Brought to you by:
Paper

MR-guided breast radiotherapy: feasibility and magnetic-field impact on skin dose

, , , and

Published 6 August 2013 © 2013 Institute of Physics and Engineering in Medicine
, , Citation Tristan C F van Heijst et al 2013 Phys. Med. Biol. 58 5917 DOI 10.1088/0031-9155/58/17/5917

0031-9155/58/17/5917

Abstract

The UMC Utrecht MRI/linac (MRL) design provides image guidance with high soft-tissue contrast, directly during radiotherapy (RT). Breast cancer patients are a potential group to benefit from better guidance in the MRL. However, due to the electron return effect, the skin dose can be increased in presence of a magnetic field. Since large skin areas are generally involved in breast RT, the purpose of this study is to investigate the effects on the skin dose, for whole-breast irradiation (WBI) and accelerated partial-breast irradiation (APBI). In ten patients with early-stage breast cancer, targets and organs at risk (OARs) were delineated on postoperative CT scans co-registered with MRI. The OARs included the skin, comprising the first 5 mm of ipsilateral-breast tissue, plus extensions. Three intensity-modulated RT techniques were considered (2× WBI, 1× APBI). Individual beam geometries were used for all patients. Specially developed MRL treatment-planning software was used. Acceptable plans were generated for 0 T, 0.35 T and 1.5 T, using a class solution. The skin dose was augmented in WBI in the presence of a magnetic field, which is a potential drawback, whereas in APBI the induced effects were negligible. This opens possibilities for developing MR-guided partial-breast treatments in the MRL.

Export citation and abstract BibTeX RIS

1. Introduction

Magnetic resonance imaging (MRI) guidance for radiotherapy (RT) has the potential of fast, high soft-tissue contrast visualization of tumours and organs at risk (OARs), directly on the treatment table. At the UMC Utrecht a hybrid MRI/linac (MRL) is constructed which integrates a 1.5 T closed-bore MRI scanner (Philips, Best, The Netherlands) with a linear accelerator (Elekta AB, Stockholm, Sweden), mounted on a ring-shaped gantry (Raaymakers et al 2009). Several other groups are currently working on MR-guided RT in cancer treatment as well (Dempsey et al 2006, Fallone et al 2009). MRI combined with RT offers possibilities for introducing high-precision MR-guided treatments for breast cancer patients.

The standard of care for early-stage breast cancer and ductal carcinoma in situ patients is breast-conserving therapy (BCT) which consists of breast-conserving surgery (BCS), followed by whole-breast irradiation (WBI). The dose is generally delivered by intensity-modulated RT (IMRT) using two tangential beams. Furthermore, in low-risk breast cancer patients, accelerated partial-breast irradiation (APBI) studies using IMRT are ongoing (Oliver et al 2007, Njeh et al 2010, Livi et al 2010, Lewin et al 2012, Saikh et al 2012).

In both WBI and APBI, image guidance is currently performed using portal imaging or cone-beam CT (Fein et al 1996, Jaffray et al 2002). Direct visualization of the target volumes is especially important when considering boosts to the target volume, which both modalities do not allow with high tissue-contrast. MR guidance is potentially useful for on-line high-contrast visualization of the tumour bed in postoperative RT, or for tumour detection in a preoperative setting (Sabine et al 2005, Whipp and Halliwell 2008, Kirby et al 2009, Lee et al 2010, Giezen et al 2011). However, apart from developing novel targeting techniques, it is of great importance to investigate the induced effects of the magnetic field itself on the dose distribution.

In the MRL a static magnetic field is always present during treatment. Charged particles moving in a magnetic field are acted upon by the Lorentz force, perpendicularly to their velocity direction. As a consequence, secondary electrons emanating from the skin into air can be bent back, resulting in a dose increase at the surface (figure 1). This electron return effect (ERE) is clearly observed at boundaries between layers with large density differences, and can induce a significant increase of skin dose, as shown by Raaijmakers et al (2005). The magnetic field also results in a shorter build-up distance which may play a role in a higher skin dose (Raaijmakers et al 2005).

Figure 1.

Figure 1. Illustration of the ERE, for left-breast WBI by means of two tangential fields. The edges of the photon beams are depicted by the blue lines. Trajectories of secondary electrons, crossing the skin–air boundary on either exit side of the irradiated breast, are represented by the arrows. The ERE may result in a higher skin dose when comparing the situation of no magnetic field (left) to that of a non-zero magnetic field directed into the plane (right).

