Published in: BMC Ophthalmology 1/2015

Open Access 01-12-2015 | Case report

A case report of progressive obstruction of Ex-PRESS miniature glaucoma shunt after transient flat anterior chamber and treatment using Nd:YAG laser

Authors: Masaki Tanito, Ichiya Sano, Akihiro Ohira

Published in: BMC Ophthalmology | Issue 1/2015

Login to get access

Abstract

Background

We report a case of Ex-PRESS miniature glaucoma shunt obstruction resulting from progressive iris synechial formation after transient anterior chamber shallowing.

Case presentation

A 68-year-old woman with pseudoexfoliation glaucoma in her right eye underwent filtration surgery with implantation of the Ex-PPESS shunt (model P-50, Alcon Japan, Tokyo, Japan) in combination with intra-surgical 0.04% mitomycin C use. After the anterior chamber injection of viscoelastic material and 100% sulfur hexafluoride gas for treatment of early postoperative over filtration, the intraocular pressure (IOP) was controlled between 9 and 12 mmHg. On postoperative day 121, gonioscopy showed that synechial formation around the shunt obstructed the axial port leaving the relief port opened. On postoperative day 274, the intraocular pressure increased to 40 mmHg and synechiae obstructed both the axial and relief ports. Dispersion of iris tissue by neodymium:yttrium-aluminium-garnet (Nd:YAG) laser (2 mJ, one shot to each port) opened both ports and immediately lowered the IOP, leaving peripheral anterior synechiae around the shunt. Up to postoperative day 400, the IOP was controlled between 13 and 15 mmHg, and the cystic bleb was maintained.

Conclusion

The synechiae formed gradually extends around the shunt’s shaft and can result in later external obstruction of the relief port. The current case requires further follow-up since synechiae remaining around the shaft can cause future obstruction. We emphasize the fact that, if the iris synechiae to the shunt once formed, it can progress and obstruct the shunt ports later.
Literature
1.
go back to reference Moisseiev E, Zunz E, Tzur R, Kurtz S, Shemesh G: Standard trabeculectomy and Ex-PRESS miniature glaucoma shunt: a comparative study and literature review.J Glaucoma 2014. [Epub ahead of print] Moisseiev E, Zunz E, Tzur R, Kurtz S, Shemesh G: Standard trabeculectomy and Ex-PRESS miniature glaucoma shunt: a comparative study and literature review.J Glaucoma 2014. [Epub ahead of print]
2.
go back to reference Wagschal LD, Trope GE, Jinapriya D, Jin YP, Buys YM: prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results.J Glaucoma 2013. [Epub ahead of print] Wagschal LD, Trope GE, Jinapriya D, Jin YP, Buys YM: prospective randomized study comparing Ex-PRESS to trabeculectomy: 1-year results.J Glaucoma 2013. [Epub ahead of print]
3.
go back to reference Maris PJ Jr, Ishida K, Netland PA: Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap.J. Glaucoma 2007, 16:14–19. 10.1097/01.ijg.0000243479.90403.cdCrossRefPubMed Maris PJ Jr, Ishida K, Netland PA: Comparison of trabeculectomy with Ex-PRESS miniature glaucoma device implanted under scleral flap.J. Glaucoma 2007, 16:14–19. 10.1097/01.ijg.0000243479.90403.cdCrossRefPubMed
4.
go back to reference Sugiyama T, Shibata M, Kojima S, Ueki M, Ikeda T: The first report on intermediate-term outcome of Ex-PRESS glaucoma filtration device implanted under scleral flap in Japanese patients.Clin. Ophthalmol. 2011, 5:1063–1066.CrossRefPubMedPubMedCentral Sugiyama T, Shibata M, Kojima S, Ueki M, Ikeda T: The first report on intermediate-term outcome of Ex-PRESS glaucoma filtration device implanted under scleral flap in Japanese patients.Clin. Ophthalmol. 2011, 5:1063–1066.CrossRefPubMedPubMedCentral
5.
go back to reference Kanner EM, Netland PA, Sarkisian SR Jr, Du H: Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery.J. Glaucoma 2009, 18:488–491. 10.1097/IJG.0b013e31818fb44eCrossRefPubMed Kanner EM, Netland PA, Sarkisian SR Jr, Du H: Ex-PRESS miniature glaucoma device implanted under a scleral flap alone or combined with phacoemulsification cataract surgery.J. Glaucoma 2009, 18:488–491. 10.1097/IJG.0b013e31818fb44eCrossRefPubMed
6.
go back to reference Bagnis A, Papadia M, Scotto R, Traverso CE: Obstruction of the Ex-PRESS miniature glaucoma device: Nd: YAG laser as a therapeutic option.J. Glaucoma 2011, 20:271.CrossRefPubMed Bagnis A, Papadia M, Scotto R, Traverso CE: Obstruction of the Ex-PRESS miniature glaucoma device: Nd: YAG laser as a therapeutic option.J. Glaucoma 2011, 20:271.CrossRefPubMed
Metadata
Title
A case report of progressive obstruction of Ex-PRESS miniature glaucoma shunt after transient flat anterior chamber and treatment using Nd:YAG laser
Authors
Masaki Tanito
Ichiya Sano
Akihiro Ohira
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2015
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-15-2

Other articles of this Issue 1/2015

BMC Ophthalmology 1/2015 Go to the issue