Published in: Archives of Orthopaedic and Trauma Surgery 10/2017

01-10-2017 | Orthopaedic Surgery

Biomechanical study of novel unilateral C1 posterior arch screws and C2 laminar screws combined with an ipsilateral crossed C1–C2 pedicle screw–rod fixation for atlantoaxial instability

Authors: Kai Shen, Zhongliang Deng, Junsong Yang, Chao Liu, Ranxi Zhang

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 10/2017

Login to get access

Abstract

Introduction

Current surgical methods to treat atlantoaxial instability pose potential risks to the surrounding blood vessels and nerves of operative approach. Therefore, more secure and highly effective methods are expected. This study sought to assess the biomechanical efficacy of a novel unilateral double screw–rod fixation system by comparing with traditional and emerging fixation methods in cadaveric models.

Materials and methods

Ligamentous cervical spines (C0–C7) from ten fresh cadaveric specimens were used to complete range of motion (ROM) test in their intact condition (control group), destabilization, and stabilization after different fixations, including unilateral C1–C2 pedicle screws (PS) with a screw–rod system (Group A), bilateral C1–C2 PS with screw–rod systems (Group B), unilateral C1 posterior arch screws (PAS) and C2 laminar screws (LS) combined with an ipsilateral paralleled C1–C2 PS–rod (Group C), and unilateral C1 PAS and C2 LS combined with an ipsilateral crossed C1–C2 PS–rod (Group D). After that, pullout strength test was performed between PS and PAS using ten isolated atlas vertebras.

Results

All fixation groups reduced flexibility in all directions compared with both control group and destabilization group. Furthermore, comparisons among different fixation groups showed that bilateral C1–C2 PS–rod (Group B), unilateral C1 PAS + C2 LS combined with an ipsilateral paralleled C1–C2 PS–rod (Group C) and unilateral C1 PAS + C2 LS combined with an ipsilateral crossed C1–C2 PS–rod (Group D) could provide a better stability, respectively, in all directions than unilateral C1–C2 PS–rod (Group A). However, no statistical significance was observed among Groups B, C, and D. Data from pullout strength test showed that both C1 PS (585 ± 53 N) and PAS (463 ± 49 N) could provide high fixed strength, although PS was better (P = 0.009).

