Can virtual monochromatic images from dual-energy CT replace low-kVp images for abdominal contrast-enhanced CT in small- and medium-sized patients?

Authors: Peijie Lv, Zhigang Zhou, Jie Liu, Yaru Chai, Huiping Zhao, Hua Guo, Daniele Marin, Jianbo Gao

Published in: European Radiology | Issue 6/2019

Login to get access

Abstract

Objective

To investigate the image quality and radiation dose of dual-energy computed tomography (DECT) with automatic spectral imaging protocol selection (ASIS) compared with those of low-kVp CT in abdominal multiphase CT.

Methods

Four groups of 60 patients each underwent abdominal scans with low-kVp CT (A, 80 kVp/300 mg I/kg, body mass index [BMI] ≤ 23.9 kg/m2; C, 100 kVp/400 mg I/kg, BMI ranging from 24 to 28.9 kg/m2) or DECT with ASIS, and the 40- to 60-keV virtual monochromatic images (VMIs) generated (B and D) were matched by age, gender, BMI, cross-sectional area, and contrast agent dose; 9 patients were excluded due to technical failures. The CT number, image noise, contrast-to-noise ratio, and subjective image quality were compared between the matched protocols (A and B or C and D) on 1.25-mm reconstructed images.

Results

VMIs at approximately 55 keV and 62 keV had CT numbers and contrast similar to those of 80-kVp and 100-kVp CT images, respectively. Compared to matched low-kVp images, VMIs at 50 keV provided a higher CT number and image noise and a similar or higher contrast and overall image quality. The radiation dose for DECT was higher than that of 80-kVp CT (increased by 10%), but was similar to that of 100-kVp CT.

Conclusion

Compared to matched low-kVp CT, VMIs at 50 keV in DECT with ASIS provided similar or higher overall image quality, with no or minimal dose penalty in small- and medium-sized patients.

