Published in: Knee Surgery, Sports Traumatology, Arthroscopy 4/2019

01-04-2019 | KNEE

Asymmetric polyethylene inserts promote favorable kinematics and better clinical outcome compared to symmetric inserts in a mobile bearing total knee arthroplasty

Authors: Gianluca Castellarin, Silvia Pianigiani, Bernardo Innocenti

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 4/2019

Login to get access

Abstract

Purpose

This study aims at comparing the effects of symmetric and asymmetric designs for the polyethylene insert currently available and also for mobile bearing total knee arthroplasty (TKA). The investigation was performed both clinically and biomechanically through finite element analysis.

Methods

303 patients, with a mobile bearing TKA, were analyzed retrospectively. All patients received the same femoral and tibial components; for the insert, 151 patients received a symmetric design (SD) and 152 an asymmetric design (AD). Additionally, a 3D finite element model of a lower leg was developed, resurfaced with the same TKAs and analysed during gait and squat activities. TKA kinematics, and bone-stresses were investigated for the two insert solutions.

Results

After surgery, patients’ average flexion improved from 105°, with 5° of preoperative extension deficit, to 120° (AD-group) and 115° (SD-group) at the latest follow-up. There was no postoperative extension deficit. No pain affected the AD-group, while an antero-lateral pain was reported in some patients of the SD-group. Patients of the AD-group presented a better ability to perform certain physical routines. Biomechanically, the SD induced higher tibial-bone stresses than the AD. Both designs replicated similar kinematics, comparable to literature. However, SD rotates more on the tray, reducing the motion between femoral and polyethylene components, while AD permits greater insert rotation.

Conclusion

The biomechanical analysis justifies the clinical findings. TKA kinematics is similar for the two designs, although the asymmetric solution shows less bone stress, thus resulting as more suitable to be cemented, avoiding lift-off issues, inducing less pain. Clinically, and biomechanically, an asymmetric mobile bearing insert could be a valid alternative to symmetric mobile bearing insert.

