Biomechanical evaluation of the primary stability of pedicle screws after augmentation with an innovative bone stabilizing system

Authors: K. Wegmann, S. Gick, C. Heidemann, D. Pennig, W. F. Neiss, L. P. Müller, P. Eysel, R. Sobottke

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 11/2013

Login to get access

Abstract

Introduction

In today’s aging population, diminished bone quality often affects the outcome of surgical treatment. This occurs especially when surgical implants must be fixed to bone, as it occurs when lumbar fusion is performed with pedicle screws. Besides Polymethylmethacrylate (PMMA) injection, several techniques have been developed to augment pedicle screws. The aim of the current study was to evaluate the primary stability of an innovative system (IlluminOss™) for the augmentation of pedicle screws in an experimental cadaveric setup. IlluminOss™ is an innovative technology featuring cement with similar biochemical characteristics to aluminum-free glass-polyalkenoate cement (GPC).

Materials and methods

IlluminOss™ was inserted transpedicularly via a balloon/catheter system in 40 human cadaveric lumbar vertebrae. For comparability, each vertebra was treated bilaterally with pedicle screws, augmented and non-augmented. The maximum failure load during pull out test was documented by a universal material testing machine.

Results

The results showed significantly higher failure loads for the augmented pedicle screws (Median 555.0 ± 261.0 N, Min. 220.0 N, Max. 1,500.0 N), compared to the native screws (Median 325.0 ± 312.1 N, Min. 29.0 N, Max. 1,400.0 N).

