01-04-2019 | Knee Arthroplasty

An accelerometer-based navigation did not improve the femoral component positioning compared to a modified conventional technique of pre-operatively planned placement of intramedullary rod in total knee arthroplasty

Authors: Tadashi Tsukeoka, Yoshikazu Tsuneizumi, Kensuke Yoshino

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 4/2019

Login to get access

Abstract

Introduction

Although the most commonly used method of femoral component alignment in total knee arthroplasty (TKA) is intramedullary (IM) guides, this method demonstrated a limited degree of accuracy. Because of the femoral anterior bowing, the tip of the guide rod will impinge on the anterior cortex if a long rod is inserted. We hypothesized that the pre-operative planned insertion depth of the rod could increase the accuracy of the femoral component positioning in conventional TKA (modified conventional technique). Accelerometer-based, portable navigation device has been postulated to have better accuracy than conventional TKA in component positioning. The purpose of this study was to compare the post-operative femoral component alignment of TKA using the modified conventional technique with the accelerometer-based navigation.

Materials and methods

Fifty-five knees underwent TKA using the modified conventional technique and femoral component positioning was compared with 55 knees performed using the accelerometer-based navigation device. The femoral component alignment was evaluated with a CT-based three-dimensional software.

Results

The mean absolute deviation from targeted alignment in the sagittal plane was significantly less in the modified conventional cohort than in the accelerometer-based navigation cohort (1.1° vs 2.6°, P < 0.001). In the modified conventional cohort, 96.4% had an alignment within 3° of a targeted angle in the coronal plane (vs 89.1% with the accelerometer-based navigation, P = 0.14), and 96.4% in the sagittal plane (vs 74.5% with the accelerometer-based navigation, P < 0.001).

