01-10-2005 | Clinical Investigation

Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry

Authors: Johannes Nepp, Katharina Krepler, Kerstin Jandrasits, Wolfgang Hauff, Georg Hanselmayer, Michaela Velikay-Parel, Karl C. Ossoinig, Andreas Wedrich

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 10/2005

Login to get access

Abstract

Background

We evaluated the validity of calculations for refractive outcome in cataract surgery in silicone oil-filled eyes. The retrosilicone space (RSS) was included in these calculations.

Methods

In a prospective study the axial length (AL) of silicone oil-filled eyes was measured. with standardized A-scan echography (SAE) and partial coherence interferometry (PCI). Meldrum’s formula was used to transform the velocity of ultrasound within the vitreous cavity. To investigate whether refractive outcome can be calculated accurately, we assessed the difference between precalculated and final refractive outcome. Furthermore, we determined the advantages and disadvantages of SAE and PCI. A minor aim was to assess whether theAL of the two eyes differed significantly.

Results

In 85% of 117 eyes the difference between precalculated and postsurgical refraction was smaller than 1 diopter spherical and statistically not significant (p>0.2). The mean AL was 24.1 mm (range 20.0–31.4 mm). The difference in outcome between the two methods was without statistical significance: the AL difference was 0.4 (±2.6) mm on measurement with SAE and 0.04 (±0.46) mm with PCI. PCI has the advantage that it can be performed more easily, without contact, while echography is advantageous in the presence of advanced cataracts. In supine position an oil-free fluid space behind the silicone oil was detected with echography. The mean dimension of this space was 1.9 (±0.67) mm and it was taken into consideration for IOL calculation. The mean AL difference between the two eyes was 0.4 mm, but the difference was greater than 1 mm in 26% of the patients.