Conclusions

Based on these data, we conclude the IlluminOss™ system can be used to augment primary screw stability regarding axial traction, compared to native screws. The IlluminOss™ monomer offers ease of control for use in biological tissues. In contrast to PMMA, no relevant heat is generated during the hardening process and there is no risk of embolism. Further studies are necessary to evaluate the usefulness of the IlluminOss™ system in the in vivo augmentation of pedicle screws in the future.
Literature
2.
3.
go back to reference Jager PL, Jonkman S, Koolhaas W, Stiekema A, Wolffenbuttel BH, Slart RH (2011) Combined vertebral fracture assessment and bone mineral density measurement: a new standard in the diagnosis of osteoporosis in academic populations. Osteoporos Int 22(4):1059–1068. doi:10.1007/s00198-010-1293-3 PubMedCrossRef Jager PL, Jonkman S, Koolhaas W, Stiekema A, Wolffenbuttel BH, Slart RH (2011) Combined vertebral fracture assessment and bone mineral density measurement: a new standard in the diagnosis of osteoporosis in academic populations. Osteoporos Int 22(4):1059–1068. doi:10.​1007/​s00198-010-1293-3 PubMedCrossRef
4.
go back to reference Shi L, Wang L, Zhang Y, Guo Z, Wu ZX, Liu D, Gao MX, Chen H, Fu SC, Lei W (2012) Improving fixation strength of pedicle screw by microarc oxidation treatment: an experimental study of osteoporotic spine in sheep. J Orthop Res. doi:10.1002/jor.22072 Shi L, Wang L, Zhang Y, Guo Z, Wu ZX, Liu D, Gao MX, Chen H, Fu SC, Lei W (2012) Improving fixation strength of pedicle screw by microarc oxidation treatment: an experimental study of osteoporotic spine in sheep. J Orthop Res. doi:10.​1002/​jor.​22072
5.
go back to reference Liu D, Wu ZX, Pan XM, Fu SC, Gao MX, Shi L, Lei W (2011) Biomechanical comparison of different techniques in primary spinal surgery in osteoporotic cadaveric lumbar vertebrae: expansive pedicle screw versus polymethylmethacrylate-augmented pedicle screw. Arch Orthop Trauma Surg 131(9):1227–1232. doi:10.1007/s00402-011-1290-9 PubMedCrossRef Liu D, Wu ZX, Pan XM, Fu SC, Gao MX, Shi L, Lei W (2011) Biomechanical comparison of different techniques in primary spinal surgery in osteoporotic cadaveric lumbar vertebrae: expansive pedicle screw versus polymethylmethacrylate-augmented pedicle screw. Arch Orthop Trauma Surg 131(9):1227–1232. doi:10.​1007/​s00402-011-1290-9 PubMedCrossRef
6.
go back to reference Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd (2001) Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J 1(2):109–114 S1529-9430(01)00020-1PubMedCrossRef Cook SD, Barbera J, Rubi M, Salkeld SL, Whitecloud TS 3rd (2001) Lumbosacral fixation using expandable pedicle screws. An alternative in reoperation and osteoporosis. Spine J 1(2):109–114 S1529-9430(01)00020-1PubMedCrossRef
7.
go back to reference Amendola L, Gasbarrini A, Fosco M, Simoes CE, Terzi S, De Iure F, Boriani S (2011) Fenestrated pedicle screws for cement-augmented purchase in patients with bone softening: a review of 21 cases. J Orthop Traumatol 12(4):193–199. doi:10.1007/s10195-011-0164-9 PubMedCrossRef Amendola L, Gasbarrini A, Fosco M, Simoes CE, Terzi S, De Iure F, Boriani S (2011) Fenestrated pedicle screws for cement-augmented purchase in patients with bone softening: a review of 21 cases. J Orthop Traumatol 12(4):193–199. doi:10.​1007/​s10195-011-0164-9 PubMedCrossRef
8.
11.
go back to reference Pinera AR, Duran C, Lopez B, Saez I, Correia E, Alvarez L (2011) Instrumented lumbar arthrodesis in elderly patients: prospective study using cannulated cemented pedicle screw instrumentation. Eur Spine J 20(Suppl 3):408–414. doi:10.1007/s00586-011-1907-2 PubMedCrossRef Pinera AR, Duran C, Lopez B, Saez I, Correia E, Alvarez L (2011) Instrumented lumbar arthrodesis in elderly patients: prospective study using cannulated cemented pedicle screw instrumentation. Eur Spine J 20(Suppl 3):408–414. doi:10.​1007/​s00586-011-1907-2 PubMedCrossRef
13.
go back to reference Folsch C, Goost H, Figiel J, Paletta JR, Schultz W, Lakemeier S (2012) Correlation of pull-out strength of cement-augmented pedicle screws with CT-volumetric measurement of cement. Biomed Tech (Berl) 57(6):473–480. doi:10.1515/bmt-2012-0012 CrossRef Folsch C, Goost H, Figiel J, Paletta JR, Schultz W, Lakemeier S (2012) Correlation of pull-out strength of cement-augmented pedicle screws with CT-volumetric measurement of cement. Biomed Tech (Berl) 57(6):473–480. doi:10.​1515/​bmt-2012-0012 CrossRef
14.
16.
17.
18.
go back to reference Chao KH, Lai YS, Chen WC, Chang CM, McClean CJ, Fan CY, Chang CH, Lin LC, Cheng CK (2013) Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. Med Eng Phys. doi:10.1016/j.medengphy.2013.04.007 S1350-4533(13)00100-8 Chao KH, Lai YS, Chen WC, Chang CM, McClean CJ, Fan CY, Chang CH, Lin LC, Cheng CK (2013) Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. Med Eng Phys. doi:10.​1016/​j.​medengphy.​2013.​04.​007 S1350-4533(13)00100-8
20.
