A Three-dimensional Method for Evaluating Changes in Acetabular Osteolytic Lesions in Response to Treatment

Authors: Hiroshi Egawa, MD, Henry Ho, MS, Cathy Huynh, BS, Robert H. Hopper Jr., PhD, C. Anderson Engh Jr., MD, Charles A. Engh, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2010

Login to get access

Abstract

The treatment of asymptomatic osteolysis among well-fixed cementless cups remains controversial. To compare the effectiveness of different treatment strategies, an objective technique for evaluating bone remodeling would be useful. By matching and comparing serial CT images with the aid of a computer-assisted imaging program, we developed a method to evaluate three-dimensional mineralization changes within osteolytic defects. Preoperative, immediate postoperative, and followup CT images were normalized based on a phantom with known densities and matched using image registration so that the same region could be analyzed on each image. New bone mineralization within the preoperative osteolytic lesion volume was quantified based on a patient-specific trabecular bone density threshold. As a pilot study, we applied this technique in 10 patients treated by polyethylene liner exchange with débridement and grafting of periacetabular osteolytic lesions using a calcium sulfate bone graft substitute. Relative to the preoperative osteolytic lesion volume, an average of 43% (range, 8%–72%) of each defect was filled with graft at revision. After resorption of the graft, an average of 24% (range, 9%–44%) of the original defect volume demonstrated evidence of new mineralization at 1-year followup. The amount of new mineralization was directly proportional (r2 = 0.70) to the defect filling achieved at revision. CT-based image analysis offers an objective method for quantifying three-dimensional bone remodeling and can be used to evaluate the effectiveness of osteolysis treatment strategies.
Level of Evidence: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Beaulé PE, LeDuff MJ, Dorey FJ, Amstutz HC. Fate of cementless acetabular components retained during revision total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2288–2293.PubMed Beaulé PE, LeDuff MJ, Dorey FJ, Amstutz HC. Fate of cementless acetabular components retained during revision total hip arthroplasty. J Bone Joint Surg Am. 2003;85:2288–2293.PubMed
2.
go back to reference Berry DJ. Periprosthetic fractures associated with osteolysis: a problem on the rise. J Arthroplasty. 2003;18(3 Suppl 1):107–111.CrossRefPubMed Berry DJ. Periprosthetic fractures associated with osteolysis: a problem on the rise. J Arthroplasty. 2003;18(3 Suppl 1):107–111.CrossRefPubMed
3.
go back to reference Borrelli J Jr, Prickett WD, Ricci WM. Treatment of nonunions and osseous defects with bone graft and calcium sulfate. Clin Orthop Relat Res. 2003;411:245–254.CrossRefPubMed Borrelli J Jr, Prickett WD, Ricci WM. Treatment of nonunions and osseous defects with bone graft and calcium sulfate. Clin Orthop Relat Res. 2003;411:245–254.CrossRefPubMed
4.
go back to reference Boucher HR, Lynch C, Young AM, Engh CA Jr, Engh C Sr. Dislocation after polyethylene liner exchange in total hip arthroplasty. J Arthroplasty. 2003;18:654–657.CrossRefPubMed Boucher HR, Lynch C, Young AM, Engh CA Jr, Engh C Sr. Dislocation after polyethylene liner exchange in total hip arthroplasty. J Arthroplasty. 2003;18:654–657.CrossRefPubMed
5.