Conclusion

The surgical technique of unilateral C1 PAS + C2 LS combined with a ipsilateral crossed C1–C2 PS–rod fixation could provide a better stability than the traditional unilateral PS–rod fixation and a same stability as bilateral PS–rod fixation, but with less risk of neurovascular injury. Therefore, this new technique may provide novel insight for an alternative of atlantoaxial instability treatment.
Literature
2.
go back to reference Nightingale RW, Winkelstein BA, Knaub KE, Richardson WJ, Luck JF, Myers BS (2002) Comparative strengths and structural properties of the upper and lower cervical spine in flexion and extension. J Biomech 35:725–732CrossRefPubMed Nightingale RW, Winkelstein BA, Knaub KE, Richardson WJ, Luck JF, Myers BS (2002) Comparative strengths and structural properties of the upper and lower cervical spine in flexion and extension. J Biomech 35:725–732CrossRefPubMed
3.
go back to reference Gautschi OP, Payer M, Corniola MV, Smoll NR, Schaller K, Tessitore E (2014) Clinically relevant complications related to posterior atlanto-axial fixation in atlanto-axial instability and their management. Clin Neurol Neurosurg 123:131–135. doi:10.1016/j.clineuro.2014.05.020 (Epub 2014 Jun 4) CrossRefPubMed Gautschi OP, Payer M, Corniola MV, Smoll NR, Schaller K, Tessitore E (2014) Clinically relevant complications related to posterior atlanto-axial fixation in atlanto-axial instability and their management. Clin Neurol Neurosurg 123:131–135. doi:10.​1016/​j.​clineuro.​2014.​05.​020 (Epub 2014 Jun 4) CrossRefPubMed
4.
go back to reference Tessitore E, Bartoli A, Schaller K, Payer M (2011) Accuracy of freehand uoroscopy-guided placement of C1 lateral mass and C2 isthmic screws in atlanto-axial instability. Acta Neurochir (Wien) 153:1417–1425. doi:10.1007/s00701-011-1039-9 (discussion 1425. Epub 2011 May 21) CrossRef Tessitore E, Bartoli A, Schaller K, Payer M (2011) Accuracy of freehand uoroscopy-guided placement of C1 lateral mass and C2 isthmic screws in atlanto-axial instability. Acta Neurochir (Wien) 153:1417–1425. doi:10.​1007/​s00701-011-1039-9 (discussion 1425. Epub 2011 May 21) CrossRef
5.
go back to reference Grob D, Crisco JJ III, Panjabi MM, Wang P, Dvorak J (1992) Biomechanical evaluation of four different posterior atlantoaxial fixation techniques. Spine 17:480–490CrossRefPubMed Grob D, Crisco JJ III, Panjabi MM, Wang P, Dvorak J (1992) Biomechanical evaluation of four different posterior atlantoaxial fixation techniques. Spine 17:480–490CrossRefPubMed
6.
go back to reference Henriques T, Cunningham BW, Olerud C, Shimamoto N, Lee GA, Larsson S, McAfee PA (2000) Biomechanical comparison of five different atlantoaxial posterior fixation techniques. Spine 25:2877–2883CrossRefPubMed Henriques T, Cunningham BW, Olerud C, Shimamoto N, Lee GA, Larsson S, McAfee PA (2000) Biomechanical comparison of five different atlantoaxial posterior fixation techniques. Spine 25:2877–2883CrossRefPubMed
7.
go back to reference Kandziora F, Kerschbaumer F, Starker M, Mittlmeier T (2000) Biomechanical assessment of transoral plate fixation for atlantoaxial instability. Spine 25:1555–1561CrossRefPubMed Kandziora F, Kerschbaumer F, Starker M, Mittlmeier T (2000) Biomechanical assessment of transoral plate fixation for atlantoaxial instability. Spine 25:1555–1561CrossRefPubMed
8.
go back to reference Kandziora F, Pflugmacher R, Ludwig K, Duda G, Mittlmeier T, Haas NP (2002) Biomechanical comparison of four anterior atlantoaxial plate systems. J Neurosurg 96:313–320PubMed Kandziora F, Pflugmacher R, Ludwig K, Duda G, Mittlmeier T, Haas NP (2002) Biomechanical comparison of four anterior atlantoaxial plate systems. J Neurosurg 96:313–320PubMed
9.
go back to reference Huang DG, Hao DJ, He BR, Wu QN, Liu TJ, Wang XD, Guo H, Fang XY (2015) Posterior atlantoaxial fixation: a review of all techniques. Spine J 15:2271–2281CrossRefPubMed Huang DG, Hao DJ, He BR, Wu QN, Liu TJ, Wang XD, Guo H, Fang XY (2015) Posterior atlantoaxial fixation: a review of all techniques. Spine J 15:2271–2281CrossRefPubMed
10.
go back to reference Harms J, Melcher RP (2001) Posterior C1–C2 fusion with polyaxial screw and rod fixation. Spine 26:2467–2471CrossRefPubMed Harms J, Melcher RP (2001) Posterior C1–C2 fusion with polyaxial screw and rod fixation. Spine 26:2467–2471CrossRefPubMed
12.
go back to reference Ferguson RL, Tencer AF, Woodard P, Allen BL Jr (1988) Biomechanical comparisons of spinal fracture models and the stabilizing effects of posterior instrumentations. Spine 13:453–460CrossRefPubMed Ferguson RL, Tencer AF, Woodard P, Allen BL Jr (1988) Biomechanical comparisons of spinal fracture models and the stabilizing effects of posterior instrumentations. Spine 13:453–460CrossRefPubMed
13.
go back to reference Dickman CA, Sonntag VK (1998) Posterior C1–C2 transarticular screw fixation for atlantoaxial arthrodesis. Neurosurgery 43:275–280CrossRefPubMed Dickman CA, Sonntag VK (1998) Posterior C1–C2 transarticular screw fixation for atlantoaxial arthrodesis. Neurosurgery 43:275–280CrossRefPubMed
14.
go back to reference Madawi A, Solanki G, Casey AT, Crockard HA (1997) Variation of the groove in the axis vertebra for the vertebral artery: implications for instrumentation. J Bone Joint Surg Br 79:820–823CrossRefPubMed Madawi A, Solanki G, Casey AT, Crockard HA (1997) Variation of the groove in the axis vertebra for the vertebral artery: implications for instrumentation. J Bone Joint Surg Br 79:820–823CrossRefPubMed
