Published in: BMC Musculoskeletal Disorders 1/2024

Open Access 01-12-2024 | Research

Biomechanical evaluation of different oblique lumbar interbody fusion constructs: a finite element analysis

Authors: Zhengquan Xu, Qingcong Zheng, Liqun Zhang, Rongsheng Chen, Zhechen Li, Weihong Xu

Published in: BMC Musculoskeletal Disorders | Issue 1/2024

Login to get access

Abstract

Background

Finite element analysis (FEA) was performed to investigate the biomechanical differences between different adjunct fixation methods for oblique lumbar interbody fusion (OLIF) and to further analyze its effect on adjacent segmental degeneration.

Methods

We built a single-segment (Si-segment) finite element model (FEM) for L4-5 and a double-segment (Do-segment) FEM for L3-5. Each complete FEM was supplemented and modified, and both developed two surgical models of OLIF with assisted internal fixation. They were OLIF with posterior bilateral percutaneous pedicle screw (TINA system) fixation (OLIF + BPS) and OLIF with lateral plate system (OLIF + LPS). The range of motion (ROM) and displacement of the vertebral body, cage stress, adjacent segment disc stress, and spinal ligament tension were recorded for the four models during flexion/extension, right/left bending, and right/left rotation by applying follower load.

Results

For the BPS and LPS systems in the six postures of flexion, extension, right/left bending, and right/left rotation, the ROM of L4 in the Si-segment FEM were 0.32°/1.83°, 0.33°/1.34°, 0.23°/0.47°, 0.24°/0.45°, 0.33°/0.79°, and 0.34°/0.62°; the ROM of L4 in the Do-segment FEM were 0.39°/2.00°, 0.37°/1.38°, 0.23°/0.47°, 0.21°/0.44°, 0.33°/0.57°, and 0.31°/0.62°, and the ROM of L3 in the Do-segment FEM were 6.03°/7.31°, 2.52°/3.50°, 4.21°/4.38°, 4.21°/4.42°, 2.09°/2.32°, and 2.07°/2.43°. BPS system had less vertebral displacement, less cage maximum stress, and less spinal ligament tension in Si/Do-segment FEM relative to the LPS system. BPS system had a smaller upper adjacent vertebral ROM, greater intervertebral disc stress in terms of left and right bending as well as left and right rotation compared to the LPS system in the L3-4 of the Do-segment FEM. There was little biomechanical difference between the same fixation system in the Si/Do-segment FEM.

Conclusions

Our finite element analysis showed that compared to OLIF + LPS, OLIF + BPS (TINA) is more effective in reducing interbody stress and spinal ligament tension, and it better maintains the stability of the target segment and provides a better fusion environment to resist cage subsidence. However, OLIF + BPS (TINA) may be more likely to cause adjacent segment degeneration than OLIF + LPS.
Literature
12.
go back to reference Magerl FP. Stabilization of the lower thoracic and lumbar spine with external skeletal fixation. Clin Orthop Relat Res. 1984;189:125–41.CrossRef Magerl FP. Stabilization of the lower thoracic and lumbar spine with external skeletal fixation. Clin Orthop Relat Res. 1984;189:125–41.CrossRef
Metadata
Title
Biomechanical evaluation of different oblique lumbar interbody fusion constructs: a finite element analysis
Authors
Zhengquan Xu
Qingcong Zheng
Liqun Zhang
Rongsheng Chen
Zhechen Li
Weihong Xu
Publication date
01-12-2024
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2024
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-024-07204-8

Other articles of this Issue 1/2024

BMC Musculoskeletal Disorders 1/2024 Go to the issue