Published in: Clinical Orthopaedics and Related Research® 11/2016

Open Access 01-11-2016 | Symposium: Advances in PEEK Technology

Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study

Authors: Ryan Siskey, MS, Lauren Ciccarelli, MS, Melissa K. C. Lui, MS, Steven M. Kurtz, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 11/2016

Login to get access

Abstract

Background

Most contemporary total disc replacements (TDRs) use conventional orthopaedic bearing couples such as ultrahigh-molecular-weight polyethylene (polyethylene) and cobalt-chromium (CoCr). Cervical total disc replacements incorporating polyetheretherketone (PEEK) bearings (specifically PEEK-on-PEEK bearings) have been previously investigated, but little is known about PEEK-on-ceramic bearings for TDR.

Questions/purposes

(1) What is the tribologic behavior of a PEEK-on-ceramic bearing for cervical TDR under idealized, clean wear test conditions? (2) How does the PEEK-on-ceramic design perform under impingement conditions? (3) How is the PEEK-on-ceramic bearing affected by abrasive wear? (4) Is the particle morphology from PEEK-on-ceramic bearings for TDRs affected by adverse wear scenarios?

Methods

PEEK-on-ceramic cervical TDR bearings were subjected to a 10 million cycle ideal wear test based on ASTM F2423 and ISO 181912-1 using a six-station spine wear simulator (MTS, Eden Prairie, MN, USA) with 5 g/L bovine serum concentration at 23° ± 2° C (ambient temperature). Validated 1 million cycle impingement and 5 million cycle abrasive tests were conducted on the PEEK-on-ceramic bearings based, in part, on retrieval analysis of a comparable bearing design as well as finite element analyses. The ceramic-on-PEEK couple was characterized for damage modes, mass and volume loss, and penetration and the lubricant was subjected to particle analysis. The resulting mass wear rate, volumetric wear rate, based on material density, and particle analysis were compared with clinically available cervical disc bearing couples.

Results

The three modes of wear (idealized, impingement, and abrasive) resulted in mean mass wear rates of 0.9 ± 0.2 mg/MC, 1.9 ± 0.5 mg/MC, and 2.8 ± 0.6 mg/MC, respectively. The mass wear rates were converted to volumetric wear rates using density and found to be 0.7 ± 0.1 mm3/MC, 1.5 ± 0.4 mm3/MC, and 2.1 ± 0.5 mm3/MC, respectively. During each test, the PEEK endplates were the primary sources of wear and demonstrated an abrasive wear mechanism. Under idealized and impingement conditions, the ceramic core also demonstrated slight polishing of the articulating surface but the change in mass was unmeasurable. During abrasive testing, the titanium transfer on the core was shown to polish over 5 MC of testing. In all cases and consistent with previous studies of other PEEK bearing couples, the particle size was primarily < 2 µm and morphology was smooth and spheroidal.

Conclusions

Overall, the idealized PEEK-on-ceramic wear rate (0.7 ± 0.1 mm3/MC) appears comparable to the published wear rates for other polymer-on-hard bearing couples (0.3–6.7 mm3/MC) and within the range of 0.2 to 1.9 mm3/MC reported for PEEK-on-PEEK cervical disc designs. The particles, based on size and morphology, also suggest the wear mechanism is comparable between the PEEK-on-ceramic couple and other polymer-on-ceramic orthopaedic couples.

