Skip to main content
Top
Published in: Surgical Endoscopy 2/2006

01-02-2006

A versatile dual-channel carbon dioxide (CO2) insufflator for various CO2 applications

The prototype

Authors: K. Nakajima, K. Yasumasa, S. Endo, T. Takahashi, A. Nishitani, R. Nezu, T. Nishida

Published in: Surgical Endoscopy | Issue 2/2006

Login to get access

Abstract

Background

Carbon dioxide (CO2), with its rapid absorptive nature, has been proven superior to atmospheric air as an insufflating agent in various clinical settings. However, CO2 insufflation has not gained wide clinical acceptance, mainly because there has been no suitable feeding system. The authors therefore have developed a versatile “dual-channel” CO2 insufflator that facilitates wider use of CO2. The objectives of this study were to introduce the authors’ prototype insufflator, to evaluate its safety and performance, and to validate CO2 application using the prototype.

Methods

The prototype insufflator provides one CO2 inlet connected to a regular CO2 gas cylinder and two CO2 outlets positioned on the front and back of the device, respectively. The CO2 gas fed from the cylinder is pressure-regulated and divided into two independent conduits inside the device. The front outlet feeds CO2 gas for pneumoperitoneum at an electronically controlled pressure and flow rate. The back channel supplies CO2 gas at a fixed flow rate, allowing manual control of insufflation for various purposes. The device was evaluated with canine models.

Results

The prototype was safe and performed well. The CO2 application (colonoscopy in this series) using the back channel was feasible while intact CO2 pneumoperitoneum was simultaneously maintained via the front channel. There were no device malfunctions. The serial abdominal x-rays indicated that intraluminal CO2 insufflation such as that used for CO2 colonoscopy caused less residual intestinal gas than conventional air insufflation.

Conclusions

The dual-channel CO2 insufflator enabled two different modes of CO2 insufflation at the same time from a single CO2 cylinder. The authors are now improving the prototype to allow safer and wider usage of CO2 in the operating room.
Literature
1.
go back to reference Bretthauer M, Thiis-Evensen E, Huppertz-Hauss G, Gisselsson L, Grotmol T, Skovlund E, Hoff G (2002) NORCCAP (Norwegian colorectal cancer prevention): a randomised trial to assess the safety and efficacy of carbon dioxide versus air insufflation in colonoscopy. Gut 50: 604–607CrossRefPubMed Bretthauer M, Thiis-Evensen E, Huppertz-Hauss G, Gisselsson L, Grotmol T, Skovlund E, Hoff G (2002) NORCCAP (Norwegian colorectal cancer prevention): a randomised trial to assess the safety and efficacy of carbon dioxide versus air insufflation in colonoscopy. Gut 50: 604–607CrossRefPubMed
2.
go back to reference Church J, Delaney C (2003) Randomized, controlled trial of carbon dioxide insufflation during colonoscopy. Dis Colon Rectum 46: 322–326PubMed Church J, Delaney C (2003) Randomized, controlled trial of carbon dioxide insufflation during colonoscopy. Dis Colon Rectum 46: 322–326PubMed
3.
go back to reference Nakajima K, Lee SW, Sonoda T, Milsom JW (2005) Intraoperative CO2 colonoscopy: a safe insufflation alternative for localization of colonic lesions during laparoscopic surgery. Surg Endosc 19: 321–325CrossRefPubMed Nakajima K, Lee SW, Sonoda T, Milsom JW (2005) Intraoperative CO2 colonoscopy: a safe insufflation alternative for localization of colonic lesions during laparoscopic surgery. Surg Endosc 19: 321–325CrossRefPubMed
4.
go back to reference Sumanac K, Zealley I, Fox BM, Rawlinson J, Salena B, Marshall JK, Stevenson GW, Hunt RH (2002) Minimizing postcolonoscopy abdominal pain by using CO2 insufflation: a prospective, randomized, double-blind, controlled trial evaluating a new commercially available CO2 delivery system. Gastrointest Endosc 56: 190–194CrossRefPubMed Sumanac K, Zealley I, Fox BM, Rawlinson J, Salena B, Marshall JK, Stevenson GW, Hunt RH (2002) Minimizing postcolonoscopy abdominal pain by using CO2 insufflation: a prospective, randomized, double-blind, controlled trial evaluating a new commercially available CO2 delivery system. Gastrointest Endosc 56: 190–194CrossRefPubMed
Metadata
Title
A versatile dual-channel carbon dioxide (CO2) insufflator for various CO2 applications
The prototype
Authors
K. Nakajima
K. Yasumasa
S. Endo
T. Takahashi
A. Nishitani
R. Nezu
T. Nishida
Publication date
01-02-2006
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 2/2006
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-005-0208-4

Other articles of this Issue 2/2006

Surgical Endoscopy 2/2006 Go to the issue