A vertical scanning Doppler probe identifies blood vessels during laparoscopic surgery

Authors: T. Ohdaira, Hideo Nagai

Published in: Surgical Endoscopy | Issue 5/2007

Login to get access

Abstract

Background

Using a conventional Doppler probe, it is difficult to identify blood vessels precisely in lymph node dissection during laparoscopic surgery.

Methods

For 26 patients undergoing laparoscopic colectomy and 26 patients undergoing laparoscopic subtotal gastrectomy, we used conventional and vertical scanning Doppler probes to compare their ability to identify invisible blood vessels through differences in power output voltage.

Results

The vertical scanning Doppler probe was 100% successful in identifying the target blood vessels regardless of the patient’s body mass index (BMI), and also was superior to the conventional Doppler probe in detection sensitivity. The vertical scanning Doppler probe was useful for image training of recently qualified surgeons through real-time monitoring of the probe position and response on the monitor. Furthermore, findings showed that the vertical scanning Doppler probe was applicable to lymph node dissection as a forceps for laparoscopic usage.

Conclusion

The vertical scanning Doppler could identify the target blood vessels regardless of the patient’s BMI. It was verified to be useful as a surgical assistive device apart from being a sensor and may serve in the education of recently qualified surgeons.
Literature
1.
go back to reference Fowler DL, White SA (1992) The use of a Doppler probe for identifying the cystic artery during a laparoscopic cholecystectomy: a pilot study. Surg Laparosc Endosc 2: 117–119PubMedCrossRef Fowler DL, White SA (1992) The use of a Doppler probe for identifying the cystic artery during a laparoscopic cholecystectomy: a pilot study. Surg Laparosc Endosc 2: 117–119PubMedCrossRef
2.
go back to reference Loughlin KR, Brooks DC (1992) The use of a Doppler probe in laparoscopic surgery. J Laparoendosc Surg 2: 191–194PubMed Loughlin KR, Brooks DC (1992) The use of a Doppler probe in laparoscopic surgery. J Laparoendosc Surg 2: 191–194PubMed
Metadata
Title
A vertical scanning Doppler probe identifies blood vessels during laparoscopic surgery
Authors
T. Ohdaira
Hideo Nagai
Publication date
01-05-2007
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 5/2007
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-006-9094-7

Other articles of this Issue 5/2007

Surgical Endoscopy 5/2007 Go to the issue