Clinical Relevance

The PEEK-on-ceramic bearing considered in this study is a novel bearing couple for use in total disc arthroplasty devices and will require clinical evaluation to fully assess the bearing couple and total disc design. However, the wear rates under idealized and adverse conditions, and particle size and morphology, suggest that PEEK-on-ceramic bearings may be a reasonable alternative to polyethylene-on-CoCr and metal-on-metal bearings currently used in cervical TDRs.
Literature
1.
go back to reference Allain J, Le Mouel S, Goutallier D, Voisin M. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.CrossRefPubMed Allain J, Le Mouel S, Goutallier D, Voisin M. Poor eight-year survival of cemented zirconia-polyethylene total hip replacements. J Bone Joint Surg Br. 1999;81:835–842.CrossRefPubMed
2.
go back to reference Anderson PA, Rouleau JP, Bryan VE, Carlson CS. Wear analysis of the Bryan Cervical Disc prosthesis. Spine. 2003;28:S186–194.CrossRefPubMed Anderson PA, Rouleau JP, Bryan VE, Carlson CS. Wear analysis of the Bryan Cervical Disc prosthesis. Spine. 2003;28:S186–194.CrossRefPubMed
3.
go back to reference Anderson PA, Rouleau JP, Toth JM, Riew KD. A comparison of simulator-tested and -retrieved cervical disc prostheses. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine. 2004;1:202–210. Anderson PA, Rouleau JP, Toth JM, Riew KD. A comparison of simulator-tested and -retrieved cervical disc prostheses. Invited submission from the Joint Section Meeting on Disorders of the Spine and Peripheral Nerves, March 2004. J Neurosurg Spine. 2004;1:202–210.
4.
go back to reference ASTM. ASTM F1877-05(2010) Standard Practice for Characterization of Particles. West Conshohocken, PA, USA: ASTM International; 2010. ASTM. ASTM F1877-05(2010) Standard Practice for Characterization of Particles. West Conshohocken, PA, USA: ASTM International; 2010.
5.
go back to reference ASTM. ASTM F2423-11 Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses. West Conshohocken, PA, USA: ASTM International; 2011. ASTM. ASTM F2423-11 Standard Guide for Functional, Kinematic, and Wear Assessment of Total Disc Prostheses. West Conshohocken, PA, USA: ASTM International; 2011.
6.
go back to reference ASTM. ASTM F2582-08 Standard Test Method for Impingement of Acetabular Prostheses. West Conshohocken, PA, USA: ASTM International; 2012. ASTM. ASTM F2582-08 Standard Test Method for Impingement of Acetabular Prostheses. West Conshohocken, PA, USA: ASTM International; 2012.
7.
go back to reference Bao Q-B, Songer M, Pimenta L, Werner D, Reyes-Sanchez A, Balsano M, Agrillo U, Coric D, Davenport K, Yuan H. Nubac disc arthroplasty: preclinical studies and preliminary safety and efficacy evaluations. SAS J. 2007;1:36–45.CrossRefPubMedPubMedCentral Bao Q-B, Songer M, Pimenta L, Werner D, Reyes-Sanchez A, Balsano M, Agrillo U, Coric D, Davenport K, Yuan H. Nubac disc arthroplasty: preclinical studies and preliminary safety and efficacy evaluations. SAS J. 2007;1:36–45.CrossRefPubMedPubMedCentral
8.
go back to reference Bragdon CR, Jasty M, Muratoglu OK, O’Connor DO, Harris WH. Third-body wear of highly cross-linked polyethylene in a hip simulator. J Arthroplasty. 2003;18:553–61.CrossRefPubMed Bragdon CR, Jasty M, Muratoglu OK, O’Connor DO, Harris WH. Third-body wear of highly cross-linked polyethylene in a hip simulator. J Arthroplasty. 2003;18:553–61.CrossRefPubMed
9.
go back to reference Brockett C, Carbone S, Fisher J, Jennings L. PEEK and CFR-PEEK as alternative bearing materials to polyethylene. Bone Joint J. 2016;98(Suppl 2):78. Brockett C, Carbone S, Fisher J, Jennings L. PEEK and CFR-PEEK as alternative bearing materials to polyethylene. Bone Joint J. 2016;98(Suppl 2):78.
10.
go back to reference Brown T, Bao Q-B. The use of self-mating PEEK as an alternative bearing material for cervical disc arthroplasty: a comparison of different simulator inputs and tribological environments. Eur Spine J. 2012;21:717–26.CrossRefPubMedCentral Brown T, Bao Q-B. The use of self-mating PEEK as an alternative bearing material for cervical disc arthroplasty: a comparison of different simulator inputs and tribological environments. Eur Spine J. 2012;21:717–26.CrossRefPubMedCentral