go back to reference Chang JD, Yoo JH, Hur M, Lee SS, Chung YK, Lee CJ. Revision total hip arthroplasty for pelvic osteolysis with well-fixed cementless cup. J Arthroplasty. 2007;22:987–992.CrossRefPubMed Chang JD, Yoo JH, Hur M, Lee SS, Chung YK, Lee CJ. Revision total hip arthroplasty for pelvic osteolysis with well-fixed cementless cup. J Arthroplasty. 2007;22:987–992.CrossRefPubMed
6.
go back to reference Chatoo M, Parfitt J, Pearse MF. Periprosthetic acetabular fracture associated with extensive osteolysis. J Arthroplasty. 1998;13:843–845.CrossRefPubMed Chatoo M, Parfitt J, Pearse MF. Periprosthetic acetabular fracture associated with extensive osteolysis. J Arthroplasty. 1998;13:843–845.CrossRefPubMed
7.
go back to reference Chiang PP, Burke DW, Freiberg AA, Rubash HE. Osteolysis of the pelvis: evaluation and treatment. Clin Orthop Relat Res. 2003;417:164–174.PubMed Chiang PP, Burke DW, Freiberg AA, Rubash HE. Osteolysis of the pelvis: evaluation and treatment. Clin Orthop Relat Res. 2003;417:164–174.PubMed
8.
go back to reference Claus AM, Totterman SM, Sychterz CJ, Tamez-Pena JG, Looney RJ, Engh CA. Computed tomography to assess pelvis osteolysis after total hip replacement. Clin Orthop Relat Res. 2004;422:167–174.CrossRefPubMed Claus AM, Totterman SM, Sychterz CJ, Tamez-Pena JG, Looney RJ, Engh CA. Computed tomography to assess pelvis osteolysis after total hip replacement. Clin Orthop Relat Res. 2004;422:167–174.CrossRefPubMed
9.
go back to reference Claus AM, Walde TA, Leung SB, Wolf RL, Engh CA Sr. Management of patients with acetabular socket wear and pelvic osteolysis. J Arthroplasty. 2003;18(Suppl 1):112–117.CrossRefPubMed Claus AM, Walde TA, Leung SB, Wolf RL, Engh CA Sr. Management of patients with acetabular socket wear and pelvic osteolysis. J Arthroplasty. 2003;18(Suppl 1):112–117.CrossRefPubMed
10.
go back to reference Deheshi BM, Allen DJ, Kim PR. Treatment of retroacetabular osteolysis with calcium sulfate and retention of original components. J Arthroplasty. 2008;23:1240.e9–1240.e12.CrossRef Deheshi BM, Allen DJ, Kim PR. Treatment of retroacetabular osteolysis with calcium sulfate and retention of original components. J Arthroplasty. 2008;23:1240.e9–1240.e12.CrossRef
11.
go back to reference Engh CA, Engh CA Jr, Claus A. The relationship between lysis and loosening in failed anatomic medullary locking components. Orthopedics. 2001;24:855–856.PubMed Engh CA, Engh CA Jr, Claus A. The relationship between lysis and loosening in failed anatomic medullary locking components. Orthopedics. 2001;24:855–856.PubMed
12.
go back to reference Engh CA Jr, Egawa H, Beykirch SE, Hopper RH Jr, Engh CA. The quality of osteolysis grafting with cementless acetabular component retention. Clin Orthop Relat Res. 2007;465:150–154.PubMed Engh CA Jr, Egawa H, Beykirch SE, Hopper RH Jr, Engh CA. The quality of osteolysis grafting with cementless acetabular component retention. Clin Orthop Relat Res. 2007;465:150–154.PubMed
13.
go back to reference Etienne G, Bezwada HP, Hungerford DS, Mont MA. The incorporation of morselized bone grafts in cementless acetabular revisions. Clin Orthop Relat Res. 2004;428:241–246.CrossRefPubMed Etienne G, Bezwada HP, Hungerford DS, Mont MA. The incorporation of morselized bone grafts in cementless acetabular revisions. Clin Orthop Relat Res. 2004;428:241–246.CrossRefPubMed
14.
