Skip to content
Licensed Unlicensed Requires Authentication Published by De Gruyter July 20, 2018

Hyperspectral imaging in perfusion and wound diagnostics – methods and algorithms for the determination of tissue parameters

  • Amadeus Holmer EMAIL logo , Jörg Marotz , Philip Wahl , Michael Dau and Peer W. Kämmerer

Abstract

Blood perfusion is the supply of tissue with blood, and oxygen is a key factor in the field of minor and major wound healing. Reduced perfusion of a wound bed or transplant often causes various complications. Reliable methods for an objective evaluation of perfusion status are still lacking, and insufficient perfusion may remain undiscovered, resulting in chronic processes and failing transplants. Hyperspectral imaging (HSI) represents a novel method with increasing importance for clinical practice. Therefore, methods, software and algorithms for a new HSI system are presented which can be used to observe tissue oxygenation and other parameters that are of importance in supervising healing processes. This could offer an improved insight into wound perfusion allowing timely intervention.

  1. Author Statement

  2. Research funding: Authors state no funding involved.

  3. Conflict of interest: The hyperspectral camera described in this publication was developed by Diaspective Vision GmbH. The first, second and third author are employees of this company. In the long-term, Diaspective Vision has proprietary interest in the development of the camera system resulting in a product for routine clinical use. We certify that the clinical investigators and the co-author have no financial interests and financial arrangements with Diaspective Vision and have received no funding for the preparation of this manuscript.

  4. Informed consent: Informed consent has been obtained from all individuals.

  5. Ethical approval: The research related to human use complied with all the relevant national regulations and institutional policies, was performed in accordance with the tenets of the Helsinki Declaration, and has been approved by the local institutional review board (A 2017-0066, Ethics Committee at the University Medical Centre Rostock, Germany) including IRB information were obtained. The hyperspectral camera TIVITA® Tissue is a CE certified medical product (Class I).

References

[1] Stremitzer S, Wild T, Hoelzenbein T. How precise is the evaluation of chronic wounds by health care professionals? Int Wound J 2007;4:156–61.10.1111/j.1742-481X.2007.00334.xSearch in Google Scholar PubMed PubMed Central

[2] Fluhr JW, Junger M. The price for treatment of chronic skin diseases: different approaches. Dermatology (Basel, Switzerland) 2007;214:6–7.10.1159/000096905Search in Google Scholar PubMed

[3] Posnett J, Gottrup F, Lundgren H, Saal G. The resource impact of wounds on health-care providers in Europe. J Wound Care 2009;18:154–61.10.12968/jowc.2009.18.4.41607Search in Google Scholar PubMed

[4] Santamaria N, Ogce F, Gorelik A. Healing rate calculation in the diabetic foot ulcer: comparing different methods. Wound Repair Regen 2012;20:786–9.10.1111/j.1524-475X.2012.00818.xSearch in Google Scholar PubMed

[5] Grambow E, Dau M, Holmer A, Lipp V, Frerich B, Klar E, et al. Hyperspectral imaging for monitoring of perfusion failure upon microvascular anastomosis in the rat hind limb. Microvas Res 2018;116:64–70.10.1016/j.mvr.2017.10.005Search in Google Scholar PubMed

[6] Lu G, Fei B. Medical hyperspectral imaging: a review. J Biomed Opt 2014;19:10901.10.1117/1.JBO.19.1.010901Search in Google Scholar PubMed PubMed Central

[7] Yudovsky D, Nouvong A, Pilon L. Hyperspectral imaging in diabetic foot wound care. J Diabetes Sci Technol 2010;4:1099–113.10.1177/193229681000400508Search in Google Scholar PubMed PubMed Central

[8] Chin MS, Freniere BB, Lo Y-C, Saleeby JH, Baker SP, Strom HM, et al. Hyperspectral imaging for early detection of oxygenation and perfusion changes in irradiated skin. J Biomed Opt 2012;17:26010.10.1117/1.JBO.17.2.026010Search in Google Scholar PubMed

[9] Zuzak KJ, Schaeberle MD, Lewis EN, Levin IW. Visible reflectance hyperspectral imaging: characterization of a noninvasive, in vivo system for determining tissue perfusion. Anal Chem 2002;74:2021–8.10.1021/ac011275fSearch in Google Scholar PubMed

[10] Calin MA, Coman T, Parasca SV, Bercaru N, Savastru R, Manea D. Hyperspectral imaging-based wound analysis using mixture-tuned matched filtering classification method. J Biomed Opt 2015;20:46004.10.1117/1.JBO.20.4.046004Search in Google Scholar PubMed

[11] Kienle A, Michels R, Schäfer J, Fugger O. Multiscale description of light propagation in biological tissue. In: Buzug TM, Holz D, Bongartz J, Kohl-Bareis M, Hartmann U, Weber S, editors. Advances in Medical Engineering. Berlin, Heidelberg: Springer Berlin Heidelberg; 2007:355–60.10.1007/978-3-540-68764-1_59Search in Google Scholar

[12] Michels R. Verständnis des mikroskopischen Ursprungs der Lichtstreuung in biologischem Gewebe. Ulm: Universität Ulm; 2010.Search in Google Scholar