11.
go back to reference Brown T, Bao Q-B, Kilpela T, Songer M. An in vitro biotribological assessment of NUBAC, a polyetheretherketone-on-polyetheretherketone articulating nucleus replacement device: methodology and results from a series of wear tests using different motion profiles, test frequencies, and environmental conditions. Spine. 2010;35:E774–E781.CrossRefPubMed Brown T, Bao Q-B, Kilpela T, Songer M. An in vitro biotribological assessment of NUBAC, a polyetheretherketone-on-polyetheretherketone articulating nucleus replacement device: methodology and results from a series of wear tests using different motion profiles, test frequencies, and environmental conditions. Spine. 2010;35:E774–E781.CrossRefPubMed
12.
go back to reference Bushelow M, Nechtow W, Hinter M, Dressel H, Kaddick C. Wear testing of a cervical total disc replacement: effect of motion and load parameters on wear rate and particle morphology. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008. Bushelow M, Nechtow W, Hinter M, Dressel H, Kaddick C. Wear testing of a cervical total disc replacement: effect of motion and load parameters on wear rate and particle morphology. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008.
13.
go back to reference Cash DJ, Khanduja V. The case for ceramic-on-polyethylene as the preferred bearing for a young adult hip replacement. Hip Int. 2014;24:421–427.CrossRefPubMed Cash DJ, Khanduja V. The case for ceramic-on-polyethylene as the preferred bearing for a young adult hip replacement. Hip Int. 2014;24:421–427.CrossRefPubMed
14.
go back to reference Chang CB, Yoo JJ, Song WS, Kim DJ, Koo KH, Kim HJ. Transfer of metallic debris from the metal surface of an acetabular cup to artificial femoral heads by scraping: Comparison between alumina and cobalt–chrome heads. J Biomed Mater Res B Appl Biomater. 2008;85:204–209.CrossRefPubMed Chang CB, Yoo JJ, Song WS, Kim DJ, Koo KH, Kim HJ. Transfer of metallic debris from the metal surface of an acetabular cup to artificial femoral heads by scraping: Comparison between alumina and cobalt–chrome heads. J Biomed Mater Res B Appl Biomater. 2008;85:204–209.CrossRefPubMed
15.
go back to reference Choudhury D, Roy T, Krupka I, Hartl M, Mootanah R. Tribological investigation of ultra-high molecular weight polyethylene against advanced ceramic surfaces in total hip joint replacement. Proc Inst Mech Eng. 2015;229:410–419.CrossRef Choudhury D, Roy T, Krupka I, Hartl M, Mootanah R. Tribological investigation of ultra-high molecular weight polyethylene against advanced ceramic surfaces in total hip joint replacement. Proc Inst Mech Eng. 2015;229:410–419.CrossRef
16.
go back to reference Coric D, Nunley PD, Guyer RD, Musante D, Carmody CN, Gordon CR, Lauryssen C, Ohnmeiss DD, Boltes MO. Prospective, randomized, multicenter study of cervical arthroplasty: 269 patients from the Kineflex|C artificial disc investigational device exemption study with a minimum 2-year follow-up: clinical article. J Neurosurg Spine. 2011;15:348–358.CrossRefPubMed Coric D, Nunley PD, Guyer RD, Musante D, Carmody CN, Gordon CR, Lauryssen C, Ohnmeiss DD, Boltes MO. Prospective, randomized, multicenter study of cervical arthroplasty: 269 patients from the Kineflex|C artificial disc investigational device exemption study with a minimum 2-year follow-up: clinical article. J Neurosurg Spine. 2011;15:348–358.CrossRefPubMed
17.
go back to reference Della Valle AG, Doty S, Gradl G, Labissiere A, Nestor BJ. Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head. J Arthroplasty. 2004;19:532–536.CrossRefPubMed Della Valle AG, Doty S, Gradl G, Labissiere A, Nestor BJ. Wear of a highly cross-linked polyethylene liner associated with metallic deposition on a ceramic femoral head. J Arthroplasty. 2004;19:532–536.CrossRefPubMed
18.
