Published in: La radiologia medica 8/2018

01-08-2018 | COMPUTED TOMOGRAPHY

Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses

Authors: Nicola Magarelli, Vincenzo De Santis, Giammaria Marziali, Amerigo Menghi, Aaron Burrofato, Luigi Pedone, Dario Del Prete, Roberto Iezzi, Chiara de Waure, Marianna D’andrea, Antonio Leone, Cesare Colosimo

Published in: La radiologia medica | Issue 8/2018

Login to get access

Abstract

Objective

The study aimed to assess image quality when using dual-energy CT (DECT) to reduce metal artifacts in subjects with knee and hip prostheses.

Methods

Twenty-two knee and 10 hip prostheses were examined in 31 patients using a DECT protocol (tube voltages 100 and 140 kVp). Monoenergetic reconstructions were extrapolated at 64, 69, 88, 105, 110, 120, 140, 170, and 190 kilo-electron volts (keV) and the optimal energy was manually selected. The B60–140 and Fast DE reconstructions were made by CT. The image quality and diagnostic value were subjectively and objectively determined. Double-blind qualitative assessment was performed by two radiologists using a Likert scale. For quantitative analysis, a circular region of interest (ROI) was placed by a third radiologist within the most evident streak artifacts on every image. Another ROI was placed in surrounding tissues without artifacts as a reference.

Results

The inter-reader agreement for the qualitative assessment was nearly 100%. The best overall image quality (37.8% rated “excellent”) was the Fast DE Siemens reconstruction, followed by B60–140 and Opt KeV (20.5 and 10.2% rated excellent). On the other hand, DECT images at 64, 69 and 88 keV had the worse scores. The number of artifacts was significantly different between monoenergetic images. Nevertheless, because of the high number of pairwise comparisons, no differences were found in the post hoc analysis except for a trend toward statistical significance when comparing the 170 and 64 keV doses.

