Bipolar sealing of lung parenchyma: tests in an ex vivo model

Authors: A. Kirschbaum, A. Clemens, T. Steinfeldt, A. Pehl, C. Meyer, D. K. Bartsch

Published in: Surgical Endoscopy | Issue 1/2015

Login to get access

Abstract

Background

Almost every pulmonary lobe resection requires cutting the lung parenchyma in the area of a lung fissure. A monopolar cutter or stapler is often used for this purpose. The seal should be absolutely airtight to prevent post-operative pulmonary fistulas. In the present study, the bipolar sealing technique was evaluated regarding air tightness of the seals during normal ventilation and its burst pressure in an ex vivo animal model.

Materials and methods

The investigations were carried out on paracardial lung lobes obtained from heart–lung preparations taken from freshly killed pigs at a slaughter house. In the laboratory, each individual lobe was perfused with Ringer’s solution at body temperature and protectively ventilated through a tube (frequency: 20 1/min, p insp = 20 mbar, PEEP +5 mbar). Non-anatomic resection was carried out in the periphery of the lung lobe. The two control groups (12 lobes per group; Group 1—stapler, Group 2—parenchyma suture) were compared to three groups in which different bipolar sealing instruments were used. They were Group 3—MARSEAL® 10 mm (KLS Martin, Tuttlingen); Group 4—MARSEAL® 5 mm; and Group 5—MARCLAMP® (KLS Martin, Tuttlingen). The SealSafe® G3 electric current was used in all cases. Ventilation was continued for 20 min following parenchymal resection. Parenchymal sealing was then judged visually in a water bath and given a score (0–3). Burst pressure (mbar) was measured by increasing the inspiration pressure stepwise. Group mean values were compared (nonparametric Mann–Whitney U test, p < 0.005).

Results

Parenchymal seals were airtight under ventilation throughout the observation period in all groups. Mean burst pressures were as follows: Group 1: 47.1 ± 6.2 mbar; Group 2: 32.9 ± 3.9 mbar; Group 3: 38.8 ± 2.2 mbar; Group 4: 25.0 ± 6.4 mbar; and Group 5: 32.9 ± 5.8 mbar. Group 1, the stapler group, thus exhibited the highest burst pressures. Burst pressures for Group 3 were significantly greater than those for Group 2 (p < 0.006). Burst pressures for groups 2 and 5 were similar (p = 0.97). However, the burst pressures for Group 4 were significantly lower than those for Group 2 (p < 0.001).