Level of evidence

Case–control study retrospective comparative study, III.
Literature
1.
go back to reference Arnout N, Vanlommel L, Vanlommel J, Luyckx JP, Labey L, Innocenti B, Victor J, Bellemans J (2015) Post cam mechanics and tibiofemoral kinematics: a dynamic in vitro analysis of eight posterior stabilized total knee designs. Knee Surg Sports Traumatol Arthrosc 23(11):3343–3353PubMedCrossRef Arnout N, Vanlommel L, Vanlommel J, Luyckx JP, Labey L, Innocenti B, Victor J, Bellemans J (2015) Post cam mechanics and tibiofemoral kinematics: a dynamic in vitro analysis of eight posterior stabilized total knee designs. Knee Surg Sports Traumatol Arthrosc 23(11):3343–3353PubMedCrossRef
2.
go back to reference Belvedere C, Leardini A, Catani F, Pianigiani S, Innocenti B (2017) In vivo kinematics of knee replacement during daily living activities: condylar and post-cam contact assessment by three-dimensional fluoroscopy and finite element analyses. J Orthop Res 35(7):1396–1403PubMedCrossRef Belvedere C, Leardini A, Catani F, Pianigiani S, Innocenti B (2017) In vivo kinematics of knee replacement during daily living activities: condylar and post-cam contact assessment by three-dimensional fluoroscopy and finite element analyses. J Orthop Res 35(7):1396–1403PubMedCrossRef
3.
go back to reference Berend ME, Davis PJ, Ritter MA, Keating EM, Faris PM, Meding JB, Malinzak RA (2010) “Thicker” polyethylene bearings are associated with higher failure rates in primary total knee arthroplasty. J Arthoplasty 25(6):17–20CrossRef Berend ME, Davis PJ, Ritter MA, Keating EM, Faris PM, Meding JB, Malinzak RA (2010) “Thicker” polyethylene bearings are associated with higher failure rates in primary total knee arthroplasty. J Arthoplasty 25(6):17–20CrossRef
4.
go back to reference Brihault J, Navacchia A, Pianigiani S, Labey L, De Corte R, Pascale V, Innocenti (2016) All-polyethylene tibial components generate higher stress and micromotions than metal-backed tibial components in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24(8):2550–2559PubMedCrossRef Brihault J, Navacchia A, Pianigiani S, Labey L, De Corte R, Pascale V, Innocenti (2016) All-polyethylene tibial components generate higher stress and micromotions than metal-backed tibial components in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24(8):2550–2559PubMedCrossRef
5.
go back to reference Caillouette JT, Anzel SH (1990) Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty. Clin Orthop Relat Res 251:198–199 Caillouette JT, Anzel SH (1990) Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty. Clin Orthop Relat Res 251:198–199
6.
go back to reference Carothers JT, Kim RH, Dennis DA, Southworth C (2011) Mobile-bearing total knee arthroplasty: a meta-analysis. J Arthroplasty 26(4):537–542PubMedCrossRef Carothers JT, Kim RH, Dennis DA, Southworth C (2011) Mobile-bearing total knee arthroplasty: a meta-analysis. J Arthroplasty 26(4):537–542PubMedCrossRef
7.
go back to reference Catani F, Innocenti B, Belvedere C, Labey L, Ensini A, Leardini A (2010) The mark coventry award: articular contact estimation in TKA using in vivo kinematics and finite element analysis. Clin Orthop Relat Res 468(1):19–28PubMedCrossRef Catani F, Innocenti B, Belvedere C, Labey L, Ensini A, Leardini A (2010) The mark coventry award: articular contact estimation in TKA using in vivo kinematics and finite element analysis. Clin Orthop Relat Res 468(1):19–28PubMedCrossRef
8.
go back to reference Castellarin G, Manili M, Tavella E (2013) A new methodological and surgical approach to total knee replacement. Sphera Med J 17:22–26 Castellarin G, Manili M, Tavella E (2013) A new methodological and surgical approach to total knee replacement. Sphera Med J 17:22–26
9.
go back to reference Dawson J, Fitzpatrick R, Murray D, Carr A (1998) Questionnaire on the perceptions of patients about total knee replacement. J Bone Jt Surg Br 80(1):63–69CrossRef Dawson J, Fitzpatrick R, Murray D, Carr A (1998) Questionnaire on the perceptions of patients about total knee replacement. J Bone Jt Surg Br 80(1):63–69CrossRef
10.
go back to reference Donahue TLH, Hull ML, Rashid MM, Jacobs CR (2003) How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint. J Biomech 36(1):19–34CrossRef Donahue TLH, Hull ML, Rashid MM, Jacobs CR (2003) How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint. J Biomech 36(1):19–34CrossRef
11.