Standard image High-resolution image

The ERE depends on the inclinations of the beams to the skin surface, with oblique angles inducing the highest increase (Raaijmakers et al 2007) and opposing beams compensating for the effect at perpendicular interfaces (Raaijmakers et al 2005). In WBI treatments, oblique beam/surface inclinations are ubiquitous and the target volume is superficial and relatively large. This results in a large irradiated area of skin. Treatments by APBI, however, are performed with several fields which do not necessarily all have oblique orientations. Additionally, APBI fields are generally smaller since they target smaller volumes and, moreover, the target volumes are not necessarily superficial. Hence, a smaller region of skin generally receives high dose, so that induced effects of the magnetic field are expected to be less prominent in APBI relative to WBI.

Current cosmetic results for WBI and APBI are good to excellent (Taylor et al 1995, Munshi et al 2009, Lewin et al 2012, Shaikh et al 2012). Further increase in skin dose could lead to a higher rate and severity of negative side effects of RT. Complications of the skin should be minimized in the treatment of breast cancer; therefore the main objective of this study is to investigate the physical effects of the magnetic field on the skin dose for WBI and for APBI.

2. Methods and materials

2.1. Patients

Ten BCT patients treated at the RT Department of the UMC Utrecht in 2011 or 2012 were enrolled in our treatment-planning study. The median age of the women was 63 years (range: 39–72 years). One patient had breast cancer on both sides and for this study both tumours were considered separately. In total, six of the tumours were right-sided and five were located in the left breast. The majority of the excised tumours were stage T1c (7), while the rest were stage T1b (2) or stage T2 (2). No patients had tumour-positive lymph nodes. The median volume of the ipsilateral breast was 1003 cc (range: 705–2373 cc).

2.2. RT techniques

The magnetic-field-induced effects were studied for both WBI and APBI. In WBI a dose of 42.56 Gy was prescribed to the breast, delivered in 16 fractions of 2.66 Gy. The first WBI technique considered was the classical tangential two-field set-up using IMRT (WBI-2). The beam angles in WBI-2 were determined such that the two beam edges inside each patient aligned, which was done for each patient individually. In order to investigate the influence of the magnetic field on multiple beam directions, also a seven-field IMRT technique was used for WBI (WBI-7) in which the beam orientations were chosen individually for each patient. For APBI, target prescription was 10 fractions of 3.85 Gy, to a total dose of 38.5 Gy, according to the 2008 IRMA study protocol of the European Organization for Research and Treatment of Cancer Radiation Oncology Group (EORTC ROG). A seven-field IMRT technique with individually determined beam angles was chosen to for the APBI approach (APBI-7).

2.3. Volumes of interest

All patients underwent a planning CT with 3 mm slice thickness, at three weeks (median: 21 days, range: 14–50 days) after surgery, in supine RT position. The palpable breast tissue was indicated with a copper wire. For WBI, the clinical target volume for the whole breast (CTVWBI) was delineated according to the RT Oncology Group (RTOG) Breast Cancer Contouring Atlas. A 5 mm margin was applied to the CTVWBI to generate the planning target volume (PTVWBI), excluding the first 5 mm under the surface (figure 2). The median PTVWBI was 741 cc (range: 517–2028 cc). Delineations of the target volumes in APBI were performed according to the 2008 IRMA study protocol of EORTC ROG. The gross tumour volume (GTVAPBI) was contoured using preoperative diagnostic imaging, preoperative MRI in treatment position, surgical clips, and possible postoperative seroma. A 15 mm margin in all directions was applied to the GTVAPBI to generate the clinical target volume (CTVAPBI), while excluding the chest wall and skin. The PTVAPBI was delineated as the CTVAPBI with a 5 mm margin, while excluding the skin (figure 2). The median PTVAPBI was 148 cc (range: 88–248 cc). For both WBI and APBI, the delineated OARs were the heart, lungs, contralateral breast, body (comprising all unspecified tissue) and the skin (figure 2). The skin considered for the analysis of the impact of the ERE was defined to be the first 5 mm under the surface of the ipsilateral breast. It was further extended up to 35 mm in anterior, posterior, medial, and lateral directions relative to the PTVWBI in order to include skin tissue expected to be irradiated in WBI-2 (figure 2). An extension of 15 mm was made in the caudal and cranial directions to include skin tissue receiving scattered dose. The median volume of the skin was 318 cc (range: 256–474 cc).

Figure 2.

Figure 2. Transversal slice of a patient CT scan with VOIs delineated: PTVWBI (red contour), GTVAPBI (purple), PTVAPBI (bright green), ipsilateral lung (yellow), contralateral lung (dark green), heart (sky blue), unspecified tissue (dark blue), contralateral breast (brown), and skin (orange, filled here with light grey). An extension made to the skin is visible here in the medial and lateral directions relative to the PTVWBI, as indicated by the white oval-shaped markings. The location of the isocentre is marked by the crosshair.

