Background

Reusable laryngoscopes have been reported to be superior to disposable laryngoscopes with plastic blades during emergent intubations. Surprisingly, at our institution a quality reporting system revealed a high number of equipment failures with reusable laryngoscopes in an emergency out-of-OR (operating room) setting. As recent studies indicated an improved quality of disposable laryngoscopes, we hypothesized that a thoroughly evaluated disposable laryngoscope would result in less equipment failure in an emergency out-of-OR setting.

Methods

To perform a more standardized and time efficient analysis, four distinct disposable laryngoscope blade/handle configurations were trialed during standard intubations (n = 4 × 30) in the OR by experienced anesthesia providers who completed a 6-question, Likert-scale/open-ended survey for product evaluation. The ‘best’ disposable blade was implemented in an emergency out-of-OR setting and equipment failure rates were monitored over a 3-year period.

Results

Different disposable laryngoscopes were equal regarding sturdiness, illumination and airway visualization. The laryngoscope with the highest overall score was significantly higher scored than the laryngoscope with the lowest overall score. All disposable laryngoscopes were more cost effective than the reusable ones, and the top scored laryngoscope demonstrated the highest 5-year cost-saving ($210 K). Implementation of the top scored disposable laryngoscope into an emergency out-of-OR setting reduced the equipment failure incidence from high 20s to 0.