[13] Hielscher AH, Alcouffe RE, Barbour RL. Comparison of finite-difference transport and diffusion calculations for photon migration in homogeneous and heterogeneous tissues. Phys Med Biol 1998;43:1285–302.10.1088/0031-9155/43/5/017Search in Google Scholar PubMed

[14] Farrell TJ, Patterson MS, Wilson B. A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo. Med Phys 1992;19:879–88.10.1118/1.596777Search in Google Scholar PubMed

[15] Bashkatov AN, Genina EA, Tuchin VV. Optical properties of skin, subcutaneous, and muscle tissues: a review. J Innov Opt Health Sci 2011;04:9–38.10.1142/S1793545811001319Search in Google Scholar

[16] Jacques SL. Optical properties of biological tissues: a review. Phys Med Biol 2013;58:R37–61.10.1088/0031-9155/58/11/R37Search in Google Scholar PubMed

[17] Prahl S. The absorption spectra of the oxygenated and deoxygenated hemoglobin molecules. graph. Available from: URL:http://omlc.ogi.edu/spectra/hemoglobin/index.html. [cited 04/2017].Search in Google Scholar

[18] Smith AM, Mancini MC, Nie S. Bioimaging: second window for in vivo imaging. Nat Nanotechnol 2009;4:710–1.10.1038/nnano.2009.326Search in Google Scholar PubMed PubMed Central

[19] Jacques S. Extinction coefficient of melanin. Available from: URL:http://omlc.ogi.edu/spectra/melanin/extcoeff.html [cited 04/2017].Search in Google Scholar

[20] Matcher SJ, Cope M, Delpy DT. Use of the water absorption spectrum to quantify tissue chromophore concentration changes in near-infrared spectroscopy. Phys Med Biol 1994;39:177–96.10.1088/0031-9155/39/1/011Search in Google Scholar PubMed

[21] van Veen RL, Sterenborg H, Pifferi A, Torricelli A, Cubeddu R. Determination of VIS- NIR absorption coefficients of mammalian fat, with time- and spatially resolved diffuse reflectance and transmission spectroscopy. In: Biomedical Topical Meeting. Washington, DC: OSA; 2004. SF4.10.1364/BIO.2004.SF4Search in Google Scholar

[22] Bashkatov AN, Genina EA, Kochubey VI, Tuchin VV. Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm. J Phys D: Appl Phys 2005;38:2543–55.10.1088/0022-3727/38/15/004Search in Google Scholar

[23] Denstedt M, Pukstad BS, Paluchowski LA, Hernandez-Palacios JE, Randeberg LL. Hyperspectral imaging as a diagnostic tool for chronic skin ulcers. Clin Hemorheol Microcirc 2017;67: 467–4.10.3233/CH-179228Search in Google Scholar PubMed

[24] Schmidt WD, Liebold K, Fassler D, Wollina U. Contact-free spectroscopy of leg ulcers: principle, technique, and calculation of spectroscopic wound scores. J Invest Dermatol 2001;116:531–5.10.1046/j.1523-1747.2001.01297.xSearch in Google Scholar PubMed

[25] Kulcke A. Device and method for recording a hyperspectral image. European patent EP2851662A2.Search in Google Scholar

[26] Kulcke A, Holmer A, Wahl P, Siemers F, Wild T, Daeschlein G. A compact hyperspectral camera for measurement of perfusion parameters in medicine. Biomed Tech (Berl) 2018. pii: /j/bmte.ahead-of-print/bmt-2017-0145/bmt-2017-0145.xml. doi: 10.1515/bmt-2017-0145. [Epub ahead of print].10.1515/bmt-2017-0145Search in Google Scholar PubMed

[27] Myers D, McGraw M, George M, Mulier K, Beilman G. Tissue hemoglobin index: a non-invasive optical measure of total tissue hemoglobin. Critical Care (London, England) 2009;13(Suppl. 5):S2.10.1186/cc8000Search in Google Scholar PubMed PubMed Central

[28] Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet (London, England) 1986;1:307–10.10.1016/j.ijnurstu.2009.10.001Search in Google Scholar

[29] Gioux S, Mazhar A, Lee BT, Lin SJ, Tobias AM, Cuccia DJ, et al. First-in-human pilot study of a spatial frequency domain oxygenation imaging system. J Biomed Opt 2011;16:86015.10.1117/1.3614566Search in Google Scholar PubMed PubMed Central

[30] Hyttel-Sorensen S, Kleiser S, Wolf M, Greisen G. Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom. Biomed Opt Express 2013;4:1662–72.10.1364/BOE.4.001662Search in Google Scholar PubMed PubMed Central

[31] Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers. Anesth Analg 2013;117:813–23.10.1213/ANE.0b013e318297d763Search in Google Scholar PubMed

Received: 2017-08-31
Accepted: 2018-06-01
Published Online: 2018-07-20
Published in Print: 2018-10-25

©2018 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 9.5.2024 from https://www.degruyter.com/document/doi/10.1515/bmt-2017-0155/html