go back to reference Eberhardt AW, McKee RT, Cuckler JM, Peterson DW, Beck PR, Lemons JE. Surface roughness of CoCr and ZrO2 femoral heads with metal transfer: a retrieval and wear simulator study. Int J Biomater. 2009;2009:185456.CrossRefPubMedPubMedCentral Eberhardt AW, McKee RT, Cuckler JM, Peterson DW, Beck PR, Lemons JE. Surface roughness of CoCr and ZrO2 femoral heads with metal transfer: a retrieval and wear simulator study. Int J Biomater. 2009;2009:185456.CrossRefPubMedPubMedCentral
19.
go back to reference Epinette JA, Manley MT. No differences found in bearing related hip survivorship at 10–12 years follow-up between patients with ceramic on highly cross-linked polyethylene bearings compared to patients with ceramic on ceramic bearings. J Arthroplasty. 2014;29:1369–1372.CrossRefPubMed Epinette JA, Manley MT. No differences found in bearing related hip survivorship at 10–12 years follow-up between patients with ceramic on highly cross-linked polyethylene bearings compared to patients with ceramic on ceramic bearings. J Arthroplasty. 2014;29:1369–1372.CrossRefPubMed
26.
go back to reference Fisher J, Firkins P, Reeves E, Hailey J, Isaac G. The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc Inst Mech Eng H. 1995;209:263–264.CrossRefPubMed Fisher J, Firkins P, Reeves E, Hailey J, Isaac G. The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc Inst Mech Eng H. 1995;209:263–264.CrossRefPubMed
27.
go back to reference Grupp TM, Meisel H-J, Cotton JA, Schwiesau J, Fritz B, Blömer W, Jansson V. Alternative bearing materials for intervertebral disc arthroplasty. Biomaterials. 2010;31:523–531.CrossRefPubMed Grupp TM, Meisel H-J, Cotton JA, Schwiesau J, Fritz B, Blömer W, Jansson V. Alternative bearing materials for intervertebral disc arthroplasty. Biomaterials. 2010;31:523–531.CrossRefPubMed
28.
go back to reference Grupp TM, Yue JJ, Garcia R Jr, Kaddick C, Fritz B, Schilling C, Schwiesau J, Blomer W. Evaluation of impingement behaviour in lumbar spinal disc arthroplasty. Eur Spine J. 2015;24:2033–2046.CrossRefPubMed Grupp TM, Yue JJ, Garcia R Jr, Kaddick C, Fritz B, Schilling C, Schwiesau J, Blomer W. Evaluation of impingement behaviour in lumbar spinal disc arthroplasty. Eur Spine J. 2015;24:2033–2046.CrossRefPubMed
29.
go back to reference ISO. ISO 18192-1:2011–Implants for Surgery–Wear of Total Intervertebral Spinal Disc Prostheses–Part 1: Loading and Displacement Parameters for Wear Testing and Corresponding Environmental Conditions for Test. Geneva, Switzerland: International Organization for Standardization; 2011. ISO. ISO 18192-1:2011–Implants for Surgery–Wear of Total Intervertebral Spinal Disc Prostheses–Part 1: Loading and Displacement Parameters for Wear Testing and Corresponding Environmental Conditions for Test. Geneva, Switzerland: International Organization for Standardization; 2011.
30.
go back to reference Kim Y-H, Ritchie A, Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg Am. 2005;87:577–582.PubMed Kim Y-H, Ritchie A, Hardaker C. Surface roughness of ceramic femoral heads after in vivo transfer of metal: correlation to polyethylene wear. J Bone Joint Surg Am. 2005;87:577–582.PubMed
31.
go back to reference Kraft M. Wear performance of a PEEK-on-PEEK bearing under cervical loading conditions and at different test frequencies. Global Spine J. 2012;2:S120–S121.CrossRef Kraft M. Wear performance of a PEEK-on-PEEK bearing under cervical loading conditions and at different test frequencies. Global Spine J. 2012;2:S120–S121.CrossRef
32.
go back to reference Kraft M. Impact of the processing history on the wear performance of a PEEK-on-PEEK bearing for cervical total disc replacement. International Journal of Mechanics and Applications. 2013;3:1–7. Kraft M. Impact of the processing history on the wear performance of a PEEK-on-PEEK bearing for cervical total disc replacement. International Journal of Mechanics and Applications. 2013;3:1–7.
33.
