Appraisal of the DIERS method for calculating postural measurements: an observational study

Authors: Brian Degenhardt, Zane Starks, Shalini Bhatia, Geoffroey-Allen Franklin

Published in: Scoliosis and Spinal Disorders | Issue 1/2017

Login to get access

Abstract

Background

Surface topography is increasingly used with postural analysis. One system, DIERS formetric 4D, measures 40 defined spine shape parameters from a 6-s scan. Through system algorithms, a set of spine shape parameter values from 1 of 12 recorded images obtained during a scan becomes the DIERS-reported value (DRV) for postural assessment. The purpose of the current study was to compare DRV with a standard average value (SAV) calculated from all 12 images to determine which method is more appropriate for assessing postural change.

Methods

One mannequin and 30 human participants were scanned over 5 days. Values from each image and the DRV for 40 defined spine shape parameters were exported, and mean DRV, mean SAV, mean DRV, and within-scan variance were calculated. Absolute difference and percent change between mean DRV and mean SAV were calculated for the mannequin and humans. Inter-method reliability was calculated for humans. Within-scan variance for each parameter was tested for significant variability.

Results

For all spine shape parameters on the mannequin, absolute difference (< 0.6 mm, 0.1°, or 0.1%) and percent change (< 2.90%) between mean DRV and mean SAV for each parameter were small. Nine parameters on human participants had a large percent change (> 7%). Absolute difference between mean DRV and mean SAV for those nine parameters was small (≤ 0.87 mm or 0.61°). Absolute difference for all other parameters ranged from 0.02 to 6.98 mm for distance measurements, from 0.01 to 1.21° for angle measurements, and from 0.15 to 0.22% for percentage measurements. Inter-method reliability between DRV and SAV was excellent (0.94–1.00). For the mannequin, within-scan variance was small (< 1.62) for all parameters. For humans, within-scan variance ranged from 0.05 to 36.04 and was different from zero for all parameters (all P < 0.001).

