A 3D finite element model to investigate prosthetic interface stresses of different posterior tibial slope

Authors: Yi Shen, Xiaomiao Li, Xiaodong Fu, Weili Wang

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2015

Login to get access

Abstract

Purpose

Posterior tibial slope that is created during proximal tibial resection in total knee arthroplasty has emerged as an important factor in the mechanics of the knee joint and the surgical outcome. But the ideal degree of posterior tibial slope for recovery of the knee joint function and preventions of complications remains controversial and should vary in different racial groups. The objective of this paper is to investigate the effects of posterior tibial slope on contact stresses in the tibial polyethylene component of total knee prostheses.

Methods

Three-dimensional finite element analysis was used to calculate contact stresses in tibial polyethylene component of total knee prostheses subjected to a compressive load. The 3D finite element model of total knee prosthesis was constructed from the images produced by 3D scanning technology. Stresses in tibial polyethylene component were calculated with four different posterior tibial slopes (0°, 3°, 6° and 9°).

Results

The 3D finite element model of total knee prosthesis we presented was well validated. We found that the stress distribution in the polythene as evaluated by the distributions of the von Mises stress, the maximum principle stress, the minimum principle stress and the Cpress were more uniform with 3° and 6° posterior tibial slopes than with 0° and 9° posterior tibial slopes. Moreover, the peaks of the above stresses and trends of changes with increasing degree of knee flexion were more ideal with 3° and 6° posterior slopes.

