Abstract

Introduction

The purpose of this study was to compare the anatomical fit of different, precontoured palmar distal radius plates.

Methods

The anatomical fit of seven different types of palmar distal radius plates [Königsee variable fixed-angle radius plate 7/3-hole, Königsee variable fixed-angle radius plate 5/3-hole (Allendorf, Germany), Medartis 2.5 Adaptive TriLock, Medartis 2.5 TriLock, Medartis 2.5 TriLock extraarticular, (Basel, Switzerland), Synthes VA-LCP distal two-column-radius, Synthes LCP extraarticular (Bettlach, Switzerland)] were investigated in 25 embalmed human cadaveric radii. An imprint of the space between the well-positioned plate and the distal radius was attained using a silicone mass and the maximum height of the silicone imprint was digitally measured. The mean maximum imprint height was compared between the seven plates using an analysis of variance with repeated measures and Bonferroni correction for multiple comparisons.

Results

The mean maximum distance between the plates and the radial cortex was <2 mm for all plates. The greatest difference was found with the Medartis Adaptive (1.99 ± 0.45 mm) and the least difference with the Synthes two-column (1.56 ± 0.76 mm), this difference being statistically significant (p = 0.005).

Conclusion

Although there was no complete congruency between the plates and the radial cortex, all distal palmar radius plates investigated in this study presented a reasonable anatomical shape. The Synthes VA-LCP distal two-column-radius plate palmar showed the best anatomical fit. A low profile and optimized anatomical precontouring minimizes irritation of the surrounding soft tissues and should be considered with plate design and implant choice.
Literature
2.
3.
go back to reference Jupiter JB, Marent-Huber M (2010) Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series. Surgical technique. J Bone Joint Surg Am 92(Suppl 1 Pt 1):96–106PubMed Jupiter JB, Marent-Huber M (2010) Operative management of distal radial fractures with 2.4-millimeter locking plates: a multicenter prospective case series. Surgical technique. J Bone Joint Surg Am 92(Suppl 1 Pt 1):96–106PubMed
4.
go back to reference Gradl G, Jupiter J, Pillukat T, Knobe M, Prommersberger KJ (2013) Corrective osteotomy of the distal radius following failed internal fixation. Arch Orthop Trauma Surg 133(8):1173–1179PubMedCrossRef Gradl G, Jupiter J, Pillukat T, Knobe M, Prommersberger KJ (2013) Corrective osteotomy of the distal radius following failed internal fixation. Arch Orthop Trauma Surg 133(8):1173–1179PubMedCrossRef
5.
go back to reference Mehling I, Muller LP, Rommens PM (2012) Comparative biomechanical studies on implant systems for the treatment of distal radius fractures: what are the conclusions for clinical practice? Handchir Mikrochir Plast Chir 44(5):300–305PubMedCrossRef Mehling I, Muller LP, Rommens PM (2012) Comparative biomechanical studies on implant systems for the treatment of distal radius fractures: what are the conclusions for clinical practice? Handchir Mikrochir Plast Chir 44(5):300–305PubMedCrossRef
6.
go back to reference Lutsky K, McKeon K, Goldfarb C, Boyer M (2009) Dorsal fixation of intra-articular distal radius fractures using 2.4-mm locking plates. Tech Hand Up Extrem Surg 13(4):187–196PubMedCrossRef Lutsky K, McKeon K, Goldfarb C, Boyer M (2009) Dorsal fixation of intra-articular distal radius fractures using 2.4-mm locking plates. Tech Hand Up Extrem Surg 13(4):187–196PubMedCrossRef
7.
go back to reference Lutsky K, Boyer M, Goldfarb C (2013) Dorsal locked plate fixation of distal radius fractures. J Hand Surg Am 38(7):1414–1422PubMedCrossRef Lutsky K, Boyer M, Goldfarb C (2013) Dorsal locked plate fixation of distal radius fractures. J Hand Surg Am 38(7):1414–1422PubMedCrossRef
8.
go back to reference Tarallo L, Mugnai R, Zambianchi F, Adani R, Catani F (2013) Volar plate fixation for the treatment of distal radius fractures: analysis of adverse events. J Orthop Trauma 27(12):740–745PubMedCrossRef Tarallo L, Mugnai R, Zambianchi F, Adani R, Catani F (2013) Volar plate fixation for the treatment of distal radius fractures: analysis of adverse events. J Orthop Trauma 27(12):740–745PubMedCrossRef