Conclusion

The modified conventional technique is a simple and equal to or more accurate method than the accelerometer-based navigation in positioning the femoral component in TKA at a mid-volume hospital.
Literature
1.
go back to reference Gilbert BM (1976) Anterior femoral curvature: its probably basis and utility as a criterion of racial assessment. Am J Phys Anthropol 45:6014CrossRef Gilbert BM (1976) Anterior femoral curvature: its probably basis and utility as a criterion of racial assessment. Am J Phys Anthropol 45:6014CrossRef
2.
go back to reference Harper MC, Carson WL (1987) Curvature of the femur and the proximal entry point for an intramedullary rod. Clin Orthop 220:155–161 Harper MC, Carson WL (1987) Curvature of the femur and the proximal entry point for an intramedullary rod. Clin Orthop 220:155–161
3.
go back to reference Singh SP, Singh S (1972) A study of anterior femoral curvature in Indian subjects. Acta Anat 83:416–425CrossRefPubMed Singh SP, Singh S (1972) A study of anterior femoral curvature in Indian subjects. Acta Anat 83:416–425CrossRefPubMed
4.
go back to reference Trudell MB (1999) Anterior femoral curvature revisited: race assessment from the femur. J Forensic Sci 44:700–707CrossRefPubMed Trudell MB (1999) Anterior femoral curvature revisited: race assessment from the femur. J Forensic Sci 44:700–707CrossRefPubMed
5.
go back to reference Tang WM, Chiu KY, Kwan MF, Ng TP, Yau WP (2005) Sagittal bowing of the distal femur in Chinese patients who require total knee arthroplasty. J Orthop Res 23(1):41–45CrossRefPubMed Tang WM, Chiu KY, Kwan MF, Ng TP, Yau WP (2005) Sagittal bowing of the distal femur in Chinese patients who require total knee arthroplasty. J Orthop Res 23(1):41–45CrossRefPubMed
6.
go back to reference Kai S, Sato T, Kog Y, Ariumi A, Tanabe Y, Seitoku N (2008) Accuracy examination for 3-D controlled insertion of intramedullary alignment rod in total knee arthroplasty. Jpn J Clin Biomech 29:285–291 Kai S, Sato T, Kog Y, Ariumi A, Tanabe Y, Seitoku N (2008) Accuracy examination for 3-D controlled insertion of intramedullary alignment rod in total knee arthroplasty. Jpn J Clin Biomech 29:285–291
7.
go back to reference Shi X, Li H, Zhou Z, Shen B, Yang J, Pei F (2016) Comparison of postoperative alignment using fixed vs individual valgus correction angle in primary total knee arthroplasty with lateral bowing. FemurJ Arthroplasty 31(5):976–983CrossRef Shi X, Li H, Zhou Z, Shen B, Yang J, Pei F (2016) Comparison of postoperative alignment using fixed vs individual valgus correction angle in primary total knee arthroplasty with lateral bowing. FemurJ Arthroplasty 31(5):976–983CrossRef
8.
go back to reference Abdelaal AH, Yamamoto N, Hayashi K, Takeuchi A, Morsy AF, Miwa S, Kajino Y, Rubio DA, Tsuchiya H (2016) Radiological assessment of the femoral bowing in Japanese population. SICOT J 2:2CrossRefPubMedPubMedCentral Abdelaal AH, Yamamoto N, Hayashi K, Takeuchi A, Morsy AF, Miwa S, Kajino Y, Rubio DA, Tsuchiya H (2016) Radiological assessment of the femoral bowing in Japanese population. SICOT J 2:2CrossRefPubMedPubMedCentral
9.
go back to reference Kim JM, Hong SH, Kim JM, Lee BS, Kim DE, Kim KA, Bin SI (2015) Femoral shaft bowing in the coronal plane has more significant effect on the coronal alignment of TKA than proximal or distal variations of femoral shape. Knee Surg Sports Traumatol Arthrosc 23(7):1936–1942CrossRefPubMed Kim JM, Hong SH, Kim JM, Lee BS, Kim DE, Kim KA, Bin SI (2015) Femoral shaft bowing in the coronal plane has more significant effect on the coronal alignment of TKA than proximal or distal variations of femoral shape. Knee Surg Sports Traumatol Arthrosc 23(7):1936–1942CrossRefPubMed
10.
go back to reference Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22(8):1097–1106CrossRefPubMed Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22(8):1097–1106CrossRefPubMed
11.
go back to reference Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20(7):1307–1322CrossRefPubMed Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20(7):1307–1322CrossRefPubMed
12.
go back to reference Mitsuhashi S, Akamatsu Y, Kobayashi H, Kusayama Y, Kumagai K, Saito T (2018) Combined CT-based and image-free navigation systems in TKA reduces postoperative outliers of rotational alignment of the tibial component. Arch Orthop Trauma Surg 138(2):259–266CrossRefPubMed Mitsuhashi S, Akamatsu Y, Kobayashi H, Kusayama Y, Kumagai K, Saito T (2018) Combined CT-based and image-free navigation systems in TKA reduces postoperative outliers of rotational alignment of the tibial component. Arch Orthop Trauma Surg 138(2):259–266CrossRefPubMed
13.
