Accuracy of combined anteversion in image-free navigated total hip arthroplasty: stem-first or cup-first technique?

Authors: Shigeo Fukunishi, Shoji Nishio, Yuki Fujihara, Shohei Okahisa, Yu Takeda, Tomokazu Fukui, Shinichi Yoshiya

Published in: International Orthopaedics | Issue 1/2016

Login to get access

Abstract

Purpose

In total hip arthroplasty (THA), combined anteversion (CA) is used as a parameter for assessment of overall prosthetic alignment. The purpose of this study was to comparatively examine the CA value in patients who underwent primary THA using the image-free navigation system either with a cup-first or stem-first technique.

Methods

Eighty-three hips undergoing primary THA using the OrthoPilot® image-free navigation system (B. Braun-Aesculap, Tuttlingen, Germany) were included in this study. The patient population was divided into two groups depending on the procedure used: cup-first technique and stem-first technique. In the cup-first group, inclination and anteversion (AV) angles were targeted at 35–45° and 15–25°, respectively, while stem antetorsion (AT) was determined for each patient based on the amount of individual native femoral AT angle. In the stem-first group, the femur was prepared first with the target angle corresponding to the native femoral AT and the cup AV was decided considering the CA calculated with Widmer’s formula (aiming at the optimal Widmer’s CA of 37.3°).

Results

Better consistency in Widmer’s CA values was attained in the stem-first group as indicated by the smaller SD values. In the assessment of overall alignment, Widmer’s CA values were within the satisfactory range (37 ± 5°) in 41.9 and 92.3 % of the subjects in the cup-first group and the stem-first group, respectively.

