Published in: BMC Anesthesiology 1/2018

Open Access 01-12-2018 | Research article

Prediction of fluid responsiveness in mechanically ventilated cardiac surgical patients: the performance of seven different functional hemodynamic parameters

Authors: Michael T. Ganter, Martin Geisen, Sonja Hartnack, Omer Dzemali, Christoph K. Hofer

Published in: BMC Anesthesiology | Issue 1/2018

Login to get access

Abstract

Background

Functional hemodynamic parameters such as stroke volume and pulse pressure variation (SVV and PPV) have been shown to be reliable predictors of fluid responsiveness in mechanically ventilated patients. Today, different minimally- and non-invasive hemodynamic monitoring systems measure functional hemodynamic parameters. Although some of these parameters are described by the same name, they differ in their measurement technique and thus may provide different results. We aimed to test the performance of seven functional hemodynamic parameters simultaneously in the same clinical setting.

Methods

Hemodynamic measurements were done in 30 cardiac surgery patients that were mechanically ventilated. Before and after a standardized intravenous fluid bolus, hemodynamics were measured by the following monitoring systems: PiCCOplus (SVVPiCCO, PPVPiCCO), LiDCOrapid (SVVLiDCO, PPVLiDCO), FloTrac (SVVFloTrac), Philips Intellivue (PPVPhilips) and Masimo pulse oximeter (pleth variability index, PVI). Prediction of fluid responsiveness was tested by calculation of receiver operating characteristic (ROC) curves including a gray zone approach and compared using Fisher’s Z-Test.

Results

Fluid administration resulted in an increase in cardiac output, while all functional hemodynamic parameters decreased. A wide range of areas under the ROC-curve (AUC’s) was observed: AUC-SVVPiCCO = 0.91, AUC-PPVPiCCO = 0.88, AUC-SVVLiDCO = 0.78, AUC-PPVLiDCO = 0.89, AUC-SVVFloTrac = 0.87, AUC-PPVPhilips = 0.92 and AUC-PVI = 0.68. Optimal threshold values for prediction of fluid responsiveness ranged between 9.5 and 17.5%. Lowest threshold values were observed for SVVLiDCO, highest for PVI.

Conclusion

All functional hemodynamic parameters tested except for PVI showed that their use allows a reliable identification of potential fluid responders. PVI however, may not be suitable after cardiac surgery to predict fluid responsiveness.

