Abstract

Purpose

The aim of this study was to develop a generic and ultra-efficient modulated arc technique for treatment with total body irradiation (TBI) without additional equipment in standard treatment rooms.

Methods

A continuous gantry arc between 300° and 70° composed of 26 subarcs (5° per subarc) using a field size of 40 × 40 cm2 was used to perform the initial beam data measurements. The profile was measured parallel to the direction of gantry rotation at a constant depth of 9 cm (phantom thickness 18 cm). Beam data were measured for single 5° subarcs, dissecting the individual contribution of each subarc to a certain measurement point. The phantom was moved to 20 measurement positions along the profile. Then profile optimization was performed manually by varying the weighting factors of all segments until calculated doses at all points were within ± 1 %. Finally, the dose distribution of the modulated arc was verified in phantom thicknesses of 18 and 28 cm.

Results

The measured profile showed a relative mean dose of 99.7 % [standard deviation (SD) 0.7 %)] over the length of 200 cm at a depth of 9 cm. The measured mean effective surface dose (at a depth of 2 cm) was 102.7 % (SD 2.1 %). The measurements in the 28 cm slab phantom revealed a mean dose of 95.9 % (SD 2.9 %) at a depth of 14 cm. The mean dose at a depth of 2 cm was 111.9 % (SD 4.1 %). Net beam-on-time for a 2 Gy fraction is approximately 8 min.

