Assessing the performance of vessel wall tracking algorithms: the importance of the test phantom

, and

Published under licence by IOP Publishing Ltd
, , Citation K V Ramnarine et al 2004 J. Phys.: Conf. Ser. 1 199 DOI 10.1088/1742-6596/1/1/043

1742-6596/1/1/199

Abstract

There is widespread clinical interest in assessing the mechanical properties of tissues and vessel walls. This study investigated the importance of the test phantom in providing a realistic assessment of clinical wall tracking performance for a variety of ultrasound modalities. B-mode, colour Doppler and Tissue Doppler Imaging (TDI) cineloop images were acquired using a Philips HDI5000 scanner and L12-5 probe. In-vivo longitudinal sections of 30 common carotid arteries and in-vitro images of pulsatile flow of a blood mimicking fluid through walled and wall-less tissue and vessel mimicking flow phantoms were analysed. Vessel wall tracking performance was assessed for our new probabilistic B-mode algorithm (PROBAL), and 3 different techniques implemented by Philips Medical Systems, based on B-mode edge detection (LDOT), colour Doppler (CVIQ) and TDI (TDIAWM). Precision (standard deviation/mean) of the peak systole dilations for respective PROBAL, LDOT, CVIQ and TDIAWM techniques were: 15.4 ± 8.4%, 23 ± 12.7%, 10 ± 10% and 10.3 ± 8.1% for the common carotid arteries; 6.4%, 22%, 11.6% and 34.5% for the wall-less flow phantom, 5.3%, 9.8%, 23.4% and 2.7% for the C-flex walled phantom and 3.9%, 2.6%, 1% and 3.2% for the latex walled phantom. The test phantom design and construction had a significant effect on the measurement of wall tracking performance.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/1742-6596/1/1/043