Published in: European Spine Journal 9/2013

01-09-2013 | Original Article

A CT-based study investigating the relationship between pedicle screw placement and stimulation threshold of compound muscle action potentials measured by intraoperative neurophysiological monitoring

Authors: Gerit Kulik, Etienne Pralong, John McManus, Damien Debatisse, Constantin Schizas

Published in: European Spine Journal | Issue 9/2013

Login to get access

Abstract

Purpose

Neurophysiological monitoring aims to improve the safety of pedicle screw placement, but few quantitative studies assess specificity and sensitivity. In this study, screw placement within the pedicle is measured (post-op CT scan, horizontal and vertical distance from the screw edge to the surface of the pedicle) and correlated with intraoperative neurophysiological stimulation thresholds.

Methods

A single surgeon placed 68 thoracic and 136 lumbar screws in 30 consecutive patients during instrumented fusion under EMG control. The female to male ratio was 1.6 and the average age was 61.3 years (SD 17.7). Radiological measurements, blinded to stimulation threshold, were done on reformatted CT reconstructions using OsiriX software. A standard deviation of the screw position of 2.8 mm was determined from pilot measurements, and a 1 mm of screw—pedicle edge distance was considered as a difference of interest (standardised difference of 0.35) leading to a power of the study of 75 % (significance level 0.05).

Results

Correct placement and stimulation thresholds above 10 mA were found in 71 % of screws. Twenty-two percent of screws caused cortical breach, 80 % of these had stimulation thresholds above 10 mA (sensitivity 20 %, specificity 90 %). True prediction of correct position of the screw was more frequent for lumbar than for thoracic screws.