Standard image High-resolution image

2.4. IMRT planning

Specially developed MRL treatment planning software (MRLTP) was used to generate IMRT plans while taking into account the presence of a magnetic field. MRLTP is a combination of GPUMCD—which is a graphics processing unit (GPU)-oriented Monte Carlo dose calculation algorithm as described by Hissoiny et al (2011)—and the central processing unit (CPU)-based fast inverse dose optimization (FIDO) as described by Goldman et al (2009). A more detailed description of this system can be found in Bol et al (2012). FIDO has the distinct advantage that the optimization process is executed within a short time frame. The MRL's characteristics are incorporated in the MRLTP. The 6 MV photon beam can rotate in full 360° around the patient with maximum field sizes of 24 cm in caudocranial direction and 56 cm in anterior–posterior direction. As a consequence, the isocentre is fixed, at 14 cm above the treatment table. The isocentre-source distance is 142.7 cm. With MRLTP, plans can be generated at any magnetic-field strength required. IMRT plans were generated for 0 T and 1.5 T, which is the field strength in the MRL. Also, plans were calculated at the intermediate value of 0.35 T. For each IMRT technique (WBI-2, WBI-7, APBI-7) a class solution of optimization objectives was developed at 0 T, which was then applied for all patients. These sets of objectives were unaltered when applied at higher field strengths. The plans were based on the fully optimized fluence maps with a 5 mm grid resolution and the acquired dose distributions consisted of 2 × 2 × 2 mm3 voxels.

2.5. Comparison of plans

For comparison of the plans, dose–volume histograms (DVHs) for all VOIs and corresponding dose parameters were calculated. For the PTV the D95%—the percentage of the PTV receiving at least 95% of the prescribed dose—to describe PTV coverage, and the D107%, describing the amount of overdose, were calculated. The starting point of the plan comparisons was a similar D95% and D107%.

Several parameters were compared for the OARs. The mean lung dose (MLD) over two lungs was computed, which was to be a maximum of 7 Gy. For the ipsilateral lung, V10 Gy—the fraction of the volume receiving 10 Gy or more—V20 Gy and V30 Gy were derived for the WBI plans while for APBI the V5 Gy, V10 Gy and V20 Gy were found. Similar parameters were calculated for the heart. Moreover, the D2cc—maximum dose encompassing 2 cc of the structure—was calculated for the heart as an indication of the maximum dose. The range 30–40 Gy is considered to be relevant for the skin dose, so the mean dose and parameters V35 Gy and D2 cc were calculated for the skin. For the V35 Gy, box-whisker plots are derived to determine how the values at different magnetic-field strengths are distributed. Also, the D2 cc of the unspecified tissue (body) and the mean dose of the contralateral breast were acquired, both of which were to be kept low. The statistical significance of differences between all dosimetric parameters was tested by performing a paired student's t-test. A value of p < 0.05 was considered to represent a statistically significant result.

Additionally, dose-difference maps for all plans were derived, which depict for the same patient the dose differences per voxel relative to the situation at 0 T, i.e. dose values at 0.35 T versus 0 T and at 1.5 T versus 0 T. This allows further analysis of the spatial distribution of any dose differences caused by the magnetic field.

3. Results

3.1. Optimization process

IMRT plans were successfully generated using MRLTP. The time required to calculate the Monte Carlo beamlets varied with size of the PTV, the number of beam angles and the magnetic-field strength (i.e. number of electron steps). For this, a computer with 4 GB RAM, a 2.27 GHz CPU, and a GeForce GX 580 GPU was used. Beamlet calculations for a WBI-2 plan took 15 min at 0 T which went up to 25 min at 1.5 T, while a typical WBI-7 set-up took 40 min for 0 T and up to an hour for a 1.5 T plan. Calculating beamlets for APBI-7 typically took 20 min for a 0 T plan, up to 25 min for a plan at 1.5 T.

For each of the three RT techniques, a class solution of objectives was employed for the optimization of all plans. Once a suitable template of objective parameters was found for one patient, an automated script was activated to apply it for optimizing the plans of all other patients. Some individual plans required additional optimization. With the use of a 32 GB RAM, 3.40 GHz, eight-core CPU, the optimization time of FIDO for a WBI-2 plan or an APBI-7 plan was typically in the order of a few seconds, while a WBI-7 plan took only half a minute to optimize.