go back to reference El-Zayat BF, Heyse TJ, Fanciullacci N, Labey L, Fuchs-Winkelmann S, Innocenti B (2016) Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study. Arch Orthop Traum Surg 136(12):1741–1752CrossRef El-Zayat BF, Heyse TJ, Fanciullacci N, Labey L, Fuchs-Winkelmann S, Innocenti B (2016) Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study. Arch Orthop Traum Surg 136(12):1741–1752CrossRef
12.
go back to reference Faul F, Erdfelder E, Buchner A, Lang AG (2009) Statistical power analyses using G*Power 3.1: tests for correlation and regression analyses. Behav Res Methods 41(4):1149–1160PubMedCrossRef Faul F, Erdfelder E, Buchner A, Lang AG (2009) Statistical power analyses using G*Power 3.1: tests for correlation and regression analyses. Behav Res Methods 41(4):1149–1160PubMedCrossRef
13.
go back to reference Faul F, Erdfelder E, Lang AG, Buchner A (2007) G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 39(2):175–191PubMedCrossRef Faul F, Erdfelder E, Lang AG, Buchner A (2007) G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 39(2):175–191PubMedCrossRef
14.
go back to reference Galbusera F, Freutel M, Dürselen L, D’Aiuto M, Croce D, Villa T, Sansone V, Innocenti B (2014) Material models and properties in the finite element analysis of knee ligaments: a literature review. Front Bioeng Biotechnol 2:54PubMedPubMedCentralCrossRef Galbusera F, Freutel M, Dürselen L, D’Aiuto M, Croce D, Villa T, Sansone V, Innocenti B (2014) Material models and properties in the finite element analysis of knee ligaments: a literature review. Front Bioeng Biotechnol 2:54PubMedPubMedCentralCrossRef
15.
go back to reference Gardiner JC, Weiss JA (2003) Subject-specific finite element analysis of the human medial collateral ligament during valgus knee loading. J Orthop Res 21(6):1098–1106PubMedCrossRef Gardiner JC, Weiss JA (2003) Subject-specific finite element analysis of the human medial collateral ligament during valgus knee loading. J Orthop Res 21(6):1098–1106PubMedCrossRef
16.
go back to reference Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ (2002) Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech 35(2):267–275PubMedCrossRef Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ (2002) Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech 35(2):267–275PubMedCrossRef
17.
go back to reference Hopley CD, Crossett LS, Chen AF (2013) Long-term clinical outcomes and survivorship after total knee arthroplasty using a rotating platform knee prosthesis: a meta-analysis. J Arthroplasty 28(1):68–77PubMedCrossRef Hopley CD, Crossett LS, Chen AF (2013) Long-term clinical outcomes and survivorship after total knee arthroplasty using a rotating platform knee prosthesis: a meta-analysis. J Arthroplasty 28(1):68–77PubMedCrossRef
18.
go back to reference Innocenti B, Truyens E, Labey L, Wong P, Victor J, Bellemans J (2009) Can medio-lateral baseplate position and load sharing induce asymptomatic local bone resorption of the proximal tibia? A finite element study. J Orthop Surg Res 4:26PubMedPubMedCentralCrossRef Innocenti B, Truyens E, Labey L, Wong P, Victor J, Bellemans J (2009) Can medio-lateral baseplate position and load sharing induce asymptomatic local bone resorption of the proximal tibia? A finite element study. J Orthop Surg Res 4:26PubMedPubMedCentralCrossRef
19.
go back to reference Innocenti B, Pianigiani S, Labey L, Victor J, Bellemans J (2011) Contact forces in several TKA designs during squatting: a numerical sensitivity analysis. J Biomech 44(8):1573–1581PubMedCrossRef Innocenti B, Pianigiani S, Labey L, Victor J, Bellemans J (2011) Contact forces in several TKA designs during squatting: a numerical sensitivity analysis. J Biomech 44(8):1573–1581PubMedCrossRef
20.
go back to reference Innocenti B, Bilgen OF, Labey L, van Lenthe GH, Sloten JV, Catani F (2014) Load sharing and ligament strains in balanced, overstuffed and understuffed UKA. A validated finite element analysis. J Arthroplasty 29(7):1491–1498PubMedCrossRef Innocenti B, Bilgen OF, Labey L, van Lenthe GH, Sloten JV, Catani F (2014) Load sharing and ligament strains in balanced, overstuffed and understuffed UKA. A validated finite element analysis. J Arthroplasty 29(7):1491–1498PubMedCrossRef
21.
go back to reference Innocenti B, Bellemans J, Catani F (2016) Deviations from optimal alignment in TKA: is there a biomechanical difference between femoral or tibial component alignment? J Arthroplasty 31(1):295–301PubMedCrossRef Innocenti B, Bellemans J, Catani F (2016) Deviations from optimal alignment in TKA: is there a biomechanical difference between femoral or tibial component alignment? J Arthroplasty 31(1):295–301PubMedCrossRef
22.