Conclusion

A screw stimulation threshold of >10 mA does not indicate correct pedicle screw placement. A hypothesised gradual decrease of screw stimulation thresholds was not observed as screw placement approaches the nerve root. Aside from a robust threshold of 2 mA indicating direct contact with nervous tissue, a secondary threshold appears to depend on patients’ pathology and surgical conditions.
Literature
1.
go back to reference Darden BV, Wood KE, Hatley MK, Owen JH, Kostuik J (1996) Evaluation of pedicle screw insertion monitored by intraoperative evoked electromyography. J Spinal Disord 9:8–16PubMed Darden BV, Wood KE, Hatley MK, Owen JH, Kostuik J (1996) Evaluation of pedicle screw insertion monitored by intraoperative evoked electromyography. J Spinal Disord 9:8–16PubMed
2.
go back to reference Whitecloud TS, Skalley TC, Cook SD, Morgan EL (1989) Roentgenographic measurement of pedicle screw penetration. Clin Orthop Relat Res 245:57–68PubMed Whitecloud TS, Skalley TC, Cook SD, Morgan EL (1989) Roentgenographic measurement of pedicle screw penetration. Clin Orthop Relat Res 245:57–68PubMed
3.
go back to reference Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K (1999) Posterior lumbar interbody fusion—a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases. Acta Orthop Scand 70:329–334PubMedCrossRef Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K (1999) Posterior lumbar interbody fusion—a retrospective study of complications after facet joint excision and pedicle screw fixation in 148 cases. Acta Orthop Scand 70:329–334PubMedCrossRef
4.
go back to reference Pihlajamaki H, Myllynen P, Bostman O (1997) Complications of transpedicular lumbosacral fixation for non-traumatic disorders. J Bone Jt Surg (British Volume) 79B:183–189CrossRef Pihlajamaki H, Myllynen P, Bostman O (1997) Complications of transpedicular lumbosacral fixation for non-traumatic disorders. J Bone Jt Surg (British Volume) 79B:183–189CrossRef
5.
go back to reference Thomsen K, Christensen FB, Eiskjaer SP, Hansen ES, Fruensgaard S, Bunger CE (1997) The effect of pedicle screw instrumentation on functional outcome and fusion rates in posterolateral lumbar spinal fusion: a prospective, randomized clinical study. Spine 22:2813–2822PubMedCrossRef Thomsen K, Christensen FB, Eiskjaer SP, Hansen ES, Fruensgaard S, Bunger CE (1997) The effect of pedicle screw instrumentation on functional outcome and fusion rates in posterolateral lumbar spinal fusion: a prospective, randomized clinical study. Spine 22:2813–2822PubMedCrossRef
6.
go back to reference Esses SI, Sachs BL, Dreyzin V (1993) Complications Associated with the technique of pedicle screw fixation—a selected survey of Abs members. Spine 18:2231–2239PubMedCrossRef Esses SI, Sachs BL, Dreyzin V (1993) Complications Associated with the technique of pedicle screw fixation—a selected survey of Abs members. Spine 18:2231–2239PubMedCrossRef
7.
go back to reference Calancie B, Madsen P, Lebwohl N (1994) Stimulus-evoked EMG monitoring during transpedicular lumbosacral spine instrumentation—initial clinical-results. Spine 19:2780–2786PubMedCrossRef Calancie B, Madsen P, Lebwohl N (1994) Stimulus-evoked EMG monitoring during transpedicular lumbosacral spine instrumentation—initial clinical-results. Spine 19:2780–2786PubMedCrossRef
8.
go back to reference Shi YB, Binette M, Martin WH, Pearson JM, Hart RA (2003) Electrical stimulation for intraoperative evaluation of thoracic pedicle screw placement. Spine 28:595–601PubMed Shi YB, Binette M, Martin WH, Pearson JM, Hart RA (2003) Electrical stimulation for intraoperative evaluation of thoracic pedicle screw placement. Spine 28:595–601PubMed
9.
go back to reference Bindal RK, Ghosh S (2007) Intraoperative electromyography monitoring in minimally invasive transforaminal lumbar interbody fusion. J Neurosurg Spine 6:126–132PubMedCrossRef Bindal RK, Ghosh S (2007) Intraoperative electromyography monitoring in minimally invasive transforaminal lumbar interbody fusion. J Neurosurg Spine 6:126–132PubMedCrossRef
10.