Conclusions

DECT with specific post-processing may reduce metal artifacts and significantly enhance the image quality and diagnostic value when evaluating metallic implants.
Literature
1.
go back to reference Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803CrossRefPubMed Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803CrossRefPubMed
2.
go back to reference White LM, Buckwalter KA (2002) Technical considerations: CT and MR imaging in the postoperative orthopedic patient. Semin Musculoskelet Radiol 6:5–17CrossRefPubMed White LM, Buckwalter KA (2002) Technical considerations: CT and MR imaging in the postoperative orthopedic patient. Semin Musculoskelet Radiol 6:5–17CrossRefPubMed
3.
4.
go back to reference Moon SG, Hong SH, Choi J-Y, Jun WS, Kang H-G, Kim H-S et al (2008) Metal artifact reduction by the alteration of technical factors in multidetector computed tomography: a 3-dimensional quantitative assessment. J Comput Assist Tomogr 32(4):630–633CrossRefPubMed Moon SG, Hong SH, Choi J-Y, Jun WS, Kang H-G, Kim H-S et al (2008) Metal artifact reduction by the alteration of technical factors in multidetector computed tomography: a 3-dimensional quantitative assessment. J Comput Assist Tomogr 32(4):630–633CrossRefPubMed
5.
go back to reference Perrella A, Lopes PML, Rocha RG, Fenyo-Pereira M, Cavalcanti MGP (2010) Influence of dental metallic artifact from multislice CT in the assessment of simulated mandibular lesions. J Appl Oral Sci. 18(2):149–154CrossRefPubMedPubMedCentral Perrella A, Lopes PML, Rocha RG, Fenyo-Pereira M, Cavalcanti MGP (2010) Influence of dental metallic artifact from multislice CT in the assessment of simulated mandibular lesions. J Appl Oral Sci. 18(2):149–154CrossRefPubMedPubMedCentral
6.
go back to reference Vande Berg B, Malghem J, Maldague B, Lecouvet F (2006) Multi-detector CT imaging in the postoperative orthopedic patient with metal hardware. Eur J Radiol 60(3):470–479CrossRefPubMed Vande Berg B, Malghem J, Maldague B, Lecouvet F (2006) Multi-detector CT imaging in the postoperative orthopedic patient with metal hardware. Eur J Radiol 60(3):470–479CrossRefPubMed
7.
go back to reference Link TM, Berning W, Scherf S, Joosten U, Joist A, Engelke K et al (2000) CT of metal implants: reduction of artifacts using an extended CT scale technique. J Comput Assist Tomogr 24(1):165–172CrossRefPubMed Link TM, Berning W, Scherf S, Joosten U, Joist A, Engelke K et al (2000) CT of metal implants: reduction of artifacts using an extended CT scale technique. J Comput Assist Tomogr 24(1):165–172CrossRefPubMed
8.
go back to reference Fishman EK, Magid D, Robertson DD, Brooker AF, Weiss P, Siegelman SS (1986) Metallic hip implants: CT with multiplanar reconstruction. Radiology 160(3):675–681CrossRefPubMed Fishman EK, Magid D, Robertson DD, Brooker AF, Weiss P, Siegelman SS (1986) Metallic hip implants: CT with multiplanar reconstruction. Radiology 160(3):675–681CrossRefPubMed
9.
go back to reference Stradiotti P, Curti A, Castellazzi G, Zerbi A (2009) Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: state of the art. Eur Spine J 18(Suppl 1):102–108CrossRefPubMedPubMedCentral Stradiotti P, Curti A, Castellazzi G, Zerbi A (2009) Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: state of the art. Eur Spine J 18(Suppl 1):102–108CrossRefPubMedPubMedCentral
10.
go back to reference Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199(5 Suppl):S9–S15CrossRefPubMed Yu L, Leng S, McCollough CH (2012) Dual-energy CT-based monochromatic imaging. AJR Am J Roentgenol 199(5 Suppl):S9–S15CrossRefPubMed
11.
go back to reference Ferrara F, Cipriani A, Rapisarda S et al (2015) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRefPubMed Ferrara F, Cipriani A, Rapisarda S et al (2015) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRefPubMed
12.
go back to reference Filograna L, Magarelli N, Leone A et al (2016) Performances of low-dose dual-energy CT in reducing artifacts from implanted metallic orthopedic devices. Skeletal Radiol 45(7):937–947CrossRefPubMed Filograna L, Magarelli N, Leone A et al (2016) Performances of low-dose dual-energy CT in reducing artifacts from implanted metallic orthopedic devices. Skeletal Radiol 45(7):937–947CrossRefPubMed
13.
go back to reference Filograna L, Magarelli N, Leone A et al (2015) Value of monoenergetic dual-energy CT (DECT) for artefact reduction from metallic orthopedic implants in post-mortem studies. Skeletal Radiol 44(9):1287–1294CrossRefPubMed Filograna L, Magarelli N, Leone A et al (2015) Value of monoenergetic dual-energy CT (DECT) for artefact reduction from metallic orthopedic implants in post-mortem studies. Skeletal Radiol 44(9):1287–1294CrossRefPubMed
14.
go back to reference Stolzmann P, Leschka S, Scheffel H et al (2010) Characterization of urinary stones with dual-energy CT: improved differentiation using a tin filter. Invest Radiol 45:1–6CrossRefPubMed Stolzmann P, Leschka S, Scheffel H et al (2010) Characterization of urinary stones with dual-energy CT: improved differentiation using a tin filter. Invest Radiol 45:1–6CrossRefPubMed
15.
go back to reference Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. EurRadiol 21:1424 Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. EurRadiol 21:1424
16.
go back to reference Meinel FG, Bischoff B, Zhang Q, Bamberg F, Reiser MF, Johnson TR (2012) Metal artifact reduction by dual-energy computed tomography using energetic extrapolation: a systematically optimized protocol. Invest Radiol 47(7):406–414CrossRefPubMed Meinel FG, Bischoff B, Zhang Q, Bamberg F, Reiser MF, Johnson TR (2012) Metal artifact reduction by dual-energy computed tomography using energetic extrapolation: a systematically optimized protocol. Invest Radiol 47(7):406–414CrossRefPubMed
17.
go back to reference Zhou C, Zhao YE, Luo S et al (2011) Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures. AcadRadiol 18:1252–1257 Zhou C, Zhao YE, Luo S et al (2011) Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures. AcadRadiol 18:1252–1257
18.
go back to reference Guggenberger R, Winklhofer S, Osterhoff G et al (2012) Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion Implants from various vendors and different spine levels. EurRadiol 22:2357–2364 Guggenberger R, Winklhofer S, Osterhoff G et al (2012) Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion Implants from various vendors and different spine levels. EurRadiol 22:2357–2364
19.
go back to reference Lin XZ, Miao F, Li JY, Dong HP, Shen Y, Chen KM (2011) High-definition CT Gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr 35:294–297CrossRefPubMed Lin XZ, Miao F, Li JY, Dong HP, Shen Y, Chen KM (2011) High-definition CT Gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr 35:294–297CrossRefPubMed
Metadata
Title
Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses
Authors
Nicola Magarelli
Vincenzo De Santis
Giammaria Marziali
Amerigo Menghi
Aaron Burrofato
Luigi Pedone
Dario Del Prete
Roberto Iezzi
Chiara de Waure
Marianna D’andrea
Antonio Leone
Cesare Colosimo
Publication date
01-08-2018
Publisher
Springer Milan
Published in
La radiologia medica / Issue 8/2018
Print ISSN: 0033-8362
Electronic ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-018-0881-8

Other articles of this Issue 8/2018

La radiologia medica 8/2018 Go to the issue