Conclusions

The results suggested that the tibial component inclination might be favourable to 7°–10° so far as the stress distribution is concerned. The range of the tibial component inclination also can decrease the wear of polyethylene. Chinese posterior tibial slope is bigger than in the West, and the current domestic use of prostheses is imported from the West, so their demands to tilt back bone cutting can lead to shorten the service life of prostheses; this experiment result is of important clinical significance, guiding orthopaedic surgeon after the best angle to cut bone.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bloebaum RD, Zou L, Bachus KN, Shea KG, Hofmann AA, Dunn HK (1997) Analysis of particles in acetabular components from patients with osteolysis. Clin Orthop Relat Res 338:109–118CrossRefPubMed Bloebaum RD, Zou L, Bachus KN, Shea KG, Hofmann AA, Dunn HK (1997) Analysis of particles in acetabular components from patients with osteolysis. Clin Orthop Relat Res 338:109–118CrossRefPubMed
2.
go back to reference Peters PC, Engh GA, Dwyer KA, Vinh TN (1992) Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am 74(7):864–876PubMed Peters PC, Engh GA, Dwyer KA, Vinh TN (1992) Osteolysis after total knee arthroplasty without cement. J Bone Joint Surg Am 74(7):864–876PubMed
3.
go back to reference Brach del Prever EM, Bistolfi A, Bracco P, Costa L (2009) UHMWPE for arthroplasty: past or future? J Orthopaed Traumatol 10:1–8CrossRef Brach del Prever EM, Bistolfi A, Bracco P, Costa L (2009) UHMWPE for arthroplasty: past or future? J Orthopaed Traumatol 10:1–8CrossRef
4.
go back to reference McKellop HA, Shen FW, Lu B (1999) Development of an extremely wear-resistant ultra-high molecular weight polyethylene for total hip replacements. J Orthop Res 17:157–167CrossRefPubMed McKellop HA, Shen FW, Lu B (1999) Development of an extremely wear-resistant ultra-high molecular weight polyethylene for total hip replacements. J Orthop Res 17:157–167CrossRefPubMed
5.
go back to reference Muratoglu OK, Bragdon CR, Jasty M (1995) The effect of radiation damage on the wear rate of UHMWPE components. In: Symposium on characterization and properties of ultra-high molecular weight polyethylene, ASTM conference Muratoglu OK, Bragdon CR, Jasty M (1995) The effect of radiation damage on the wear rate of UHMWPE components. In: Symposium on characterization and properties of ultra-high molecular weight polyethylene, ASTM conference
6.
go back to reference Akagi M, Asano T, Clarke IC, Niiyama N, Kyomoto N, Nakamura T (2006) Wear and toughness of crosslinked polyethylene for total knee replacements: a study using a simulator and small-punch testing. J Orthop Res 24:2021–2027CrossRefPubMed Akagi M, Asano T, Clarke IC, Niiyama N, Kyomoto N, Nakamura T (2006) Wear and toughness of crosslinked polyethylene for total knee replacements: a study using a simulator and small-punch testing. J Orthop Res 24:2021–2027CrossRefPubMed
7.
go back to reference Blunn GW, Walker PS, Joshi A, Hardinge K (1991) The dominance of cyclic sliding in producing wear in total knee replacements. Clin Orthop Relat Res 273:253–260PubMed Blunn GW, Walker PS, Joshi A, Hardinge K (1991) The dominance of cyclic sliding in producing wear in total knee replacements. Clin Orthop Relat Res 273:253–260PubMed
8.
go back to reference Lee HY, Kim SJ, Kang KT, Kim SH, Park KK (2012) The effect of tibial posterior slope on contact force and ligaments stresses in posterior-stabilized total knee arthroplasty-explicit finite element analysis. Knee Surg Relat Res 24(2):91–98PubMedCentralCrossRefPubMed Lee HY, Kim SJ, Kang KT, Kim SH, Park KK (2012) The effect of tibial posterior slope on contact force and ligaments stresses in posterior-stabilized total knee arthroplasty-explicit finite element analysis. Knee Surg Relat Res 24(2):91–98PubMedCentralCrossRefPubMed
9.
go back to reference Panni AS, Cerciello S, Vasso M, Tartarone M (2009) Stiffness in total knee arthroplasty. J Orthop Traumatol 10(3):111–118CrossRef Panni AS, Cerciello S, Vasso M, Tartarone M (2009) Stiffness in total knee arthroplasty. J Orthop Traumatol 10(3):111–118CrossRef
10.
go back to reference Matsuda S, Miura H, Nagamine R, Urabe K, Ikenoue T, Okazaki K, Iwamoto Y (1999) Posterior tibial slope in the normal and varus knee. Am J Knee Surg 12(3):165–168PubMed Matsuda S, Miura H, Nagamine R, Urabe K, Ikenoue T, Okazaki K, Iwamoto Y (1999) Posterior tibial slope in the normal and varus knee. Am J Knee Surg 12(3):165–168PubMed
11.