Conclusion

MARSEAL® 10 mm and MARCLAMP® achieved adequate burst pressures compared to the two control groups and thus might be suitable for clinical use.
Literature
2.
go back to reference Abolhoda A et al (1998) Prolonged air leak following radical upper lobectomy: an analysis of incidence and possible risk factors. Chest 113(6):1507–1510PubMedCrossRef Abolhoda A et al (1998) Prolonged air leak following radical upper lobectomy: an analysis of incidence and possible risk factors. Chest 113(6):1507–1510PubMedCrossRef
3.
go back to reference Seely AJ et al (2010) Systematic classification of morbidity and mortality after thoracic surgery. Ann Thorac Surg 90(3):936–942; discussion 942PubMedCrossRef Seely AJ et al (2010) Systematic classification of morbidity and mortality after thoracic surgery. Ann Thorac Surg 90(3):936–942; discussion 942PubMedCrossRef
4.
go back to reference Cerfolio RJ et al (2002) Predictors and treatment of persistent air leaks. Ann Thorac Surg 73(6):1727–1730; discussion 1730–1731PubMedCrossRef Cerfolio RJ et al (2002) Predictors and treatment of persistent air leaks. Ann Thorac Surg 73(6):1727–1730; discussion 1730–1731PubMedCrossRef
5.
go back to reference Bardell T, Petsikas D (2003) What keeps postpulmonary resection patients in hospital? Can Respir J 10(2):86–89PubMed Bardell T, Petsikas D (2003) What keeps postpulmonary resection patients in hospital? Can Respir J 10(2):86–89PubMed
6.
go back to reference Liang S et al (2013) Quantifying the incidence and impact of postoperative prolonged alveolar air leak after pulmonary resection. J Thorac Cardiovasc Surg 145(4):948–954PubMedCrossRef Liang S et al (2013) Quantifying the incidence and impact of postoperative prolonged alveolar air leak after pulmonary resection. J Thorac Cardiovasc Surg 145(4):948–954PubMedCrossRef
7.
go back to reference Droghetti A et al (2008) A prospective randomized trial comparing completion technique of fissures for lobectomy: stapler versus precision dissection and sealant. J Thorac Cardiovasc Surg 136(2):383–391PubMedCrossRef Droghetti A et al (2008) A prospective randomized trial comparing completion technique of fissures for lobectomy: stapler versus precision dissection and sealant. J Thorac Cardiovasc Surg 136(2):383–391PubMedCrossRef
9.
go back to reference Filosso PL et al (2013) Efficacy and safety of human fibrinogen-thrombin patch (TachoSil(R)) in the treatment of postoperative air leakage in patients submitted to redo surgery for lung malignancies: a randomized trial. Interact CardioVasc Thorac Surg 16(5):661–666PubMedCentralPubMedCrossRef Filosso PL et al (2013) Efficacy and safety of human fibrinogen-thrombin patch (TachoSil(R)) in the treatment of postoperative air leakage in patients submitted to redo surgery for lung malignancies: a randomized trial. Interact CardioVasc Thorac Surg 16(5):661–666PubMedCentralPubMedCrossRef
10.
go back to reference Anegg U et al (2007) Efficiency of fleece-bound sealing (TachoSil) of air leaks in lung surgery: a prospective randomised trial. Eur J Cardiothorac Surg 31(2):198–202PubMedCrossRef Anegg U et al (2007) Efficiency of fleece-bound sealing (TachoSil) of air leaks in lung surgery: a prospective randomised trial. Eur J Cardiothorac Surg 31(2):198–202PubMedCrossRef
11.
go back to reference Zhang R et al (2012) TissuePatch as a novel synthetic sealant for repair of superficial lung defect: in vitro tests results. Ann Surg Innov Res 6(1):12PubMedCentralPubMedCrossRef Zhang R et al (2012) TissuePatch as a novel synthetic sealant for repair of superficial lung defect: in vitro tests results. Ann Surg Innov Res 6(1):12PubMedCentralPubMedCrossRef
12.
go back to reference Brunelli A et al (2004) Predictors of prolonged air leak after pulmonary lobectomy. Ann Thorac Surg 77(4):1205–1210; discussion 1210PubMedCrossRef Brunelli A et al (2004) Predictors of prolonged air leak after pulmonary lobectomy. Ann Thorac Surg 77(4):1205–1210; discussion 1210PubMedCrossRef
13.
go back to reference Ahmed A, Page RD (2008) The utility of intrapleural instillation of autologous blood for prolonged air leak after lobectomy. Curr Opin Pulm Med 14(4):343–347PubMedCrossRef Ahmed A, Page RD (2008) The utility of intrapleural instillation of autologous blood for prolonged air leak after lobectomy. Curr Opin Pulm Med 14(4):343–347PubMedCrossRef
14.
go back to reference Athanassiadi K, Bagaev E, Haverich A (2009) Autologous blood pleurodesis for persistent air leak. Thorac Cardiovasc Surg 57(8):476–479PubMedCrossRef Athanassiadi K, Bagaev E, Haverich A (2009) Autologous blood pleurodesis for persistent air leak. Thorac Cardiovasc Surg 57(8):476–479PubMedCrossRef
15.
go back to reference Lang-Lazdunski L, Coonar AS (2004) A prospective study of autologous ‘blood patch’ pleurodesis for persistent air leak after pulmonary resection. Eur J Cardiothorac Surg 26(5):897–900PubMedCrossRef Lang-Lazdunski L, Coonar AS (2004) A prospective study of autologous ‘blood patch’ pleurodesis for persistent air leak after pulmonary resection. Eur J Cardiothorac Surg 26(5):897–900PubMedCrossRef
16.