3.2. Induced effects for WBI-2

At the three different magnetic-field strengths 0 T, 0.35 T and 1.5 T, similar target coverage is achieved. Furthermore, no target overdose is observed (table 1).

Table 1. Dose parameters calculated for WBI-2 (columns 2–4) and WBI-7 (columns 5–7), at 0 T, 0.35 T, and 1.5 T, respectively, for all breast tumours (n = 11). Prescription was 16 × 2.66 Gy = 42.56 Gy. Values are denoted as: 'mean (standard deviation)'.

  WBI-2 WBI-7
Dose parameter 0 T 0.35 T 1.5 T 0 T 0.35 T 1.5 T
PTV            
D95% (%) 96.5 (0.5) 96.7 (0.7) 96.7 (0.6) 95.9 (0.6) 96.1 (0.7) 96.1 (0.7)
D107% (%) 0.0 (0.0) 0.0 (0.1) 0.0 (0.1) 0.1 (0.3) 0.1 (0.3) 0.1 (0.2)
Ips. lung            
MLD (Gy) 2.9 (0.9) 3.0 (0.9) 3.0 (0.9) 5.2 (1.1) 5.8 (1.1) 5.9 (1.1)
V10 Gy (%) 16.7 (4.9) 16.2 (4.8) 16.2 (4.7) 36.0 (7.7) 35.5 (7.6) 36.6 (7.8)
V20 Gy (%) 13.6 (4.5) 13.5 (4.5) 13.6 (4.4) 17.3 (7.3) 17.2 (7.0) 17.4 (6.9)
V30 Gy (%) 7.1 (3.7) 8.1 (3.8) 9.3 (3.8) 3.5 (3.1) 3.7 (3.2) 3.9 (3.1)
Heart            
D2 cc (Gy) 20.5 (18.6) 20.2 (18.5) 19.9 (18.5) 28.3 (9.2) 28.3 (9.1) 28.2 (9.0)
V25 Gy (%) 1.5 (2.0) 1.4 (2.0) 1.4 (2.0) 3.4 (4.3) 3.1 (3.9) 3.2 (4.1)
Unspecified tissue            
D2 cc (Gy) 46.0 (2.7) 46.7 (2.7) 47.2 (2.8) 45.0 (2.3) 45.0 (2.4) 45.2 (2.4)
Contr. breast            
Mean (Gy) 0.2 (0.3) 0.2 (0.3) 0.2 (0.3) 3.3 (1.7) 3.1 (1.6) 3.2 (1.7)
Skin            
D2 cc (Gy) 43.3 (0.6) 44.6 (1.0) 45.6 (1.1) 43.3 (0.3) 44.4 (1.4) 45.6 (2.9)
Mean (Gy) 29.5 (1.4) 32.3 (1.6) 33.2 (1.7) 27.9 (1.4) 30.2 (1.7) 29.8 (1.3)

D95% = fraction of the volume receiving at least 95% of the prescribed dose, or 40.43 Gy. D107% = fraction of the volume receiving at least 107% of the prescribed dose, or 45.54 Gy. Ips. lung = ipsilateral lung. MLD = mean lung dose (over both lungs). VXGy = fraction of volume receiving at least X Gy. D2 cc = dose received by 2cc of the structure. Contr. breast = contralateral breast.

Relative to the situation at 0 T, the mean skin dose in WBI-2 is raised by 9.5% (p < 0.01) and 12.5% (p < 0.01) at 0.35 T and 1.5 T, respectively (table 1). This dose increase is clearly visible in a typical DVH and dose-difference maps as shown in figures 3(a) and 4(a), (b), respectively. It can also be observed in figures 4(a), (b) that the increase in skin dose is especially manifested at the most superficial voxels of the skin. The most significant change in dose distribution in the skin is found in the range of around 30–40 Gy (figure 3(a)). Further analysis shows that the ERE causes the skin volume receiving more than 35 Gy (V35 Gy) at both non-zero magnetic-field strengths to increase considerably. This can be observed in the box-whisker plot in figure 5(a), where the whiskers of the boxes of V35 Gy at both 0.35 T and 1.5 T do not overlap with those at 0 T, indicating significant differences (p < 0.01). Also, the maximum skin dose (D2 cc) is slightly raised, with an increase of 2.3 Gy at 1.5 T compared to no magnetic field (table 1).

Figure 3.

