01-09-2005 | Physics

Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CT

Authors: Sadi Gündoğdu, Kamran Mahmutyazıcıoğlu, Hüseyin Özdemir, Ahmet Savranlar, Kıyasettin Asil

Published in: European Radiology | Issue 9/2005

Login to get access

Abstract

The purpose of this study was to analyze the effect of various tube current settings (mAs) and optimize the image quality and dose for adult cranial CT protocol. Sixty adult patients who underwent a cranial CT scanning for different indications were subdivided into three subgroups. Subjective image and noise quality scores and quantitative noise measurements were selectively studied on three reference levels (cerebellar, basal ganglia and centrum semiovale levels). For each subgroup, only one level was studied. Head circumference (HC) and the maximum anteroposterior diameter (MAPD) of each patient were measured. At 50% decreased dose protocol, there was no poor quality score at any level. At nearly 60% decreased dose protocol, the incidence of poor quality scores was much higher at the cerebellar level than at the other two levels. For the same protocol number, quantitative noise measurements were higher at the cerebellar level than the other two supratentorial levels. The correlation was found to be significant between HC, MAPD and quantitative noise measurements, and there was a non-significant correlation between HC and subjective noise scores. In adult cranial CT, depending on the level, a dose reduction of up to 60% may be possible while maintaining image quality.
Literature
1.
go back to reference van Unnik J, Broerse JJ, Geleijns J, Jansen JT, Zoetelief J, Zweers D (1997) Survey of CT techniques and absorbed dose in various Dutch hospitals. Br J Radiol 70:367–371PubMed van Unnik J, Broerse JJ, Geleijns J, Jansen JT, Zoetelief J, Zweers D (1997) Survey of CT techniques and absorbed dose in various Dutch hospitals. Br J Radiol 70:367–371PubMed
2.
go back to reference Wade JP, Weyman JC, Goldstone KE (1997) CT standard protocols are of limited value in assessing actual patient dose. Br J Radiol 70:1146–1151PubMed Wade JP, Weyman JC, Goldstone KE (1997) CT standard protocols are of limited value in assessing actual patient dose. Br J Radiol 70:1146–1151PubMed
3.
go back to reference Wong ETH, Yu SK, Lai M, Wong YC, Lau PC (2000) MAPD—an objective way to select mAs for paediatric brain CT. Br J Radiol 74:932–937 Wong ETH, Yu SK, Lai M, Wong YC, Lau PC (2000) MAPD—an objective way to select mAs for paediatric brain CT. Br J Radiol 74:932–937
4.
go back to reference Karabulut N, Toru M, Gelebek V, Gulsun M, Ariyurek OM (2002) Comparison of low-dose and standard-dose helical CT in the evaluation of pulmonary nodules. Eur Radiol 12:2764–2769PubMed Karabulut N, Toru M, Gelebek V, Gulsun M, Ariyurek OM (2002) Comparison of low-dose and standard-dose helical CT in the evaluation of pulmonary nodules. Eur Radiol 12:2764–2769PubMed
5.
go back to reference Wilting JE, Zwartkruis A, van Leeuwen MS et al (2001) A rational approach to dose reduction in CT: individualized scan protocols. Eur Radiol 11:2627–2632CrossRefPubMed Wilting JE, Zwartkruis A, van Leeuwen MS et al (2001) A rational approach to dose reduction in CT: individualized scan protocols. Eur Radiol 11:2627–2632CrossRefPubMed
6.
go back to reference Verdun FR, Lepori D, Monnin P, Valley JF, Schnyder P, Gudinchet F (2004) Management of patient dose and image noise in routine pediatric CT abdominal examinations. Eur Radiol 14:835–8411CrossRefPubMed Verdun FR, Lepori D, Monnin P, Valley JF, Schnyder P, Gudinchet F (2004) Management of patient dose and image noise in routine pediatric CT abdominal examinations. Eur Radiol 14:835–8411CrossRefPubMed
7.
go back to reference Kalra MK, Prasad S, Saini S et al (2002) Clinical comparison of standard-dose and 50% reduced-dose abdominal CT: effect on image quality. Am J Roentgenol 179:1101–1106 Kalra MK, Prasad S, Saini S et al (2002) Clinical comparison of standard-dose and 50% reduced-dose abdominal CT: effect on image quality. Am J Roentgenol 179:1101–1106
8.
go back to reference Cohnen M, Fischer H, Hamacher J et al (2000) CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. Am J Neuroradiol 21:1654–1660PubMed Cohnen M, Fischer H, Hamacher J et al (2000) CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. Am J Neuroradiol 21:1654–1660PubMed
9.
go back to reference Chan CY, Wong YC, Chau LF, Yu SK, Lau PC (1999) Radiation dose reduction in paediatric cranial CT. Pediatr Radiol 29:770–775CrossRefPubMed Chan CY, Wong YC, Chau LF, Yu SK, Lau PC (1999) Radiation dose reduction in paediatric cranial CT. Pediatr Radiol 29:770–775CrossRefPubMed
11.
go back to reference Takahashi M, Maguire WM, Ashtari M et al (1998) Low-dose spiral computed tomography of the thorax: comparison with the standard-dose technique. Invest Radiol 33:68–73CrossRefPubMed Takahashi M, Maguire WM, Ashtari M et al (1998) Low-dose spiral computed tomography of the thorax: comparison with the standard-dose technique. Invest Radiol 33:68–73CrossRefPubMed
12.
go back to reference Marmolya G, Wiesen EJ, Yagan R, Haria CD, Shah AC (1991) Paranasal sinuses: low-dose CT. Radiology 181:689–691PubMed Marmolya G, Wiesen EJ, Yagan R, Haria CD, Shah AC (1991) Paranasal sinuses: low-dose CT. Radiology 181:689–691PubMed
13.
go back to reference Kamel RI, Hemandez JR, Martin EJ et al (1994) Radiation dose reduction in CT of pediatric pelvis. Radiology 190:683–687PubMed Kamel RI, Hemandez JR, Martin EJ et al (1994) Radiation dose reduction in CT of pediatric pelvis. Radiology 190:683–687PubMed
14.
go back to reference Zeman RK, Fox SH, Silverman PM, Davros WJ, Carter LM, Griego D (1993) Helical (spiral) CT of the abdomen. Am J Roentgenol 160:719–725 Zeman RK, Fox SH, Silverman PM, Davros WJ, Carter LM, Griego D (1993) Helical (spiral) CT of the abdomen. Am J Roentgenol 160:719–725
Metadata
Title
Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CT
Authors
Sadi Gündoğdu
Kamran Mahmutyazıcıoğlu
Hüseyin Özdemir
Ahmet Savranlar
Kıyasettin Asil
Publication date
01-09-2005
Publisher
Springer-Verlag
Published in
European Radiology / Issue 9/2005
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-004-2550-7

Other articles of this Issue 9/2005

European Radiology 9/2005 Go to the issue