9.
go back to reference Soong M, van Leerdam R, Guitton TG et al (2011) Fracture of the distal radius: risk factors for complications after locked volar plate fixation. J Hand Surg Am 36(1):3–9PubMedCrossRef Soong M, van Leerdam R, Guitton TG et al (2011) Fracture of the distal radius: risk factors for complications after locked volar plate fixation. J Hand Surg Am 36(1):3–9PubMedCrossRef
10.
go back to reference Kitay A, Swanstrom M, Schreiber JJ et al (2013) Volar plate position and flexor tendon rupture following distal radius fracture fixation. J Hand Surg Am 38(6):1091–1096PubMedCrossRef Kitay A, Swanstrom M, Schreiber JJ et al (2013) Volar plate position and flexor tendon rupture following distal radius fracture fixation. J Hand Surg Am 38(6):1091–1096PubMedCrossRef
11.
go back to reference Souer JS, Ring D, Matschke S et al (2010) Comparison of functional outcome after volar plate fixation with 2.4-mm titanium versus 3.5-mm stainless-steel plate for extra-articular fracture of distal radius. J Hand Surg Am 35(3):398–405PubMedCrossRef Souer JS, Ring D, Matschke S et al (2010) Comparison of functional outcome after volar plate fixation with 2.4-mm titanium versus 3.5-mm stainless-steel plate for extra-articular fracture of distal radius. J Hand Surg Am 35(3):398–405PubMedCrossRef
12.
13.
go back to reference Kamath AF, Zurakowski D, Day CS (2006) Low-profile dorsal plating for dorsally angulated distal radius fractures: an outcomes study. J Hand Surg Am 31(7):1061–1067PubMedCrossRef Kamath AF, Zurakowski D, Day CS (2006) Low-profile dorsal plating for dorsally angulated distal radius fractures: an outcomes study. J Hand Surg Am 31(7):1061–1067PubMedCrossRef
14.
go back to reference Rozental TD, Beredjiklian PK, Bozentka DJ (2003) Functional outcome and complications following two types of dorsal plating for unstable fractures of the distal part of the radius. J Bone Joint Surg Am 85-A(10):1956–1960PubMed Rozental TD, Beredjiklian PK, Bozentka DJ (2003) Functional outcome and complications following two types of dorsal plating for unstable fractures of the distal part of the radius. J Bone Joint Surg Am 85-A(10):1956–1960PubMed
15.
go back to reference Yu YR, Makhni MC, Tabrizi S et al (2011) Complications of low-profile dorsal versus volar locking plates in the distal radius: a comparative study. J Hand Surg Am 36(7):1135–1141PubMedCrossRef Yu YR, Makhni MC, Tabrizi S et al (2011) Complications of low-profile dorsal versus volar locking plates in the distal radius: a comparative study. J Hand Surg Am 36(7):1135–1141PubMedCrossRef
16.
go back to reference Rikli DA, Regazzoni P (1996) Fractures of the distal end of the radius treated by internal fixation and early function. A preliminary report of 20 cases. J Bone Joint Surg Br 78(4):588–592PubMed Rikli DA, Regazzoni P (1996) Fractures of the distal end of the radius treated by internal fixation and early function. A preliminary report of 20 cases. J Bone Joint Surg Br 78(4):588–592PubMed
17.
go back to reference Fowler JR, Ilyas AM (2013) Prospective evaluation of distal radius fractures treated with variable-angle volar locking plates. J Hand Surg Am 38(11):2198–2203PubMedCrossRef Fowler JR, Ilyas AM (2013) Prospective evaluation of distal radius fractures treated with variable-angle volar locking plates. J Hand Surg Am 38(11):2198–2203PubMedCrossRef
18.
go back to reference Kainz H, Dall’ara E, Antoni A et al (2013) Calcium phosphate cement augmentation after volar locking plating of distal radius fracture significantly increases stability. Eur J Orthop Surg Traumatol 24(6):869–875PubMedCrossRef Kainz H, Dall’ara E, Antoni A et al (2013) Calcium phosphate cement augmentation after volar locking plating of distal radius fracture significantly increases stability. Eur J Orthop Surg Traumatol 24(6):869–875PubMedCrossRef
19.
go back to reference Högel F, Mair S, Eberle S et al (2013) Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model. Arch Orthop Trauma Surg 133(1):51–57PubMedCrossRef Högel F, Mair S, Eberle S et al (2013) Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model. Arch Orthop Trauma Surg 133(1):51–57PubMedCrossRef
20.
