Published in: La radiologia medica 8/2014

01-08-2014 | Radiobiology And Safety

Assessment of dose exposure and image quality in coronary angiography performed by 640-slice CT: a comparison between adaptive iterative and filtered back-projection algorithm by propensity analysis

Authors: Ernesto Di Cesare, Antonio Gennarelli, Alessandra Di Sibio, Valentina Felli, Alessandra Splendiani, Giovanni Luca Gravina, Antonio Barile, Carlo Masciocchi

Published in: La radiologia medica | Issue 8/2014

Login to get access

Abstract

Purpose

This study was performed to confirm, by propensity score matching, whether the use of adaptive–iterative dose reduction (AIDR 3D) with a built-in automatic exposure control system provides clinical and dosimetric advantages with respect to the traditional filtered back-projection (FBP) algorithm without automatic exposure modulation.

Materials and methods

A total of 200 consecutive patients undergoing coronary computed tomography (CT) angiography on a 640-slice CT scanner were studied. A protocol with exposure parameters based on patient body mass index (BMI) and with images reconstructed using FBP (group A) was compared with a protocol with images acquired using tube current decided by an automatic exposure control system and reconstructed using AIDR (group B). Mean effective dose and image quality with both objective and subjective measurements were assessed.

Results

Mean effective dose was 23.6 % lower in group B than in group A (2.56 versus 3.34 mSv; p < 0.0001). Noise was significantly lower in group B with consequent higher signal-to-noise (SNR) and contrast-to-noise (CNR) (p < 0.0001) compared with group A. Subjective quality parameters were also significantly higher in group B.