go back to reference Griffin WL, Fehring TK, Mason JB, McCoy TH, Odum S, Terefenko CS. Early morbidity of modular exchange for polyethylene wear and osteolysis. J Arthroplasty. 2004;19(7 Suppl 2):61–66.CrossRefPubMed Griffin WL, Fehring TK, Mason JB, McCoy TH, Odum S, Terefenko CS. Early morbidity of modular exchange for polyethylene wear and osteolysis. J Arthroplasty. 2004;19(7 Suppl 2):61–66.CrossRefPubMed
15.
go back to reference Helfet DL, Ali A. Periprosthetic fractures of the acetabulum. Instr Course Lect. 2004;53:93–98.PubMed Helfet DL, Ali A. Periprosthetic fractures of the acetabulum. Instr Course Lect. 2004;53:93–98.PubMed
16.
go back to reference Hozack WJ, Mesa JJ, Carey C, Rothman RH. Relationship between polyethylene wear, pelvic osteolysis, and clinical symptomatology in patients with cementless acetabular components: a framework for decision making. J Arthroplasty. 1996;11:769–772.CrossRefPubMed Hozack WJ, Mesa JJ, Carey C, Rothman RH. Relationship between polyethylene wear, pelvic osteolysis, and clinical symptomatology in patients with cementless acetabular components: a framework for decision making. J Arthroplasty. 1996;11:769–772.CrossRefPubMed
17.
go back to reference Kelly CM, Wilkins RM. Treatment of benign bone lesions with an injectable calcium sulfate-based bone graft substitute. Orthopedics. 2004;27(1 Suppl):s131–s135.PubMed Kelly CM, Wilkins RM. Treatment of benign bone lesions with an injectable calcium sulfate-based bone graft substitute. Orthopedics. 2004;27(1 Suppl):s131–s135.PubMed
18.
go back to reference Kelly CM, Wilkins RM, Gitelis S, Hartjen C, Watson JT, Kim PT. The use of a surgical grade calcium sulfate as a bone graft substitute: results of a multicenter trial. Clin Orthop Relat Res. 2001;382:42–50.CrossRefPubMed Kelly CM, Wilkins RM, Gitelis S, Hartjen C, Watson JT, Kim PT. The use of a surgical grade calcium sulfate as a bone graft substitute: results of a multicenter trial. Clin Orthop Relat Res. 2001;382:42–50.CrossRefPubMed
19.
go back to reference Kitamura N, Leung SB, Engh CA Sr. Characteristics of pelvic osteolysis on computed tomography after total hip arthroplasty. Clin Orthop Relat Res. 2005;441:291–297.CrossRefPubMed Kitamura N, Leung SB, Engh CA Sr. Characteristics of pelvic osteolysis on computed tomography after total hip arthroplasty. Clin Orthop Relat Res. 2005;441:291–297.CrossRefPubMed
20.
go back to reference Kitamura N, Pappedemos PC, Duffy PR, Stepniewski AS, Hopper RH, Engh CA Jr, Engh CA. The value of anteroposterior pelvic radiographs for evaluating pelvis osteolysis. Clin Orthop Relat Res. 2006;453:239–245.CrossRefPubMed Kitamura N, Pappedemos PC, Duffy PR, Stepniewski AS, Hopper RH, Engh CA Jr, Engh CA. The value of anteroposterior pelvic radiographs for evaluating pelvis osteolysis. Clin Orthop Relat Res. 2006;453:239–245.CrossRefPubMed
21.
go back to reference Leopold SS, Jacobs JJ, Rosenberg AG. Cancellous allograft in revision total hip arthroplasty: a clinical review. Clin Orthop Relat Res. 2000;371:86–97.CrossRefPubMed Leopold SS, Jacobs JJ, Rosenberg AG. Cancellous allograft in revision total hip arthroplasty: a clinical review. Clin Orthop Relat Res. 2000;371:86–97.CrossRefPubMed
22.
go back to reference Leung S, Naudie D, Kitamura N, Walde T, Engh CA. Computed tomography in the assessment of periacetabular osteolysis. J Bone Joint Surg Am. 2005;87:592–597.CrossRefPubMed Leung S, Naudie D, Kitamura N, Walde T, Engh CA. Computed tomography in the assessment of periacetabular osteolysis. J Bone Joint Surg Am. 2005;87:592–597.CrossRefPubMed