go back to reference Kubo K, Clarke IC, Sorimachi T, Williams PA, Donaldson TK, Yamamoto K. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: a hip simulator model. Wear. 2009;267:734–742.CrossRef Kubo K, Clarke IC, Sorimachi T, Williams PA, Donaldson TK, Yamamoto K. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: a hip simulator model. Wear. 2009;267:734–742.CrossRef
34.
go back to reference Kurtz SM. Total disc arthroplasty. In: Kurtz SM, Edidin AA, eds. Spine Technology Handbook. New York, NY, USA: Academic Press; 2006:1–10.CrossRef Kurtz SM. Total disc arthroplasty. In: Kurtz SM, Edidin AA, eds. Spine Technology Handbook. New York, NY, USA: Academic Press; 2006:1–10.CrossRef
35.
go back to reference Kurtz SM, Ciccarelli L, Harper ML, Siskey R, Shorez J, Chan FW. Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements. Int J Spine Surg. 2012;6:145–156.CrossRefPubMedPubMedCentral Kurtz SM, Ciccarelli L, Harper ML, Siskey R, Shorez J, Chan FW. Comparison of in vivo and simulator-retrieved metal-on-metal cervical disc replacements. Int J Spine Surg. 2012;6:145–156.CrossRefPubMedPubMedCentral
36.
go back to reference Kurtz SM, Toth JM, Siskey R, Ciccarelli L, MacDonald D, Isaza J, Lanman T, Punt I, Steinbeck M, Goffin J, van Ooij A. The latest lessons learned from retrieval analyses of ultra-high molecular weight polyethylene, metal-on-metal, and alternative bearing total disc replacements. Semin Spine Surg. 2012;24:57–70.CrossRefPubMedPubMedCentral Kurtz SM, Toth JM, Siskey R, Ciccarelli L, MacDonald D, Isaza J, Lanman T, Punt I, Steinbeck M, Goffin J, van Ooij A. The latest lessons learned from retrieval analyses of ultra-high molecular weight polyethylene, metal-on-metal, and alternative bearing total disc replacements. Semin Spine Surg. 2012;24:57–70.CrossRefPubMedPubMedCentral
37.
go back to reference Lebl DR, Cammisa FP, Girardi FP, Wright T, Abjornson C. In vivo functional performance of failed Prodisc-L devices: retrieval analysis of lumbar total disc replacements. Spine (Phila Pa 1976). 2012;37:E1209–1217. Lebl DR, Cammisa FP, Girardi FP, Wright T, Abjornson C. In vivo functional performance of failed Prodisc-L devices: retrieval analysis of lumbar total disc replacements. Spine (Phila Pa 1976). 2012;37:E1209–1217.
38.
go back to reference Lübbeke A, Gonzalez A, Garavaglia G, Roussos C, Bonvin A, Stern R, Peter R, Hoffmeyer P. A comparative assessment of small-head metal-on-metal and ceramic-on-polyethylene total hip replacements. Bone Joint J. 2014;96:868–875.CrossRefPubMed Lübbeke A, Gonzalez A, Garavaglia G, Roussos C, Bonvin A, Stern R, Peter R, Hoffmeyer P. A comparative assessment of small-head metal-on-metal and ceramic-on-polyethylene total hip replacements. Bone Joint J. 2014;96:868–875.CrossRefPubMed
39.
go back to reference Luchetti WT, Copley LA, Vresilovic EJ, Black J, Steinberg ME. Drain entrapment and titanium to ceramic head deposition. J Arthroplasty. 1998;13:713–717.CrossRefPubMed Luchetti WT, Copley LA, Vresilovic EJ, Black J, Steinberg ME. Drain entrapment and titanium to ceramic head deposition. J Arthroplasty. 1998;13:713–717.CrossRefPubMed
40.
go back to reference Meftah M, Klingenstein GG, Yun RJ, Ranawat AS, Ranawat CS. Long-term performance of ceramic and metal femoral heads on conventional polyethylene in young and active patients. J Bone Joint Surg Am. 2013;95:1193–1197.CrossRefPubMed Meftah M, Klingenstein GG, Yun RJ, Ranawat AS, Ranawat CS. Long-term performance of ceramic and metal femoral heads on conventional polyethylene in young and active patients. J Bone Joint Surg Am. 2013;95:1193–1197.CrossRefPubMed
41.
go back to reference Moussa ME, Esposito CI, Elpers ME, Wright TM, Padgett DE. Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals. J Arthroplasty. 2015;30:713–717.CrossRefPubMed Moussa ME, Esposito CI, Elpers ME, Wright TM, Padgett DE. Hip dislocation increases roughness of oxidized zirconium femoral heads in total hip arthroplasty: an analysis of 59 retrievals. J Arthroplasty. 2015;30:713–717.CrossRefPubMed
42.
