01-08-2006 | Original Article

A multicentered comparison of measurements obtained with microtip and external water pressure transducers

Authors: Andrew F. Hundley, Morton B. Brown, Linda Brubaker, Geoffrey W. Cundiff, Karl Kreder, Peter Lotze, Holly E. Richter, Halina Zyczynski, Anne M. Weber, Anthony G. Visco

Published in: International Urogynecology Journal | Issue 4/2006

Login to get access

Abstract

This study compared simultaneous intravesical pressure readings obtained with catheter-mounted microtip transducers and external water pressure transducer catheters during filling cystometry. Women undergoing multichannel urodynamic testing were randomly assigned to one of three groups: two microtip catheters, two external water pressure transducer catheters, or one of each type. Intravesical pressure was measured simultaneously for each transducer combination in each subject for minimal and maximal Valsalva effort and minimal, moderate, and maximal cough effort at two sequential bladder volumes (150 and 300 ml). Paired t tests were used to compare the means of the intravesical pressure obtained by the two types of catheters. The largest mean differences were observed when comparing microtip and water pressure transducers. Correlations of maximum pressure were consistently high between two microtip transducers and two water pressure transducers but lower for the microtip–water combination. Excellent reproducibility was demonstrated with transducers of similar types for intravesical pressures recorded during Valsalva and cough in women without prolapse. However, considerable variability was seen in pressures recorded by different transducers, particularly dependent on the water catheter manufacturer, indicating that intravesical pressure recordings from microtip and water-based systems are not interchangeable.
Literature
1.
go back to reference Abrams P, Cardozo L, Fall M et al (2002) The standardisation of terminology of lower urinary tract function: report from the Standardisation Sub-committee of the International Continence Society. Am J Obstet Gynecol 187:116–126CrossRefPubMed Abrams P, Cardozo L, Fall M et al (2002) The standardisation of terminology of lower urinary tract function: report from the Standardisation Sub-committee of the International Continence Society. Am J Obstet Gynecol 187:116–126CrossRefPubMed
2.
go back to reference Bump RC, Elser DM, Theofrastous JP, McClish DK (1995) Valsalva leak point pressures in women with genuine stress incontinence: reproducibility, effect of catheter caliber, and correlations with other measures of urethral resistance. Continence Program for Women Research Group. Am J Obstet Gynecol 173:551–557CrossRefPubMed Bump RC, Elser DM, Theofrastous JP, McClish DK (1995) Valsalva leak point pressures in women with genuine stress incontinence: reproducibility, effect of catheter caliber, and correlations with other measures of urethral resistance. Continence Program for Women Research Group. Am J Obstet Gynecol 173:551–557CrossRefPubMed
3.
go back to reference Swift SE, Utrie JW (1996) The need for standardization of the Valsalva leak-point pressure. Int Urogynecol J Pelvic Floor Dysfunct 7:227–230CrossRefPubMed Swift SE, Utrie JW (1996) The need for standardization of the Valsalva leak-point pressure. Int Urogynecol J Pelvic Floor Dysfunct 7:227–230CrossRefPubMed
4.
go back to reference Swift SE, Yoon EA (1999) Test–retest reliability of the cough stress test in the evaluation of urinary incontinence. Obstet Gynecol 94:99–102CrossRefPubMed Swift SE, Yoon EA (1999) Test–retest reliability of the cough stress test in the evaluation of urinary incontinence. Obstet Gynecol 94:99–102CrossRefPubMed
5.
go back to reference Miklos JR, Sze EH, Karram MM (1995) A critical appraisal of the methods of measuring leak-point pressures in women with stress incontinence. Obstet Gynecol 86:349–352CrossRefPubMed Miklos JR, Sze EH, Karram MM (1995) A critical appraisal of the methods of measuring leak-point pressures in women with stress incontinence. Obstet Gynecol 86:349–352CrossRefPubMed
6.
go back to reference Wang AC, Chen MC (2002) A comparison of urethral pressure profilometry using microtip and double-lumen perfusion catheters in women with genuine stress incontinence. BJOG 109:322–326PubMedCrossRef Wang AC, Chen MC (2002) A comparison of urethral pressure profilometry using microtip and double-lumen perfusion catheters in women with genuine stress incontinence. BJOG 109:322–326PubMedCrossRef
7.
go back to reference Hundley AF, Visco AG (2004) Comparison of measurements obtained with microtip and external water pressure transducers. Int Urogynecol J Pelvic Floor Dysfunct 15:276–280PubMed Hundley AF, Visco AG (2004) Comparison of measurements obtained with microtip and external water pressure transducers. Int Urogynecol J Pelvic Floor Dysfunct 15:276–280PubMed
8.
go back to reference Rowan D, James ED, Kramer AE, Sterling AM, Suhel PF (1987) Urodynamic equipment: technical aspects. Produced by the International Continence Society Working Party on Urodynamic Equipment. J Med Eng Technol 11:57–64PubMedCrossRef Rowan D, James ED, Kramer AE, Sterling AM, Suhel PF (1987) Urodynamic equipment: technical aspects. Produced by the International Continence Society Working Party on Urodynamic Equipment. J Med Eng Technol 11:57–64PubMedCrossRef
Metadata
Title
A multicentered comparison of measurements obtained with microtip and external water pressure transducers
Authors
Andrew F. Hundley
Morton B. Brown
Linda Brubaker
Geoffrey W. Cundiff
Karl Kreder
Peter Lotze
Holly E. Richter
Halina Zyczynski
Anne M. Weber
Anthony G. Visco
Publication date
01-08-2006
Publisher
Springer-Verlag
Published in
International Urogynecology Journal / Issue 4/2006
Print ISSN: 0937-3462
Electronic ISSN: 1433-3023
DOI
https://doi.org/10.1007/s00192-005-0027-0

Other articles of this Issue 4/2006

International Urogynecology Journal 4/2006 Go to the issue