go back to reference Orbay JL, Touhami A (2006) Current concepts in volar fixed-angle fixation of unstable distal radius fractures. Clin Orthop Relat Res 445:58–67PubMed Orbay JL, Touhami A (2006) Current concepts in volar fixed-angle fixation of unstable distal radius fractures. Clin Orthop Relat Res 445:58–67PubMed
21.
go back to reference Hershman SH, Immerman I, Bechtel C et al (2013) The effects of pronator quadratus repair on outcomes after volar plating of distal radius fractures. J Orthop Trauma 27(3):130–133PubMedCrossRef Hershman SH, Immerman I, Bechtel C et al (2013) The effects of pronator quadratus repair on outcomes after volar plating of distal radius fractures. J Orthop Trauma 27(3):130–133PubMedCrossRef
22.
go back to reference Brown EN, Lifchez SD (2011) Flexor pollicis longus tendon rupture after volar plating of a distal radius fracture: pronator quadratus plate coverage may not adequately protect tendons. Eplasty 11:461–466 Brown EN, Lifchez SD (2011) Flexor pollicis longus tendon rupture after volar plating of a distal radius fracture: pronator quadratus plate coverage may not adequately protect tendons. Eplasty 11:461–466
23.
go back to reference Tosti R, Ilyas AM (2013) Prospective evaluation of pronator quadratus repair following volar plate fixation of distal radius fractures. J Hand Surg Am 38(9):1678–1684PubMedCrossRef Tosti R, Ilyas AM (2013) Prospective evaluation of pronator quadratus repair following volar plate fixation of distal radius fractures. J Hand Surg Am 38(9):1678–1684PubMedCrossRef
24.
go back to reference Buzzel JE, Weikert DR, Watson JT, Lee DH (2008) Precontoured fixed-angle volar distal radius plates: a comparison of anatomic fit. J Hand Surg Am 33(7):1144–1152CrossRef Buzzel JE, Weikert DR, Watson JT, Lee DH (2008) Precontoured fixed-angle volar distal radius plates: a comparison of anatomic fit. J Hand Surg Am 33(7):1144–1152CrossRef
25.
go back to reference Imatani J, Akita K, Yamaguchi K et al (2012) An anatomical study of the watershed line on the volar, distal aspect of the radius: implications for plate placement and avoidance of tendon ruptures. J Hand Surg Am 37(8):1550–1554PubMedCrossRef Imatani J, Akita K, Yamaguchi K et al (2012) An anatomical study of the watershed line on the volar, distal aspect of the radius: implications for plate placement and avoidance of tendon ruptures. J Hand Surg Am 37(8):1550–1554PubMedCrossRef
26.
go back to reference Limthongthang R, Bachoura A, Jacoby SM, Osterman AL (2014) Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus. J Hand Surg Am 39(5):852–860PubMedCrossRef Limthongthang R, Bachoura A, Jacoby SM, Osterman AL (2014) Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus. J Hand Surg Am 39(5):852–860PubMedCrossRef
27.
go back to reference Klug RA, Press CM, Gonzalez MH (2007) Rupture of the flexor pollicis longus tendon after volar fixed-angle plating of a distal radius fracture: a case report. J Hand Surg Am 32(7):984–988PubMedCrossRef Klug RA, Press CM, Gonzalez MH (2007) Rupture of the flexor pollicis longus tendon after volar fixed-angle plating of a distal radius fracture: a case report. J Hand Surg Am 32(7):984–988PubMedCrossRef
28.
go back to reference Cross AW, Schmidt CC (2008) Flexor tendon injuries following locked volar plating of distal radius fractures. J Hand Surg Am 33(2):164–167PubMedCrossRef Cross AW, Schmidt CC (2008) Flexor tendon injuries following locked volar plating of distal radius fractures. J Hand Surg Am 33(2):164–167PubMedCrossRef
29.
go back to reference Arora R, Lutz M, Hennerbichler A et al (2007) Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma 21(5):316–322PubMedCrossRef Arora R, Lutz M, Hennerbichler A et al (2007) Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma 21(5):316–322PubMedCrossRef
30.
go back to reference Esenwein P, Sonderegger J, Gruenert J et al (2013) Complications following palmar plate fixation of distal radius fractures: a review of 665 cases. Arch Orthop Trauma Surg 133(8):1155–1162PubMedCrossRef Esenwein P, Sonderegger J, Gruenert J et al (2013) Complications following palmar plate fixation of distal radius fractures: a review of 665 cases. Arch Orthop Trauma Surg 133(8):1155–1162PubMedCrossRef
31.