Conclusion

This highly efficient modulated arc technique for TBI can replace conventional treatment techniques, providing a homogeneous dose distribution, dosimetric robustness, extremely fast delivery, and applicability in small treatment rooms, with no need for additional equipment.
Literature
2.
go back to reference Quast U (1987) Total body irradiation—review of treatment techniques in Europe. Radiother Oncol 9:91–106CrossRefPubMed Quast U (1987) Total body irradiation—review of treatment techniques in Europe. Radiother Oncol 9:91–106CrossRefPubMed
3.
go back to reference Heublein AC (1932) Preliminary report on continuous irradiation of the entire body. Radiol 18:1051–1062CrossRef Heublein AC (1932) Preliminary report on continuous irradiation of the entire body. Radiol 18:1051–1062CrossRef
4.
5.
go back to reference Yao R, Bernard D, Turian J et al (2012) A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity. Med Phys 39:2239–2248CrossRefPubMed Yao R, Bernard D, Turian J et al (2012) A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity. Med Phys 39:2239–2248CrossRefPubMed
6.
go back to reference Held M, Kirby N, Morin O et al (2012) Dosimetric aspects of inverse-planned modulated-arc total-body irradiation. Med Phys 39:5263–5271CrossRefPubMed Held M, Kirby N, Morin O et al (2012) Dosimetric aspects of inverse-planned modulated-arc total-body irradiation. Med Phys 39:5263–5271CrossRefPubMed
7.
go back to reference Kirby N, Held M, Morin O et al (2012) Inverse-planned modulated-arc total-body irradiation. Med Phys 39:2761–2764CrossRefPubMed Kirby N, Held M, Morin O et al (2012) Inverse-planned modulated-arc total-body irradiation. Med Phys 39:2761–2764CrossRefPubMed
8.
go back to reference Chretien M, Cote C, Blais R et a (2000) A variable speed translating couch technique for total body irradiation. Med Phys 27:1127–1130CrossRefPubMed Chretien M, Cote C, Blais R et a (2000) A variable speed translating couch technique for total body irradiation. Med Phys 27:1127–1130CrossRefPubMed
9.
go back to reference Connors S, Scrimger J, Logus W et al (1988) Development of a translating bed for total body irradiation. Med Dosim 13:195–199PubMed Connors S, Scrimger J, Logus W et al (1988) Development of a translating bed for total body irradiation. Med Dosim 13:195–199PubMed
10.
go back to reference Gerig LH, Szanto J, Bichay T et al (1994) A translating-bed technique for total-body irradiation. Phys Med Biol 39:19–35CrossRefPubMed Gerig LH, Szanto J, Bichay T et al (1994) A translating-bed technique for total-body irradiation. Phys Med Biol 39:19–35CrossRefPubMed
11.
go back to reference Zhuang T, Wu Q (2010) Generating arbitrary one-dimensional dose profiles using rotational therapy. Phys Med Biol 55:6263–6277CrossRefPubMed Zhuang T, Wu Q (2010) Generating arbitrary one-dimensional dose profiles using rotational therapy. Phys Med Biol 55:6263–6277CrossRefPubMed
12.
go back to reference Jahnke A, Jahnke L, Molina-Duran F et al (2014) Arc therapy for total body irradiation—a robust novel treatment technique for standard treatment rooms. Radiother Oncol 110:553–557. (2014/01/21 ed.)CrossRefPubMed Jahnke A, Jahnke L, Molina-Duran F et al (2014) Arc therapy for total body irradiation—a robust novel treatment technique for standard treatment rooms. Radiother Oncol 110:553–557. (2014/01/21 ed.)CrossRefPubMed
13.
go back to reference Mancosu P, Navarria P, Castagna L et al (2013) Interplay effects between dose distribution quality and positioning accuracy in total marrow irradiation with volumetric modulated arc therapy. Med Phys 40:111713CrossRefPubMed Mancosu P, Navarria P, Castagna L et al (2013) Interplay effects between dose distribution quality and positioning accuracy in total marrow irradiation with volumetric modulated arc therapy. Med Phys 40:111713CrossRefPubMed
14.
go back to reference Gruen A, Ebell W, Wlodarczyk W et al (2013) Total body irradiation (TBI) using helical tomotherapy in children and young adults undergoing stem cell transplantation. Radiat Oncol 8:92PubMedCentralCrossRefPubMed Gruen A, Ebell W, Wlodarczyk W et al (2013) Total body irradiation (TBI) using helical tomotherapy in children and young adults undergoing stem cell transplantation. Radiat Oncol 8:92PubMedCentralCrossRefPubMed
15.
go back to reference Penagaricano JA, Chao M, Van Rhee F, Ratanatharathorn V et al (2011) Clinical feasibility of TBI with helical tomotherapy. Bone Marrow Transplant 46:929–935CrossRefPubMed Penagaricano JA, Chao M, Van Rhee F, Ratanatharathorn V et al (2011) Clinical feasibility of TBI with helical tomotherapy. Bone Marrow Transplant 46:929–935CrossRefPubMed
16.
go back to reference Wong JY, Rosenthal J, Liu A et al (2009) Image-guided total-marrow irradiation using helical tomotherapy in patients with multiple myeloma and acute leukemia undergoing hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys 73:273–279PubMedCentralCrossRefPubMed Wong JY, Rosenthal J, Liu A et al (2009) Image-guided total-marrow irradiation using helical tomotherapy in patients with multiple myeloma and acute leukemia undergoing hematopoietic cell transplantation. Int J Radiat Oncol Biol Phys 73:273–279PubMedCentralCrossRefPubMed
17.
go back to reference Schultheiss T.E, Wong J, Liu A et al (2007) Image-guided total marrow and total lymphatic irradiation using helical tomotherapy. Int J Radiat Oncol Biol Phys 67:1259–1267CrossRefPubMed Schultheiss T.E, Wong J, Liu A et al (2007) Image-guided total marrow and total lymphatic irradiation using helical tomotherapy. Int J Radiat Oncol Biol Phys 67:1259–1267CrossRefPubMed
18.
go back to reference Surucu M, Yeginer M, Kavak G et al (2012) Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom. Med Phys 39:281–288CrossRefPubMed Surucu M, Yeginer M, Kavak G et al (2012) Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom. Med Phys 39:281–288CrossRefPubMed
19.
go back to reference Sampath S, Schultheiss TE, Wong J (2005) Dose response and factors related to interstitial pneumonitis after bone marrow transplant. Int J Radiat Oncol Biol Phys 63:876–884CrossRefPubMed Sampath S, Schultheiss TE, Wong J (2005) Dose response and factors related to interstitial pneumonitis after bone marrow transplant. Int J Radiat Oncol Biol Phys 63:876–884CrossRefPubMed
20.
go back to reference Van Dyk J, Galvin JM, Glasgow GP et al (1986) AAPM report 17, Task group 29 of the Radiation Therapy Committee, American Association of Physicists in Medicine. Van Dyk J, Galvin JM, Glasgow GP et al (1986) AAPM report 17, Task group 29 of the Radiation Therapy Committee, American Association of Physicists in Medicine.
Metadata
Title
Accelerating total body irradiation with large field modulated arc therapy in standard treatment rooms without additional equipment
Authors
Martin Polednik, Ph.D.
Frank Lohr, M.D.
Michael Ehmann, M.D.
Frederik Wenz, M.D.
Publication date
01-11-2015
Publisher
Springer Berlin Heidelberg
Published in
Strahlentherapie und Onkologie / Issue 11/2015
Print ISSN: 0179-7158
Electronic ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-015-0883-6

Other articles of this Issue 11/2015

Strahlentherapie und Onkologie 11/2015 Go to the issue