Non-invasive in vivo near-infrared optical measurement of the penetration depth in the neonatal head

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation F Faris et al 1991 Clin. Phys. Physiol. Meas. 12 353 DOI 10.1088/0143-0815/12/4/005

0143-0815/12/4/353

Abstract

The non-invasive optical technique of near-infrared spectroscopy (NIRS) was used to measure the depths of light penetration at four wavelengths in the NIR region. Near-infrared absorbance measurements were carried out on ten preterm infants. The apparent absorbance data (in optical density units) collected at 775, 805, 845, and 904 nm were measured at different positions on the head. Linear relationships that satisfy the Lambert law were obtained when the apparent absorbance at a given wavelength was plotted against the inter-optrode distance. From the slopes of the resulting straight lines, the depth of penetration of the NIR light was calculated and found to be independent of the position of the probes on the head. Calculated average values of the depth of penetration of near-infrared light in the whole neonatal head ranged between 6.3 and 8.5 mm.

Export citation and abstract BibTeX RIS

10.1088/0143-0815/12/4/005