go back to reference Elhao M et al (2009) Efficacy of using electrosurgical bipolar vessel sealing during vaginal hysterectomy in patients with different degrees of operative difficulty: a randomised controlled trial. Eur J Obstet Gynecol Reprod Biol 147(1):86–90PubMedCrossRef Elhao M et al (2009) Efficacy of using electrosurgical bipolar vessel sealing during vaginal hysterectomy in patients with different degrees of operative difficulty: a randomised controlled trial. Eur J Obstet Gynecol Reprod Biol 147(1):86–90PubMedCrossRef
17.
go back to reference Ashkenazi D, Mazzawi S, Rakover Y (2006) Hemostasis and bleeding control in thyroid surgery using electrothermal system (Ligasure): our experience in 317 patients. Harefuah 145(8): 561–564, 632, 631 Ashkenazi D, Mazzawi S, Rakover Y (2006) Hemostasis and bleeding control in thyroid surgery using electrothermal system (Ligasure): our experience in 317 patients. Harefuah 145(8): 561–564, 632, 631
18.
go back to reference Macario A et al (2008) Operative time and other outcomes of the electrothermal bipolar vessel sealing system (LigaSure) versus other methods for surgical hemostasis: a meta-analysis. Surg Innov 15(4):284–291PubMedCrossRef Macario A et al (2008) Operative time and other outcomes of the electrothermal bipolar vessel sealing system (LigaSure) versus other methods for surgical hemostasis: a meta-analysis. Surg Innov 15(4):284–291PubMedCrossRef
19.
go back to reference Manouras A et al (2005) Electrothermal bipolar vessel sealing system is a safe and time-saving alternative to classic suture ligation in total thyroidectomy. Head Neck 27(11):959–962PubMedCrossRef Manouras A et al (2005) Electrothermal bipolar vessel sealing system is a safe and time-saving alternative to classic suture ligation in total thyroidectomy. Head Neck 27(11):959–962PubMedCrossRef
20.
go back to reference Shigemura N et al (2004) A new tissue-sealing technique using the Ligasure system for nonanatomical pulmonary resection: preliminary results of sutureless and stapleless thoracoscopic surgery. Ann Thorac Surg 77(4):1415–1418; discussion 1419PubMedCrossRef Shigemura N et al (2004) A new tissue-sealing technique using the Ligasure system for nonanatomical pulmonary resection: preliminary results of sutureless and stapleless thoracoscopic surgery. Ann Thorac Surg 77(4):1415–1418; discussion 1419PubMedCrossRef
21.
go back to reference Tirabassi MV et al (2004) Quantitation of lung sealing in the survival swine model. J Pediatr Surg 39(3):387–390PubMedCrossRef Tirabassi MV et al (2004) Quantitation of lung sealing in the survival swine model. J Pediatr Surg 39(3):387–390PubMedCrossRef
22.
go back to reference Relave F et al (2010) Thoracoscopic lung biopsies in heaves-affected horses using a bipolar tissue sealing system. Vet Surg 39(7):839–846PubMedCrossRef Relave F et al (2010) Thoracoscopic lung biopsies in heaves-affected horses using a bipolar tissue sealing system. Vet Surg 39(7):839–846PubMedCrossRef
23.
go back to reference Mayhew PD et al (2012) Use of the Ligasure vessel-sealing device for thoracoscopic peripheral lung biopsy in healthy dogs. Vet Surg 41(4):523–528PubMedCrossRef Mayhew PD et al (2012) Use of the Ligasure vessel-sealing device for thoracoscopic peripheral lung biopsy in healthy dogs. Vet Surg 41(4):523–528PubMedCrossRef
24.
go back to reference Santini M et al (2006) Use of an electrothermal bipolar tissue sealing system in lung surgery. Eur J Cardiothorac Surg 29(2):226–230PubMedCrossRef Santini M et al (2006) Use of an electrothermal bipolar tissue sealing system in lung surgery. Eur J Cardiothorac Surg 29(2):226–230PubMedCrossRef
25.
go back to reference Macchiarini P et al (1999) Experimental and clinical evaluation of a new synthetic, absorbable sealant to reduce air leaks in thoracic operations. J Thorac Cardiovasc Surg 117(4):751–758PubMedCrossRef Macchiarini P et al (1999) Experimental and clinical evaluation of a new synthetic, absorbable sealant to reduce air leaks in thoracic operations. J Thorac Cardiovasc Surg 117(4):751–758PubMedCrossRef
26.
go back to reference Tan C et al (2011) A prospective randomized controlled study to assess the effectiveness of CoSeal(R) to seal air leaks in lung surgery. Eur J Cardiothorac Surg 40(2):304–308PubMed Tan C et al (2011) A prospective randomized controlled study to assess the effectiveness of CoSeal(R) to seal air leaks in lung surgery. Eur J Cardiothorac Surg 40(2):304–308PubMed
27.
go back to reference Sugi K et al (2003) Use of the bipolar vessel sealing system in lung resection. Kyobu Geka 56(7):551–554PubMed Sugi K et al (2003) Use of the bipolar vessel sealing system in lung resection. Kyobu Geka 56(7):551–554PubMed
Metadata
Title
Bipolar sealing of lung parenchyma: tests in an ex vivo model
Authors
A. Kirschbaum
A. Clemens
T. Steinfeldt
A. Pehl
C. Meyer
D. K. Bartsch
Publication date
01-01-2015
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 1/2015
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-3664-x

Other articles of this Issue 1/2015

Surgical Endoscopy 1/2015 Go to the issue