Skip to main content
Top
Published in: Journal of Orthopaedic Surgery and Research 1/2019

Open Access 01-12-2019 | Research article

A comparative study of the effect of drilling depth on generation of compressive force by headless compression screws using conical and cylindrical type of drill bit

Authors: Hyung-Sik Kim, Ho-Jung Kang, Yun-Rak Choi, Won-Taek Oh, Il-Hyun Koh

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2019

Login to get access

Abstract

Background

This study was conducted to measure the effect of different drilling depths on compression forces generated by two commonly used headless compression screws using the two different types of drill bit, the Acutrak® mini (conical type drill bit) and the Synthes 3.0 HCS® (cylindrical type drill bit).

Methods

A load cell was placed between two Sawbone blocks, which were 12 mm and 40 mm in thickness, respectively. After placing the guide pin into the center of the block, the drilling depth of the Acutrak® mini and Synthes HCS® screws ranged from 16 to 28 mm and 22 to 28 mm, respectively. The 24-mm screws were inserted and the compression force was measured immediately and at 30 min post-insertion.

Results

The Acutrak® mini generated greater compression force compared to the Synthes 3.0 HCS® when drilled to a depth of less than 24 mm. The compression force of the Acutrak® mini showed a strong inverse correlation with the drilling depth. There was no significant inverse correlation observed between the compression force of the Synthes HCS® and the drilling depth.

Conclusions

If the screw length and the drill depth are the same, the Synthes 3.0 HCS® (cylindrical type drill bit) is safer and easier to use as it has no change in the compression force even when over-drilling because the compression force of the two screws is similar. As for the Acutrak® mini (conical type drill bit), while it is technically demanding due to varying compression force according to the drill depth, it can be used in certain cases because it can give stronger compression force through under-drilling.
Literature
1.
go back to reference Fowler JR, Ilyas AM. Headless compression screw fixation of scaphoid fractures. Hand Clin. 2010;26:351–61 vi.CrossRef Fowler JR, Ilyas AM. Headless compression screw fixation of scaphoid fractures. Hand Clin. 2010;26:351–61 vi.CrossRef
2.
go back to reference Gardner AW, Yew YT, Neo PY, Lau CC, Tay SC. Interfragmentary compression profile of 4 headless bone screws: an analysis of the compression lost on reinsertion. J Hand Surg Am. 2012;37:1845–51.CrossRef Gardner AW, Yew YT, Neo PY, Lau CC, Tay SC. Interfragmentary compression profile of 4 headless bone screws: an analysis of the compression lost on reinsertion. J Hand Surg Am. 2012;37:1845–51.CrossRef
3.
go back to reference Crawford LA, Powell ES, Trail IA. The fixation strength of scaphoid bone screws: an in vitro investigation using polyurethane foam. J Hand Surg Am. 2012;37:255–60.CrossRef Crawford LA, Powell ES, Trail IA. The fixation strength of scaphoid bone screws: an in vitro investigation using polyurethane foam. J Hand Surg Am. 2012;37:255–60.CrossRef
4.
go back to reference Gruszka D, Herr R, Hely H, Hofmann P, Klitscher D, Hofmann A, et al. Impact of different screw designs on durability of fracture fixation: in vitro study with cyclic loading of scaphoid bones. PLoS One. 2016;11:e0145949.CrossRef Gruszka D, Herr R, Hely H, Hofmann P, Klitscher D, Hofmann A, et al. Impact of different screw designs on durability of fracture fixation: in vitro study with cyclic loading of scaphoid bones. PLoS One. 2016;11:e0145949.CrossRef
5.
go back to reference Koh IH, Kang HJ, Kim JS, Park SJ, Choi YR. A central threadless shaft screw is better than a fully threaded variable pitch screw for unstable scaphoid nonunion: a biomechanical study. Injury. 2015;46:638–42.CrossRef Koh IH, Kang HJ, Kim JS, Park SJ, Choi YR. A central threadless shaft screw is better than a fully threaded variable pitch screw for unstable scaphoid nonunion: a biomechanical study. Injury. 2015;46:638–42.CrossRef
6.
go back to reference Sugathan HK, Kilpatrick M, Joyce TJ, Harrison JW. A biomechanical study on variation of compressive force along the Acutrak 2 screw. Injury. 2012;43:205–8.CrossRef Sugathan HK, Kilpatrick M, Joyce TJ, Harrison JW. A biomechanical study on variation of compressive force along the Acutrak 2 screw. Injury. 2012;43:205–8.CrossRef
7.
go back to reference Assari S, Darvish K, Ilyas AM. Biomechanical analysis of second-generation headless compression screws. Injury. 2012;43:1159–65.CrossRef Assari S, Darvish K, Ilyas AM. Biomechanical analysis of second-generation headless compression screws. Injury. 2012;43:1159–65.CrossRef
8.
go back to reference Grewal R, Assini J, Sauder D, Ferreira L, Johnson J, Faber K. A comparison of two headless compression screws for operative treatment of scaphoid fractures. J Orthop Surg Res. 2011;6:27.CrossRef Grewal R, Assini J, Sauder D, Ferreira L, Johnson J, Faber K. A comparison of two headless compression screws for operative treatment of scaphoid fractures. J Orthop Surg Res. 2011;6:27.CrossRef
9.
go back to reference Hart A, Harvey EJ, Rabiei R, Barthelat F, Martineau PA. Fixation strength of four headless compression screws. Med Eng Phys. 2016;38:1037–43.CrossRef Hart A, Harvey EJ, Rabiei R, Barthelat F, Martineau PA. Fixation strength of four headless compression screws. Med Eng Phys. 2016;38:1037–43.CrossRef
10.
go back to reference Roebke AJ, Roebke LJ, Goyal KS. Fracture gap reduction with variable-pitch headless screws. J Hand Surg Am. 2018;43:385 e1–8.CrossRef Roebke AJ, Roebke LJ, Goyal KS. Fracture gap reduction with variable-pitch headless screws. J Hand Surg Am. 2018;43:385 e1–8.CrossRef
Metadata
Title
A comparative study of the effect of drilling depth on generation of compressive force by headless compression screws using conical and cylindrical type of drill bit
Authors
Hyung-Sik Kim
Ho-Jung Kang
Yun-Rak Choi
Won-Taek Oh
Il-Hyun Koh
Publication date
01-12-2019
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2019
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-018-1044-2

Other articles of this Issue 1/2019

Journal of Orthopaedic Surgery and Research 1/2019 Go to the issue