Abstract

Purpose

Radiofrequency ablation for liver tumors (liver RFA) is widely performed under ultrasound guidance. However, discriminating between the tumor and the needle is often difficult because of cavitation caused by RFA-induced coagulation. An unclear ultrasound image can lead to complications and tumor residue. Therefore, image-guided navigation systems based on fiducial registration have been developed. Fiducial points are usually set on a patient’s skin. But the use of internal fiducial points can improve the accuracy of navigation. In this study, a new device is introduced to use internal fiducial points using 2D US.

Methods

3D Slicer as the navigation software, Polaris Vicra as the position sensor, and two target tumors in a 3D abdominal phantom as puncture targets were used. Also, a new device that makes it possible to obtain tracking coordinates in the body was invented. First, two-dimensional reslice images from the CT images using 3D Slicer were built. A virtual needle was displayed on the two-dimensional reslice image, reflecting the movement of the actual needle after fiducial registration. A phantom experiment using three sets of fiducial point configurations: one conventional case using only surface points, and two cases in which the center of the target tumor was selected as a fiducial point was performed. For each configuration, one surgeon punctured each target tumor ten times under guidance from the 3D Slicer display. Finally, a statistical analysis examining the puncture error was performed.

Results

The puncture error for each target tumor decreased significantly when the center of the target tumor was included as one of the fiducial points, compared with when only surface points were used.