23.
go back to reference Lombardi AV Jr, Berend KR. Isolated acetabular liner exchange. J Am Acad Orthop Surg. 2008;16:243–248.PubMed Lombardi AV Jr, Berend KR. Isolated acetabular liner exchange. J Am Acad Orthop Surg. 2008;16:243–248.PubMed
24.
go back to reference Maloney W, Rosenberg A. Implant Wear Symposium 2007 Clinical Work Group. What is the outcome of treatment for osteolysis? J Am Acad Orthop Surg. 2008;16(Suppl 1):S26–S32.PubMed Maloney W, Rosenberg A. Implant Wear Symposium 2007 Clinical Work Group. What is the outcome of treatment for osteolysis? J Am Acad Orthop Surg. 2008;16(Suppl 1):S26–S32.PubMed
25.
go back to reference Maloney WJ, Herzwurm P, Paprosky W, Rubash HE, Engh CA. Treatment of pelvic osteolysis with a stable acetabular component inserted without cement as part of a total hip replacement. J Bone Joint Surg Am. 1997;79:1628–1634.PubMed Maloney WJ, Herzwurm P, Paprosky W, Rubash HE, Engh CA. Treatment of pelvic osteolysis with a stable acetabular component inserted without cement as part of a total hip replacement. J Bone Joint Surg Am. 1997;79:1628–1634.PubMed
26.
go back to reference Maloney WJ, Paprosky W, Engh CA Sr, Rubash H. Surgical treatment of pelvic osteolysis. Clin Orthop Relat Res. 2001;393:78–84.CrossRefPubMed Maloney WJ, Paprosky W, Engh CA Sr, Rubash H. Surgical treatment of pelvic osteolysis. Clin Orthop Relat Res. 2001;393:78–84.CrossRefPubMed
27.
go back to reference Mehin R, Yuan X, Haydon C, Rorabeck CH, Bourne RB, McCalden RW, MacDonald SJ. Retroacetabular osteolysis: when to operate? Clin Orthop Relat Res. 2004;428:247–255.CrossRefPubMed Mehin R, Yuan X, Haydon C, Rorabeck CH, Bourne RB, McCalden RW, MacDonald SJ. Retroacetabular osteolysis: when to operate? Clin Orthop Relat Res. 2004;428:247–255.CrossRefPubMed
28.
go back to reference Mirzayan R, Panossian V, Avedian R, Forrester DM, Menendez LR. The use of calcium sulfate in the treatment of benign bone lesions: a preliminary report. J Bone Joint Surg Am. 2001;83:355–358.CrossRefPubMed Mirzayan R, Panossian V, Avedian R, Forrester DM, Menendez LR. The use of calcium sulfate in the treatment of benign bone lesions: a preliminary report. J Bone Joint Surg Am. 2001;83:355–358.CrossRefPubMed
29.
go back to reference Naudie DD, Engh CA Sr. Surgical management of polyethylene wear and pelvic osteolysis with modular uncemented acetabular components. J Arthroplasty. 2004;19(Suppl 1):124–129.CrossRefPubMed Naudie DD, Engh CA Sr. Surgical management of polyethylene wear and pelvic osteolysis with modular uncemented acetabular components. J Arthroplasty. 2004;19(Suppl 1):124–129.CrossRefPubMed
30.
go back to reference Peters CL, Erickson JA, Dunn HK. Revision of well-fixed cementless acetabular components for polyethylene failure. Clin Orthop Relat Res. 2003;414:129–135.CrossRefPubMed Peters CL, Erickson JA, Dunn HK. Revision of well-fixed cementless acetabular components for polyethylene failure. Clin Orthop Relat Res. 2003;414:129–135.CrossRefPubMed
31.
go back to reference Phillips BC, Barnes CL. MIIG HV case analysis. J Surg Orthop Adv. 2008;17:58–61.PubMed Phillips BC, Barnes CL. MIIG HV case analysis. J Surg Orthop Adv. 2008;17:58–61.PubMed
32.