Conclusions

The minimal variability observed in the mannequin suggested the DIERS formetric 4D instrument had high within-scan reliability. The DRV and SAV provided comparable spine shape parameter values. Because within-scan variability is not reported with the DRV, the clinical usefulness of current DRV values is limited. Establishing an estimate of variance with the SAV will allow clinicians to better identify a clinically meaningful change.
Literature
1.
go back to reference Drerup B, Hierholzer E. Automatic localization of anatomical landmarks on the back surface and construction of a body-fixed coordinate system. J Biomech. 1987;20:961–70.CrossRefPubMed Drerup B, Hierholzer E. Automatic localization of anatomical landmarks on the back surface and construction of a body-fixed coordinate system. J Biomech. 1987;20:961–70.CrossRefPubMed
2.
go back to reference Frerich JM, Hertzler K, Knott P, Mardjetko S. Comparison of radiographic and surface topography measurements in adolescents with idiopathic scoliosis. Open Orthop J. 2012;6:261–5.CrossRefPubMedPubMedCentral Frerich JM, Hertzler K, Knott P, Mardjetko S. Comparison of radiographic and surface topography measurements in adolescents with idiopathic scoliosis. Open Orthop J. 2012;6:261–5.CrossRefPubMedPubMedCentral
3.
go back to reference Hackenberg L, Hierholzer E, Pötzl W, Götze C, Liljenqvist U. Rasterstereographic back shape analysis in idiopathic scoliosis after anterior correction and fusion. Clin Biomech (Bristol, Avon). 2003;18:1–8.CrossRef Hackenberg L, Hierholzer E, Pötzl W, Götze C, Liljenqvist U. Rasterstereographic back shape analysis in idiopathic scoliosis after anterior correction and fusion. Clin Biomech (Bristol, Avon). 2003;18:1–8.CrossRef
4.
go back to reference Hackenberg L, Hierholzer E, Pötzl W, Götze C, Liljenqvist U. Rasterstereographic back shape analysis in idiopathic scoliosis after posterior correction and fusion. Clin Biomech (Bristol, Avon). 2003;18:883–9.CrossRef Hackenberg L, Hierholzer E, Pötzl W, Götze C, Liljenqvist U. Rasterstereographic back shape analysis in idiopathic scoliosis after posterior correction and fusion. Clin Biomech (Bristol, Avon). 2003;18:883–9.CrossRef
5.
go back to reference Knott P, Sturm P, Lonner B, Cahill P, Betsch M, McCarthy R, et al. Multicenter comparison of 3D spinal measurements using surface topography with those from conventional radiography. Spine Deform. 2016;4:98–103.CrossRefPubMed Knott P, Sturm P, Lonner B, Cahill P, Betsch M, McCarthy R, et al. Multicenter comparison of 3D spinal measurements using surface topography with those from conventional radiography. Spine Deform. 2016;4:98–103.CrossRefPubMed
6.
go back to reference Tabard-Fougere A, Bonnefoy-Mazure A, Hanquinet S, Lascombes P, Armand S, Dayer R. Validity and reliability of spine rasterstereography in patients with adolescent idiopathic scoliosis. Spine (Phila Pa 1976). 2017;42:98–105.CrossRef Tabard-Fougere A, Bonnefoy-Mazure A, Hanquinet S, Lascombes P, Armand S, Dayer R. Validity and reliability of spine rasterstereography in patients with adolescent idiopathic scoliosis. Spine (Phila Pa 1976). 2017;42:98–105.CrossRef
7.
go back to reference Guidetti L, Bonavolonta V, Tito A, Reis VM, Gallotta MC, Baldari C. Intra- and interday reliability of spine rasterstereography. Biomed Res Int. 2013;2013:745480.CrossRefPubMedPubMedCentral Guidetti L, Bonavolonta V, Tito A, Reis VM, Gallotta MC, Baldari C. Intra- and interday reliability of spine rasterstereography. Biomed Res Int. 2013;2013:745480.CrossRefPubMedPubMedCentral
8.
go back to reference Lason G, Peeters L, Vandenberghe K, Byttebier G, Comhaire F. Reassessing the accuracy and reproducibility of Diers formetric measurements in healthy volunteers. Int J Osteopath Med. 2015;18:247–54.CrossRef Lason G, Peeters L, Vandenberghe K, Byttebier G, Comhaire F. Reassessing the accuracy and reproducibility of Diers formetric measurements in healthy volunteers. Int J Osteopath Med. 2015;18:247–54.CrossRef
9.
go back to reference Mohokum M, Mendoza S, Udo W, Sitter H, Paletta JR, Skwara A. Reproducibility of rasterstereography for kyphotic and lordotic angles, trunk length, and trunk inclination: a reliability study. Spine (Phila Pa 1976). 2010;35:1353–8.CrossRef Mohokum M, Mendoza S, Udo W, Sitter H, Paletta JR, Skwara A. Reproducibility of rasterstereography for kyphotic and lordotic angles, trunk length, and trunk inclination: a reliability study. Spine (Phila Pa 1976). 2010;35:1353–8.CrossRef
10.
go back to reference Schroeder J, Reer R, Braumann KM. Video raster stereography back shape reconstruction: a reliability study for sagittal, frontal, and transversal plane parameters. Eur Spine J. 2015;24:262–9.CrossRefPubMed Schroeder J, Reer R, Braumann KM. Video raster stereography back shape reconstruction: a reliability study for sagittal, frontal, and transversal plane parameters. Eur Spine J. 2015;24:262–9.CrossRefPubMed
11.
go back to reference Drerup B, Hierholzer E. Objective determination of anatomical landmarks on the body surface: measurement of the vertebra prominens from surface curvature. J Biomech. 1985;18:467–74.CrossRefPubMed Drerup B, Hierholzer E. Objective determination of anatomical landmarks on the body surface: measurement of the vertebra prominens from surface curvature. J Biomech. 1985;18:467–74.CrossRefPubMed
12.
go back to reference Goh S, Price RI, Leedman PJ, Singer KP. Rasterstereographic analysis of the thoracic sagittal curvature: a reliability study. J Musculoskelet Res. 1999;03:137–42.CrossRef Goh S, Price RI, Leedman PJ, Singer KP. Rasterstereographic analysis of the thoracic sagittal curvature: a reliability study. J Musculoskelet Res. 1999;03:137–42.CrossRef
13.
go back to reference Drerup B, Hierholzer E. Movement of the human pelvis and displacement of related anatomical landmarks on the body surface. J Biomech. 1987;20:971–7.CrossRefPubMed Drerup B, Hierholzer E. Movement of the human pelvis and displacement of related anatomical landmarks on the body surface. J Biomech. 1987;20:971–7.CrossRefPubMed
14.
go back to reference Drerup B, Ellger B, Meyer zu Bentrup F, Hierholzer E. Functional rasterstereographic images. A new method for biomechanical analysis of skeletal geometry [in German]. Orthopade. 2001;30:242–50.CrossRefPubMed Drerup B, Ellger B, Meyer zu Bentrup F, Hierholzer E. Functional rasterstereographic images. A new method for biomechanical analysis of skeletal geometry [in German]. Orthopade. 2001;30:242–50.CrossRefPubMed
15.
go back to reference Drerup B, Hierholzer E. Back shape measurement using video rasterstereography and three-dimensional reconstruction of spinal shape. Clin Biomech (Bristol, Avon). 1994;9:28–36.CrossRef Drerup B, Hierholzer E. Back shape measurement using video rasterstereography and three-dimensional reconstruction of spinal shape. Clin Biomech (Bristol, Avon). 1994;9:28–36.CrossRef
Metadata
Title
Appraisal of the DIERS method for calculating postural measurements: an observational study
Authors
Brian Degenhardt
Zane Starks
Shalini Bhatia
Geoffroey-Allen Franklin
Publication date
01-12-2017
Publisher
BioMed Central
Published in
Scoliosis and Spinal Disorders / Issue 1/2017
Electronic ISSN: 2397-1789
DOI
https://doi.org/10.1186/s13013-017-0134-y

Other articles of this Issue 1/2017

Scoliosis and Spinal Disorders 1/2017 Go to the issue