Conclusions

Comparative analysis by propensity score matching confirms that AIDR 3D with automatic exposure control is able to reduce significantly the mean radiation dose and improve the image quality compared with traditional FBP without exposure modulation.
Literature
1.
go back to reference Lloyd-Jones D, Adams RJ, Brown TM et al (2010) Executive summary: heart disease and stroke statistics—2010 update: a report from the American Heart Association. Circulation 121:948–954PubMedCrossRef Lloyd-Jones D, Adams RJ, Brown TM et al (2010) Executive summary: heart disease and stroke statistics—2010 update: a report from the American Heart Association. Circulation 121:948–954PubMedCrossRef
2.
go back to reference Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323PubMedCrossRef Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323PubMedCrossRef
3.
go back to reference Hamon M, Biondi-Zoccai GG, Malagutti P et al (2006) Diagnostic performance of multislice spiral computed tomography of coronary arteries as compared with conventional invasive coronary angiography: a meta-analysis. J Am Coll Cardiol 48:1896–1910PubMedCrossRef Hamon M, Biondi-Zoccai GG, Malagutti P et al (2006) Diagnostic performance of multislice spiral computed tomography of coronary arteries as compared with conventional invasive coronary angiography: a meta-analysis. J Am Coll Cardiol 48:1896–1910PubMedCrossRef
4.
go back to reference Di Cesare E, Carbone I, Carriero A et al (2012) Clinical indications for cardiac computed tomography. From the Working Group of the Cardiac Radiology Section of the Italian Society of Medical Radiology (SIRM). Radiol Med 117:901–938PubMedCrossRef Di Cesare E, Carbone I, Carriero A et al (2012) Clinical indications for cardiac computed tomography. From the Working Group of the Cardiac Radiology Section of the Italian Society of Medical Radiology (SIRM). Radiol Med 117:901–938PubMedCrossRef
5.
go back to reference Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507PubMedCrossRef Hausleiter J, Meyer T, Hermann F et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507PubMedCrossRef
6.
go back to reference Labounty TM, Earls JP, Leipsic J et al (2010) Effect of a standardized quality-improvement protocol onradiation dose in coronary computed tomographic angiography. Am J Cardiol 106:1663–1667PubMedCrossRef Labounty TM, Earls JP, Leipsic J et al (2010) Effect of a standardized quality-improvement protocol onradiation dose in coronary computed tomographic angiography. Am J Cardiol 106:1663–1667PubMedCrossRef
7.
go back to reference Tomizawa N, Nojo T, Akahane M et al (2012) Adaptive iterative dose reduction in coronary CT angiography using 320-row CT: assessment of radiation dose reduction and image quality. J Cardiovasc Comput Tomogr 6:318–324PubMedCrossRef Tomizawa N, Nojo T, Akahane M et al (2012) Adaptive iterative dose reduction in coronary CT angiography using 320-row CT: assessment of radiation dose reduction and image quality. J Cardiovasc Comput Tomogr 6:318–324PubMedCrossRef
8.
go back to reference Yoo RE, Park EA, Lee W et al (2013) Image quality of adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection. Int J Cardiovasc Imaging 29:669–676PubMedCrossRef Yoo RE, Park EA, Lee W et al (2013) Image quality of adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection. Int J Cardiovasc Imaging 29:669–676PubMedCrossRef
9.
go back to reference Williams MC, Weir NW, Mirsadraee S et al (2013) Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography. Clin Radiol 68:e570–e577PubMedCentralPubMedCrossRef Williams MC, Weir NW, Mirsadraee S et al (2013) Iterative reconstruction and individualized automatic tube current selection reduce radiation dose while maintaining image quality in 320-multidetector computed tomography coronary angiography. Clin Radiol 68:e570–e577PubMedCentralPubMedCrossRef
10.
go back to reference Chun EJ, Lee W, Choi YH et al (2008) Effects of nitroglycerin on the diagnostic accuracy of electrocardiogram-gated coronary computed tomography angiography. J Comput Assist Tomogr 32:86–92PubMedCrossRef Chun EJ, Lee W, Choi YH et al (2008) Effects of nitroglycerin on the diagnostic accuracy of electrocardiogram-gated coronary computed tomography angiography. J Comput Assist Tomogr 32:86–92PubMedCrossRef
11.
go back to reference Austen WG, Edwards JE, Roe BB et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMedCrossRef Austen WG, Edwards JE, Roe BB et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMedCrossRef
12.
go back to reference Leipsic J, Labounty TM, Heilbron B et al (2010) Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol 195:649–654PubMedCrossRef Leipsic J, Labounty TM, Heilbron B et al (2010) Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol 195:649–654PubMedCrossRef
13.
go back to reference Park EA, Lee W, Kim KW et al (2011) Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose. Int J Cardiovasc Imaging 28:1775–1786PubMedCrossRef Park EA, Lee W, Kim KW et al (2011) Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose. Int J Cardiovasc Imaging 28:1775–1786PubMedCrossRef
14.
go back to reference Bittencourt MS, Schmidt B, Seltmann M et al (2011) Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience. Int J Cardiovasc Imaging 27:1081–1087PubMedCrossRef Bittencourt MS, Schmidt B, Seltmann M et al (2011) Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience. Int J Cardiovasc Imaging 27:1081–1087PubMedCrossRef
15.
go back to reference Hara AK, Paden RG, Silva AC et al (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef Hara AK, Paden RG, Silva AC et al (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef
16.
go back to reference Bongartz G, Golding SJ, Jurik AG et al (2004) European Commission Website. European guidelines for multislice computed tomography: appendix C. Funded by the European Commission, March 2004. Contract No. FIGM-CT2000-20078-CT TIP Bongartz G, Golding SJ, Jurik AG et al (2004) European Commission Website. European guidelines for multislice computed tomography: appendix C. Funded by the European Commission, March 2004. Contract No. FIGM-CT2000-20078-CT TIP
17.
go back to reference Rubin DB (1997) Estimating causal effects from large data sets using propensity scores. Ann Intern Med 127(8 Pt 2):757–763PubMedCrossRef Rubin DB (1997) Estimating causal effects from large data sets using propensity scores. Ann Intern Med 127(8 Pt 2):757–763PubMedCrossRef
18.
go back to reference Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46CrossRef
19.
go back to reference Hoe J, Toh KH (2009) First experience with 320-row multidetector CT coronary angiography scanning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr 3:257–261PubMedCrossRef Hoe J, Toh KH (2009) First experience with 320-row multidetector CT coronary angiography scanning with prospective electrocardiogram gating to reduce radiation dose. J Cardiovasc Comput Tomogr 3:257–261PubMedCrossRef
20.
go back to reference Dewey M, Zimmermann E, Hamm B et al (2009) Non invasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to-head pilot investigation. Circulation 120:867–875PubMedCrossRef Dewey M, Zimmermann E, Hamm B et al (2009) Non invasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to-head pilot investigation. Circulation 120:867–875PubMedCrossRef
21.
go back to reference Lee CH, Goo JM, Ye HJ et al (2008) Radiation dose modulation techniques in the multidetector CT era: from basics to practice. Radiographics 28:1451–1459PubMedCrossRef Lee CH, Goo JM, Ye HJ et al (2008) Radiation dose modulation techniques in the multidetector CT era: from basics to practice. Radiographics 28:1451–1459PubMedCrossRef
22.
go back to reference Chen MY, Steigner ML, Leung SW et al (2013) Simulated 50% radiation dose reduction in coronary CT angiography using adaptative iterative dose reduction in three-dimensions. Int J Cardiovasc Imaging 29:1167–1175PubMedCentralPubMedCrossRef Chen MY, Steigner ML, Leung SW et al (2013) Simulated 50% radiation dose reduction in coronary CT angiography using adaptative iterative dose reduction in three-dimensions. Int J Cardiovasc Imaging 29:1167–1175PubMedCentralPubMedCrossRef
23.
go back to reference Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptative statistical iterative reconstruction algorithm. AJR 194:191–199PubMedCrossRef Silva AC, Lawder HJ, Hara A et al (2010) Innovations in CT dose reduction strategy: application of the adaptative statistical iterative reconstruction algorithm. AJR 194:191–199PubMedCrossRef
24.
go back to reference Cademartiri F, La Grutta L, Runza G et al (2007) Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography. Eur Radiol 17:1842–1849PubMedCentralPubMedCrossRef Cademartiri F, La Grutta L, Runza G et al (2007) Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography. Eur Radiol 17:1842–1849PubMedCentralPubMedCrossRef
25.
go back to reference La Grutta L, Galia M, Gentile G et al (2013) Comparison of iodinated contrast media for the assessment of atherosclerotic plaque attenuation values by CT coronary angiography: observations in an ex vivo model. Br J Radiol 86:20120238PubMedCrossRef La Grutta L, Galia M, Gentile G et al (2013) Comparison of iodinated contrast media for the assessment of atherosclerotic plaque attenuation values by CT coronary angiography: observations in an ex vivo model. Br J Radiol 86:20120238PubMedCrossRef
Metadata
Title
Assessment of dose exposure and image quality in coronary angiography performed by 640-slice CT: a comparison between adaptive iterative and filtered back-projection algorithm by propensity analysis
Authors
Ernesto Di Cesare
Antonio Gennarelli
Alessandra Di Sibio
Valentina Felli
Alessandra Splendiani
Giovanni Luca Gravina
Antonio Barile
Carlo Masciocchi
Publication date
01-08-2014
Publisher
Springer Milan
Published in
La radiologia medica / Issue 8/2014
Print ISSN: 0033-8362
Electronic ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-014-0382-3

Other articles of this Issue 8/2014

La radiologia medica 8/2014 Go to the issue