Conclusion

The AL of eyes filled with silicone oil can be measured reliably with SAE and PCI. In supine position the RSS has to be considered to obtain more accurate IOL calculations.
Literature
1.
go back to reference Byrne SF, Green RL (1992) Axial eye length measurements. In: Ultrasound of the eye and orbit. Mosby, St Louis, pp 214–242 Byrne SF, Green RL (1992) Axial eye length measurements. In: Ultrasound of the eye and orbit. Mosby, St Louis, pp 214–242
2.
go back to reference Dick B, Pavlovic S, Jacobi FK, Eisenmann D, Jacobi KW (1996) Objektive Bestimmung der Refraktionsänderungen bei silikongefüllten Augen. Einfluss der Kopfposition auf die Refraktion. Klien Monatsbl Augenheilkd 208:459–466 Dick B, Pavlovic S, Jacobi FK, Eisenmann D, Jacobi KW (1996) Objektive Bestimmung der Refraktionsänderungen bei silikongefüllten Augen. Einfluss der Kopfposition auf die Refraktion. Klien Monatsbl Augenheilkd 208:459–466
3.
go back to reference Drexler W, Findl O, Menapace R (1998) PCI: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534CrossRefPubMed Drexler W, Findl O, Menapace R (1998) PCI: a novel approach to biometry in cataract surgery. Am J Ophthalmol 126:524–534CrossRefPubMed
4.
go back to reference Effert R, Lommatzsch A, Wessing A (1996) Klinische Erfahrungen nach Implantation verschiedener Linsentypen bei Silikonöltamponade. Klin Monatsbl Augenheilkd 208:467–471PubMed Effert R, Lommatzsch A, Wessing A (1996) Klinische Erfahrungen nach Implantation verschiedener Linsentypen bei Silikonöltamponade. Klin Monatsbl Augenheilkd 208:467–471PubMed
5.
go back to reference Ghoraba HH, El-Dorghamy AA, Atia AF, Ismail-Yassin AA (2002) The problems of biometry in combined silicone oil removal and cataract extraction. Clinical trial. Retina 22:589–596PubMedCrossRef Ghoraba HH, El-Dorghamy AA, Atia AF, Ismail-Yassin AA (2002) The problems of biometry in combined silicone oil removal and cataract extraction. Clinical trial. Retina 22:589–596PubMedCrossRef
6.
go back to reference Hauff W (1983) Die Biometrie—eine exakte Methode zur Messung der Achsenlänge des Auges. Wien Klin Wochenschr 95:271–274PubMed Hauff W (1983) Die Biometrie—eine exakte Methode zur Messung der Achsenlänge des Auges. Wien Klin Wochenschr 95:271–274PubMed
7.
go back to reference Hitzenberger CK, Drexler W, Baumgartner A, Lexer F, Sattmann H, Esslinger M, Kulhavy M, Fercher AF (1997) Optical measurements of intraocular distances: a comparison of methods. Lasers Light Ophthalmol 8:85–95 Hitzenberger CK, Drexler W, Baumgartner A, Lexer F, Sattmann H, Esslinger M, Kulhavy M, Fercher AF (1997) Optical measurements of intraocular distances: a comparison of methods. Lasers Light Ophthalmol 8:85–95
8.
go back to reference Lai PC, Savage HI, Payman, AS (2003) Refractive surprise after contact ultrasonography. ARVO Annual Meeting Abstract Search and Program Planner, Abstract 212 Lai PC, Savage HI, Payman, AS (2003) Refractive surprise after contact ultrasonography. ARVO Annual Meeting Abstract Search and Program Planner, Abstract 212
9.
go back to reference Lege BA, Haigs W (2004) Laser interference biometry versus ultrasound biometry in certain clinical conditions. Graefes Arch Clin Exp Ophthalmol 242(1):8–12CrossRefPubMed Lege BA, Haigs W (2004) Laser interference biometry versus ultrasound biometry in certain clinical conditions. Graefes Arch Clin Exp Ophthalmol 242(1):8–12CrossRefPubMed
10.
go back to reference Lucke K, Laqua H (1990) Anatomic and functional success. In: Silicone oil in the treatment of complicated retinal detachments: techniques, results and complications. Springer, Berlin Heidelberg New York, pp 45–48 Lucke K, Laqua H (1990) Anatomic and functional success. In: Silicone oil in the treatment of complicated retinal detachments: techniques, results and complications. Springer, Berlin Heidelberg New York, pp 45–48
11.
go back to reference Meldrum ML, Aaberg TM, Patel A, Davis J (1996) Cataract extraction after silicone oil repair of retinal detachments due to necrotizing retinitis. Arch Ophthalmol 114:885–892PubMed Meldrum ML, Aaberg TM, Patel A, Davis J (1996) Cataract extraction after silicone oil repair of retinal detachments due to necrotizing retinitis. Arch Ophthalmol 114:885–892PubMed
12.
go back to reference Moisseiev J, Bartov E, Cahane M, Blumenthal M, Treister G (1992) Cataract extraction in eyes filled with silicone oil. Arch Ophthalmol 110:1649–1651PubMed Moisseiev J, Bartov E, Cahane M, Blumenthal M, Treister G (1992) Cataract extraction in eyes filled with silicone oil. Arch Ophthalmol 110:1649–1651PubMed
13.
go back to reference Murray DC, Durrani OM, Good P, Benson MT, Kirby GR (2002) Biometry of the silicone oil-filled eye. II. Eye 16 (6):727–730CrossRefPubMed Murray DC, Durrani OM, Good P, Benson MT, Kirby GR (2002) Biometry of the silicone oil-filled eye. II. Eye 16 (6):727–730CrossRefPubMed
14.
go back to reference Ossoinig K-C (1979) Standardized echography: basic principles, clinical applications and results. Int Ophthalmol Clin 19:127–210PubMedCrossRef Ossoinig K-C (1979) Standardized echography: basic principles, clinical applications and results. Int Ophthalmol Clin 19:127–210PubMedCrossRef
15.
go back to reference Ossoinig KC (1984) How to obtain maximum measuring accuracies with standardized A-scan. In: Hillman JS, Le May MM (eds) Ophthalmic ultrasonography. Documenta ophthalmologica. Proceedings series, vol 38. Junk, The Hague, pp 197–216 Ossoinig KC (1984) How to obtain maximum measuring accuracies with standardized A-scan. In: Hillman JS, Le May MM (eds) Ophthalmic ultrasonography. Documenta ophthalmologica. Proceedings series, vol 38. Junk, The Hague, pp 197–216
16.
go back to reference Smith RC, Smith GT, Wong D (1990) Refractive changes in silicone oil filled eyes. Eye 4:230–234PubMed Smith RC, Smith GT, Wong D (1990) Refractive changes in silicone oil filled eyes. Eye 4:230–234PubMed
17.
go back to reference Stefansson E, Tiedemann J (1988) Optics of the eye with air or silicone oil. Retina 8:10–19PubMed Stefansson E, Tiedemann J (1988) Optics of the eye with air or silicone oil. Retina 8:10–19PubMed
18.
go back to reference Takei K, Sekine Y, Okamoto F, Hommura S (2002) Measurement of axial length of eyes with incomplete filling of silicone oil in the vitreous cavity using x ray computed tomography. Br J Ophthalmol 86:47–50CrossRefPubMed Takei K, Sekine Y, Okamoto F, Hommura S (2002) Measurement of axial length of eyes with incomplete filling of silicone oil in the vitreous cavity using x ray computed tomography. Br J Ophthalmol 86:47–50CrossRefPubMed
19.
go back to reference Tehrani M et al (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29(4):741–746CrossRefPubMed Tehrani M et al (2003) Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg 29(4):741–746CrossRefPubMed
20.
go back to reference Till P (1976) Solid tissue model for the standardization of the echo ophthalmograph 7200 MA (Kretztechnik) Doc Ophthalmol 41:205–240CrossRefPubMed Till P (1976) Solid tissue model for the standardization of the echo ophthalmograph 7200 MA (Kretztechnik) Doc Ophthalmol 41:205–240CrossRefPubMed
21.
go back to reference Till P (1980) Testung von Schallköpfen auf ihre Eignung zur Gewebsdifferenzierung mit Hilfe des Festkörper-Gewebsphantoms. Klin Monatsbl Augenheilkd 176:337–340PubMedCrossRef Till P (1980) Testung von Schallköpfen auf ihre Eignung zur Gewebsdifferenzierung mit Hilfe des Festkörper-Gewebsphantoms. Klin Monatsbl Augenheilkd 176:337–340PubMedCrossRef
22.
go back to reference Verbeek AM, Bayer AL, Thijssen JM (1981). Echographic diagnosis after intraocular silicone oil injection. Doc Ophthalmol Proc Ser 29:59–66 Verbeek AM, Bayer AL, Thijssen JM (1981). Echographic diagnosis after intraocular silicone oil injection. Doc Ophthalmol Proc Ser 29:59–66
Metadata
Title
Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry
Authors
Johannes Nepp
Katharina Krepler
Kerstin Jandrasits
Wolfgang Hauff
Georg Hanselmayer
Michaela Velikay-Parel
Karl C. Ossoinig
Andreas Wedrich
Publication date
01-10-2005
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 10/2005
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-004-1117-2

Other articles of this Issue 10/2005

Graefe's Archive for Clinical and Experimental Ophthalmology 10/2005 Go to the issue

Announcements

Announcements