Conclusion

This study introduces a new device to use internal fiducial points and suggests that the accuracy of image-guided navigation systems for liver RFA can be improved by using the new device.
Literature
1.
go back to reference McGahan JP, Browning PD, Brock JM, Tesluk H (1990) Hepatic ablation using radiofrequency electrocautery. lnvest Radiol 25:267–270CrossRef McGahan JP, Browning PD, Brock JM, Tesluk H (1990) Hepatic ablation using radiofrequency electrocautery. lnvest Radiol 25:267–270CrossRef
2.
go back to reference Rossi S, Fornari F, Pathies C (1990) Thermal lesions induced by 480 KHz localized current. Field in gunea pig and pig liver. Tumori 76:54–57PubMed Rossi S, Fornari F, Pathies C (1990) Thermal lesions induced by 480 KHz localized current. Field in gunea pig and pig liver. Tumori 76:54–57PubMed
3.
go back to reference Rossi S, Di Stasi M, Buscarini E (1995) Percutaneous radiofrequency interstitial thermal ablation in the treatment of small hepatocellular carcinoma. Cancer J Sci Am 1:73–81PubMed Rossi S, Di Stasi M, Buscarini E (1995) Percutaneous radiofrequency interstitial thermal ablation in the treatment of small hepatocellular carcinoma. Cancer J Sci Am 1:73–81PubMed
4.
go back to reference Hasegawa K, Kokudo N, Makuuchi M, Izumi N, Ichida T, Kudo M, Ku Y, Sakamoto M, Nakashima O, Matsui O (2013) Comparison of resection and ablation for hepatocellular carcinoma: a cohort study based on a Japanese nationwide survey. J Hepatol 58:724–729CrossRefPubMed Hasegawa K, Kokudo N, Makuuchi M, Izumi N, Ichida T, Kudo M, Ku Y, Sakamoto M, Nakashima O, Matsui O (2013) Comparison of resection and ablation for hepatocellular carcinoma: a cohort study based on a Japanese nationwide survey. J Hepatol 58:724–729CrossRefPubMed
5.
go back to reference Livraghi T, Solbiati L, Meloni MF, Gazelle GS, Halpern EF, Goldberg SN (2003) Treatment of focal liver tumors with percutaneous radio frequency ablation: complications encountered in a multicenter study. Radiology 226:441–451CrossRefPubMed Livraghi T, Solbiati L, Meloni MF, Gazelle GS, Halpern EF, Goldberg SN (2003) Treatment of focal liver tumors with percutaneous radio frequency ablation: complications encountered in a multicenter study. Radiology 226:441–451CrossRefPubMed
6.
go back to reference Nakazawa T, Kokubu S, Shibuya A, Ono K, Watanabe M, Hidaka H (2007) Radiofrequency ablation of hepatocellular carcinoma: correlation between local tumor progression after ablation and ablative margin. Am J Roentgenol 188:480–488CrossRef Nakazawa T, Kokubu S, Shibuya A, Ono K, Watanabe M, Hidaka H (2007) Radiofrequency ablation of hepatocellular carcinoma: correlation between local tumor progression after ablation and ablative margin. Am J Roentgenol 188:480–488CrossRef
7.
go back to reference Livraghi T, Goldberg SN, Lazzaroni S, Meloni F, Ierace T, Solbiati L (2000) Hepatocellular carcinoma: radiofrequency ablation of medium and large lesions. Radiology 214:761–768CrossRefPubMed Livraghi T, Goldberg SN, Lazzaroni S, Meloni F, Ierace T, Solbiati L (2000) Hepatocellular carcinoma: radiofrequency ablation of medium and large lesions. Radiology 214:761–768CrossRefPubMed
8.
go back to reference Stoll J, Ren H, Dupont PE (2012) Passive markers for tracking surgical instruments in real-time 3D ultrasound imaging. IEEE Trans Med Imaging 31(3):563–575 Stoll J, Ren H, Dupont PE (2012) Passive markers for tracking surgical instruments in real-time 3D ultrasound imaging. IEEE Trans Med Imaging 31(3):563–575
9.
go back to reference Banovac F, Tang J, Xu S (2005) Precision targeting of liver lesions using a novel electromagnetic navigation device in physiologic phantom and swine. Med Phys 32:2698–705CrossRefPubMed Banovac F, Tang J, Xu S (2005) Precision targeting of liver lesions using a novel electromagnetic navigation device in physiologic phantom and swine. Med Phys 32:2698–705CrossRefPubMed
10.
go back to reference Krücker J, Xu S, Glossop N (2007) Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy. J Vasc Interv Radiol 18:1141–1150CrossRefPubMedPubMedCentral Krücker J, Xu S, Glossop N (2007) Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy. J Vasc Interv Radiol 18:1141–1150CrossRefPubMedPubMedCentral
11.
go back to reference Krücker J, Xu S, Venkatesan A, Locklin JK, Amalo H, Glossop N, Wood BJ (2011) Clinical utility of real-time fusion guidance for biopsy and ablation. J Vasc Interv Radiol 22:515–524CrossRefPubMedPubMedCentral Krücker J, Xu S, Venkatesan A, Locklin JK, Amalo H, Glossop N, Wood BJ (2011) Clinical utility of real-time fusion guidance for biopsy and ablation. J Vasc Interv Radiol 22:515–524CrossRefPubMedPubMedCentral
12.
go back to reference Nicolau SA, Pennec X, Soler L (2009) An augmented reality system for liver thermal ablation: design and evaluation on clinical cases. Med Image Anal 13:494–506CrossRefPubMed Nicolau SA, Pennec X, Soler L (2009) An augmented reality system for liver thermal ablation: design and evaluation on clinical cases. Med Image Anal 13:494–506CrossRefPubMed
13.
go back to reference Hong J, Nakashima H, Konishi K (2006) Interventional navigation for abdominal therapy based on simultaneous use of MRI and ultrasound. Med Biol Eng Comput 44:1127–1134CrossRefPubMed Hong J, Nakashima H, Konishi K (2006) Interventional navigation for abdominal therapy based on simultaneous use of MRI and ultrasound. Med Biol Eng Comput 44:1127–1134CrossRefPubMed
14.
go back to reference Marescaux J, Diana M, Soler L (2013) Augmented reality and minimally invasive surgery. J Gastroenterol Hepatol Res 2(5):555–560 Marescaux J, Diana M, Soler L (2013) Augmented reality and minimally invasive surgery. J Gastroenterol Hepatol Res 2(5):555–560
15.