go back to reference Rodriguez-Olaverri JC, Zimick NC, Merola A, Vicente J, Rodriguez J, Tabuenca A, Loste A, Sunen E, Burgos J, Hevia E, Piza-Vallespir G (2008) Comparing the clinical and radiological outcomes of pedicular transvertebral screw fixation of the lumbosacral spine in spondylolisthesis versus unilateral transforaminal lumbar interbody fusion (TLIF) with posterior fixation using anterior cages. Spine 33:1977–1981PubMedCrossRef Rodriguez-Olaverri JC, Zimick NC, Merola A, Vicente J, Rodriguez J, Tabuenca A, Loste A, Sunen E, Burgos J, Hevia E, Piza-Vallespir G (2008) Comparing the clinical and radiological outcomes of pedicular transvertebral screw fixation of the lumbosacral spine in spondylolisthesis versus unilateral transforaminal lumbar interbody fusion (TLIF) with posterior fixation using anterior cages. Spine 33:1977–1981PubMedCrossRef
11.
go back to reference Glassman SD, Dimar JR, Puno RM, Johnson JR, Shields CB, Linden RD (1995) A prospective analysis of intraoperative electromyographic monitoring of pedicle screw placement with computed tomographic scan confirmation. Spine 20:1375–1379PubMed Glassman SD, Dimar JR, Puno RM, Johnson JR, Shields CB, Linden RD (1995) A prospective analysis of intraoperative electromyographic monitoring of pedicle screw placement with computed tomographic scan confirmation. Spine 20:1375–1379PubMed
12.
go back to reference Gonzalez AA, Jeyanandarajan D, Hansen C, Zada G, Hsieh PC (2009) Intraoperative neurophysiological monitoring during spine surgery: a review. Neurosurg Focus 27:E6PubMedCrossRef Gonzalez AA, Jeyanandarajan D, Hansen C, Zada G, Hsieh PC (2009) Intraoperative neurophysiological monitoring during spine surgery: a review. Neurosurg Focus 27:E6PubMedCrossRef
13.
go back to reference Donohue ML, Murtagh-Schaffer C, Basta J, Moquin RR, Bashir A, Calancie B (2008) Pulse-train stimulation for detecting medial malpositioning of thoracic pedicle screws. Spine 33:E378–E385PubMedCrossRef Donohue ML, Murtagh-Schaffer C, Basta J, Moquin RR, Bashir A, Calancie B (2008) Pulse-train stimulation for detecting medial malpositioning of thoracic pedicle screws. Spine 33:E378–E385PubMedCrossRef
14.
go back to reference Bose B, Wierzbowski LR, Sestokas AK (2002) Neurophysiologic monitoring of spinal nerve root function during instrumented posterior lumbar spine surgery. Spine 27:1444–1450PubMedCrossRef Bose B, Wierzbowski LR, Sestokas AK (2002) Neurophysiologic monitoring of spinal nerve root function during instrumented posterior lumbar spine surgery. Spine 27:1444–1450PubMedCrossRef
15.
go back to reference Gabriel C, Peyman A, Grant EH (2009) Electrical conductivity of tissue at frequencies below 1 MHz. Phys Med Biol 54:4863–4878PubMedCrossRef Gabriel C, Peyman A, Grant EH (2009) Electrical conductivity of tissue at frequencies below 1 MHz. Phys Med Biol 54:4863–4878PubMedCrossRef
16.
go back to reference Anderson DG, Wierzbowski LR, Schwartz DM, Hilibrand AS, Vaccaro AR, Albert TJ (2002) Pedicle screws with high electrical resistance: a potential source of error with stimulus-evoked EMG. Spine 27:1577–1581PubMedCrossRef Anderson DG, Wierzbowski LR, Schwartz DM, Hilibrand AS, Vaccaro AR, Albert TJ (2002) Pedicle screws with high electrical resistance: a potential source of error with stimulus-evoked EMG. Spine 27:1577–1581PubMedCrossRef
17.
go back to reference Holland NR (1998) Intraoperative electromyography during thoracolumbar spinal surgery. Spine 23:1915–1922PubMedCrossRef Holland NR (1998) Intraoperative electromyography during thoracolumbar spinal surgery. Spine 23:1915–1922PubMedCrossRef
18.
go back to reference Fehlings MG, Brodke DS, Norvell DC, Dettori JR (2010) The evidence for intraoperative neurophysiological monitoring in spine surgery does it make a difference? Spine 35:S37–S46PubMedCrossRef Fehlings MG, Brodke DS, Norvell DC, Dettori JR (2010) The evidence for intraoperative neurophysiological monitoring in spine surgery does it make a difference? Spine 35:S37–S46PubMedCrossRef
19.
go back to reference Sutter M, Eggspuehler A, Grob D, Jeszenszky D, Benini A, Porchet F, Mueller A, Dvorak J (2007) The diagnostic value of multimodal intraoperative monitoring (MIOM) during spine surgery: a prospective study of 1,017 patients. Eur Spine J 16:S162–S170PubMedCrossRef Sutter M, Eggspuehler A, Grob D, Jeszenszky D, Benini A, Porchet F, Mueller A, Dvorak J (2007) The diagnostic value of multimodal intraoperative monitoring (MIOM) during spine surgery: a prospective study of 1,017 patients. Eur Spine J 16:S162–S170PubMedCrossRef