15.
go back to reference Abumi K, Takada T, Shono Y, Kaneda K, Fujiya M (1999) Posterior occipitocervical reconstruction using cervical pedicle screws and plate-rod systems. Spine 24:1425–1434CrossRefPubMed Abumi K, Takada T, Shono Y, Kaneda K, Fujiya M (1999) Posterior occipitocervical reconstruction using cervical pedicle screws and plate-rod systems. Spine 24:1425–1434CrossRefPubMed
16.
go back to reference Kuroki H, Rengachary SS, Goel VK, Holekamp SA, Pitkanen V, Ebraheim NA (2005) Biomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation. Neurosurgery 56:151–159 (discussion 151–159) PubMed Kuroki H, Rengachary SS, Goel VK, Holekamp SA, Pitkanen V, Ebraheim NA (2005) Biomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation. Neurosurgery 56:151–159 (discussion 151–159) PubMed
17.
go back to reference Claybrooks R, Kayanja M, Milks R, Benzel E (2007) Atlantoaxial fusion: a biomechanical analysis of two C1–C2 fusion techniques. Spine J 7:682–688CrossRefPubMed Claybrooks R, Kayanja M, Milks R, Benzel E (2007) Atlantoaxial fusion: a biomechanical analysis of two C1–C2 fusion techniques. Spine J 7:682–688CrossRefPubMed
19.
go back to reference Aota Y, Honda A, Uesugi M, Yamashita T, Baba N, Niwa T, Saito T (2006) Vertebral artery injury in C-1 lateral mass screw fixation. Case illustration. J Neurosurg Spine 5:554CrossRefPubMed Aota Y, Honda A, Uesugi M, Yamashita T, Baba N, Niwa T, Saito T (2006) Vertebral artery injury in C-1 lateral mass screw fixation. Case illustration. J Neurosurg Spine 5:554CrossRefPubMed
20.
go back to reference Rocha R, Safavi-Abbasi S, Reis C, Theodore N, Bambakidis N, de Oliveira E, Sonntag VK, Crawford NR (2007) Working area, safety zones, and angles of approach for posterior C-1 lateral mass screw placement: a quantitative anatomical and morphometric evaluation. J Neurosurg Spine 6:247–254CrossRefPubMed Rocha R, Safavi-Abbasi S, Reis C, Theodore N, Bambakidis N, de Oliveira E, Sonntag VK, Crawford NR (2007) Working area, safety zones, and angles of approach for posterior C-1 lateral mass screw placement: a quantitative anatomical and morphometric evaluation. J Neurosurg Spine 6:247–254CrossRefPubMed
21.
go back to reference Paramore CG, Dickman CA, Sonntag VKH (1996) The anatomical suitability of the C1–C2 complex for transarticular screw fixation. J Neurosurg 85:221–224CrossRefPubMed Paramore CG, Dickman CA, Sonntag VKH (1996) The anatomical suitability of the C1–C2 complex for transarticular screw fixation. J Neurosurg 85:221–224CrossRefPubMed
22.
go back to reference Fujii T, Oda T, Kato Y, Fujita S, Tanaka M (2000) Accuracy of atlantoaxial transartivular screw fixaion. Spine 25:1760–1764CrossRef Fujii T, Oda T, Kato Y, Fujita S, Tanaka M (2000) Accuracy of atlantoaxial transartivular screw fixaion. Spine 25:1760–1764CrossRef
24.
go back to reference Miyakoshi N, Hongo M, Kobayashi T, Suzuki T, Abe E, Shimada Y (2014) Comparison between bilateral C2 pedicle screwing and unilateral C2 pedicle screwing, combined with contralateral C2 laminar screwing, for atlantoaxial posterior fixation. Asian Spine J 8:777–785. doi:10.4184/asj.2014.8.6.777 (Epub 2014 Dec 17) CrossRefPubMedPubMedCentral Miyakoshi N, Hongo M, Kobayashi T, Suzuki T, Abe E, Shimada Y (2014) Comparison between bilateral C2 pedicle screwing and unilateral C2 pedicle screwing, combined with contralateral C2 laminar screwing, for atlantoaxial posterior fixation. Asian Spine J 8:777–785. doi:10.​4184/​asj.​2014.​8.​6.​777 (Epub 2014 Dec 17) CrossRefPubMedPubMedCentral
25.
go back to reference Meyer D, Meyer F, Kretschmer T, Börm W (2012) Translaminar screws of the axis–an alternative technique for rigid screw fixation in upper cervical spine instability. Neurosurg Rev 35:255–261. doi:10.1007/s10143-011-0358-x (discussion 261. Epub 2011 Nov 17) CrossRefPubMed Meyer D, Meyer F, Kretschmer T, Börm W (2012) Translaminar screws of the axis–an alternative technique for rigid screw fixation in upper cervical spine instability. Neurosurg Rev 35:255–261. doi:10.​1007/​s10143-011-0358-x (discussion 261. Epub 2011 Nov 17) CrossRefPubMed
28.
go back to reference Jin GX, Wang H (2016) Unilateral C-1 posterior arch screws and C-2 laminar screws combined with a 1-side C1–2 pedicle screw system as salvage fixation for atlantoaxial instability. J Neurosurg Spine 24:315–320CrossRef Jin GX, Wang H (2016) Unilateral C-1 posterior arch screws and C-2 laminar screws combined with a 1-side C1–2 pedicle screw system as salvage fixation for atlantoaxial instability. J Neurosurg Spine 24:315–320CrossRef
Metadata
Title
Biomechanical study of novel unilateral C1 posterior arch screws and C2 laminar screws combined with an ipsilateral crossed C1–C2 pedicle screw–rod fixation for atlantoaxial instability
Authors
Kai Shen
Zhongliang Deng
Junsong Yang
Chao Liu
Ranxi Zhang
Publication date
01-10-2017
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 10/2017
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-017-2781-0

Other articles of this Issue 10/2017

Archives of Orthopaedic and Trauma Surgery 10/2017 Go to the issue