005) Influence of pupil and optical zone diameter on higher-order aberrations after wavefront-guided myopic LASIK. J Cataract Refract Surg 31:2272–2280PubMedCrossRef Bühren J, Kühne C, Kohnen T (2005) Influence of pupil and optical zone diameter on higher-order aberrations after wavefront-guided myopic LASIK. J Cataract Refract Surg 31:2272–2280PubMedCrossRef
52.
go back to reference Holladay JT, Janes JA (2002) Topographic changes in corneal asphericity and effective optical zone after laser in situ keratomileusis. J Cataract Refract Surg 28:942–947PubMedCrossRef Holladay JT, Janes JA (2002) Topographic changes in corneal asphericity and effective optical zone after laser in situ keratomileusis. J Cataract Refract Surg 28:942–947PubMedCrossRef
53.
go back to reference Tahzib NG, Bootsma SJ, Eggink FA, Nuijts RM (2006) Functional outcome and patient satisfaction after Artisan phakic intraocular lens implantation for the correction of myopia. Am J Ophthalmol 142:31–39PubMedCrossRef Tahzib NG, Bootsma SJ, Eggink FA, Nuijts RM (2006) Functional outcome and patient satisfaction after Artisan phakic intraocular lens implantation for the correction of myopia. Am J Ophthalmol 142:31–39PubMedCrossRef
54.
go back to reference Pop M, Payette Y (2004) Risk factors for night vision complaints after LASIK for myopia. Ophthalmology 111:3–10PubMedCrossRef Pop M, Payette Y (2004) Risk factors for night vision complaints after LASIK for myopia. Ophthalmology 111:3–10PubMedCrossRef
55.
go back to reference Alió JL, Elkady B, Ortiz D, Bernabeu G (2008) Clinical outcomes and intraocular optical quality of a diffractive multifocal intraocular lens with asymmetrical light distribution. J Cataract Refract Surg 34:942–948PubMedCrossRef Alió JL, Elkady B, Ortiz D, Bernabeu G (2008) Clinical outcomes and intraocular optical quality of a diffractive multifocal intraocular lens with asymmetrical light distribution. J Cataract Refract Surg 34:942–948PubMedCrossRef
56.
go back to reference de Vries NE, Webers CA, Touwslager WR, Bauer NJ, de Brabander J, Berendschot TT, Nuijts RM (2011) Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg 37:859–865PubMedCrossRef de Vries NE, Webers CA, Touwslager WR, Bauer NJ, de Brabander J, Berendschot TT, Nuijts RM (2011) Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg 37:859–865PubMedCrossRef
Metadata
Title
Can we measure mesopic pupil size with the cobalt blue light slit-lamp biomicroscopy method?
Authors
Miguel J. Maldonado
Alberto López-Miguel
David P. Piñero
José R. Juberías
Juan C. Nieto
Jorge L. Alió
Publication date
01-11-2012
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 11/2012
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-011-1909-0

Other articles of this Issue 11/2012

Graefe's Archive for Clinical and Experimental Ophthalmology 11/2012 Go to the issue