20.
go back to reference Hilibrand AS, Schwartz DM, Sethuraman V, Vaccaro AR, Albert TJ (2004) Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery. J Bone Jt Surg (American Volume) 86A:1248–1253 Hilibrand AS, Schwartz DM, Sethuraman V, Vaccaro AR, Albert TJ (2004) Comparison of transcranial electric motor and somatosensory evoked potential monitoring during cervical spine surgery. J Bone Jt Surg (American Volume) 86A:1248–1253
21.
go back to reference Rodriguez-Olaverri JC, Zimick NC, Merola A, De Blas G, Burgos J, Piza-Vallespir G, Hevia E, Vicente J, Sanper I, Domenech P, Regidor I (2008) Using triggered electromyographic threshold in the intercostal muscles to evaluate the accuracy of upper thoracic pedicle screw placement (T3–T6). Spine 33:E194–E197PubMedCrossRef Rodriguez-Olaverri JC, Zimick NC, Merola A, De Blas G, Burgos J, Piza-Vallespir G, Hevia E, Vicente J, Sanper I, Domenech P, Regidor I (2008) Using triggered electromyographic threshold in the intercostal muscles to evaluate the accuracy of upper thoracic pedicle screw placement (T3–T6). Spine 33:E194–E197PubMedCrossRef
22.
go back to reference Kim DH, Zaremski J, Kwon B, Jenis L, Woodard E, Bode R, Banco RJ (2007) Risk factors for false positive transcranial motor evoked potential monitoring alerts during surgical treatment of cervical myelopathy. Spine 32:3041–3046PubMedCrossRef Kim DH, Zaremski J, Kwon B, Jenis L, Woodard E, Bode R, Banco RJ (2007) Risk factors for false positive transcranial motor evoked potential monitoring alerts during surgical treatment of cervical myelopathy. Spine 32:3041–3046PubMedCrossRef
23.
go back to reference Beatty RM, Mcguire P, Moroney JM, Holladay FP (1995) Continuous intraoperative electromyographic recording during spinal surgery. J Neurosurg 82:401–405PubMedCrossRef Beatty RM, Mcguire P, Moroney JM, Holladay FP (1995) Continuous intraoperative electromyographic recording during spinal surgery. J Neurosurg 82:401–405PubMedCrossRef
24.
go back to reference Lehman RA, Potter BK, Kuklo TR, Chang AS, Polly DW, Shawen SB, Orchowski JR (2004) Probing for thoracic pedicle screw tract violation(s)—is it valid? J Spinal Disord Tech 17:277–283PubMedCrossRef Lehman RA, Potter BK, Kuklo TR, Chang AS, Polly DW, Shawen SB, Orchowski JR (2004) Probing for thoracic pedicle screw tract violation(s)—is it valid? J Spinal Disord Tech 17:277–283PubMedCrossRef
25.
go back to reference Raynor BL, Lenke LG, Kim Y, Hanson DS, Wilson-Holden TJ, Bridwell KH, Padberg AM (2002) Can triggered electromyograph thresholds predict safe thoracic pedicle screw placement? Spine 27:2030–2035PubMedCrossRef Raynor BL, Lenke LG, Kim Y, Hanson DS, Wilson-Holden TJ, Bridwell KH, Padberg AM (2002) Can triggered electromyograph thresholds predict safe thoracic pedicle screw placement? Spine 27:2030–2035PubMedCrossRef
26.
go back to reference Bolger C, Brayda-Bruno M, Kaelin A et al (2003) A new device to detect iatrogenic initial vertebral cortex perforation: first clinical results [abstract]. Eur Spine J 12(1):S18 Bolger C, Brayda-Bruno M, Kaelin A et al (2003) A new device to detect iatrogenic initial vertebral cortex perforation: first clinical results [abstract]. Eur Spine J 12(1):S18
27.
go back to reference Chaput CD, George K, Samdani AF, Williams JI, Gaughan J, Betz RR (2012) Reduction in radiation (fluoroscopy) while maintaining safe placement of pedicle screws during lumbar spine fusion. Spine 37(21):E1305–E1309PubMedCrossRef Chaput CD, George K, Samdani AF, Williams JI, Gaughan J, Betz RR (2012) Reduction in radiation (fluoroscopy) while maintaining safe placement of pedicle screws during lumbar spine fusion. Spine 37(21):E1305–E1309PubMedCrossRef
Metadata
Title
A CT-based study investigating the relationship between pedicle screw placement and stimulation threshold of compound muscle action potentials measured by intraoperative neurophysiological monitoring
Authors
Gerit Kulik
Etienne Pralong
John McManus
Damien Debatisse
Constantin Schizas
Publication date
01-09-2013
Publisher
Springer Berlin Heidelberg
Published in
European Spine Journal / Issue 9/2013
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-013-2754-0

Other articles of this Issue 9/2013

European Spine Journal 9/2013 Go to the issue