go back to reference Burge R, Dawson-Hughes B, Solomon DH, Wong JB, King A, Tosteson A (2007) Incidence and economic burden of osteoporosis-related fractures in the United States, 2005–2025. J Bone Miner Res 22(3):465–475. doi:10.1359/jbmr.061113 PubMedCrossRef Burge R, Dawson-Hughes B, Solomon DH, Wong JB, King A, Tosteson A (2007) Incidence and economic burden of osteoporosis-related fractures in the United States, 2005–2025. J Bone Miner Res 22(3):465–475. doi:10.​1359/​jbmr.​061113 PubMedCrossRef
21.
go back to reference Dempster DW (2011) Osteoporosis and the burden of osteoporosis-related fractures. Am J Manag Care 17(Suppl 6):S164–S169. doi:49414 PubMed Dempster DW (2011) Osteoporosis and the burden of osteoporosis-related fractures. Am J Manag Care 17(Suppl 6):S164–S169. doi:49414 PubMed
22.
go back to reference DeWald CJ, Stanley T (2006) Instrumentation-related complications of multilevel fusions for adult spinal deformity patients over age 65: surgical considerations and treatment options in patients with poor bone quality. Spine (Phila Pa 1976) 31(19 Suppl):S144–S151. doi:10.1097/01.brs.0000236893.65878.39 CrossRef DeWald CJ, Stanley T (2006) Instrumentation-related complications of multilevel fusions for adult spinal deformity patients over age 65: surgical considerations and treatment options in patients with poor bone quality. Spine (Phila Pa 1976) 31(19 Suppl):S144–S151. doi:10.​1097/​01.​brs.​0000236893.​65878.​39 CrossRef
24.
25.
go back to reference Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96(3 Suppl):309–312PubMed Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96(3 Suppl):309–312PubMed
26.
go back to reference Pare PE, Chappuis JL, Rampersaud R, Agarwala AO, Perra JH, Erkan S, Wu C (2011) Biomechanical evaluation of a novel fenestrated pedicle screw augmented with bone cement in osteoporotic spines. Spine (Phila Pa 1976) 36(18):E1210–E1214. doi:10.1097/BRS.0b013e318205e3af CrossRef Pare PE, Chappuis JL, Rampersaud R, Agarwala AO, Perra JH, Erkan S, Wu C (2011) Biomechanical evaluation of a novel fenestrated pedicle screw augmented with bone cement in osteoporotic spines. Spine (Phila Pa 1976) 36(18):E1210–E1214. doi:10.​1097/​BRS.​0b013e318205e3af​ CrossRef
27.
go back to reference Deramond H, Wright NT, Belkoff SM (1999) Temperature elevation caused by bone cement polymerization during vertebroplasty. Bone 25(2 Suppl):17S–21SPubMedCrossRef Deramond H, Wright NT, Belkoff SM (1999) Temperature elevation caused by bone cement polymerization during vertebroplasty. Bone 25(2 Suppl):17S–21SPubMedCrossRef
29.
go back to reference Petridis AK, Maslehaty H, Eichenhofer T, Gillner S, Scholz M (2013) Pulmonary embolism associated with poly(methyl methacrylate) (PMMA) after vertebroplasty. A complication more often than one thinks. Acta Neurochir (Wien) 155(2):319–320. doi:10.1007/s00701-012-1584-x CrossRef Petridis AK, Maslehaty H, Eichenhofer T, Gillner S, Scholz M (2013) Pulmonary embolism associated with poly(methyl methacrylate) (PMMA) after vertebroplasty. A complication more often than one thinks. Acta Neurochir (Wien) 155(2):319–320. doi:10.​1007/​s00701-012-1584-x CrossRef
30.
32.
33.
go back to reference Burkhart KJ, Nowak TE, Blum J, Kuhn S, Welker M, Sternstein W, Mueller LP, Rommens PM (2010) Influence of formalin fixation on the biomechanical properties of human diaphyseal bone. Biomed Tech (Berl) 55(6):361–365. doi:10.1515/BMT.2010.043 CrossRef Burkhart KJ, Nowak TE, Blum J, Kuhn S, Welker M, Sternstein W, Mueller LP, Rommens PM (2010) Influence of formalin fixation on the biomechanical properties of human diaphyseal bone. Biomed Tech (Berl) 55(6):361–365. doi:10.​1515/​BMT.​2010.​043 CrossRef
34.
go back to reference Liljenqvist U, Hackenberg L, Link T, Halm H (2001) Pullout strength of pedicle screws versus pedicle and laminar hooks in the thoracic spine. Acta Orthop Belg 67(2):157–163PubMed Liljenqvist U, Hackenberg L, Link T, Halm H (2001) Pullout strength of pedicle screws versus pedicle and laminar hooks in the thoracic spine. Acta Orthop Belg 67(2):157–163PubMed
35.
go back to reference Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ (2011) Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord 12:33. doi:10.1186/1471-2474-12-33 PubMedCrossRef Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ (2011) Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord 12:33. doi:10.​1186/​1471-2474-12-33 PubMedCrossRef
37.
go back to reference Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ (2009) Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon) 24(8):613–618. doi:10.1016/j.clinbiomech.2009.05.002 CrossRef Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ (2009) Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon) 24(8):613–618. doi:10.​1016/​j.​clinbiomech.​2009.​05.​002 CrossRef
Metadata
Title
Biomechanical evaluation of the primary stability of pedicle screws after augmentation with an innovative bone stabilizing system
Authors
K. Wegmann
S. Gick
C. Heidemann
D. Pennig
W. F. Neiss
L. P. Müller
P. Eysel
R. Sobottke
Publication date
01-11-2013
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 11/2013
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-013-1842-2

Other articles of this Issue 11/2013

Archives of Orthopaedic and Trauma Surgery 11/2013 Go to the issue