go back to reference Bai B, Baez J, Testa N, Kummer FJ (2000) Effect of posterior cut angle on tibial component loading. J Arthroplasty 15(7):916–920CrossRefPubMed Bai B, Baez J, Testa N, Kummer FJ (2000) Effect of posterior cut angle on tibial component loading. J Arthroplasty 15(7):916–920CrossRefPubMed
13.
go back to reference Baltzopoulos V (1995) Muscular and tibiofemoral joint forces during isokinetic concentric knee extension. Clin Biomech (Bristol, Avon) 10(4):208–214CrossRef Baltzopoulos V (1995) Muscular and tibiofemoral joint forces during isokinetic concentric knee extension. Clin Biomech (Bristol, Avon) 10(4):208–214CrossRef
14.
go back to reference Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ (2002) Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech 35(2):267–275CrossRefPubMed Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ (2002) Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech 35(2):267–275CrossRefPubMed
15.
go back to reference Kellis E (2001) Tibiofemoral joint forces during maximal isokinetic eccentric and concentric efforts of the knee flexors. Clin Biomech (Bristol, Avon) 16(3):229–236CrossRef Kellis E (2001) Tibiofemoral joint forces during maximal isokinetic eccentric and concentric efforts of the knee flexors. Clin Biomech (Bristol, Avon) 16(3):229–236CrossRef
16.
go back to reference Piazza SJ, Delp SL (2001) Three-dimensional dynamic simulation of total knee replacement motion during a step-up task. J Biomech Eng 123(6):599–606CrossRefPubMed Piazza SJ, Delp SL (2001) Three-dimensional dynamic simulation of total knee replacement motion during a step-up task. J Biomech Eng 123(6):599–606CrossRefPubMed
17.
go back to reference Taylor M, Barrett DS (2003) Explicit finite element simulation of eccentric loading in total knee replacement. Clin Orthop Relat Res 414:162–171CrossRefPubMed Taylor M, Barrett DS (2003) Explicit finite element simulation of eccentric loading in total knee replacement. Clin Orthop Relat Res 414:162–171CrossRefPubMed
18.
go back to reference Zheng N, Fleisig GS, Escamilla RF, Barrentine SW (1998) An analytical model of the knee for estimation of internal forces during exercise. J Biomech 31(10):963–967CrossRefPubMed Zheng N, Fleisig GS, Escamilla RF, Barrentine SW (1998) An analytical model of the knee for estimation of internal forces during exercise. J Biomech 31(10):963–967CrossRefPubMed
19.
go back to reference Godest AC, de Cloke CS, Taylor M, Gregson PJ, Keane AJ, Sathasivan S, Walker PS (2000) A computational model for the prediction of total knee replacement kinematics in the sagittal plane. J Biomech 33(4):435–442CrossRefPubMed Godest AC, de Cloke CS, Taylor M, Gregson PJ, Keane AJ, Sathasivan S, Walker PS (2000) A computational model for the prediction of total knee replacement kinematics in the sagittal plane. J Biomech 33(4):435–442CrossRefPubMed
20.
go back to reference Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon) 17(2):140–146CrossRef Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol, Avon) 17(2):140–146CrossRef
21.
go back to reference Rawlinson JJ, Bartel DL (2002) Flat medial-lateral conformity in total knee replacements does not minimize contact stresses. J Biomech 35(1):27–34CrossRefPubMed Rawlinson JJ, Bartel DL (2002) Flat medial-lateral conformity in total knee replacements does not minimize contact stresses. J Biomech 35(1):27–34CrossRefPubMed
22.
go back to reference Sathasivam S, Walker PS (1997) A computer model with surface friction for the prediction of total knee kinematics. J Biomech 30(2):177–184CrossRefPubMed Sathasivam S, Walker PS (1997) A computer model with surface friction for the prediction of total knee kinematics. J Biomech 30(2):177–184CrossRefPubMed
23.
go back to reference Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R (2004) Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations. J Biomech 37(1):45–53CrossRefPubMed Villa T, Migliavacca F, Gastaldi D, Colombo M, Pietrabissa R (2004) Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations. J Biomech 37(1):45–53CrossRefPubMed
24.
go back to reference Yang RS, Lin HJ (2001) Contact stress on polyethylene components of a new rotating hinge with a spherical contact surface. Clin Biomech (Bristol, Avon) 16(6):540–546CrossRef Yang RS, Lin HJ (2001) Contact stress on polyethylene components of a new rotating hinge with a spherical contact surface. Clin Biomech (Bristol, Avon) 16(6):540–546CrossRef
25.
