Visualisation of a coronary stent presenting late mechanical failure in an aberrant left circumflex artery with coronary computed tomography angiography and optical coherence tomography

Gyula Szántó1, MD; Attila Thury1*, MD, PhD; Gábor Volford2, MD; Péter Hausinger1, MD; Ágnes Séllei2, MD; Imre Ungi1, MD, PhD

A 56-year-old male was admitted with NSTEMI and underwent percutaneous coronary intervention with bare metal (cobalt-chromium) stent implantation (3.0×24 mm, 16 atm, post-dilated with 3.0×20 mm non-compliant balloon, 20 atm) in the mid segment of an anomalous circumflex artery (retroaortic course from the right sinus of Valsalva). The patient remained asymptomatic for five years post-procedure, when he presented with mild atypical chest pain. Myocardial stress perfusion scintigraphy detected no transient ischaemia; however, due to his persisting symptoms, a coronary computed tomography angiography was performed. This revealed restenosis and chronic recoil in the implanted stent (Online Figure 1). On follow-up angiography prominent hinge motion at the restenotic part of the stent was also demonstrated (Moving image 1). Optical coherence tomography (OCT) examination was performed (Moving image 2). This identified the segment with chronic recoil apart from neointimal hyperplasia (Figure 1). Within this segment we detected different abnormalities of the structure of struts, which points to possible mechanical limitations of current metallic rigid stent designs. In order to alleviate his symptoms, repeat PCI with a drug-eluting balloon was performed with angiographic success.

Figure 1. Optical coherence tomography at follow-up. OCT (C7-XR Fourier domain; St. Jude Medical, St. Paul, MN, USA) of the stented anomalous circumflex was performed (20 mm/s, 100 frames/s; longitudinal view of the pullback (E) after a predilatation with a 2.0×15 mm balloon (8 atm). At distal edge (A) the stent structure seems to be intact. The cross-section at minimal lumen area (B) shows marked concentric neointimal hyperplasia and is suggestive of chronic recoil. At the cross-section 3.6 mm more proximal (C) uneven circumferential distribution of struts can be seen. 3-D reconstruction of the pullback with delineation of struts and lumen border (D). Maximum intensity projection of polar transformation (F) (methodology based on Wang et al1) shows stent pseudo-fracture (strut separation related to open-cell stent design; similar to that found by Bartorelli et al2) at cross-sections marked with arrows; the upper location corresponds with uneven strut distribution shown on (C). However, real disruption of strut integrity was also visualised (blue circle) confirming stent fracture.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Online data supplement

Moving image 1. Follow-up angiography of the anomalous LCx (AP view).

Moving image 2. OCT pullback.

Online Figure 1. Coronary computed tomography angiography at follow-up. 3-D reconstructed view of epicardial arteries. A) Volume-rendered 3-D image clearly shows the anomalous origin of the LCx artery from the right coronary sinus (RCS) near the ostium of the right coronary artery (RCA). The LCx takes a retroaortic course (NCS: non-coronary sinus); its mid-proximal part was stented at index procedure. B) The stent appears bent. C) The stented segment shows apparent restenosis on curved multiplanar reconstruction of LCx. D) The enlarged image is suggestive of chronic recoil of the restenotic part of the stent, black spots inside the stent (arrowheads) suggestive of neointimal hyperplasia, whereas gaps are also detected (arrows).

Volume 8 Number 12
Apr 19, 2013
Volume 8 Number 12
View full issue


Key metrics

On the same subject

Editorial

10.4244/EIJ-E-24-00022 Jun 3, 2024
Epicardial inflow versus myocardial distribution: average regional transmural coronary flow is not enough
Johnson NP and Gould K
free

Editorial

10.4244/EIJ-E-24-00018 Jun 3, 2024
Left atrial appendage occlusion: can CCT see the leaks?
Reddy M and Shahab A
free

Viewpoint

10.4244/EIJ-D-24-00137 Jun 3, 2024
Left atrial appendage occlusion: are we targeting the right patients?
Holmes DR and Alkhouli M
free

Original Research

10.4244/EIJ-D-23-00713 Jun 3, 2024
Impact of hyperaemic stenosis resistance on long-term outcomes of stable angina in the ILIAS Registry
Boerhout C et al

Flashlight

10.4244/EIJ-D-23-00927 Jun 3, 2024
Chimney/snorkel stenting during TAV-in-TAV: bedside to bench
Brown CL et al

Original Research

10.4244/EIJ-D-23-00778 Jun 3, 2024
Effects of age on microvascular function in patients with normal coronary arteries
Jansen TP et al
Trending articles
313.48

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
42.95

Clinical research

10.4244/EIJ-D-23-00282 Jul 17, 2023
Early safety and feasibility of a first-in-class biomimetic transcatheter aortic valve - DurAVR
Kodali SK et al
30.35

EXPERT REVIEW

10.4244/EIJ-D-17-00437 Feb 20, 2018
Catheter-based therapies in acute pulmonary embolism
Schultz J et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved