The Contour—Early Human Experience of a Novel Aneurysm Occlusion Device

Authors: P. Bhogal, I. Lylyk, J. Chudyk, N. Perez, C. Bleise, P. Lylyk

Published in: Clinical Neuroradiology | Issue 1/2021

Login to get access

Abstract

Background and Purpose

Endosaccular flow disruption is a recognized treatment options for treating both unruptured and ruptured aneurysms. The Contour device is designed to target the neck of an aneurysm and cause flow disruption within the aneurysm hence promoting thrombosis and neo-endothelialization at the neck. This article presents initial experiences with the Contour.

Methods

The prospectively maintained database was retrospectively reviewed to identify patients treated with the Contour device. Demographic data, aneurysm characteristics, clinical result, and clinical and radiological follow-up information were recorded.

Results

The review identified 3 patients (2 female), with 3 unruptured aneurysms, of average age 67 ± 8.7 years (range 62–77 years). The aneurysms were all located in the anterior circulation including one pericallosal, one at the A1–2 junction and one on the ICA bifurcation. The mean average dome height was 7.6 ± 0.62 mm (range 7.1–8.3 mm), dome width 5.7 ± 2 mm (range 3.5–7.5 mm), and neck width 3.6 ± 0.95 mm (range 2.5–4.2 mm). At follow-up angiography two of the aneurysms were completely occluded and one device had displaced into the aneurysm sac due to inappropriate positioning of the device. Of the patients one had minor stroke during the postoperative period but returned to baseline neurology. All patients were mRS 0 at last follow-up.

Conclusion

The Contour is a promising new aneurysm occlusion device. Further studies with longer term follow-up are required to determine the efficacy of this novel device.
Literature
1.
go back to reference Pierot L, Gubucz I, Buhk JH, Holtmannspötter M, Herbreteau D, Stockx L et al. Safety and efficacy of aneurysm treatment with the WEB: results of the WEBCAST 2 study. AJNR Am J Neuroradiol. 2017;38:1151–5.CrossRef Pierot L, Gubucz I, Buhk JH, Holtmannspötter M, Herbreteau D, Stockx L et al. Safety and efficacy of aneurysm treatment with the WEB: results of the WEBCAST 2 study. AJNR Am J Neuroradiol. 2017;38:1151–5.CrossRef
2.
go back to reference Pierot L, Costalat V, Moret J, Szikora I, Klisch J, Herbreteau D et al. Safety and efficacy of aneurysm treatment with WEB: results of the WEBCAST study. J Neurosurg. 2016;124:1250–6.CrossRef Pierot L, Costalat V, Moret J, Szikora I, Klisch J, Herbreteau D et al. Safety and efficacy of aneurysm treatment with WEB: results of the WEBCAST study. J Neurosurg. 2016;124:1250–6.CrossRef
3.
go back to reference Pierot L, Spelle L, Molyneux A, Byrne J, WEBCAST and French Observatory Investigators. Clinical and anatomical follow-up in patients with aneurysms treated with the WEB device: 1‑year follow-up report in the cumulated population of 2 prospective, multicenter series (WEBCAST and French observatory). Neurosurgery. 2016;78:133–41.CrossRef Pierot L, Spelle L, Molyneux A, Byrne J, WEBCAST and French Observatory Investigators. Clinical and anatomical follow-up in patients with aneurysms treated with the WEB device: 1‑year follow-up report in the cumulated population of 2 prospective, multicenter series (WEBCAST and French observatory). Neurosurgery. 2016;78:133–41.CrossRef
4.
go back to reference Fiorella D, Molyneux A, Coon A, Szikora I, Saatci I, Baltacioglu F et al. Demographic, procedural and 30-day safety results from the WEB intra-saccular therapy study (WEB-IT). J Neurointerv Surg. 2017;9:1191–6.CrossRef Fiorella D, Molyneux A, Coon A, Szikora I, Saatci I, Baltacioglu F et al. Demographic, procedural and 30-day safety results from the WEB intra-saccular therapy study (WEB-IT). J Neurointerv Surg. 2017;9:1191–6.CrossRef
5.
go back to reference van Rooij SBT, van Rooij WJ, Peluso JP, Sluzewski M, Bechan RS, Kortman HG et al. WEB treatment of ruptured Intracranial aneurysms: a single-center cohort of 100 patients. AJNR Am J Neuroradiol. 2017;38:2282–7.CrossRef van Rooij SBT, van Rooij WJ, Peluso JP, Sluzewski M, Bechan RS, Kortman HG et al. WEB treatment of ruptured Intracranial aneurysms: a single-center cohort of 100 patients. AJNR Am J Neuroradiol. 2017;38:2282–7.CrossRef
6.
go back to reference Lv X, Zhang Y, Jiang W. Systematic review of WEB for wide-necked bifurcation aneurysms: complications adequate occlusion rate, morbidity and mortality. World Neurosurg. 2017;110:20–5.CrossRef Lv X, Zhang Y, Jiang W. Systematic review of WEB for wide-necked bifurcation aneurysms: complications adequate occlusion rate, morbidity and mortality. World Neurosurg. 2017;110:20–5.CrossRef
7.
go back to reference Raymond J, Guilbert F, Weill A, Georganos SA, Juravsky L, Lambert A et al. Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke. 2003;34:1398–403.CrossRef Raymond J, Guilbert F, Weill A, Georganos SA, Juravsky L, Lambert A et al. Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke. 2003;34:1398–403.CrossRef
8.
go back to reference Wiebers DO, Whisnant JP, Huston J 3rd, Meissner I, Brown RD Jr, Piepgras DG et al. Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet. 2003;362:103–10.CrossRef Wiebers DO, Whisnant JP, Huston J 3rd, Meissner I, Brown RD Jr, Piepgras DG et al. Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet. 2003;362:103–10.CrossRef
9.
go back to reference Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J et al. International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet. 2002;360:1267–74.CrossRef Molyneux A, Kerr R, Stratton I, Sandercock P, Clarke M, Shrimpton J et al. International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet. 2002;360:1267–74.CrossRef
10.
go back to reference Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA et al. International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet. 2005;366:809–17.CrossRef Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA et al. International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet. 2005;366:809–17.CrossRef
11.
go back to reference Kadirvel R, Ding YH, Dai D, Rezek I, Lewis DA, Kallmes DF. Cellular mechanisms of aneurysm occlusion after treatment with a flow diverter. Radiology. 2014;270:394–9.CrossRef Kadirvel R, Ding YH, Dai D, Rezek I, Lewis DA, Kallmes DF. Cellular mechanisms of aneurysm occlusion after treatment with a flow diverter. Radiology. 2014;270:394–9.CrossRef
13.
go back to reference Ding YH, Dai D, Schroeder D, Kadirvel R, Kallmes DF. Experimental testing of the dual-layer woven endobridge device using an elastase-induced aneurysm model in rabbits. Interv Neuroradiol. 2016;22:299–303.CrossRef Ding YH, Dai D, Schroeder D, Kadirvel R, Kallmes DF. Experimental testing of the dual-layer woven endobridge device using an elastase-induced aneurysm model in rabbits. Interv Neuroradiol. 2016;22:299–303.CrossRef
14.
go back to reference Ding YH, Lewis DA, Kadirvel R, Dai D, Kallmes DF. The woven endobridge: a new aneurysm occlusion device. AJNR Am J Neuroradiol. 2011;32:607–11.CrossRef Ding YH, Lewis DA, Kadirvel R, Dai D, Kallmes DF. The woven endobridge: a new aneurysm occlusion device. AJNR Am J Neuroradiol. 2011;32:607–11.CrossRef
15.
go back to reference Turk AS, Maia O, Ferreira CC, Freitas D, Mocco J, Hanel R. Periprocedural safety of aneurysm embolization with the Medina coil system: the early human experience. J Neurointerv Surg. 2016;8:168–72.CrossRef Turk AS, Maia O, Ferreira CC, Freitas D, Mocco J, Hanel R. Periprocedural safety of aneurysm embolization with the Medina coil system: the early human experience. J Neurointerv Surg. 2016;8:168–72.CrossRef
16.
go back to reference Aguilar Perez M, Bhogal P, Martinez Moreno R, Bäzner H, Ganslandt O, Henkes H. The Medina embolic device: early clinical experience from a single center. J Neurointerv Surg. 2017;9:77–87.CrossRef Aguilar Perez M, Bhogal P, Martinez Moreno R, Bäzner H, Ganslandt O, Henkes H. The Medina embolic device: early clinical experience from a single center. J Neurointerv Surg. 2017;9:77–87.CrossRef
17.
go back to reference Bhogal P, Brouwer PA, Yeo L, Svensson M, Söderman M. The Medina embolic device: Karolinska experience. Interv Neuroradiol. 2018;24:4–13.CrossRef Bhogal P, Brouwer PA, Yeo L, Svensson M, Söderman M. The Medina embolic device: Karolinska experience. Interv Neuroradiol. 2018;24:4–13.CrossRef
18.
go back to reference Sourour NA, Vande Perre S, Maria FD, Papagiannaki C, Gabrieli J, Pistocchi S et al. Medina® embolization device for the treatment of Intracranial aneurysms: safety and angiographic effectiveness at 6 months. Neurosurgery. 2018;82:155–62.CrossRef Sourour NA, Vande Perre S, Maria FD, Papagiannaki C, Gabrieli J, Pistocchi S et al. Medina® embolization device for the treatment of Intracranial aneurysms: safety and angiographic effectiveness at 6 months. Neurosurgery. 2018;82:155–62.CrossRef
19.
go back to reference Frölich AM, Nawka MT, Ernst M, Frischmuth I, Fiehler J, Buhk JH. Intra-aneurysmal flow disruption after implantation of the Medina® embolization device depends on aneurysm neck coverage. PLoS ONE. 2018;13:e191975.CrossRef Frölich AM, Nawka MT, Ernst M, Frischmuth I, Fiehler J, Buhk JH. Intra-aneurysmal flow disruption after implantation of the Medina® embolization device depends on aneurysm neck coverage. PLoS ONE. 2018;13:e191975.CrossRef
20.
go back to reference Clajus C, Strasilla C, Fiebig T, Sychra V, Fiorella D, Klisch J. Initial and mid-term results from 108 consecutive patients with cerebral aneurysms treated with the WEB device. J Neurointerv Surg. 2017;9:411–7.CrossRef Clajus C, Strasilla C, Fiebig T, Sychra V, Fiorella D, Klisch J. Initial and mid-term results from 108 consecutive patients with cerebral aneurysms treated with the WEB device. J Neurointerv Surg. 2017;9:411–7.CrossRef
21.
