Abstract

Background

The aim of this study is to see how the sutureless, stentless, Perceval S aortic valves behave when implanted in elderly patients with small aortic root and the comparison with a second group of patients with similar characteristics where a conventional stented bioprosthesis was implanted. This is a prospective randomized institutional study.

Methods

Our material is composed from 25 patients who underwent aortic valve replacement with sutureless self-anchoring Perceval S valve implantation (LivaNova), compared with 25 patients with conventional stented biological prosthesis implanted (soprano LivaNova group). The two groups of patients have similar demographic and medical characteristics with severe aortic stenosis. The study was conducted from January 2012 to June 2014. Preoperative, intraoperative and postoperative parameters were studied in order to investigate the utility of the Perceval S valves in this group of patients.

Results

The Perceval S valve implantation seems to be an interesting biological valve with good hemodynamic characteristics as compared with the typical biological prosthesis providing shorter ischemia time (40 ± 5.50 min vs 86 ± 15.86 min; p < 0.001), shorter extracorporeal circulation time (73.75 ± 8.12 min vs 120.36 ± 28.31 min p < 0.001), less operation time (149.38 ± 15.22 min vs 206.64 ± 42.85 min; p < 0.001) and better postoperative recovery. The postoperative gradients were 23.5 ± 19.20 mmHg vs 24.5 ± 19.90 mmHg respectively. The postoperative effective orifice area in these two groups were respectively 1.5 =/-0.19 cm2 vs 1.1=/-0.5 cm2 (p 0.002). Among the 25 patients of the Soprano stented valve, 3 (12 %) came back in 6 months with New York Heart Association (NYHA) 3. The PPM of these patients was the cause of readmission in the Hospital required diuresis and supplementary treatment.