Conclusions

The stem-first technique with image-free navigated THA could effectively achieve accurate and consistent control of the CA value and thus is expected to improve the surgical outcome.
Literature
1.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220PubMed
2.
go back to reference Wixson RL, MacDonald MA (2005) Total hip arthroplasty through a minimal posterior approach using imageless computer–assisted hip navigation. J Arthroplasty 20(7):51–56PubMedCrossRef Wixson RL, MacDonald MA (2005) Total hip arthroplasty through a minimal posterior approach using imageless computer–assisted hip navigation. J Arthroplasty 20(7):51–56PubMedCrossRef
3.
go back to reference Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stöckl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87(6):762–769PubMedCrossRef Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stöckl B (2005) Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br 87(6):762–769PubMedCrossRef
4.
go back to reference Hayashi S, Nishiyama T, Fujishiro T, Hashimoto S, Kanzaki N, Nishida K, Kuroda R, Kurosaka M (2013) Evaluation of the accuracy of femoral component orientation by the CT-based fluoro-matched navigation system. Int Orthop 37(6):1063–1068PubMedPubMedCentralCrossRef Hayashi S, Nishiyama T, Fujishiro T, Hashimoto S, Kanzaki N, Nishida K, Kuroda R, Kurosaka M (2013) Evaluation of the accuracy of femoral component orientation by the CT-based fluoro-matched navigation system. Int Orthop 37(6):1063–1068PubMedPubMedCentralCrossRef
5.
go back to reference Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22:815–821PubMedCrossRef Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22:815–821PubMedCrossRef
6.
go back to reference Fukunishi S, Fukui T, Imamura F, Nishio S, Shibanuma N, Yoshiya S (2008) Assessment of accuracy of acetabular cup orientation in CT-free navigated total hip arthroplasty. Orthopedics 31:987 Fukunishi S, Fukui T, Imamura F, Nishio S, Shibanuma N, Yoshiya S (2008) Assessment of accuracy of acetabular cup orientation in CT-free navigated total hip arthroplasty. Orthopedics 31:987
7.
go back to reference Fukui T, Fukunishi S, Nishio S, Shibanuma N, Yoshiya S (2010) Use of image-free navigation in determination of acetabular cup orientation: analysis of factors affecting precision. Orthopedics 33:38–42PubMedCrossRef Fukui T, Fukunishi S, Nishio S, Shibanuma N, Yoshiya S (2010) Use of image-free navigation in determination of acetabular cup orientation: analysis of factors affecting precision. Orthopedics 33:38–42PubMedCrossRef
8.
go back to reference Fukunishi S, Fukui T, Nishio S, Fujihara Y, Yoshiya S (2012) Combined anteversion of the total hip arthroplasty implanted with image-free cup navigation and without stem navigation. Orthop Rev (Pavia) 4:e33CrossRef Fukunishi S, Fukui T, Nishio S, Fujihara Y, Yoshiya S (2012) Combined anteversion of the total hip arthroplasty implanted with image-free cup navigation and without stem navigation. Orthop Rev (Pavia) 4:e33CrossRef
9.
go back to reference Amuwa C, Dorr LD (2008) The combined anteversion technique for acetabular component anteversion. J Arthroplasty 23(7):1068–1070PubMedCrossRef Amuwa C, Dorr LD (2008) The combined anteversion technique for acetabular component anteversion. J Arthroplasty 23(7):1068–1070PubMedCrossRef
10.
11.
go back to reference Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg Br 75(2):228–232PubMed Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg Br 75(2):228–232PubMed
12.
go back to reference Ranawac CS, Maynard MJ (1991) Modern techniques of cemented total hip arthroplasty. Tech Orthop 6(3):17–25CrossRef Ranawac CS, Maynard MJ (1991) Modern techniques of cemented total hip arthroplasty. Tech Orthop 6(3):17–25CrossRef
13.
go back to reference Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17(3):282–288PubMedCrossRef Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17(3):282–288PubMedCrossRef
14.
go back to reference Yoshimine F (2006) The safe-zones for combined cup and neck anteversions that fulfill the essential range of motion and their optimum combination in total hip replacements. J Biomech 39:1315–1323PubMedCrossRef Yoshimine F (2006) The safe-zones for combined cup and neck anteversions that fulfill the essential range of motion and their optimum combination in total hip replacements. J Biomech 39:1315–1323PubMedCrossRef
15.
go back to reference Hisatome T, Doi H (2011) Theoretically optimum position of the prosthesis in total hip arthroplasty to fulfill the severe range of motion criteria due to neck impingement. J Orthop Sci 16:229–237PubMedCrossRef Hisatome T, Doi H (2011) Theoretically optimum position of the prosthesis in total hip arthroplasty to fulfill the severe range of motion criteria due to neck impingement. J Orthop Sci 16:229–237PubMedCrossRef
16.
go back to reference Barsoum WK, Patterson RW, Higuera C, Klika AK, Krebs VE, Molloy R (2007) A computer model of the position of the combined component in the prevention of impingement in total hip replacement. J Bone Joint Surg Br 89(6):839–845PubMedCrossRef Barsoum WK, Patterson RW, Higuera C, Klika AK, Krebs VE, Molloy R (2007) A computer model of the position of the combined component in the prevention of impingement in total hip replacement. J Bone Joint Surg Br 89(6):839–845PubMedCrossRef
