01-04-2013 | Original Article

Appropriate collimators in a small animal SPECT scanner with CZT detector

Authors: Yusuke Higaki, Masato Kobayashi, Tomoya Uehara, Hirofumi Hanaoka, Yasushi Arano, Keiichi Kawai

Published in: Annals of Nuclear Medicine | Issue 3/2013

Login to get access

Abstract

Objective

Almost all small animal SPECT is performed with pinhole collimators (PH), including single-PH (SPH) and multi-PH (MPH). In the clinical study, not only PH but also parallel-hole collimator (PAH) is often used in planar and SPECT imaging. However, there have been no comparative studies on image quality with various collimators on the small animal imaging. This study compared the basic characteristics of PH and PAH in small animal imaging.

Methods

Performance of planar and SPECT images was evaluated using 99mTcO4 and SPH, MPH and PAH with low energy and high resolution on the SPECT/CT scanner FX3200. We measured sensitivity, resolution, concentration linearity and uniformity. Planar imaging of mice with 99mTc-labeled mercaptoacetyltriglycine (99mTc-MAG3) was performed using SPH and PAH. SPECT imaging with 99mTc-methylene diphosphonate (99mTc-MDP) was performed using all collimators.

Results

With SPH, MPH and PAH, sensitivity was 43.5, 211.2 and 926.5 cps/MBq, respectively, and spatial resolution was 0.60/0.56, non/0.96, 5.20/5.34 mm full-width half maximum (planar/SPECT), respectively. There were marked correlations between the radioactivity counts on images and radioactivity with all collimators. Values of % standard deviation on planar imaging showed small differences between the SPH and PAH, while the values were the smallest on SPECT imaging with MPH. On imaging of mice, SPH yielded high-quality 99mTc-MAG3-planar images when compared with PAH. MPH yielded sharper 99mTc-MDP-SPECT images than SPH and PAH.