Key Points

• Virtual monochromatic images at approximately 55 keV and 62 keV have CT numbers and contrast similar to those of 80-kVp and 100-kVp CT images, respectively, with a given noise index.
• The radiation dose in dual-energy CT with automatic spectral imaging protocol selection was slightly higher than that of 80-kVp CT (increased by 10%) but was similar to that of 100-kVp CT.
• Dual-energy CT may be able to replace l00-kVp CT for routine clinical abdominal contrast-enhanced CT scans.
Literature
1.
go back to reference Lee KH, Lee JM, Moon SK et al (2012) Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 265:437–447CrossRefPubMed Lee KH, Lee JM, Moon SK et al (2012) Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 265:437–447CrossRefPubMed
2.
go back to reference Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology 254:145–153CrossRefPubMed Marin D, Nelson RC, Schindera ST et al (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology 254:145–153CrossRefPubMed
3.
go back to reference Marin D, Nelson RC, Samei E et al (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection--initial clinical experience. Radiology 251:771–779CrossRefPubMed Marin D, Nelson RC, Samei E et al (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection--initial clinical experience. Radiology 251:771–779CrossRefPubMed
4.
go back to reference Schindera ST, Diedrichsen L, Müller HC et al (2011) Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study. Radiology 260:454–462CrossRefPubMed Schindera ST, Diedrichsen L, Müller HC et al (2011) Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study. Radiology 260:454–462CrossRefPubMed
5.
go back to reference Marin D, Nelson RC, Barnhart H et al (2010) Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique--preliminary results. Radiology 256:450–459CrossRefPubMed Marin D, Nelson RC, Barnhart H et al (2010) Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique--preliminary results. Radiology 256:450–459CrossRefPubMed
6.
go back to reference Lv P, Lin XZ, Chen K, Gao J (2012) Spectral CT in patients with small HCC: investigation of image quality and diagnostic accuracy. Eur Radiol 22:2117–2124CrossRefPubMed Lv P, Lin XZ, Chen K, Gao J (2012) Spectral CT in patients with small HCC: investigation of image quality and diagnostic accuracy. Eur Radiol 22:2117–2124CrossRefPubMed
7.
go back to reference Noda Y, Kanematsu M, Goshima S et al (2015) Reducing iodine load in hepatic CT for patients with chronic liver disease with a combination of low-tube-voltage and adaptive statistical iterative reconstruction. Eur J Radiol l84:11–18CrossRef Noda Y, Kanematsu M, Goshima S et al (2015) Reducing iodine load in hepatic CT for patients with chronic liver disease with a combination of low-tube-voltage and adaptive statistical iterative reconstruction. Eur J Radiol l84:11–18CrossRef
8.
go back to reference Yamada Y, Jinzaki M, Hosokawa T, Tanami Y, Abe T, Kuribayashi S (2014) Abdominal CT: an intra-individual comparison between virtual monochromatic spectral and polychromatic 120-kVp images obtained during the same examination. Eur J Radiol 83:1715–1722CrossRefPubMed Yamada Y, Jinzaki M, Hosokawa T, Tanami Y, Abe T, Kuribayashi S (2014) Abdominal CT: an intra-individual comparison between virtual monochromatic spectral and polychromatic 120-kVp images obtained during the same examination. Eur J Radiol 83:1715–1722CrossRefPubMed
9.
go back to reference Marin D, Boll DT, Mileto A, Nelson RC (2014) State of the art: dual-energy CT of the abdomen. Radiology 271:327–342CrossRefPubMed Marin D, Boll DT, Mileto A, Nelson RC (2014) State of the art: dual-energy CT of the abdomen. Radiology 271:327–342CrossRefPubMed
10.
go back to reference Solomon J, Mileto A, Nelson RC, Roy Choudhury K, Samei E (2016) Quantitative features of liver lesions, lung nodules, and renal stones at multi-detector row CT examinations: dependency on radiation dose and reconstruction algorithm. Radiology 279:185–194 Solomon J, Mileto A, Nelson RC, Roy Choudhury K, Samei E (2016) Quantitative features of liver lesions, lung nodules, and renal stones at multi-detector row CT examinations: dependency on radiation dose and reconstruction algorithm. Radiology 279:185–194
11.
go back to reference Lv P, Liu J, Yan X et al (2017) CT spectral imaging for monitoring the therapeutic efficacy of VEGF receptor kinase inhibitor AG-013736 in rabbit VX2 liver tumours. Eur Radiol 27:918–926CrossRefPubMed Lv P, Liu J, Yan X et al (2017) CT spectral imaging for monitoring the therapeutic efficacy of VEGF receptor kinase inhibitor AG-013736 in rabbit VX2 liver tumours. Eur Radiol 27:918–926CrossRefPubMed
12.
go back to reference Marin D, Davis D, Roy Choudhury K et al (2017) Characterization of small focal renal lesions: diagnostic accuracy with single-phase contrast-enhanced dual-energy CT with material attenuation analysis compared with conventional attenuation measurements. Radiology 284:737–747 Marin D, Davis D, Roy Choudhury K et al (2017) Characterization of small focal renal lesions: diagnostic accuracy with single-phase contrast-enhanced dual-energy CT with material attenuation analysis compared with conventional attenuation measurements. Radiology 284:737–747