go back to reference Innocenti B, Robledo Yagüe H, Alario Bernabé R, Pianigiani S (2015) Investigation on the effects induced by TKA features on tibio-femoral mechanics part I: femoral component designs. J Mech Med Biol 15(2):1540034CrossRef Innocenti B, Robledo Yagüe H, Alario Bernabé R, Pianigiani S (2015) Investigation on the effects induced by TKA features on tibio-femoral mechanics part I: femoral component designs. J Mech Med Biol 15(2):1540034CrossRef
23.
go back to reference Innocenti B, Salandra P, Pascale W, Pianigiani S (2016) How accurate and reproducible are the identification of cruciate and collateral ligament insertions using MRI? Knee 23(4):575–581PubMedCrossRef Innocenti B, Salandra P, Pascale W, Pianigiani S (2016) How accurate and reproducible are the identification of cruciate and collateral ligament insertions using MRI? Knee 23(4):575–581PubMedCrossRef
24.
go back to reference Innocenti B, Pianigiani S, Ramundo G, Thienpont E (2016) Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplasty 31(12):2685–2691PubMedCrossRef Innocenti B, Pianigiani S, Ramundo G, Thienpont E (2016) Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplasty 31(12):2685–2691PubMedCrossRef
25.
go back to reference Kayabasi O, Ekici B (2007) The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method. Mater Des 28(8):2269–2277CrossRef Kayabasi O, Ekici B (2007) The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method. Mater Des 28(8):2269–2277CrossRef
26.
go back to reference Kurtz SM, Ong KL, Lau E, Widmer M, Maravic M, Gómez-Barrena E, de Pina Mde F, Manno V, Torre M, Walter WL, de Steiger R, Geesink RG, Peltola M, Röder C (2011) International survey of primary and revision total knee replacement. Int Orthop 35(12):1783–1789PubMedPubMedCentralCrossRef Kurtz SM, Ong KL, Lau E, Widmer M, Maravic M, Gómez-Barrena E, de Pina Mde F, Manno V, Torre M, Walter WL, de Steiger R, Geesink RG, Peltola M, Röder C (2011) International survey of primary and revision total knee replacement. Int Orthop 35(12):1783–1789PubMedPubMedCentralCrossRef
27.
go back to reference Kwak JY, Jeong JH, Lee SH, Jung HJ, Jung YB (2012) Comparison of the Clinical outcomes after total knee arthroplasty with the LCS rotating platform mobile bearing knee system and the PFC sigma RP-F mobile bearing knee system. Clin Orthop Surg 4(4):256–262PubMedPubMedCentralCrossRef Kwak JY, Jeong JH, Lee SH, Jung HJ, Jung YB (2012) Comparison of the Clinical outcomes after total knee arthroplasty with the LCS rotating platform mobile bearing knee system and the PFC sigma RP-F mobile bearing knee system. Clin Orthop Surg 4(4):256–262PubMedPubMedCentralCrossRef
28.
go back to reference Leppin J, O’Sullivan P, Winston K (2016) Effect size—large medium and small. Perspect Med Educ 5(6):347–349CrossRef Leppin J, O’Sullivan P, Winston K (2016) Effect size—large medium and small. Perspect Med Educ 5(6):347–349CrossRef
29.
go back to reference Li YL, Wu Q, Ning GZ, Feng SQ, Wu QL, Li Y, Hao Y (2014) No difference in clinical outcome between fixed- and mobile-bearing TKA: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 22(3):565–575PubMedCrossRef Li YL, Wu Q, Ning GZ, Feng SQ, Wu QL, Li Y, Hao Y (2014) No difference in clinical outcome between fixed- and mobile-bearing TKA: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 22(3):565–575PubMedCrossRef
30.
go back to reference Luyckx T, Didden K, Vandenneucker H, Labey L, Innocenti B, Bellemans J (2009) Is there a biomechanical explanation for anterior knee pain in patients with patella alta? Influence of patellar height on patellofemoral contact force, contact area and contact pressure. J Bone Jt Surg (Br) 91(3):344–350CrossRef Luyckx T, Didden K, Vandenneucker H, Labey L, Innocenti B, Bellemans J (2009) Is there a biomechanical explanation for anterior knee pain in patients with patella alta? Influence of patellar height on patellofemoral contact force, contact area and contact pressure. J Bone Jt Surg (Br) 91(3):344–350CrossRef
31.
go back to reference McNamara BP, Cristofolini L, Toni A, Taylor D (1997) Relationship between bone-prosthesis bonding and load transfer in total hip reconstruction. J Biomech 30(6):621–630PubMedCrossRef McNamara BP, Cristofolini L, Toni A, Taylor D (1997) Relationship between bone-prosthesis bonding and load transfer in total hip reconstruction. J Biomech 30(6):621–630PubMedCrossRef
32.