go back to reference Puri L, Wixson RL, Stern SH, Kohli J, Hendrix RW, Stulberg SD. Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:609–614.PubMed Puri L, Wixson RL, Stern SH, Kohli J, Hendrix RW, Stulberg SD. Use of helical computed tomography for the assessment of acetabular osteolysis after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:609–614.PubMed
33.
go back to reference Saleh KJ, Thongtrangan I, Schwarz EM. Osteolysis: medical and surgical approaches. Clin Orthop Relat Res. 2004;427:138–147.CrossRefPubMed Saleh KJ, Thongtrangan I, Schwarz EM. Osteolysis: medical and surgical approaches. Clin Orthop Relat Res. 2004;427:138–147.CrossRefPubMed
34.
go back to reference Sanchez-Sotelo J, McGrory BJ, Berry DJ. Acute periprosthetic fracture of the acetabulum associated with osteolytic pelvic lesions: a report of 3 cases. J Arthroplasty. 2000;15:126–130.CrossRefPubMed Sanchez-Sotelo J, McGrory BJ, Berry DJ. Acute periprosthetic fracture of the acetabulum associated with osteolytic pelvic lesions: a report of 3 cases. J Arthroplasty. 2000;15:126–130.CrossRefPubMed
35.
go back to reference Schmalzried TP, Fowble VA, Amstutz HC. The fate of pelvic osteolysis after reoperations: no recurrence with lesional treatment. Clin Orthop Relat Res. 1998;350:128–137.CrossRefPubMed Schmalzried TP, Fowble VA, Amstutz HC. The fate of pelvic osteolysis after reoperations: no recurrence with lesional treatment. Clin Orthop Relat Res. 1998;350:128–137.CrossRefPubMed
36.
go back to reference Stulberg SD, Wixson RL, Adams AD, Hendrix RW, Bernfield JB. Monitoring pelvis osteolysis following total hip replacement surgery: an algorithm for surveillance. J Bone Joint Surg Am. 2002;84:116–122.CrossRefPubMed Stulberg SD, Wixson RL, Adams AD, Hendrix RW, Bernfield JB. Monitoring pelvis osteolysis following total hip replacement surgery: an algorithm for surveillance. J Bone Joint Surg Am. 2002;84:116–122.CrossRefPubMed
37.
go back to reference Wright JM, Pellicci PM, Salvati EA, Ghelman B, Roberts MM, Koh JL. Bone density adjacent to press-fit acetabular components: a prospective analysis with quantitative computed tomography. J Bone Joint Surg Am. 2001;83:529–536.CrossRefPubMed Wright JM, Pellicci PM, Salvati EA, Ghelman B, Roberts MM, Koh JL. Bone density adjacent to press-fit acetabular components: a prospective analysis with quantitative computed tomography. J Bone Joint Surg Am. 2001;83:529–536.CrossRefPubMed
38.
go back to reference Yazdi M, Beaulieu L. Artifacts in spiral x-ray CT scanners: problems and solutions. Proc World Acad Sci Eng Technol. 2007;26:376–380. Yazdi M, Beaulieu L. Artifacts in spiral x-ray CT scanners: problems and solutions. Proc World Acad Sci Eng Technol. 2007;26:376–380.
Metadata
Title
A Three-dimensional Method for Evaluating Changes in Acetabular Osteolytic Lesions in Response to Treatment
Authors
Hiroshi Egawa, MD
Henry Ho, MS
Cathy Huynh, BS
Robert H. Hopper Jr., PhD
C. Anderson Engh Jr., MD
Charles A. Engh, MD
Publication date
01-02-2010
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2010
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-009-1050-0

Other articles of this Issue 2/2010

Clinical Orthopaedics and Related Research® 2/2010 Go to the issue

Letter to the Editor

Reply to Letter to the Editor

Symposium: Papers Presented at the Hip Society Meetings 2009

Is There Faster Recovery With an Anterior or Anterolateral THA? A Pilot Study

Symposium: Papers Presented at the Hip Society Meetings 2009

Biographical Sketch: Virgil Pendelton Gibney, MD, 1847–1927