go back to reference Beck JD, Harness NG, Spencer HT (2014) Volar plate fixation failure for volar shearing distal radius fractures with small lunate facet fragments. J Hand Surg Am 39(4):670–678PubMedCrossRef Beck JD, Harness NG, Spencer HT (2014) Volar plate fixation failure for volar shearing distal radius fractures with small lunate facet fragments. J Hand Surg Am 39(4):670–678PubMedCrossRef
32.
go back to reference Bassi RS, Krishnan KM, Dhillon SS, Deshmukh SC (2003) Palmar cortical angle of the distal radius: a radiological study. J Hand Surg Br 28(2):163–164PubMedCrossRef Bassi RS, Krishnan KM, Dhillon SS, Deshmukh SC (2003) Palmar cortical angle of the distal radius: a radiological study. J Hand Surg Br 28(2):163–164PubMedCrossRef
33.
go back to reference Gasse N, Lepage D, Pem R et al (2011) Anatomical and radiological study applied to distal radius surgery. Arch Orthop Trauma Surg 33(6):485–490 Gasse N, Lepage D, Pem R et al (2011) Anatomical and radiological study applied to distal radius surgery. Arch Orthop Trauma Surg 33(6):485–490
34.
go back to reference Evans S, Ramasamy A, Deshmukh SC (2014) Distal volar radial plates: How anatomical are they? Orthop Traumatol Surg Res 100(3):293–295PubMedCrossRef Evans S, Ramasamy A, Deshmukh SC (2014) Distal volar radial plates: How anatomical are they? Orthop Traumatol Surg Res 100(3):293–295PubMedCrossRef
35.
go back to reference Schneppendahl J, Windolf J, Kaufmann R (2012) Distal radius fractures: current concepts. J Hand Surg Am 37(8):1718–1725PubMedCrossRef Schneppendahl J, Windolf J, Kaufmann R (2012) Distal radius fractures: current concepts. J Hand Surg Am 37(8):1718–1725PubMedCrossRef
36.
go back to reference Matschke S, Wentzensen A, Ring D et al (2011) Comparison of angle stable plate fixation approaches for distal radius fractures. Injury 42(4):385–392PubMedCrossRef Matschke S, Wentzensen A, Ring D et al (2011) Comparison of angle stable plate fixation approaches for distal radius fractures. Injury 42(4):385–392PubMedCrossRef
37.
go back to reference Lerch S, Sextro HG, Wilken F, Wittenberg CE (2009) Clinical and radiological results after distal radius fracture: intramedullary locking nail versus volar locking plate osteosynthesis. Z Orthop Unfall 147(5):547–552PubMed Lerch S, Sextro HG, Wilken F, Wittenberg CE (2009) Clinical and radiological results after distal radius fracture: intramedullary locking nail versus volar locking plate osteosynthesis. Z Orthop Unfall 147(5):547–552PubMed
38.
go back to reference Figl M, Weninger P, Liska M, Hofbauer M, Leixnering M (2009) Volar fixed-angle plate osteosynthesis of unstable distal radius fractures: 12 months results. Arch Orthop Trauma Surg 129(5):661–669PubMedCrossRef Figl M, Weninger P, Liska M, Hofbauer M, Leixnering M (2009) Volar fixed-angle plate osteosynthesis of unstable distal radius fractures: 12 months results. Arch Orthop Trauma Surg 129(5):661–669PubMedCrossRef
39.
go back to reference Mehling I, Meier M, Schlör U, Krimmer H (2007) Multidirectional palmar fixed-angle plate fixation for unstable distal radius fracture. Handchir Mikrochir Plast Chir 39(1):29–33PubMedCrossRef Mehling I, Meier M, Schlör U, Krimmer H (2007) Multidirectional palmar fixed-angle plate fixation for unstable distal radius fracture. Handchir Mikrochir Plast Chir 39(1):29–33PubMedCrossRef
40.
go back to reference Drobetz H, Weninger P, Grant C et al (2013) More is not necessarily better. A biomechanical study on distal screw numbers in volar locking distal radius plates. Injury 44(4):535–539PubMedCrossRef Drobetz H, Weninger P, Grant C et al (2013) More is not necessarily better. A biomechanical study on distal screw numbers in volar locking distal radius plates. Injury 44(4):535–539PubMedCrossRef
Metadata
Title
Anatomical fit of seven different palmar distal radius plates
Authors
Johannes Oppermann
Max Wacker
Gregor Stein
Hans-Philipp Springorum
Wolfram Friedrich Neiss
Klaus J. Burkhart
Peer Eysel
Jens Dargel
Publication date
01-10-2014
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 10/2014
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-014-2072-y

Other articles of this Issue 10/2014

Archives of Orthopaedic and Trauma Surgery 10/2014 Go to the issue