Trial registration

NCT02571465, registered on October 7th, 2015 (retrospectively registered).
Literature
1.
go back to reference Myles PS, Andrews S, Nicholson J, Lobo DN, Mythen M. Contemporary approaches to perioperative IV fluid therapy. World J Surg. 2017;41(10):2457–63.CrossRefPubMed Myles PS, Andrews S, Nicholson J, Lobo DN, Mythen M. Contemporary approaches to perioperative IV fluid therapy. World J Surg. 2017;41(10):2457–63.CrossRefPubMed
4.
go back to reference Min JJ, Gil NS, Lee JH, Ryu DK, Kim CS, Lee SM. Predictor of fluid responsiveness in the 'grey zone': augmented pulse pressure variation through a temporary increase in tidal volume. Br J Anaesth. 2017;119(1):50–6.CrossRefPubMed Min JJ, Gil NS, Lee JH, Ryu DK, Kim CS, Lee SM. Predictor of fluid responsiveness in the 'grey zone': augmented pulse pressure variation through a temporary increase in tidal volume. Br J Anaesth. 2017;119(1):50–6.CrossRefPubMed
5.
go back to reference Ansari BM, Zochios V, Falter F, Klein AA. Physiological controversies and methods used to determine fluid responsiveness: a qualitative systematic review. Anaesthesia. 2016;71(1):94–105.CrossRefPubMed Ansari BM, Zochios V, Falter F, Klein AA. Physiological controversies and methods used to determine fluid responsiveness: a qualitative systematic review. Anaesthesia. 2016;71(1):94–105.CrossRefPubMed
6.
go back to reference Hofer CK, Cecconi M, Marx G, della Rocca G. Minimally invasive haemodynamic monitoring. Eur J Anaesthesiol. 2009;26(12):996–1002.CrossRefPubMed Hofer CK, Cecconi M, Marx G, della Rocca G. Minimally invasive haemodynamic monitoring. Eur J Anaesthesiol. 2009;26(12):996–1002.CrossRefPubMed
8.
go back to reference Teboul JL, Saugel B, Cecconi M, De Backer D, Hofer CK, Monnet X, Perel A, Pinsky MR, Reuter DA, Rhodes A, et al. Less invasive hemodynamic monitoring in critically ill patients. Intensive Care Med. 2016;42(9):1350–9.CrossRefPubMed Teboul JL, Saugel B, Cecconi M, De Backer D, Hofer CK, Monnet X, Perel A, Pinsky MR, Reuter DA, Rhodes A, et al. Less invasive hemodynamic monitoring in critically ill patients. Intensive Care Med. 2016;42(9):1350–9.CrossRefPubMed
10.
go back to reference Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez JC, Muller M. pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinf. 2011;12:77.CrossRef Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez JC, Muller M. pROC: an open-source package for R and S+ to analyze and compare ROC curves. BMC Bioinf. 2011;12:77.CrossRef
11.
go back to reference Cannesson M, Le Manach Y, Hofer CK, Goarin JP, Lehot JJ, Vallet B, Tavernier B. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a "gray zone" approach. Anesthesiology. 2011;115(2):231–41.CrossRefPubMed Cannesson M, Le Manach Y, Hofer CK, Goarin JP, Lehot JJ, Vallet B, Tavernier B. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a "gray zone" approach. Anesthesiology. 2011;115(2):231–41.CrossRefPubMed
12.
go back to reference Vos JJ, Poterman M, Salm PP, Van Amsterdam K, Struys MM, Scheeren TW, Kalmar AF. Noninvasive pulse pressure variation and stroke volume variation to predict fluid responsiveness at multiple thresholds: a prospective observational study. Can J Anaesth. 2015;62(11):1153–60.CrossRefPubMedPubMedCentral Vos JJ, Poterman M, Salm PP, Van Amsterdam K, Struys MM, Scheeren TW, Kalmar AF. Noninvasive pulse pressure variation and stroke volume variation to predict fluid responsiveness at multiple thresholds: a prospective observational study. Can J Anaesth. 2015;62(11):1153–60.CrossRefPubMedPubMedCentral
13.
go back to reference Gogbashian A, Sedrakyan A, Treasure T. EuroSCORE: a systematic review of international performance. Eur Journal Cardiothorac Surg. 2004;25(5):695–700.CrossRef Gogbashian A, Sedrakyan A, Treasure T. EuroSCORE: a systematic review of international performance. Eur Journal Cardiothorac Surg. 2004;25(5):695–700.CrossRef
14.
go back to reference Osman D, Ridel C, Ray P, Monnet X, Anguel N, Richard C, Teboul JL. Cardiac filling pressures are not appropriate to predict hemodynamic response to volume challenge. Crit Care Med. 2007;35(1):64–8.CrossRefPubMed Osman D, Ridel C, Ray P, Monnet X, Anguel N, Richard C, Teboul JL. Cardiac filling pressures are not appropriate to predict hemodynamic response to volume challenge. Crit Care Med. 2007;35(1):64–8.CrossRefPubMed
15.
go back to reference Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37(9):2642–7.CrossRefPubMed Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37(9):2642–7.CrossRefPubMed
16.