Conclusions

The characteristics of PH and PAH differed on small animal imaging. Although sensitivity was higher with PAH, PH showed higher resolution. Among the PH collimators, SPH was more appropriate for planar imaging, and MPH was more suitable for SPECT imaging in a small animal imaging scanner with CZT detector.
Literature
1.
go back to reference Weisenberger AG, Wojcik R, Bradley EL, Brewer P, Majewski S, Qian J, et al. SPECT-CT system for small animal imaging. IEEE Trans Nucl Sci. 2003;1:74–9.CrossRef Weisenberger AG, Wojcik R, Bradley EL, Brewer P, Majewski S, Qian J, et al. SPECT-CT system for small animal imaging. IEEE Trans Nucl Sci. 2003;1:74–9.CrossRef
2.
go back to reference Franc BL, Acton PD, Mari C, Hasegawa BH. Small-animal SPECT and SPECT/CT: important tools for preclinical investigation. J Nucl Med. 2008;49:1651–63.PubMedCrossRef Franc BL, Acton PD, Mari C, Hasegawa BH. Small-animal SPECT and SPECT/CT: important tools for preclinical investigation. J Nucl Med. 2008;49:1651–63.PubMedCrossRef
3.
go back to reference de Kemp RA, Epstein FH, Catana C, Tsui BM, Ritman EL. Small-animal molecular imaging methods. J Nucl Med. 2010;51:18S–32S.PubMedCrossRef de Kemp RA, Epstein FH, Catana C, Tsui BM, Ritman EL. Small-animal molecular imaging methods. J Nucl Med. 2010;51:18S–32S.PubMedCrossRef
4.
go back to reference Mueller B, O’Connor MK, Blevis I, Rhodes DJ, Smith R, Collins DA, et al. Evaluation of a small cadmium zinc telluride detector for scintimammography. J Nucl Med. 2003;44:602–9.PubMed Mueller B, O’Connor MK, Blevis I, Rhodes DJ, Smith R, Collins DA, et al. Evaluation of a small cadmium zinc telluride detector for scintimammography. J Nucl Med. 2003;44:602–9.PubMed
5.
go back to reference Funk T, Parnham KB, Patt BE, Li J, Iwanczyk JS, Iwata K. A new CdZnTe-based gamma camera for high resolution pinhole SPECT. Conf Record IEEE Nucl Sci Symp Med Imag. 2003;4:2320–4. Funk T, Parnham KB, Patt BE, Li J, Iwanczyk JS, Iwata K. A new CdZnTe-based gamma camera for high resolution pinhole SPECT. Conf Record IEEE Nucl Sci Symp Med Imag. 2003;4:2320–4.
6.
go back to reference Kim H, Furenlid LR, Crawford MJ, Wilson DW, Barber HB, Peterson TE, et al. SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. Med Phys. 2006;33:465–74.PubMedCrossRef Kim H, Furenlid LR, Crawford MJ, Wilson DW, Barber HB, Peterson TE, et al. SemiSPECT: a small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays. Med Phys. 2006;33:465–74.PubMedCrossRef
7.
go back to reference Scheiber C. CdTe and CdZnTe detectors in nuclear medicine. Nucl Instr Meth Phys Res A. 2010;448:513–24.CrossRef Scheiber C. CdTe and CdZnTe detectors in nuclear medicine. Nucl Instr Meth Phys Res A. 2010;448:513–24.CrossRef
8.
go back to reference Mok GS, Yu J, Du Y, Wang Y, Tsui Bm. Evaluation of a multi-pinhole collimator for imaging small animals with different sizes. Mol Imag Biol. 2012;14:60–9.CrossRef Mok GS, Yu J, Du Y, Wang Y, Tsui Bm. Evaluation of a multi-pinhole collimator for imaging small animals with different sizes. Mol Imag Biol. 2012;14:60–9.CrossRef
9.
go back to reference Beekman FJ, van der Have F, Vastenhouw B, van der Linden AJ, van Rijk PP, Burbach JP, et al. U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. J Nucl Med. 2005;46:1194–200.PubMed Beekman FJ, van der Have F, Vastenhouw B, van der Linden AJ, van Rijk PP, Burbach JP, et al. U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. J Nucl Med. 2005;46:1194–200.PubMed
10.
go back to reference van der Have F, Vastenhouw B, Ramakers RM, Branderhorst W, Krah JO, Ji C, et al. U-SPECT-II: An ultra-high-resolution device for molecular small-animal imaging. J Nucl Med. 2009;50:599–605.PubMedCrossRef van der Have F, Vastenhouw B, Ramakers RM, Branderhorst W, Krah JO, Ji C, et al. U-SPECT-II: An ultra-high-resolution device for molecular small-animal imaging. J Nucl Med. 2009;50:599–605.PubMedCrossRef
11.
go back to reference Umeda IO, Tani K, Tsuda K, Kobayashi M, Ogata M, Kimura S, et al. High resolution SPECT imaging for visualization of intratumoral heterogeneity using a SPECT/CT scanner dedicated for small animal imaging. Ann Nucl Med. 2012;1:67–76.CrossRef Umeda IO, Tani K, Tsuda K, Kobayashi M, Ogata M, Kimura S, et al. High resolution SPECT imaging for visualization of intratumoral heterogeneity using a SPECT/CT scanner dedicated for small animal imaging. Ann Nucl Med. 2012;1:67–76.CrossRef
12.
