Clinical impact of a new cone beam CT angiography respiratory motion artifact reduction algorithm during hepatic intra-arterial interventions

Authors: Marco Dioguardi Burgio, Thomas Benseghir, Vincent Roche, Carmela Garcia Alba, Jean Baptiste Debry, Annie Sibert, Valérie Vilgrain, Maxime Ronot

Published in: European Radiology | Issue 1/2020

Login to get access

Abstract

Objectives

To assess the impact of recently developed respiratory motion correction software on contrast-enhanced cone beam CT angiography (CBCT-a) for intraprocedural image guidance during intra-arterial liver-directed therapy.

Methods

From 2015 to 2017, two groups of patients who underwent intra-arterial liver-directed therapy with (breathing, n = 30) or without (still, n = 30) significant respiratory motion artifacts were retrospectively included. All CBCT-a were processed with and without dedicated respiratory motion correction software. Four readers independently assessed the following in both reconstructions (motion correction ON and OFF): (1) overall image quality on a 0-to-5 point scale, and (2) presence of relevant peri-procedural information on tumor and vasculature (overall vessel geometry, visibility of extrahepatic vessels, target tumor conspicuity, visibility of tumor feeders).

Results

Motion correction increased the average image quality in the breathing group from 2.0 ± 0.9 to 2.9 ± 1.0 (p < 0.01). The visibility of vessel geometry, extrahepatic vessels, and tumor feeders was significantly improved for all readers, and tumor conspicuity was improved for three readers. The average image quality was not significantly different between reconstructions in the still group (motion correction ON and OFF), for any of the readers (4.0 ± 0.6 vs 4.2 ± 0.6; p = 0.12). There was no change in the visibility of vessel geometry, extrahepatic vessels, tumor feeders, or tumor conspicuity for the four readers using the respiratory motion correction software in this group.

Conclusions

Using the dedicated respiratory motion correction software during intra-arterial liver-directed procedures increases the visualization of relevant peri-procedural information and image quality in CBCT-a corrupted by respiratory motion artifacts without affecting these elements in still CBCT-a.

Key Points

• The use of respiratory motion correction software could reduce the need for cone beam CT angiography acquisition retake.
• Motion correction software significantly increases the visibility of vessel geometry, extrahepatic vessels, and tumor feeders, as well as tumor conspicuity in cone beam CT angiography corrupted by respiratory motion artifacts.
• The use of respiratory motion correction software on cone beam CT angiography uncorrupted by respiratory motion artifact does not result in decreased image quality.
Appendix
Available only for authorised users
Literature
1.
go back to reference Raj S, Irani FG, Tay KH, Tan BS (2013) C-arm cone beam computed tomography: a new tool in the interventional suite. Ann Acad Med Singapore 42:585–592PubMed Raj S, Irani FG, Tay KH, Tan BS (2013) C-arm cone beam computed tomography: a new tool in the interventional suite. Ann Acad Med Singapore 42:585–592PubMed
15.
go back to reference Sindel A, Bögel M, Maier A, Fahrig R, Hornegger J, Dörfler A (2015) Respiratory motion compensation for C-arm CT liver imaging. In: Handels H, Deserno TM, Meinzer HP, Tolxdorff T (eds) Bildverarbeitung für die Medizin 2015. Springer Berlin Heidelberg, Berlin, Heidelberg, pp 221–226 Sindel A, Bögel M, Maier A, Fahrig R, Hornegger J, Dörfler A (2015) Respiratory motion compensation for C-arm CT liver imaging. In: Handels H, Deserno TM, Meinzer HP, Tolxdorff T (eds) Bildverarbeitung für die Medizin 2015. Springer Berlin Heidelberg, Berlin, Heidelberg, pp 221–226
Metadata
Title
Clinical impact of a new cone beam CT angiography respiratory motion artifact reduction algorithm during hepatic intra-arterial interventions
Authors
Marco Dioguardi Burgio
Thomas Benseghir
Vincent Roche
Carmela Garcia Alba
Jean Baptiste Debry
Annie Sibert
Valérie Vilgrain
Maxime Ronot
Publication date
01-01-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 1/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06355-w

Other articles of this Issue 1/2020

European Radiology 1/2020 Go to the issue