Skip to main content
Top
Published in: Journal of Artificial Organs 3/2013

01-09-2013 | Original Article

Blood viscometer applying electromagnetically spinning method

Authors: Kazuyoshi Fukunaga, Masaya Onuki, Yoshinori Ohtsuka, Taichi Hirano, Keiji Sakai, Yasuharu Ohgoe, Ayako Katoh, Toshiyuki Yaguchi, Akio Funakubo, Yasuhiro Fukui

Published in: Journal of Artificial Organs | Issue 3/2013

Login to get access

Abstract

Viscosity is an important parameter which affects hemodynamics during extracorporeal circulation and long-term cardiac support. In this study, we have aimed to develop a novel viscometer with which we can easily measure blood viscosity by applying the electromagnetically spinning (EMS) method. In the EMS method, we can rotate an aluminum ball 2 mm in diameter indirectly in a test tube with 0.3 ml sample of a liquid by utilizing the moment caused by the Lorentz force as well as separate the test tube from the viscometer body. First, we calibrated the EMS viscometer by means of liquid samples with known viscosities and computational fluid dynamics. Then, when we measured the viscosity of 9.4 mPa s silicone oil in order to evaluate the performance of the EMS viscometer, the mean viscosity was found to be 9.55 ± 0.10 mPa s at available shear rates from 10 to 240 s−1. Finally, we measured the viscosity of bovine blood. We prepared four blood samples whose hematocrit levels were adjusted to 23, 45, 50, and 70 % and a plasma sample without hemocyte components. As a result, the measurements of blood viscosities showed obedience to Casson’s equation. We found that the viscosity was approximately constant in Newtonian silicone oil, whereas the viscosity decreased with increasing shear rate in non-Newtonian bovine blood. These results suggest that the EMS viscometer will be useful to measure blood viscosity at the clinical site.
Literature
1.
go back to reference Yarnell WG, Baker PM, Sweetnam PM, Bainton D, O’Brien JR, Whitehead PJ, Elwood PC. Fibrinogen, viscosity, and white blood cell count are major risk factors for ischemic heart disease. The Caerphilly and Speedwell collaborative heart disease studies. Circulation. 2001;83:836–44.CrossRef Yarnell WG, Baker PM, Sweetnam PM, Bainton D, O’Brien JR, Whitehead PJ, Elwood PC. Fibrinogen, viscosity, and white blood cell count are major risk factors for ischemic heart disease. The Caerphilly and Speedwell collaborative heart disease studies. Circulation. 2001;83:836–44.CrossRef
2.
go back to reference Shinmyozu K, Hashiguchi K, Maruyama I, Igata A, Ohkatsu Y. Studies on the parameters of blood viscosity and platelet function in patients with transient ischemic attack. Jpn J Stoke. 1983;5:12–9 (in Japanese).CrossRef Shinmyozu K, Hashiguchi K, Maruyama I, Igata A, Ohkatsu Y. Studies on the parameters of blood viscosity and platelet function in patients with transient ischemic attack. Jpn J Stoke. 1983;5:12–9 (in Japanese).CrossRef
3.
go back to reference Zimm BH, Crothers DM. Simplified rotating cylinder viscometer for DNA. Proc Natl Acad Sci USA. 1962;48:905–11.PubMedCrossRef Zimm BH, Crothers DM. Simplified rotating cylinder viscometer for DNA. Proc Natl Acad Sci USA. 1962;48:905–11.PubMedCrossRef
4.
go back to reference Martin JL, Sinusas R, Jacobs RM, Copley AL. Transistorized falling ball viscometer for rapid testing of biological and technological system. Biorheology. 1974;11:449–52.PubMed Martin JL, Sinusas R, Jacobs RM, Copley AL. Transistorized falling ball viscometer for rapid testing of biological and technological system. Biorheology. 1974;11:449–52.PubMed
5.
go back to reference Harwood CF, Pilpel N. Rotational viscometer for powders and granules. Lab Pract. 1968;17:1236–8.PubMed Harwood CF, Pilpel N. Rotational viscometer for powders and granules. Lab Pract. 1968;17:1236–8.PubMed
6.
go back to reference Alexy T, Pais E, Wenby RB, Hogenauer W, Toth K, Meiselman HJ, Kensey KR. Measurement of whole blood viscosity profiles via an automated viscometer : technical details and clinical relevance. Clin Lab. 2005;51:523–9.PubMed Alexy T, Pais E, Wenby RB, Hogenauer W, Toth K, Meiselman HJ, Kensey KR. Measurement of whole blood viscosity profiles via an automated viscometer : technical details and clinical relevance. Clin Lab. 2005;51:523–9.PubMed
7.
go back to reference Eguchi Y, Karino T. Measurement of rheologic property of blood by a falling-ball blood viscometer. Ann Biomed Eng. 2008;36:543–53.CrossRef Eguchi Y, Karino T. Measurement of rheologic property of blood by a falling-ball blood viscometer. Ann Biomed Eng. 2008;36:543–53.CrossRef
8.
go back to reference Sakai K, Hirano T, Hosoda M. Electromagnetically spinning sphere viscometer. Jpn J Appl Phys. 2010;3:016602. Sakai K, Hirano T, Hosoda M. Electromagnetically spinning sphere viscometer. Jpn J Appl Phys. 2010;3:016602.
9.
go back to reference Yasuda M, Kurauchi N, Nakamura M, Hirano T, Sakai K. Application of electro-magnetically spinning sphere viscometer for variety of viscosity measurement. J Soc Rheol Jpn. 2011;39:29–35 (in Japanese).CrossRef Yasuda M, Kurauchi N, Nakamura M, Hirano T, Sakai K. Application of electro-magnetically spinning sphere viscometer for variety of viscosity measurement. J Soc Rheol Jpn. 2011;39:29–35 (in Japanese).CrossRef
10.
go back to reference Scott Blair GW. An equation for the flow of blood, plasma and serum through glass capillaries. Nature. 1959;183:613–4.CrossRef Scott Blair GW. An equation for the flow of blood, plasma and serum through glass capillaries. Nature. 1959;183:613–4.CrossRef
11.
go back to reference Copley AL, Scott Blair GW, Glover FA, Thorley RS. Capillary flow and wall adherence of bovine blood, plasma and serum in contact with glass and fibrin surfaces. Kolloid-Zeitschrift. 1960;168:101–7.CrossRef Copley AL, Scott Blair GW, Glover FA, Thorley RS. Capillary flow and wall adherence of bovine blood, plasma and serum in contact with glass and fibrin surfaces. Kolloid-Zeitschrift. 1960;168:101–7.CrossRef
12.
go back to reference Wells RE, Merrill EW. Influence of flow properties of blood upon viscosity-hematocrit relationships. J Clin Invest. 1962;41:1591–8.PubMedCrossRef Wells RE, Merrill EW. Influence of flow properties of blood upon viscosity-hematocrit relationships. J Clin Invest. 1962;41:1591–8.PubMedCrossRef
Metadata
Title
Blood viscometer applying electromagnetically spinning method
Authors
Kazuyoshi Fukunaga
Masaya Onuki
Yoshinori Ohtsuka
Taichi Hirano
Keiji Sakai
Yasuharu Ohgoe
Ayako Katoh
Toshiyuki Yaguchi
Akio Funakubo
Yasuhiro Fukui
Publication date
01-09-2013
Publisher
Springer Japan
Published in
Journal of Artificial Organs / Issue 3/2013
Print ISSN: 1434-7229
Electronic ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-013-0707-3

Other articles of this Issue 3/2013

Journal of Artificial Organs 3/2013 Go to the issue