go back to reference Lescher S, du Mesnil de Rochemont R, Berkefeld J. Woven endobridge (WEB) device for endovascular treatment of complex unruptured aneurysms—a single center experience. Neuroradiology. 2016;58:383–90.CrossRef Lescher S, du Mesnil de Rochemont R, Berkefeld J. Woven endobridge (WEB) device for endovascular treatment of complex unruptured aneurysms—a single center experience. Neuroradiology. 2016;58:383–90.CrossRef
22.
go back to reference Lawson A, Molyneux A, Sellar R, Lamin S, Thomas A, Gholkar A et al. Safety results from the treatment of 109 cerebral aneurysms using the woven endobridge technique: preliminary results in the united kingdom. J Neurosurg. 2018;128:144–53.CrossRef Lawson A, Molyneux A, Sellar R, Lamin S, Thomas A, Gholkar A et al. Safety results from the treatment of 109 cerebral aneurysms using the woven endobridge technique: preliminary results in the united kingdom. J Neurosurg. 2018;128:144–53.CrossRef
23.
go back to reference Popielski J, Berlis A, Weber W, Fischer S. Two-center experience in the endovascular treatment of ruptured and unruptured intracranial aneurysms using the WEB device: a retrospective analysis. AJNR Am J Neuroradiol. 2018;39:111–7.CrossRef Popielski J, Berlis A, Weber W, Fischer S. Two-center experience in the endovascular treatment of ruptured and unruptured intracranial aneurysms using the WEB device: a retrospective analysis. AJNR Am J Neuroradiol. 2018;39:111–7.CrossRef
24.
go back to reference Caroff J, Mihalea C, Dargento F, Neki H, Ikka L, Benachour N et al. Woven endobridge (WEB) device for endovascular treatment of ruptured intracranial wide-neck aneurysms: a single-center experience. Neuroradiology. 2014;56:755–61.CrossRef Caroff J, Mihalea C, Dargento F, Neki H, Ikka L, Benachour N et al. Woven endobridge (WEB) device for endovascular treatment of ruptured intracranial wide-neck aneurysms: a single-center experience. Neuroradiology. 2014;56:755–61.CrossRef
25.
go back to reference Behme D, Berlis A, Weber W. Woven endobridge Intrasaccular flow disrupter for the treatment of ruptured and unruptured wide-neck cerebral aneurysms: report of 55 cases. AJNR Am J Neuroradiol. 2015;36:1501–6.CrossRef Behme D, Berlis A, Weber W. Woven endobridge Intrasaccular flow disrupter for the treatment of ruptured and unruptured wide-neck cerebral aneurysms: report of 55 cases. AJNR Am J Neuroradiol. 2015;36:1501–6.CrossRef
26.
go back to reference Herbreteau D, Bibi R, Narata AP, Janot K, Papagiannaki C, Soize S et al. Are anatomic results influenced by WEB shape modification? Analysis in a prospective, single-center series of 39 patients with aneurysms treated with the WEB. AJNR Am J Neuroradiol. 2016;37:2280–6.CrossRef Herbreteau D, Bibi R, Narata AP, Janot K, Papagiannaki C, Soize S et al. Are anatomic results influenced by WEB shape modification? Analysis in a prospective, single-center series of 39 patients with aneurysms treated with the WEB. AJNR Am J Neuroradiol. 2016;37:2280–6.CrossRef
27.
go back to reference Limbucci N, Leone G, Rosi A, Consoli A, Renieri L, Laiso A et al. Endovascular treatment of unruptured Intracranial aneurysms by the woven endobridge device (WEB): are there any aspects influencing aneurysm occlusion? World Neurosurg. 2018;109:e183–93.CrossRef Limbucci N, Leone G, Rosi A, Consoli A, Renieri L, Laiso A et al. Endovascular treatment of unruptured Intracranial aneurysms by the woven endobridge device (WEB): are there any aspects influencing aneurysm occlusion? World Neurosurg. 2018;109:e183–93.CrossRef
28.
go back to reference Cognard C, Januel AC. Remnants and recurrences after the use of the WEB intrasaccular device in large-neck bifurcation aneurysms. Neurosurgery. 2015;76:522–30. discussion 530.CrossRef Cognard C, Januel AC. Remnants and recurrences after the use of the WEB intrasaccular device in large-neck bifurcation aneurysms. Neurosurgery. 2015;76:522–30. discussion 530.CrossRef
Metadata
Title
The Contour—Early Human Experience of a Novel Aneurysm Occlusion Device
Authors
P. Bhogal
I. Lylyk
J. Chudyk
N. Perez
C. Bleise
P. Lylyk
Publication date
01-03-2021
Publisher
Springer Berlin Heidelberg
Published in
Clinical Neuroradiology / Issue 1/2021
Print ISSN: 1869-1439
Electronic ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-020-00876-4

Other articles of this Issue 1/2021

Clinical Neuroradiology 1/2021 Go to the issue