Ablation time efficiency and lesion volume - in vitro comparison of 4 mm, non irrigated, gold- and platinum-iridium-tip radiofrequency ablation catheters

Authors: Tibor Balázs, Romola Laczkó, Eszter Bognár, Serkan Akman, Péter Nagy, Endre Zima, János Dobránszky, Tamás Szili-Török

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2013

Login to get access

Abstract

Purpose

We investigated the development of in vitro lesion formation in relation to ablation time and power delivery using 4-mm, non-irrigated, platinum–iridium (Pt/Ir), and gold-tip catheters.

Methods

Radiofrequency catheter ablation was performed on porcine liver preparations in a 5–60-s time range with 5 s increments applied at different lesion sites. Each series was repeated four times for both catheters. The lesion volume, the delivered energy, and the ablation power curves were analyzed in 96 ablations.

Results

The delivered energy was higher and the lesion volume was significantly larger when using a gold-tip catheter. The lesion volume that was created with the Pt/Ir-tip electrode within 60 s was created after a shorter 40 s ablation period with the gold-tip electrode (Au, 40 s, 345 ± 75 mm3 vs. Pt/Ir, 60 s, 328 ± 30 mm3).

Conclusion

The gold-tip catheters had a lower tip temperature resulting in higher energy delivery due to the better thermo-conductivity of the gold alloy. A significantly higher lesion volume and a faster lesion creation were found for gold catheters as compared to Pt/Ir catheters. Gold-tip catheters can create the same adequate lesion within 33.3 % less ablation time due to the higher values of delivered energy as compared to those observed with the use of Pt/Ir catheters.
Literature
1.
go back to reference Blomstrom-Lundqvist, C., Scheinman, M. M., Aliot, E. M., Alpert, J. S., Calkins, H., Camm, A. J., et al. (2003). ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias—executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (writing committee to develop guidelines for the management of patients with supraventricular arrhythmias). Circulation, 108, 1871–1909.PubMedCrossRef Blomstrom-Lundqvist, C., Scheinman, M. M., Aliot, E. M., Alpert, J. S., Calkins, H., Camm, A. J., et al. (2003). ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias—executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (writing committee to develop guidelines for the management of patients with supraventricular arrhythmias). Circulation, 108, 1871–1909.PubMedCrossRef
2.
go back to reference Yokoyama, K., Nakagawa, H., Wittkampf, F. H., Pitha, J. V., Lazzara, R., & Jackman, W. M. (2006). Comparison of electrode cooling between internal and open irrigation in radiofrequency ablation lesion depth and incidence of thrombus and steam pop. Circulation, 1(13), 11–19. Yokoyama, K., Nakagawa, H., Wittkampf, F. H., Pitha, J. V., Lazzara, R., & Jackman, W. M. (2006). Comparison of electrode cooling between internal and open irrigation in radiofrequency ablation lesion depth and incidence of thrombus and steam pop. Circulation, 1(13), 11–19.
3.
go back to reference Haines, D. E. (1991). Determinants of lesion size during radiofrequency catheter ablation: the role of electrode-tissue contact pressure and duration of energy delivery. Journal of Cardiovascular Electrophysiology, 2, 509–515.CrossRef Haines, D. E. (1991). Determinants of lesion size during radiofrequency catheter ablation: the role of electrode-tissue contact pressure and duration of energy delivery. Journal of Cardiovascular Electrophysiology, 2, 509–515.CrossRef
4.
go back to reference Thomas, S. P., Guy, D. J. R., Boyd, A. C., Eipper, V. E., Ross, D. L., & Chard, R. B. (2003). Comparison of epicardial and endocardial linear ablation using handheld probes. The Annals of Thoracic Surgery, 75, 543–548. Thomas, S. P., Guy, D. J. R., Boyd, A. C., Eipper, V. E., Ross, D. L., & Chard, R. B. (2003). Comparison of epicardial and endocardial linear ablation using handheld probes. The Annals of Thoracic Surgery, 75, 543–548.
5.
go back to reference Scavee, C., Jais, P., Hsu, L. F., Sanders, P., Hocini, M., Weerasooriya, R., et al. (2004). Prospective randomized comparison of irrigated-tip and large-tip catheter ablation of cavotricuspid isthmus-dependent atrial flutter. European Heart Journal, 25, 963–969. Scavee, C., Jais, P., Hsu, L. F., Sanders, P., Hocini, M., Weerasooriya, R., et al. (2004). Prospective randomized comparison of irrigated-tip and large-tip catheter ablation of cavotricuspid isthmus-dependent atrial flutter. European Heart Journal, 25, 963–969.
6.
go back to reference Robles de Medina, E. O., Wittkampf, F. H., & Hauer, R. N. (1989). Control of radiofrequency lesion size by power regulation. Circulation, 80, 962–968.PubMedCrossRef Robles de Medina, E. O., Wittkampf, F. H., & Hauer, R. N. (1989). Control of radiofrequency lesion size by power regulation. Circulation, 80, 962–968.PubMedCrossRef
7.
go back to reference Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., et al. (1995). Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation, 91, 2264–2273.PubMedCrossRef Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., et al. (1995). Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation, 91, 2264–2273.PubMedCrossRef
8.
go back to reference Wittkampf, F. H., Nakagawa, H., Foresti, S., Aoyama, H., & Jackman, W. M. (2005). Saline-irrigated radiofrequency ablation electrode with external cooling. Journal of Cardiovascular Electrophysiology, 16, 323–328.PubMedCrossRef Wittkampf, F. H., Nakagawa, H., Foresti, S., Aoyama, H., & Jackman, W. M. (2005). Saline-irrigated radiofrequency ablation electrode with external cooling. Journal of Cardiovascular Electrophysiology, 16, 323–328.PubMedCrossRef
9.
go back to reference Linhart, M., Mollnau, H., Bitzen, A., Wurtz, S., Schrickel, J. W., Andrie, R., et al. (2009). In vitro comparison of platinum–iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters. Europace, 11, 565–570. doi:10.1093/europace/eup040.PubMedCrossRef Linhart, M., Mollnau, H., Bitzen, A., Wurtz, S., Schrickel, J. W., Andrie, R., et al. (2009). In vitro comparison of platinum–iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters. Europace, 11, 565–570. doi:10.​1093/​europace/​eup040.PubMedCrossRef
Metadata
Title
Ablation time efficiency and lesion volume - in vitro comparison of 4 mm, non irrigated, gold- and platinum-iridium-tip radiofrequency ablation catheters
Authors
Tibor Balázs
Romola Laczkó
Eszter Bognár
Serkan Akman
Péter Nagy
Endre Zima
János Dobránszky
Tamás Szili-Török
Publication date
01-01-2013
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2013
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9743-9

Other articles of this Issue 1/2013

Journal of Interventional Cardiac Electrophysiology 1/2013 Go to the issue