13.
go back to reference Lv P, Liu J, Chai Y, Yan X, Gao J, Dong J (2017) Automatic spectral imaging protocol selection and iterative reconstruction in abdominal CT with reduced contrast agent dose: initial experience. Eur Radiol 27:374–383CrossRefPubMed Lv P, Liu J, Chai Y, Yan X, Gao J, Dong J (2017) Automatic spectral imaging protocol selection and iterative reconstruction in abdominal CT with reduced contrast agent dose: initial experience. Eur Radiol 27:374–383CrossRefPubMed
14.
go back to reference De Cecco CN, Darnell A, Macías N et al (2013) Second-generation dual-energy computed tomography of the abdomen: radiation dose comparison with 64- and 128-row single-energy acquisition. J Comput Assist Tomogr 37:543–546CrossRefPubMed De Cecco CN, Darnell A, Macías N et al (2013) Second-generation dual-energy computed tomography of the abdomen: radiation dose comparison with 64- and 128-row single-energy acquisition. J Comput Assist Tomogr 37:543–546CrossRefPubMed
15.
go back to reference Purysko AS, Primak AN, Baker ME et al (2014) Comparison of radiation dose and image quality from single-energy and dual-energy CT examinations in the same patients screened for hepatocellular carcinoma. Clin Radiol 69:e538–e544CrossRefPubMed Purysko AS, Primak AN, Baker ME et al (2014) Comparison of radiation dose and image quality from single-energy and dual-energy CT examinations in the same patients screened for hepatocellular carcinoma. Clin Radiol 69:e538–e544CrossRefPubMed
16.
go back to reference Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379CrossRefPubMedPubMedCentral Yu L, Christner JA, Leng S, Wang J, Fletcher JG, McCollough CH (2011) Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality. Med Phys 38:6371–6379CrossRefPubMedPubMedCentral
17.
go back to reference Chen YA, Prabhudesai V, Castel H, Gupta S (2016) CT scan does not differentiate patients with hepatopulmonary syndrome from other patients with liver disease. PLoS One 11:e0158637CrossRefPubMedPubMedCentral Chen YA, Prabhudesai V, Castel H, Gupta S (2016) CT scan does not differentiate patients with hepatopulmonary syndrome from other patients with liver disease. PLoS One 11:e0158637CrossRefPubMedPubMedCentral
18.
go back to reference Yedibela S, Elad L, Wein A et al (2005) Neoadjuvant chemotherapy does not increase postoperative complication rate after resection of colorectal liver metastases. Eur J Surg Oncol 31:141–146CrossRefPubMed Yedibela S, Elad L, Wein A et al (2005) Neoadjuvant chemotherapy does not increase postoperative complication rate after resection of colorectal liver metastases. Eur J Surg Oncol 31:141–146CrossRefPubMed
19.
go back to reference WHO Expert Consultation (2004) Appropriate body-mass index for Asian populations and its implications for policy and intervention strategies. Lancet 363:157–163 WHO Expert Consultation (2004) Appropriate body-mass index for Asian populations and its implications for policy and intervention strategies. Lancet 363:157–163
20.
go back to reference Ichikawa T, Erturk SM, Araki T (2006) Multiphasic contrast-enhanced multidetector-row CT of liver: contrast-enhancement theory and practical scan protocol with a combination of fixed injection duration and patients’ body-weight-tailored dose of contrast material. Eur J Radiol 58:165–176CrossRefPubMed Ichikawa T, Erturk SM, Araki T (2006) Multiphasic contrast-enhanced multidetector-row CT of liver: contrast-enhancement theory and practical scan protocol with a combination of fixed injection duration and patients’ body-weight-tailored dose of contrast material. Eur J Radiol 58:165–176CrossRefPubMed
21.
go back to reference Gervaise A, Naulet P, Beuret F et al (2014) Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: impact of body mass index. AJR Am J Roentgenol 202:553–560CrossRefPubMed Gervaise A, Naulet P, Beuret F et al (2014) Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: impact of body mass index. AJR Am J Roentgenol 202:553–560CrossRefPubMed
22.
go back to reference Zhu Z, Zhao XM, Zhao YF, Wang XY, Zhou CW (2015) Feasibility study of using Gemstone Spectral Imaging (GSI) and adaptive statistical iterative reconstruction (ASIR) for reducing radiation and iodine contrast dose in abdominal CT patients with high BMI values. PLoS One 10:e0129201CrossRefPubMedPubMedCentral Zhu Z, Zhao XM, Zhao YF, Wang XY, Zhou CW (2015) Feasibility study of using Gemstone Spectral Imaging (GSI) and adaptive statistical iterative reconstruction (ASIR) for reducing radiation and iodine contrast dose in abdominal CT patients with high BMI values. PLoS One 10:e0129201CrossRefPubMedPubMedCentral
23.
go back to reference Awai K, Takada K, Onishi H, Hori S (2002) Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT. Radiology 224:757–763CrossRefPubMed Awai K, Takada K, Onishi H, Hori S (2002) Aortic and hepatic enhancement and tumor-to-liver contrast: analysis of the effect of different concentrations of contrast material at multi-detector row helical CT. Radiology 224:757–763CrossRefPubMed