go back to reference Cecconi M, De Backer D, Antonelli M, Beale R, Bakker J, Hofer C, Jaeschke R, Mebazaa A, Pinsky MR, Teboul JL, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014;40(12):1795–815.CrossRefPubMedPubMedCentral Cecconi M, De Backer D, Antonelli M, Beale R, Bakker J, Hofer C, Jaeschke R, Mebazaa A, Pinsky MR, Teboul JL, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014;40(12):1795–815.CrossRefPubMedPubMedCentral
17.
go back to reference Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology. 2010;113(5):1220–35.CrossRefPubMed Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology. 2010;113(5):1220–35.CrossRefPubMed
18.
go back to reference Hadian M, Kim HK, Severyn DA, Pinsky MR. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care. 2010;14(6):R212.CrossRefPubMedPubMedCentral Hadian M, Kim HK, Severyn DA, Pinsky MR. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care. 2010;14(6):R212.CrossRefPubMedPubMedCentral
19.
go back to reference Ganter MT, Alhashemi JA, Al-Shabasy AM, Schmid UM, Schott P, Shalabi SA, Badri AM, Hartnack S, Hofer CK. Continuous cardiac output measurement by un-calibrated pulse wave analysis and pulmonary artery catheter in patients with septic shock. J Clin Monit Comput. 2016;30(1):13–22.CrossRefPubMed Ganter MT, Alhashemi JA, Al-Shabasy AM, Schmid UM, Schott P, Shalabi SA, Badri AM, Hartnack S, Hofer CK. Continuous cardiac output measurement by un-calibrated pulse wave analysis and pulmonary artery catheter in patients with septic shock. J Clin Monit Comput. 2016;30(1):13–22.CrossRefPubMed
20.
go back to reference Monnet X, Picard F, Lidzborski E, Mesnil M, Duranteau J, Richard C, Teboul JL. The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge. Crit Care. 2012;16(5):R212.CrossRefPubMedPubMedCentral Monnet X, Picard F, Lidzborski E, Mesnil M, Duranteau J, Richard C, Teboul JL. The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge. Crit Care. 2012;16(5):R212.CrossRefPubMedPubMedCentral
21.
go back to reference Marik PE. Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth. 2013;27(1):121–34.CrossRefPubMed Marik PE. Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth. 2013;27(1):121–34.CrossRefPubMed
23.
go back to reference Lamia B, Kim HK, Severyn DA, Pinsky MR. Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter. J Clin Monit Comput. 2018;32(1):33–43. Lamia B, Kim HK, Severyn DA, Pinsky MR. Cross-comparisons of trending accuracies of continuous cardiac-output measurements: pulse contour analysis, bioreactance, and pulmonary-artery catheter. J Clin Monit Comput. 2018;32(1):33–43.
24.
go back to reference Sandroni C, Cavallaro F, Marano C, Falcone C, De Santis P, Antonelli M. Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis. Intensive Care Med. 2012;38(9):1429–37.CrossRefPubMed Sandroni C, Cavallaro F, Marano C, Falcone C, De Santis P, Antonelli M. Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis. Intensive Care Med. 2012;38(9):1429–37.CrossRefPubMed
25.
go back to reference Monnet X, Guerin L, Jozwiak M, Bataille A, Julien F, Richard C, Teboul JL. Pleth variability index is a weak predictor of fluid responsiveness in patients receiving norepinephrine. Br J Anaesth. 2013;110(2):207–13.CrossRefPubMed Monnet X, Guerin L, Jozwiak M, Bataille A, Julien F, Richard C, Teboul JL. Pleth variability index is a weak predictor of fluid responsiveness in patients receiving norepinephrine. Br J Anaesth. 2013;110(2):207–13.CrossRefPubMed
26.
go back to reference Biais M, Ehrmann S, Mari A, Conte B, Mahjoub Y, Desebbe O, Pottecher J, Lakhal K, Benzekri-Lefevre D, Molinari N, et al. Clinical relevance of pulse pressure variations for predicting fluid responsiveness in mechanically ventilated intensive care unit patients: the grey zone approach. Crit Care. 2014;18(6):587.CrossRefPubMedPubMedCentral Biais M, Ehrmann S, Mari A, Conte B, Mahjoub Y, Desebbe O, Pottecher J, Lakhal K, Benzekri-Lefevre D, Molinari N, et al. Clinical relevance of pulse pressure variations for predicting fluid responsiveness in mechanically ventilated intensive care unit patients: the grey zone approach. Crit Care. 2014;18(6):587.CrossRefPubMedPubMedCentral
Metadata
Title
Prediction of fluid responsiveness in mechanically ventilated cardiac surgical patients: the performance of seven different functional hemodynamic parameters
Authors
Michael T. Ganter
Martin Geisen
Sonja Hartnack
Omer Dzemali
Christoph K. Hofer
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2018
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-018-0520-x

Other articles of this Issue 1/2018

BMC Anesthesiology 1/2018 Go to the issue