go back to reference Magota K, Kubo N, Kuge Y, Nishijima K, Zhao S, Tamaki N. Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging. Eur J Nucl Med Mol Imaging. 2011;38:742–52.PubMedCrossRef Magota K, Kubo N, Kuge Y, Nishijima K, Zhao S, Tamaki N. Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging. Eur J Nucl Med Mol Imaging. 2011;38:742–52.PubMedCrossRef
13.
go back to reference Weber DA, Ivanovic M, Franceschi D, Strand SE, Eriandsson K, Franceschi M, et al. Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals. J Nucl Med. 1994;35:342–8.PubMed Weber DA, Ivanovic M, Franceschi D, Strand SE, Eriandsson K, Franceschi M, et al. Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals. J Nucl Med. 1994;35:342–8.PubMed
14.
go back to reference Song TY, Choi Y, Chung YH, Jung JH, Min BJ, Hong KJ, et al. Performance amelioration for small animal SPECT using optimized pinhole collimator and image correction technique. IEEE Trans Nucl Sci. 2004;6:3458–62. Song TY, Choi Y, Chung YH, Jung JH, Min BJ, Hong KJ, et al. Performance amelioration for small animal SPECT using optimized pinhole collimator and image correction technique. IEEE Trans Nucl Sci. 2004;6:3458–62.
15.
go back to reference Qian J, Bradley EL, Majewski S, Popov V, Saha MS, Smith MF, et al. A small-animal imaging system capable of multipinhole circular/helical SPECT and parallel-hole SPECT. Nucl Instr Meth Phys Res A. 2008;594:102–10.CrossRef Qian J, Bradley EL, Majewski S, Popov V, Saha MS, Smith MF, et al. A small-animal imaging system capable of multipinhole circular/helical SPECT and parallel-hole SPECT. Nucl Instr Meth Phys Res A. 2008;594:102–10.CrossRef
16.
go back to reference Zingermana Y, Golan H, Gersten A, Moalem A. A compact CT/SPECT system for small-object imaging. Nucl Instr Meth Phys Res A. 2008;584:135–48.CrossRef Zingermana Y, Golan H, Gersten A, Moalem A. A compact CT/SPECT system for small-object imaging. Nucl Instr Meth Phys Res A. 2008;584:135–48.CrossRef
17.
go back to reference Figueroa SD, Winkelmann CT, Volkert WA, Hoffman TJ. Performance characteristics of an integrated small animal SPECT/CT unit. Conf Record IEEE Nucl Sci Symp Med Imag. 2005;1752–6. Figueroa SD, Winkelmann CT, Volkert WA, Hoffman TJ. Performance characteristics of an integrated small animal SPECT/CT unit. Conf Record IEEE Nucl Sci Symp Med Imag. 2005;1752–6.
18.
go back to reference Herrler T, Wang H, Tischer A, Bartenstein P, Jauch KW, Guba M, et al. 99mTc-MAG3 scintigraphy for the longitudinal follow-up of kidney function in a mouse model of renal ischemia-reperfusion injury. EJNMMI Res. 2012;2:2.PubMedCrossRef Herrler T, Wang H, Tischer A, Bartenstein P, Jauch KW, Guba M, et al. 99mTc-MAG3 scintigraphy for the longitudinal follow-up of kidney function in a mouse model of renal ischemia-reperfusion injury. EJNMMI Res. 2012;2:2.PubMedCrossRef
19.
go back to reference Roberts J, Chen B, Curtis LM, Agarwal A, Sanders PW, Zinn KR. Detection of early changes in renal function using 99mTc-MAG3 imaging in a murine model of ischemia-reperfusion injury. Am J Physiol Renal Physiol. 2007;293:F1408–12.PubMedCrossRef Roberts J, Chen B, Curtis LM, Agarwal A, Sanders PW, Zinn KR. Detection of early changes in renal function using 99mTc-MAG3 imaging in a murine model of ischemia-reperfusion injury. Am J Physiol Renal Physiol. 2007;293:F1408–12.PubMedCrossRef
20.
go back to reference Arveschoug AK, Bertelsen H, Vammen B. Presurgical localization of abnormal parathyroid glands using a single injection of Tc-99m sestamibi: comparison of high-resolution parallel-hole and pinhole collimators, and interobserver and intraobserver variation. Clin Nucl Med. 2002;27:249–54.PubMedCrossRef Arveschoug AK, Bertelsen H, Vammen B. Presurgical localization of abnormal parathyroid glands using a single injection of Tc-99m sestamibi: comparison of high-resolution parallel-hole and pinhole collimators, and interobserver and intraobserver variation. Clin Nucl Med. 2002;27:249–54.PubMedCrossRef
Metadata
Title
Appropriate collimators in a small animal SPECT scanner with CZT detector
Authors
Yusuke Higaki
Masato Kobayashi
Tomoya Uehara
Hirofumi Hanaoka
Yasushi Arano
Keiichi Kawai
Publication date
01-04-2013
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 3/2013
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-012-0681-5

Other articles of this Issue 3/2013

Annals of Nuclear Medicine 3/2013 Go to the issue