go back to reference Ren H, Campos-Nanez E, Yaniv Z, Banovac F, Abeledo H, Hata N, Cleary K (2014) Treatment planning and image guidance for radiofrequency ablation of large tumors. IEEE J Biomed Health Inform 18(3):920–928CrossRefPubMed Ren H, Campos-Nanez E, Yaniv Z, Banovac F, Abeledo H, Hata N, Cleary K (2014) Treatment planning and image guidance for radiofrequency ablation of large tumors. IEEE J Biomed Health Inform 18(3):920–928CrossRefPubMed
16.
go back to reference Fitzpatrick JM, West JB, Maurer CR Jr (1998) Predicting error in rigid-body point-based registration. IEEE Trans Med Imaging 17:694–702CrossRefPubMed Fitzpatrick JM, West JB, Maurer CR Jr (1998) Predicting error in rigid-body point-based registration. IEEE Trans Med Imaging 17:694–702CrossRefPubMed
17.
go back to reference Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M, Buatti J, Aylward S, Miller JV, Pieper S, Kikinis R (2012) 3D Slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging 30(9):1323–1341. doi:10.1016/j.mri.2012.05.001 CrossRefPubMedPubMedCentral Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M, Buatti J, Aylward S, Miller JV, Pieper S, Kikinis R (2012) 3D Slicer as an image computing platform for the quantitative imaging network. Magn Reson Imaging 30(9):1323–1341. doi:10.​1016/​j.​mri.​2012.​05.​001 CrossRefPubMedPubMedCentral
18.
go back to reference Lasso A, Heffter T, Rankin A, Pinter C, Fichtinger G (2014) PLUS: open-source toolkit for ultrasound-guided intervention systems. IEEE Trans Biomed Eng 61(10):2527–2537CrossRefPubMedPubMedCentral Lasso A, Heffter T, Rankin A, Pinter C, Fichtinger G (2014) PLUS: open-source toolkit for ultrasound-guided intervention systems. IEEE Trans Biomed Eng 61(10):2527–2537CrossRefPubMedPubMedCentral
19.
go back to reference Tokuda J, Fischer GS, Papademetris X, Yaniv Z, Ibanez L, Cheng P, Liu H, Blevins J, Arata J, Golby AJ, Kapur T, Pieper S, Burdette EC, Fichtinger G, Tempany CM, Hata N (2009) OpenlGTLink: an open network protocol for image-guided therapy environment. Int J Med Robot Comput Assist Surg 5(4):423–434CrossRef Tokuda J, Fischer GS, Papademetris X, Yaniv Z, Ibanez L, Cheng P, Liu H, Blevins J, Arata J, Golby AJ, Kapur T, Pieper S, Burdette EC, Fichtinger G, Tempany CM, Hata N (2009) OpenlGTLink: an open network protocol for image-guided therapy environment. Int J Med Robot Comput Assist Surg 5(4):423–434CrossRef
20.
go back to reference Ordas S, Yaniv Z, Cheng P, Tokuda J, Liu H, Hata N, Cleary K (2009) Interfacing proprietary hardware with the Image-Guided Surgery Toolkit (IGSTK): a case for the OpenIGTLink protocol. In: Proceedings of SPIE medical imaging, image processing, p 7264 Ordas S, Yaniv Z, Cheng P, Tokuda J, Liu H, Hata N, Cleary K (2009) Interfacing proprietary hardware with the Image-Guided Surgery Toolkit (IGSTK): a case for the OpenIGTLink protocol. In: Proceedings of SPIE medical imaging, image processing, p 7264
21.
go back to reference Wiles AD, Thompson DG, Donald D (2004) Accuracy assessment and interpretation for optical tracking systems. Proc SPIE 5367:421–432CrossRef Wiles AD, Thompson DG, Donald D (2004) Accuracy assessment and interpretation for optical tracking systems. Proc SPIE 5367:421–432CrossRef
22.
go back to reference Porter BC, Rubens DJ, Strang JG, Smith J, Totterma S, Parker KJ (2001) Three-dimensional registration and fusion of ultrasound and MRI using major vessels as fiducial markers. IEEE Trans Med Imaging 20(4):354–359CrossRefPubMed Porter BC, Rubens DJ, Strang JG, Smith J, Totterma S, Parker KJ (2001) Three-dimensional registration and fusion of ultrasound and MRI using major vessels as fiducial markers. IEEE Trans Med Imaging 20(4):354–359CrossRefPubMed
23.
go back to reference Penney GP, Blackall JM, Hamady MS, Sabharwal T, Adam A, Hawkes DJ (2004) Registration of freehand 3D ultrasound and magnetic resonance liver images. Med Image Anal 8(1):81–91CrossRefPubMed Penney GP, Blackall JM, Hamady MS, Sabharwal T, Adam A, Hawkes DJ (2004) Registration of freehand 3D ultrasound and magnetic resonance liver images. Med Image Anal 8(1):81–91CrossRefPubMed
24.
go back to reference Wein W, Brunke S, Khamene A, Callstrom MR, Navab N (2008) Automatic CT-ultrasound registration for diagnostic imaging and image-guided intervention. Med Image Anal 12(5):577–585CrossRefPubMed Wein W, Brunke S, Khamene A, Callstrom MR, Navab N (2008) Automatic CT-ultrasound registration for diagnostic imaging and image-guided intervention. Med Image Anal 12(5):577–585CrossRefPubMed
25.
go back to reference Lange T, Papenberg N, Heldmann S, Modersitzki J, Fischer B, Lamecker H (2009) Schlag PM.3D ultrasound-CT registration of the liver using combined landmark-intensity. Information 4(1):79–88 Lange T, Papenberg N, Heldmann S, Modersitzki J, Fischer B, Lamecker H (2009) Schlag PM.3D ultrasound-CT registration of the liver using combined landmark-intensity. Information 4(1):79–88
26.
go back to reference Weon C, Hyun Nam W, Lee D, Lee JY, Ra JB (2015) Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images. Med Phys 42(1):335–347CrossRefPubMed Weon C, Hyun Nam W, Lee D, Lee JY, Ra JB (2015) Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images. Med Phys 42(1):335–347CrossRefPubMed
Metadata
Title
A new device for fiducial registration of image-guided navigation system for liver RFA
Authors
Nobutaka Doba
Hiroyuki Fukuda
Kazushi Numata
Yoshiteru Hao
Kouji Hara
Akito Nozaki
Masaaki Kondo
Makoto Chuma
Katsuaki Tanaka
Shigeo Takebayashi
Norihiro Koizumi
Akira Kobayashi
Junichi Tokuda
Shin Maeda
Publication date
01-01-2018
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 1/2018
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-017-1647-9

Other articles of this Issue 1/2018

International Journal of Computer Assisted Radiology and Surgery 1/2018 Go to the issue