go back to reference Oh KJ, Pandher DS, Lee SH, Sung Joon SD Jr, Lee ST (2009) Meta-analysis comparing outcomes of fixed-bearing and mobile-bearing prostheses in total knee arthroplasty. J Arthroplasty 24(6):873–884PubMedCrossRef Oh KJ, Pandher DS, Lee SH, Sung Joon SD Jr, Lee ST (2009) Meta-analysis comparing outcomes of fixed-bearing and mobile-bearing prostheses in total knee arthroplasty. J Arthroplasty 24(6):873–884PubMedCrossRef
33.
go back to reference Peña E, Calvo B, Martinez MA, Doblaré M (2006) A three-dimensional finite element analysis of the combined behavior of ligaments and menisci in the healthy human knee joint. J Biomech 39(9):1686–1701PubMedCrossRef Peña E, Calvo B, Martinez MA, Doblaré M (2006) A three-dimensional finite element analysis of the combined behavior of ligaments and menisci in the healthy human knee joint. J Biomech 39(9):1686–1701PubMedCrossRef
34.
go back to reference Pianigiani S, Chevalier Y, Labey L, Pascale V, Innocenti B (2012) Tibio-femoral kinematics in different total knee arthroplasty designs during a loaded squat: a numerical sensitivity study. J Biomech 45(13):2315–2323PubMedCrossRef Pianigiani S, Chevalier Y, Labey L, Pascale V, Innocenti B (2012) Tibio-femoral kinematics in different total knee arthroplasty designs during a loaded squat: a numerical sensitivity study. J Biomech 45(13):2315–2323PubMedCrossRef
35.
go back to reference Pianigiani S, Alario Bernabé R, Robledo Yagüe H, Innocenti B (2015) Investigation on the effects induced by TKA features on tibio-femoral mechanics part II: tibial insert designs. J Mech Med Biol 15(2):1540035CrossRef Pianigiani S, Alario Bernabé R, Robledo Yagüe H, Innocenti B (2015) Investigation on the effects induced by TKA features on tibio-femoral mechanics part II: tibial insert designs. J Mech Med Biol 15(2):1540035CrossRef
36.
go back to reference Pianigiani S, Vander Sloten J, Pascale W, Labey L, Innocenti B (2015) A new graphical method to display data sets representing biomechanical knee behaviour. J Exp Orthop 2(1):18PubMedPubMedCentralCrossRef Pianigiani S, Vander Sloten J, Pascale W, Labey L, Innocenti B (2015) A new graphical method to display data sets representing biomechanical knee behaviour. J Exp Orthop 2(1):18PubMedPubMedCentralCrossRef
37.
go back to reference Pianigiani S, Labey L, Pascale W, Innocenti B (2016) Knee kinetics and kinematics: what are the effects of TKA malconfigurations? Knee Surg Sports Traumatol Arthrosc 24(8):2415–2421PubMedCrossRef Pianigiani S, Labey L, Pascale W, Innocenti B (2016) Knee kinetics and kinematics: what are the effects of TKA malconfigurations? Knee Surg Sports Traumatol Arthrosc 24(8):2415–2421PubMedCrossRef
38.
go back to reference Ramaniraka NA, Terrier A, Theumann N, Siegrist O (2005) Effects of the posterior cruciate ligament reconstruction on the biomechanics of the knee joint: a finite element analysis. Clin Biomech 20(4):434–442CrossRef Ramaniraka NA, Terrier A, Theumann N, Siegrist O (2005) Effects of the posterior cruciate ligament reconstruction on the biomechanics of the knee joint: a finite element analysis. Clin Biomech 20(4):434–442CrossRef
39.
go back to reference Raut VV, Stone MH, Wroblewski BM (1993) Reduction of postoperative blood loss after press-fit condylar knee arthroplasty with use of a femoral intramedullary plug. J Bone Jt Surg Am 75(9):1356–1357CrossRef Raut VV, Stone MH, Wroblewski BM (1993) Reduction of postoperative blood loss after press-fit condylar knee arthroplasty with use of a femoral intramedullary plug. J Bone Jt Surg Am 75(9):1356–1357CrossRef
40.
go back to reference Rho YJ, Kuhn-Spearing L, Zioupos P (1998) Mechanical properties and the hierarchical structure of bone. Med Eng Phys 20(2):92–102PubMedCrossRef Rho YJ, Kuhn-Spearing L, Zioupos P (1998) Mechanical properties and the hierarchical structure of bone. Med Eng Phys 20(2):92–102PubMedCrossRef
41.
go back to reference Sarathi Kopparti P, Lewis G (2007) Influence of three variables on the stresses in a three-dimensional model of a proximal tibia-total knee implant construct. Biomed Mater Eng 17(1):19–28PubMed Sarathi Kopparti P, Lewis G (2007) Influence of three variables on the stresses in a three-dimensional model of a proximal tibia-total knee implant construct. Biomed Mater Eng 17(1):19–28PubMed
42.
go back to reference Scuderi GR, Bourne RB, Noble PC, Benjamin JB, Lonner JH, Scott WN (2012) The new knee society knee scoring system. Clin Orthop Relat Res 470(1):3–19PubMedCrossRef Scuderi GR, Bourne RB, Noble PC, Benjamin JB, Lonner JH, Scott WN (2012) The new knee society knee scoring system. Clin Orthop Relat Res 470(1):3–19PubMedCrossRef
43.