Conclusion

Disposable laryngoscopes are cost effective and superior to reusable laryngoscopes in an emergency out-of-OR setting. We demonstrate that the implementation of a disposable laryngoscope in the emergency out-of-OR setting resulted in a near elimination of equipment related quality submissions which ultimately enhances patient safety.
Literature
1.
go back to reference Higgs A, McGrath BA, Goddard C, Rangasami J, Suntharalingam G, Gale R, et al. Guidelines for the management of tracheal intubation in critically ill adults. Br J Anaesth. 2018;120(2):323–52.CrossRef Higgs A, McGrath BA, Goddard C, Rangasami J, Suntharalingam G, Gale R, et al. Guidelines for the management of tracheal intubation in critically ill adults. Br J Anaesth. 2018;120(2):323–52.CrossRef
2.
go back to reference Cook TM, Woodall N, Harper J, Benger J. Major complications of airway management in the UK: results of the fourth National Audit Project of the Royal College of Anaesthetists and the difficult airway society. Part 2: intensive care and emergency departments. Br J Anaesth. 2011;106(5):632–42.CrossRef Cook TM, Woodall N, Harper J, Benger J. Major complications of airway management in the UK: results of the fourth National Audit Project of the Royal College of Anaesthetists and the difficult airway society. Part 2: intensive care and emergency departments. Br J Anaesth. 2011;106(5):632–42.CrossRef
3.
go back to reference Cooper JB, Newbower RS, Kitz RJ. An analysis of major errors and equipment failures in anesthesia management: considerations for prevention and detection. Anesthesiology. 1984;60(1):34–42.CrossRef Cooper JB, Newbower RS, Kitz RJ. An analysis of major errors and equipment failures in anesthesia management: considerations for prevention and detection. Anesthesiology. 1984;60(1):34–42.CrossRef
4.
go back to reference Jefferson P, Perkins V, Edwards VA, Ball DR. Problems with disposable laryngoscope blades. Anaesthesia. 2003;58(4):385–6 discussion 386.CrossRef Jefferson P, Perkins V, Edwards VA, Ball DR. Problems with disposable laryngoscope blades. Anaesthesia. 2003;58(4):385–6 discussion 386.CrossRef
5.
go back to reference Amour J, Le Manach YL, Borel M, Lenfant F, Nicolas-Robin A, Carillion A, et al. Comparison of single-use and reusable metal laryngoscope blades for orotracheal intubation during rapid sequence induction of anesthesia: a multicenter cluster randomized study. Anesthesiology. 2010;112(2):325–32.CrossRef Amour J, Le Manach YL, Borel M, Lenfant F, Nicolas-Robin A, Carillion A, et al. Comparison of single-use and reusable metal laryngoscope blades for orotracheal intubation during rapid sequence induction of anesthesia: a multicenter cluster randomized study. Anesthesiology. 2010;112(2):325–32.CrossRef
6.
go back to reference Jabre P, Leroux B, Brohon S, Penet C, Lockey D, Adnet F, et al. A comparison of plastic single-use with metallic reusable laryngoscope blades for out-of-hospital tracheal intubation. Ann Emerg Med. 2007;50(3):258–63.CrossRef Jabre P, Leroux B, Brohon S, Penet C, Lockey D, Adnet F, et al. A comparison of plastic single-use with metallic reusable laryngoscope blades for out-of-hospital tracheal intubation. Ann Emerg Med. 2007;50(3):258–63.CrossRef
7.
go back to reference Esler MD, Baines LC, Wilkinson DJ, Langford RM. Decontamination of laryngoscopes: a survey of national practice. Anaesthesia. 1999;54(6):587–92.CrossRef Esler MD, Baines LC, Wilkinson DJ, Langford RM. Decontamination of laryngoscopes: a survey of national practice. Anaesthesia. 1999;54(6):587–92.CrossRef
8.
go back to reference Walczak R, Arnold M, Grewal J, Yuan X, Suryadevara A, Marzouk H. Reusable vs disposable nasopharyngolaryngoscopes: cost analysis and resident survey. Laryngoscope Investig Otolaryngol. 2021;6(1):88–93.CrossRef Walczak R, Arnold M, Grewal J, Yuan X, Suryadevara A, Marzouk H. Reusable vs disposable nasopharyngolaryngoscopes: cost analysis and resident survey. Laryngoscope Investig Otolaryngol. 2021;6(1):88–93.CrossRef
9.
go back to reference Jabre P, Galinski M, Ricard-Hibon A, Devaud ML, Ruscev M, Kulstad E, et al. Out-of-hospital tracheal intubation with single-use versus reusable metal laryngoscope blades: a multicenter randomized controlled trial. Ann Emerg Med. 2011;57(3):225–31.CrossRef Jabre P, Galinski M, Ricard-Hibon A, Devaud ML, Ruscev M, Kulstad E, et al. Out-of-hospital tracheal intubation with single-use versus reusable metal laryngoscope blades: a multicenter randomized controlled trial. Ann Emerg Med. 2011;57(3):225–31.CrossRef
10.
go back to reference Ogrinc G, Davies L, Goodman D, Batalden P, Davidoff F, Stevens D. SQUIRE 2.0 (<em>Standards for QUality Improvement Reporting Excellence)</em>: revised publication guidelines from a detailed consensus process. BMJ Quality & Safety. 2016;25(12):986–92.CrossRef Ogrinc G, Davies L, Goodman D, Batalden P, Davidoff F, Stevens D. SQUIRE 2.0 (<em>Standards for QUality Improvement Reporting Excellence)</em>: revised publication guidelines from a detailed consensus process. BMJ Quality & Safety. 2016;25(12):986–92.CrossRef
11.
go back to reference Sherman JD, Raibley LAT, Eckelman MJ. Life cycle assessment and costing methods for device procurement: comparing reusable and single-use disposable laryngoscopes. Anesth Analg. 2018;127(2):434–43.CrossRef Sherman JD, Raibley LAT, Eckelman MJ. Life cycle assessment and costing methods for device procurement: comparing reusable and single-use disposable laryngoscopes. Anesth Analg. 2018;127(2):434–43.CrossRef
12.
go back to reference Landrigan PJ, Fuller R, Acosta NJR, Adeyi O, Arnold R, Basu NN, et al. The lancet commission on pollution and health. Lancet. 2018;391(10119):462–512.CrossRef Landrigan PJ, Fuller R, Acosta NJR, Adeyi O, Arnold R, Basu NN, et al. The lancet commission on pollution and health. Lancet. 2018;391(10119):462–512.CrossRef
13.
go back to reference Hopf HW, Chesebro BB, Sherman JD. Health care sustainability science: the next frontier for clinician scientists in anesthesiology. ASA Monitor. 2020;84(4):54–5. Hopf HW, Chesebro BB, Sherman JD. Health care sustainability science: the next frontier for clinician scientists in anesthesiology. ASA Monitor. 2020;84(4):54–5.
14.
go back to reference Rowley E, Dingwall R. The use of single-use devices in anaesthesia: balancing the risks to patient safety. Anaesthesia. 2007;62(6):569–74.CrossRef Rowley E, Dingwall R. The use of single-use devices in anaesthesia: balancing the risks to patient safety. Anaesthesia. 2007;62(6):569–74.CrossRef
Metadata
Title
Disposable laryngoscope intubation to reduce equipment failure in an emergency out of OR setting - a quality control case study
Authors
Colby G. Simmons
Tobias Eckle
Dustin Rogers
Jason D. Williams
Jason C. Brainard
Publication date
01-12-2023
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2023
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-022-01956-3

Other articles of this Issue 1/2023

BMC Anesthesiology 1/2023 Go to the issue