Abstract

Shoulder arthroscopy has become a common procedure in today’s orthopedic practice. The safety of this procedure has been well established, but there are some complications associated with every surgical procedure both minor and major. In the present era, with advanced arthroscopic instruments, it is rare to encounter the problem of instrument breakage during arthroscopic surgery. Here, we report an unusual case in which we found a detached arthroscopic lens within the shoulder joint. A 58-year-old male patient who was previously operated for shoulder arthroscopy for the treatment of impingement syndrome combined with shoulder stiffness. We performed shoulder arthroscopy again and removed the detached lens arthroscopically. This case warrants the need for the surgeon and the operating room staff to be well acquainted with the arthroscopic instruments and to check the instruments properly before and after the completion of the procedure. If the operating room staff would have identified the damage to the scope, encountered during the primary operation, we could have avoided the second procedure to remove the lens.
Literature
2.
go back to reference Gambarjella RA, Tibone JE (1983) Knife blade in the knee joint: a complication of arthroscopic surgery. Am J Sports Med 11:267–268CrossRef Gambarjella RA, Tibone JE (1983) Knife blade in the knee joint: a complication of arthroscopic surgery. Am J Sports Med 11:267–268CrossRef
3.
go back to reference Small NC (1989) Complications in arthroscopy performed by experienced arthroscopists. Arthroscopy 4(3):215–221CrossRef Small NC (1989) Complications in arthroscopy performed by experienced arthroscopists. Arthroscopy 4(3):215–221CrossRef
4.
go back to reference Dick W, Glinz W, Henche HR, Ruckstuhl J, Wruhs O, Zollinger H (1978) Complications of arthroscopy. A review of 3714 cases (author’s transl) [in German]. Arch Orthop Trauma Surg 92(1):69–73. doi:10.1007/BF00381643 CrossRefPubMed Dick W, Glinz W, Henche HR, Ruckstuhl J, Wruhs O, Zollinger H (1978) Complications of arthroscopy. A review of 3714 cases (author’s transl) [in German]. Arch Orthop Trauma Surg 92(1):69–73. doi:10.​1007/​BF00381643 CrossRefPubMed
Metadata
Title
A detached arthroscopic lens within the shoulder joint: a case report
Authors
Jong-Hun Ji
Jae-Jung Jeong
Karnav Panchal
Yong Park
Publication date
01-08-2014
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 8/2014
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-014-2011-y

Other articles of this Issue 8/2014

Archives of Orthopaedic and Trauma Surgery 8/2014 Go to the issue