17.
go back to reference Shoji T, Yasunaga Y, Yamasaki T, Izumi S, Hachisuka S, Ochi M (2015) Low femoral antetorsion and total hip arthroplasty: a risk factor. Int Orthop 39(1):7–12PubMedCrossRef Shoji T, Yasunaga Y, Yamasaki T, Izumi S, Hachisuka S, Ochi M (2015) Low femoral antetorsion and total hip arthroplasty: a risk factor. Int Orthop 39(1):7–12PubMedCrossRef
18.
go back to reference Wassilew GI, Perka C, Koenig K, Janz V, Asbach P, Hasart O (2010) 3D CT analysis of combined cup and stem anteversion in cases of cup navigation hip arthroplasty. Orthopedics 33:48–51PubMedCrossRef Wassilew GI, Perka C, Koenig K, Janz V, Asbach P, Hasart O (2010) 3D CT analysis of combined cup and stem anteversion in cases of cup navigation hip arthroplasty. Orthopedics 33:48–51PubMedCrossRef
19.
go back to reference Sendtner E, Tibor S, Winkler R, Wörner M, Grifka J, Renkawitz T (2010) Stem torsion in total hip replacement. CT measurements in 60 patients. Acta Orthop 81(5):579–582PubMedPubMedCentralCrossRef Sendtner E, Tibor S, Winkler R, Wörner M, Grifka J, Renkawitz T (2010) Stem torsion in total hip replacement. CT measurements in 60 patients. Acta Orthop 81(5):579–582PubMedPubMedCentralCrossRef
20.
go back to reference Wines AP, McNicol D (2006) Computed tomography measurement of the accuracy of component version in total hip arthroplasty. J Arthroplasty 21(5):696–701PubMedCrossRef Wines AP, McNicol D (2006) Computed tomography measurement of the accuracy of component version in total hip arthroplasty. J Arthroplasty 21(5):696–701PubMedCrossRef
21.
go back to reference Bargar WL, Jamali AA, Nejad AH (2010) Femoral anteversion in THA and its lack of correction with native acetabular anteversion. Clin Orthop Relat Res 468(2):527–532PubMedPubMedCentralCrossRef Bargar WL, Jamali AA, Nejad AH (2010) Femoral anteversion in THA and its lack of correction with native acetabular anteversion. Clin Orthop Relat Res 468(2):527–532PubMedPubMedCentralCrossRef
23.
go back to reference Husmann O, Rubin PJ, Leyvraz PF, de Roguin B, Argenson JN (1997) Three-dimensional morphology of the proximal femur. J Arthroplasty 12(4):444–450PubMedCrossRef Husmann O, Rubin PJ, Leyvraz PF, de Roguin B, Argenson JN (1997) Three-dimensional morphology of the proximal femur. J Arthroplasty 12(4):444–450PubMedCrossRef
24.
go back to reference Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P (2009) A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am 91(11):2598–2604PubMedCrossRef Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P (2009) A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am 91(11):2598–2604PubMedCrossRef
25.
go back to reference Kitada M, Nakamura N, Iwana D, Kakimoto A, Nishi T, Sugano N (2011) Evaluation of the accuracy of computed tomography-based navigation for femoral stem orientation and leg length discrepancy. J Arthroplasty 26(5):674–679PubMedCrossRef Kitada M, Nakamura N, Iwana D, Kakimoto A, Nishi T, Sugano N (2011) Evaluation of the accuracy of computed tomography-based navigation for femoral stem orientation and leg length discrepancy. J Arthroplasty 26(5):674–679PubMedCrossRef
26.
go back to reference Sugano N, Noble PC, Kamaric E (1998) A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr 22(4):610–614PubMedCrossRef Sugano N, Noble PC, Kamaric E (1998) A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr 22(4):610–614PubMedCrossRef
27.
go back to reference Nakashima Y, Hirata M, Akiyama M, Itokawa T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Iwamoto Y (2014) Combined anteversion technique reduced the dislocation in cementless total hip arthroplasty. Int Orthop 38(1):27–32PubMedPubMedCentralCrossRef Nakashima Y, Hirata M, Akiyama M, Itokawa T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Iwamoto Y (2014) Combined anteversion technique reduced the dislocation in cementless total hip arthroplasty. Int Orthop 38(1):27–32PubMedPubMedCentralCrossRef
28.
go back to reference Zhang J, Wang L, Mao Y, Li H, Ding H, Zhu Z (2014) The use of combined anteversion in total hip arthroplasty for patients with developmental dysplasia of the hip. J Arthroplasty 29(3):621–625PubMedCrossRef Zhang J, Wang L, Mao Y, Li H, Ding H, Zhu Z (2014) The use of combined anteversion in total hip arthroplasty for patients with developmental dysplasia of the hip. J Arthroplasty 29(3):621–625PubMedCrossRef
Metadata
Title
Accuracy of combined anteversion in image-free navigated total hip arthroplasty: stem-first or cup-first technique?
Authors
Shigeo Fukunishi
Shoji Nishio
Yuki Fujihara
Shohei Okahisa
Yu Takeda
Tomokazu Fukui
Shinichi Yoshiya
Publication date
01-01-2016
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 1/2016
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-2784-9

Other articles of this Issue 1/2016

International Orthopaedics 1/2016 Go to the issue