go back to reference Fernandes PR, Folgado J, Jacobs C, Pellegrini V (2002) A contact model with ingrowth control for bone remodeling around cementless stems. J Biomech 35(2):167–176CrossRefPubMed Fernandes PR, Folgado J, Jacobs C, Pellegrini V (2002) A contact model with ingrowth control for bone remodeling around cementless stems. J Biomech 35(2):167–176CrossRefPubMed
26.
go back to reference Van Rietbergen B, Huiskes R, Weinans H, Sumner DR, Turner TM, Galante JO (1993) ESB Research Award 1992. The mechanism of bone remodeling and resorption around press-fitted THA stems. J Biomech 26(4–5):369–382CrossRefPubMed Van Rietbergen B, Huiskes R, Weinans H, Sumner DR, Turner TM, Galante JO (1993) ESB Research Award 1992. The mechanism of bone remodeling and resorption around press-fitted THA stems. J Biomech 26(4–5):369–382CrossRefPubMed
27.
go back to reference Jia X, Zhang M, Li X, Lee WC (2005) A quasi-dynamic nonlinear finite element model to investigate prosthetic interface stresses during walking for trans-tibial amputees. Clin Biomech (Bristol, Avon) 20(6):630–635CrossRef Jia X, Zhang M, Li X, Lee WC (2005) A quasi-dynamic nonlinear finite element model to investigate prosthetic interface stresses during walking for trans-tibial amputees. Clin Biomech (Bristol, Avon) 20(6):630–635CrossRef
28.
go back to reference Lee WC, Zhang M, Jia X, Cheung JT (2004) Finite element modeling of the contact interface between trans-tibial residual limb and prosthetic socket. Med Eng Phys 26(8):655–662CrossRefPubMed Lee WC, Zhang M, Jia X, Cheung JT (2004) Finite element modeling of the contact interface between trans-tibial residual limb and prosthetic socket. Med Eng Phys 26(8):655–662CrossRefPubMed
29.
go back to reference Nyman JS, Hazelwood SJ, Rodrigo JJ, Martin RB, Yeh OC (2004) Long stemmed total knee arthroplasty with interlocking screws: a computational bone adaptation study. J Orthop Res 22(1):51–57CrossRefPubMed Nyman JS, Hazelwood SJ, Rodrigo JJ, Martin RB, Yeh OC (2004) Long stemmed total knee arthroplasty with interlocking screws: a computational bone adaptation study. J Orthop Res 22(1):51–57CrossRefPubMed
30.
go back to reference Genin P, Weill G, Julliard R (1993) The tibial slope. Proposal for a measurement method. J Radiol 74(1):27–33PubMed Genin P, Weill G, Julliard R (1993) The tibial slope. Proposal for a measurement method. J Radiol 74(1):27–33PubMed
31.
go back to reference Seo SS, Kim CW, Kim JH, Min YK (2013) Clinical results associated with changes of posterior tibial slope in total knee arthroplasty. Knee Surg Relat Res 25(1):25–29PubMedCentralCrossRefPubMed Seo SS, Kim CW, Kim JH, Min YK (2013) Clinical results associated with changes of posterior tibial slope in total knee arthroplasty. Knee Surg Relat Res 25(1):25–29PubMedCentralCrossRefPubMed
32.
go back to reference Brazier J, Migaud H, Gougeon F, Cotten A, Fontaine C, Duquennoy A (1996) Evaluation of methods for radiographic measurement of the tibial slope. A study of 83 healthy knees. Rev Chir Orthop Reparatrice Appar Mot 82(3):195–200PubMed Brazier J, Migaud H, Gougeon F, Cotten A, Fontaine C, Duquennoy A (1996) Evaluation of methods for radiographic measurement of the tibial slope. A study of 83 healthy knees. Rev Chir Orthop Reparatrice Appar Mot 82(3):195–200PubMed
33.
go back to reference Soyer J, Iborra JP, Pries P, Clarac JP (2001) Mid-term behavior of the bone fixation in non-cemented Miller-Galante 1 total knee arthroplasty. Rev Chir Orthop Reparatrice Appar Mot 87(1):40–49PubMed Soyer J, Iborra JP, Pries P, Clarac JP (2001) Mid-term behavior of the bone fixation in non-cemented Miller-Galante 1 total knee arthroplasty. Rev Chir Orthop Reparatrice Appar Mot 87(1):40–49PubMed
34.
go back to reference Wang ZW, Liu YL, Lin KJ, Qu TB, Dong X, Cheng CK, Hai Y (2012) The effects of implantation of tibio-femoral components in hyperextension on kinematics of TKA. Knee Surg Sports Traumatol Arthrosc 20(10):2032–2038CrossRefPubMed Wang ZW, Liu YL, Lin KJ, Qu TB, Dong X, Cheng CK, Hai Y (2012) The effects of implantation of tibio-femoral components in hyperextension on kinematics of TKA. Knee Surg Sports Traumatol Arthrosc 20(10):2032–2038CrossRefPubMed
Metadata
Title
A 3D finite element model to investigate prosthetic interface stresses of different posterior tibial slope
Authors
Yi Shen
Xiaomiao Li
Xiaodong Fu
Weili Wang
Publication date
01-11-2015
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2015
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-014-3144-9

Other articles of this Issue 11/2015

Knee Surgery, Sports Traumatology, Arthroscopy 11/2015 Go to the issue