Conclusions

Aortic valve replacement with Perceval aortic valves in geriatric patients with comorbidities and small aortic annulus seems to be an alternative, safe and “fast” intervention with excellent short and mid-term results which provides a better effective orifice area.
Literature
1.
go back to reference Folliguet TA, Laborde F, Zannis K, Gorayeb G, Haverich A, Shrestha M. Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers. Ann Thorac Surg. 2012;93:1483–8.CrossRefPubMed Folliguet TA, Laborde F, Zannis K, Gorayeb G, Haverich A, Shrestha M. Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers. Ann Thorac Surg. 2012;93:1483–8.CrossRefPubMed
2.
go back to reference Shrestha M, Foliiguet T, Meuris B, Dibie A, Bara C, Herregods MC, Meuris B, et al. Suturuless Perceval S aortic valve replacement: a multicenter, prospective pilot trial. J Heart Valve Dis. 2009;18:698–702.PubMed Shrestha M, Foliiguet T, Meuris B, Dibie A, Bara C, Herregods MC, Meuris B, et al. Suturuless Perceval S aortic valve replacement: a multicenter, prospective pilot trial. J Heart Valve Dis. 2009;18:698–702.PubMed
3.
go back to reference Reposini A, Kolelnikov I, Bouchikhi R, Torre T, Passaretti B, Parodi O, et al. Single suture line placement of a pericardial stentless valve. J Thorac Cardiovasc Surg. 2005;130:1265–9.CrossRef Reposini A, Kolelnikov I, Bouchikhi R, Torre T, Passaretti B, Parodi O, et al. Single suture line placement of a pericardial stentless valve. J Thorac Cardiovasc Surg. 2005;130:1265–9.CrossRef
4.
go back to reference David TE, Feindel CM, Bos J, Sun Z, Scully HE, Rakowski H. Aortic Valve replacement with a stentless porcine aortic valve. A six-year experience. J Cardiovasc Surg. 1994;108:1030–6. David TE, Feindel CM, Bos J, Sun Z, Scully HE, Rakowski H. Aortic Valve replacement with a stentless porcine aortic valve. A six-year experience. J Cardiovasc Surg. 1994;108:1030–6.
5.
go back to reference Apostolakis E, Baikoussis NG, Papakonstantinou NA, Goudevenos J. Patient-prosthesis mismatch and strategies to prevent it during aortic valve replacement. Hellenic J Cardiol. 2011;52(1):41–51.PubMed Apostolakis E, Baikoussis NG, Papakonstantinou NA, Goudevenos J. Patient-prosthesis mismatch and strategies to prevent it during aortic valve replacement. Hellenic J Cardiol. 2011;52(1):41–51.PubMed
7.
go back to reference Thierry C, Englberger L, Stalder M. Recent developments for surgical aortic valve replacement: The concept of sutureless valve technology. Switzerland: Clinic for Cardiovascular Surgery, University Hospital Berne. Volume 4. 2013. ISSN: 2075-9010. Thierry C, Englberger L, Stalder M. Recent developments for surgical aortic valve replacement: The concept of sutureless valve technology. Switzerland: Clinic for Cardiovascular Surgery, University Hospital Berne. Volume 4. 2013. ISSN: 2075-9010.
8.
go back to reference Langanay T, Flécher E, Fouquet O, Ruggieri VG, De La Tour B, Félix C, et al. Aortic valve replacement in the elderly: the real life. Ann Thorac Surg. 2010;93:70–8.CrossRef Langanay T, Flécher E, Fouquet O, Ruggieri VG, De La Tour B, Félix C, et al. Aortic valve replacement in the elderly: the real life. Ann Thorac Surg. 2010;93:70–8.CrossRef
9.
go back to reference Al-Sarraf N, Thalib L, Hughes A, Houlihan M, Tolan M, Young V, et al. Cross-clamp time is a independent predictor of mortality and morbidity in low- and high – risk cardiac patients. Int J Surg. 2011;9:104–9.CrossRefPubMed Al-Sarraf N, Thalib L, Hughes A, Houlihan M, Tolan M, Young V, et al. Cross-clamp time is a independent predictor of mortality and morbidity in low- and high – risk cardiac patients. Int J Surg. 2011;9:104–9.CrossRefPubMed
10.
go back to reference Shrestha M, Maeding I, Höffler K, Koigeldiyev N, Marsch G, Siemeni T, et al. Valve replacement in geriatric patients with small aortic roots: are sutureless valves the future? Interact Cardiovasc Thorac Surg. 2013;17(5):778–82.CrossRefPubMedPubMedCentral Shrestha M, Maeding I, Höffler K, Koigeldiyev N, Marsch G, Siemeni T, et al. Valve replacement in geriatric patients with small aortic roots: are sutureless valves the future? Interact Cardiovasc Thorac Surg. 2013;17(5):778–82.CrossRefPubMedPubMedCentral
11.
go back to reference Santarpino G, Pfeiffer S, Concistrè G, Fischlein T. A supra-annular malposition of the Perceval S sutureless aortic valve: the 'χ-movement' removal technique and subsequent reimplantation. Interact Cardiovasc Thorac Surg. 2012;15(2):280–1.CrossRefPubMedPubMedCentral Santarpino G, Pfeiffer S, Concistrè G, Fischlein T. A supra-annular malposition of the Perceval S sutureless aortic valve: the 'χ-movement' removal technique and subsequent reimplantation. Interact Cardiovasc Thorac Surg. 2012;15(2):280–1.CrossRefPubMedPubMedCentral
12.
go back to reference Santarpino G, Pfeiffer S, Concistrè G, Fischlein T. Perceval S aortic valve implantation in mini-invasive surgery: the simple sutureless solution. Interact Cardiovasc Thorac Surg. 2012;15(3):357–60.CrossRefPubMedPubMedCentral Santarpino G, Pfeiffer S, Concistrè G, Fischlein T. Perceval S aortic valve implantation in mini-invasive surgery: the simple sutureless solution. Interact Cardiovasc Thorac Surg. 2012;15(3):357–60.CrossRefPubMedPubMedCentral
13.
go back to reference Meuris B, Flameng WJ, Laborde F, Folliguet TA, Haverich A, Shrestha M. Five-year results of the pilot trial of a sutureless valve. J Thorac Cardiovasc Surg. 2015;150(1):84–8.CrossRefPubMed Meuris B, Flameng WJ, Laborde F, Folliguet TA, Haverich A, Shrestha M. Five-year results of the pilot trial of a sutureless valve. J Thorac Cardiovasc Surg. 2015;150(1):84–8.CrossRefPubMed
14.
go back to reference Zannis K, Joffre J, Czitrom D, Folliguet T, Noghin M, Lansac MN, et al. Aortic valve replacement with the perceval S bioprosthesis: single-center experience in 143 patients. J Heart Valve Dis. 2014;23(6):795–802.PubMed Zannis K, Joffre J, Czitrom D, Folliguet T, Noghin M, Lansac MN, et al. Aortic valve replacement with the perceval S bioprosthesis: single-center experience in 143 patients. J Heart Valve Dis. 2014;23(6):795–802.PubMed
15.
go back to reference Villa E, Messina A, Laborde F, Shrestha M, Troise G, Zannis K, et al. Challenge for perceval: aortic valve replacement with small sutureless valves--a multicenter study. Ann Thorac Surg. 2015;99(4):1248–54.CrossRefPubMed Villa E, Messina A, Laborde F, Shrestha M, Troise G, Zannis K, et al. Challenge for perceval: aortic valve replacement with small sutureless valves--a multicenter study. Ann Thorac Surg. 2015;99(4):1248–54.CrossRefPubMed
Metadata
Title
Aortic valve replacement in elderly with small aortic root and low body surface area; the Perceval S valve and its impact in effective orifice area
Authors
Panagiotis Dedeilias
Nikolaos G. Baikoussis
Efstathia Prappa
Dimitrios Asvestas
Michalis Argiriou
Christos Charitos
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2016
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-016-0438-7

Other articles of this Issue 1/2016

Journal of Cardiothoracic Surgery 1/2016 Go to the issue