go back to reference Müller FA, Hagymási M, Greil P, Zeiler G, Schuh A. Transfer of metallic debris after dislocation of ceramic femoral heads in hip prostheses. Arch Orthop Trauma Surg. 2006;126:174–180.CrossRefPubMed Müller FA, Hagymási M, Greil P, Zeiler G, Schuh A. Transfer of metallic debris after dislocation of ceramic femoral heads in hip prostheses. Arch Orthop Trauma Surg. 2006;126:174–180.CrossRefPubMed
43.
go back to reference Nechtow W, Bushelow M, Hinter M, Ochs A, Kaddick C. Cervical disc prosthesis polyethylene wear following the ISO cervical test. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008. Nechtow W, Bushelow M, Hinter M, Ochs A, Kaddick C. Cervical disc prosthesis polyethylene wear following the ISO cervical test. 54th Annual Meeting of the Orthopaedic Research Society, San Francisco, CA, USA; 2008.
45.
go back to reference Rajpura A, Kendoff D, Board T. The current state of bearing surfaces in total hip replacement. Bone Joint J. 2014;96:147–156.CrossRefPubMed Rajpura A, Kendoff D, Board T. The current state of bearing surfaces in total hip replacement. Bone Joint J. 2014;96:147–156.CrossRefPubMed
46.
go back to reference Rajpura A, Wroblewski B, Siney P, Board T, Jones HW. 28-year clinical and radiological follow-up of cross-linked polyethylene on alumina ceramic: a prospective study of modified Charnley low frictional torque arthroplasty. Bone Joint J. 2015;97(Suppl 12):43. Rajpura A, Wroblewski B, Siney P, Board T, Jones HW. 28-year clinical and radiological follow-up of cross-linked polyethylene on alumina ceramic: a prospective study of modified Charnley low frictional torque arthroplasty. Bone Joint J. 2015;97(Suppl 12):43.
47.
go back to reference Shkolnikov YP, Bowden A, MacDonald D, Kurtz SM. Wear pattern observations from TDR retrievals using autoregistration of voxel data. J Biomed Mater Res B Appl Biomater. 2010;94:312–317.PubMedPubMedCentral Shkolnikov YP, Bowden A, MacDonald D, Kurtz SM. Wear pattern observations from TDR retrievals using autoregistration of voxel data. J Biomed Mater Res B Appl Biomater. 2010;94:312–317.PubMedPubMedCentral
48.
go back to reference Sorimachi T, Clarke I, Williams P, Gustafson A, Yamamoto K. Third-body abrasive wear challenge of 32 mm conventional and 44 mm highly crosslinked polyethylene liners in a hip simulator model. Proc Inst Mech Eng H. 2009;223:607–23.CrossRefPubMed Sorimachi T, Clarke I, Williams P, Gustafson A, Yamamoto K. Third-body abrasive wear challenge of 32 mm conventional and 44 mm highly crosslinked polyethylene liners in a hip simulator model. Proc Inst Mech Eng H. 2009;223:607–23.CrossRefPubMed
49.
go back to reference Tipper J, Firkins P, Besong A, Barbour P, Nevelos J, Stone M, Ingham E, Fisher J. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear. 2001;250:120–128.CrossRef Tipper J, Firkins P, Besong A, Barbour P, Nevelos J, Stone M, Ingham E, Fisher J. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear. 2001;250:120–128.CrossRef
50.
go back to reference Wang A, Dumbleton JH. Ultra-high-molecular-weight polyethylene (UHMWPE) as a bearing material in hip joint replacements. In: Wang QJ, Chung Y-W, eds. Encyclopedia of Tribology. Boston, MA, USA: Springer; 2013:3933–3939.CrossRef Wang A, Dumbleton JH. Ultra-high-molecular-weight polyethylene (UHMWPE) as a bearing material in hip joint replacements. In: Wang QJ, Chung Y-W, eds. Encyclopedia of Tribology. Boston, MA, USA: Springer; 2013:3933–3939.CrossRef
51.
go back to reference Xin H, Shepherd D, Dearn K. A tribological assessment of a PEEK based self-mating total cervical disc replacement. Wear. 2013;303:473–479.CrossRef Xin H, Shepherd D, Dearn K. A tribological assessment of a PEEK based self-mating total cervical disc replacement. Wear. 2013;303:473–479.CrossRef
Metadata
Title
Are PEEK-on-Ceramic Bearings an Option for Total Disc Arthroplasty? An In Vitro Tribology Study
Authors
Ryan Siskey, MS
Lauren Ciccarelli, MS
Melissa K. C. Lui, MS
Steven M. Kurtz, PhD
Publication date
01-11-2016
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 11/2016
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-5041-7