24.
go back to reference Lv P, Lin XZ, Li J, Li W, Chen K (2011) Differentiation of small hepatic hemangioma from small hepatocellular carcinoma: recently introduced spectral CT method. Radiology 259:720–729CrossRefPubMed Lv P, Lin XZ, Li J, Li W, Chen K (2011) Differentiation of small hepatic hemangioma from small hepatocellular carcinoma: recently introduced spectral CT method. Radiology 259:720–729CrossRefPubMed
25.
go back to reference Wichmann JL, Hardie AD, Schoepf UJ et al (2017) Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT. Eur Radiol 27(2):642–650CrossRefPubMed Wichmann JL, Hardie AD, Schoepf UJ et al (2017) Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT. Eur Radiol 27(2):642–650CrossRefPubMed
26.
go back to reference Nakayama Y, Awai K, Funama Y et al (2005) Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. Radiology 237:945–951CrossRefPubMed Nakayama Y, Awai K, Funama Y et al (2005) Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. Radiology 237:945–951CrossRefPubMed
27.
go back to reference Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection publication 103 or dual-energy scanning. AJR Am J Roentgenol 194:881–889CrossRefPubMed Christner JA, Kofler JM, McCollough CH (2010) Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection publication 103 or dual-energy scanning. AJR Am J Roentgenol 194:881–889CrossRefPubMed
28.
go back to reference Roch P, Aubert B (2013) French diagnostic reference levels in diagnostic radiology, computed tomography and nuclear medicine: 2004-2008 review. Radiat Prot Dosimetry 154:52–75CrossRefPubMed Roch P, Aubert B (2013) French diagnostic reference levels in diagnostic radiology, computed tomography and nuclear medicine: 2004-2008 review. Radiat Prot Dosimetry 154:52–75CrossRefPubMed
29.
go back to reference Simantirakis G, Hourdakis CJ, Economides S et al (2015) Diagnostic reference levels and patient doses in computed tomography examinations in Greece. Radiat Prot Dosimetry 163:319–324CrossRefPubMed Simantirakis G, Hourdakis CJ, Economides S et al (2015) Diagnostic reference levels and patient doses in computed tomography examinations in Greece. Radiat Prot Dosimetry 163:319–324CrossRefPubMed
30.
go back to reference Treier R, Aroua A, Verdun FR, Samara E, Stuessi A, Trueb PR (2010) Patient doses in CT examinations in Switzerland: implementation of national diagnostic reference levels. Radiat Prot Dosimetry 142:244–254CrossRefPubMed Treier R, Aroua A, Verdun FR, Samara E, Stuessi A, Trueb PR (2010) Patient doses in CT examinations in Switzerland: implementation of national diagnostic reference levels. Radiat Prot Dosimetry 142:244–254CrossRefPubMed
31.
go back to reference Palorini F, Origgi D, Granata C, Matranga D, Salerno S (2014) Adult exposures from MDCT including multiphase studies: first Italian nationwide survey. Eur Radiol 24:469–483CrossRefPubMed Palorini F, Origgi D, Granata C, Matranga D, Salerno S (2014) Adult exposures from MDCT including multiphase studies: first Italian nationwide survey. Eur Radiol 24:469–483CrossRefPubMed
32.
go back to reference Agrawal MD, Pinho DF, Kulkarni NM, Hahn PF, Guimaraes AR, Sahani DV (2014) Oncologic applications of dual-energy CT in the abdomen. Radiographics 34:589–612CrossRefPubMed Agrawal MD, Pinho DF, Kulkarni NM, Hahn PF, Guimaraes AR, Sahani DV (2014) Oncologic applications of dual-energy CT in the abdomen. Radiographics 34:589–612CrossRefPubMed
33.
go back to reference Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262CrossRefPubMed Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262CrossRefPubMed
34.
go back to reference Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199:S9–S15CrossRefPubMed
35.
go back to reference Simons D, Kachelriess M, Schlemmer HP (2014) Recent developments of dual-energy CT in oncology. Eur Radiol 24:930–939CrossRefPubMed Simons D, Kachelriess M, Schlemmer HP (2014) Recent developments of dual-energy CT in oncology. Eur Radiol 24:930–939CrossRefPubMed
36.
go back to reference Chai Y, Xing J, Gao J et al (2016) Feasibility of virtual nonenhanced images derived from single-source fast kVp-switching dual-energy CT in evaluating gastric tumors. Eur J Radiol 85:366–372CrossRefPubMed Chai Y, Xing J, Gao J et al (2016) Feasibility of virtual nonenhanced images derived from single-source fast kVp-switching dual-energy CT in evaluating gastric tumors. Eur J Radiol 85:366–372CrossRefPubMed
Metadata
Title
Can virtual monochromatic images from dual-energy CT replace low-kVp images for abdominal contrast-enhanced CT in small- and medium-sized patients?
Authors
Peijie Lv
Zhigang Zhou
Jie Liu
Yaru Chai
Huiping Zhao
Hua Guo
Daniele Marin
Jianbo Gao
Publication date
01-06-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 6/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5850-z

Other articles of this Issue 6/2019

European Radiology 6/2019 Go to the issue