go back to reference Nam D, Nawabi DH, Cross MB, Heyse TJ, Mayman DJ (2012) Accelerometer-based computer navigation for performing the distal femoral resection in total knee arthroplasty. J Arthroplasty 27(9):1717–1722CrossRefPubMed Nam D, Nawabi DH, Cross MB, Heyse TJ, Mayman DJ (2012) Accelerometer-based computer navigation for performing the distal femoral resection in total knee arthroplasty. J Arthroplasty 27(9):1717–1722CrossRefPubMed
14.
go back to reference Nam D, Weeks KD, Reinhardt KR et al (2013) Accelerometer-based, portable navigation vs imageless, large-console computer-assisted navigation in total knee arthroplasty: a comparison of radiographic results. J Arthroplasty 28(2):255–261CrossRefPubMed Nam D, Weeks KD, Reinhardt KR et al (2013) Accelerometer-based, portable navigation vs imageless, large-console computer-assisted navigation in total knee arthroplasty: a comparison of radiographic results. J Arthroplasty 28(2):255–261CrossRefPubMed
15.
go back to reference Huang EH, Copp SN, Bugbee WD (2015) Accuracy of a handheld accelerometer-based navigation system for femoral and tibial resection in total knee arthroplasty. J Arthroplasty 30(11):1906–1910CrossRefPubMed Huang EH, Copp SN, Bugbee WD (2015) Accuracy of a handheld accelerometer-based navigation system for femoral and tibial resection in total knee arthroplasty. J Arthroplasty 30(11):1906–1910CrossRefPubMed
16.
go back to reference Steinhaus ME, McLawhorn AS, Richardson SS, Maher P, Mayman DJ (2016) Handheld navigation device and patient-specific cutting guides result in similar coronal alignment for primary total knee arthroplasty: a retrospective matched cohort study. HSS J 12(3):224–234CrossRefPubMedPubMedCentral Steinhaus ME, McLawhorn AS, Richardson SS, Maher P, Mayman DJ (2016) Handheld navigation device and patient-specific cutting guides result in similar coronal alignment for primary total knee arthroplasty: a retrospective matched cohort study. HSS J 12(3):224–234CrossRefPubMedPubMedCentral
17.
go back to reference Fujimoto E, Sasashige Y, Nakata K, Yokota G, Omoto T, Ochi M (2017) Technical considerations and accuracy improvement of accelerometer-based portable computer navigation for performing distal femoral resection in total knee arthroplasty. J Arthroplasty 32(1):53–60CrossRefPubMed Fujimoto E, Sasashige Y, Nakata K, Yokota G, Omoto T, Ochi M (2017) Technical considerations and accuracy improvement of accelerometer-based portable computer navigation for performing distal femoral resection in total knee arthroplasty. J Arthroplasty 32(1):53–60CrossRefPubMed
18.
go back to reference Gharaibeh MA, Solayar GN, Harris IA, Chen DB, MacDessi SJ (2017) Accelerometer-based, portable navigation (kneealign) vs conventional instrumentation for total knee arthroplasty: a prospective randomized comparative trial. J Arthroplasty 32(3):777–782CrossRefPubMed Gharaibeh MA, Solayar GN, Harris IA, Chen DB, MacDessi SJ (2017) Accelerometer-based, portable navigation (kneealign) vs conventional instrumentation for total knee arthroplasty: a prospective randomized comparative trial. J Arthroplasty 32(3):777–782CrossRefPubMed
19.
go back to reference Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Birkenbach AM, Grifka J, Keshmiri A (2016) Appropriate sagittal femoral component alignment cannot be ensured by intramedullary alignment rods. Knee Surg Sports Traumatol Arthrosc 24(8):2453–2460CrossRefPubMed Maderbacher G, Schaumburger J, Baier C, Zeman F, Springorum HR, Birkenbach AM, Grifka J, Keshmiri A (2016) Appropriate sagittal femoral component alignment cannot be ensured by intramedullary alignment rods. Knee Surg Sports Traumatol Arthrosc 24(8):2453–2460CrossRefPubMed
20.
go back to reference Loh B, Chen JY, Yew AKS, Pang HN, Tay DKJ, Chia SL, Lo NN, Yeo SJ (2017) The accuracy of a hand-held navigation system in total knee arthroplasty. Arch Orthop Trauma Surg 137(3):381–386CrossRefPubMed Loh B, Chen JY, Yew AKS, Pang HN, Tay DKJ, Chia SL, Lo NN, Yeo SJ (2017) The accuracy of a hand-held navigation system in total knee arthroplasty. Arch Orthop Trauma Surg 137(3):381–386CrossRefPubMed
Metadata
Title
An accelerometer-based navigation did not improve the femoral component positioning compared to a modified conventional technique of pre-operatively planned placement of intramedullary rod in total knee arthroplasty
Authors
Tadashi Tsukeoka
Yoshikazu Tsuneizumi
Kensuke Yoshino
Publication date
01-04-2019
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 4/2019
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-019-03147-1

Other articles of this Issue 4/2019

Archives of Orthopaedic and Trauma Surgery 4/2019 Go to the issue