go back to reference Smith TO, Ejtehadi F, Nichols R, Davies L, Donell ST, Hing CB (2010) Clinical and radiological outcomes of fixed-versus mobile-bearing total knee replacement: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 18(3):325–340PubMedCrossRef Smith TO, Ejtehadi F, Nichols R, Davies L, Donell ST, Hing CB (2010) Clinical and radiological outcomes of fixed-versus mobile-bearing total knee replacement: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 18(3):325–340PubMedCrossRef
44.
go back to reference Smith H, Jan M, Mahomed NN, Davey JR, Gandhi R (2011) Meta-analysis and systematic review of clinical outcomes comparing mobile bearing and fixed bearing total knee arthroplasty. J Arthroplasty 26(8):1205–1213PubMedCrossRef Smith H, Jan M, Mahomed NN, Davey JR, Gandhi R (2011) Meta-analysis and systematic review of clinical outcomes comparing mobile bearing and fixed bearing total knee arthroplasty. J Arthroplasty 26(8):1205–1213PubMedCrossRef
45.
go back to reference Sobieraj MC, Rimnac CM (2009) Ultra high molecular weight polyethylene: mechanics, morphology, and clinical behavior. J Mech Behav Biomed Mater 2(5):433–443PubMedCrossRef Sobieraj MC, Rimnac CM (2009) Ultra high molecular weight polyethylene: mechanics, morphology, and clinical behavior. J Mech Behav Biomed Mater 2(5):433–443PubMedCrossRef
46.
go back to reference Soenen M, Baracchi M, De Corte R, Labey L, Innocenti B (2013) Stemmed TKA in a femur with a total hip arthroplasty: is there a safe distance between the stem tips? J Arthroplasty 28(8):1437–1445PubMedCrossRef Soenen M, Baracchi M, De Corte R, Labey L, Innocenti B (2013) Stemmed TKA in a femur with a total hip arthroplasty: is there a safe distance between the stem tips? J Arthroplasty 28(8):1437–1445PubMedCrossRef
47.
go back to reference Verdini F, Zara C, Leo T, Mengarelli A, Cardarelli S, Innocenti B (2017) Assessment of patient functional performance in different knee arthroplasty designs during unconstrained squat. Muscles Ligaments Tendons J 7(3):514–523PubMedCrossRef Verdini F, Zara C, Leo T, Mengarelli A, Cardarelli S, Innocenti B (2017) Assessment of patient functional performance in different knee arthroplasty designs during unconstrained squat. Muscles Ligaments Tendons J 7(3):514–523PubMedCrossRef
48.
go back to reference Viceconti M, Casali M, Massari B, Cristofolini L, Bassini S, Toni A (1996) The ‘standardized femur program’. Proposal for a reference geometry to be used for the creation of finite element models of the femur. J Biomech 29(9):1241PubMedCrossRef Viceconti M, Casali M, Massari B, Cristofolini L, Bassini S, Toni A (1996) The ‘standardized femur program’. Proposal for a reference geometry to be used for the creation of finite element models of the femur. J Biomech 29(9):1241PubMedCrossRef
49.
go back to reference Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16(5):358–365PubMedCrossRef Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16(5):358–365PubMedCrossRef
50.
go back to reference Victor J, Labey L, Wong P, Innocenti B, Bellemans J (2010) The Influence of muscle load on tibio-femoral knee kinematics. J Orthop Res 28(4):419–428PubMed Victor J, Labey L, Wong P, Innocenti B, Bellemans J (2010) The Influence of muscle load on tibio-femoral knee kinematics. J Orthop Res 28(4):419–428PubMed
51.
go back to reference Wen Y, Liu D, Huang Y, Li B (2011) A meta-analysis of the fixed-bearing and mobile-bearing prostheses in total knee arthroplasty. Arch Orthop Trauma Surg 131(10):1341–1350PubMedCrossRef Wen Y, Liu D, Huang Y, Li B (2011) A meta-analysis of the fixed-bearing and mobile-bearing prostheses in total knee arthroplasty. Arch Orthop Trauma Surg 131(10):1341–1350PubMedCrossRef
52.
go back to reference Yau WP, Ng TP, Chiu KY (2001) Unusual complication associated with femoral intramedullary alignment guide in total knee arthroplasty. J Arthroplasty 16(2):247–249PubMedCrossRef Yau WP, Ng TP, Chiu KY (2001) Unusual complication associated with femoral intramedullary alignment guide in total knee arthroplasty. J Arthroplasty 16(2):247–249PubMedCrossRef
Metadata
Title
Asymmetric polyethylene inserts promote favorable kinematics and better clinical outcome compared to symmetric inserts in a mobile bearing total knee arthroplasty
Authors
Gianluca Castellarin
Silvia Pianigiani
Bernardo Innocenti
Publication date
01-04-2019
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 4/2019
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-018-5207-9

Other articles of this Issue 4/2019

Knee Surgery, Sports Traumatology, Arthroscopy 4/2019 Go to the issue