Figure 3. DVHs at different field strengths, 0 T (full), 0.35 T (dash-dotted), and 1.5 T (dashed), acquired from the plans for one typical patient. For each separate RT technique, (a) WBI-2; (b) WBI-7; (c) APBI-7, the PTV and OARs, including the skin, are shown. See legend for colour specifications. Here, 'ips.' = ipsilateral, 'contr.' = contralateral, 'unspec. tissue' = all unspecified tissue. For WBI, the skin at different magnetic-field strengths is also indicated by arrows with corresponding text boxes. Prescribed dose for WBI and APBI was 42.56 Gy and 38.5 Gy, respectively.

Standard image High-resolution image
Figure 4.

Figure 4. Maps of dose differences (in Gy) per voxel relative to the situation of no magnetic field. Examples of transversal slices are depicted in each consecutive row for each RT technique on three different patients, while the different magnetic-field strengths are arranged per column, i.e. (a) WBI-2 at 0.35 T versus 0 T; (b) WBI-2 at 1.5 T versus 0 T; (c) WBI-7 at 0.35 T versus 0 T; (d) WBI-7 at 1.5 T versus 0 T; (e) APBI-7 at 0.35 T versus 0 T; (f) APBI-7 at 1.5 T versus 0 T. Voxel size is 2 × 2 × 2 mm3. Differences range from −5 Gy (dark blue) to +10 Gy (dark red).

Standard image High-resolution image
Figure 5.

Figure 5. Box-whisker plots of the skin dose parameter V35 Gy, as obtained from all plans (11 breast tumours). The box whiskers are represented for (a) WBI-2; (b) WBI-7; (c) APBI-7. They are evaluated for field strengths 0 T (red), 0.35 T (orange), and 1.5 T (blue). The line in each box represents the median; the lower and upper quartiles determine the height of each box—or interquartile range; the minimum and maximum values are represented by the whiskers. The p-values acquired from the student's paired t-test are depicted to indicate significances relative to 0 T.

Standard image High-resolution image

The MLD in WBI-2 is low (∼3.0 Gy) and unaffected by the magnetic field (table 1). This is also true for the V10 Gy of the ipsilateral lung. However, it receives more high dose at increasing field strength as indicated by the V30 Gy going up by 2.2% (p < 0.01) at 1.5 T relative to 0 T. This high-dose volume can also be observed in the dose-difference maps of figures 4(a), (b). The heart dose is unaffected by the presence of the magnetic field in WBI-2. Both the V25 Gy and D2 cc of the heart remain constant. This is also true for the contralateral-breast mean dose, which remains very low. The unspecified tissue receives a higher dose locally as indicated by the D2 cc which is on average 1.2 Gy more (p = 0.02) at 1.5 T than at 0 T (table 1).

3.3. Induced effects for WBI-7

Similar target coverage is achieved for all patients in WBI-7 at all field strengths and, again, no target overdoses are observed (table 1).

Although the mean skin dose in WBI-7 is lower than in WBI-2, it is raised by the ERE. This augmentation, relative to the 0 T situation, is on average 8.2% (p < 0.01) and 6.8% (p < 0.01) at 0.35 T and 1.5 T, respectively (table 1). The fact that this increase is less severe than in WBI-2 can be seen in the typical DVH example in figure 3(b) and the dose-difference maps of figures 4(c), (d). The principle change in skin dose is in the range of around 30–40 Gy, which was also observed for WBI-2. Indeed, V35 Gy is increased considerably (p < 0.01) at 0.35 T and at 1.5 T. This can also be seen in the box-whisker plot in figure 5(b), where the interquartile ranges (height of the boxes) of V35 Gy, at both 0.35 T and at 1.5 T, do not overlap with those at 0 T, indicating that there is a significant difference between the medians. It nevertheless shows a smaller increase compared to the tangential technique. The D2 cc of the skin is similar to that in WBI-2 and is in presence of a magnetic field raised by a similar amount (table 1).

The MLD in WBI-7 is higher than in WBI-2 and is increased with respectively 0.6 Gy and 0.7 Gy at 0.35 T and 1.5 T (table 1). The volumes of the ipsilateral lung receiving at least 10 Gy and 20 Gy, respectively, are larger than in WBI-2 and remain almost constant. However, the V30 Gy is smaller relative to WBI-2, and is only enlarged from on average 3.5% (p = 0.17) at 0 T to 3.9% (p = 0.11) at 1.5 T. Due to the beam orientation in WBI-7, the V25 Gy and D2 cc for the heart and mean dose of the contralateral breast are higher than for WBI-2, but they remain constant with increasing magnetic-field strengths. Also, the maximum dose indicated by D2 cc of the unspecified tissue is increased with on average only 0.2 Gy more (p = 0.02) at 1.5 T than at 0 T (table 1). This effect is smaller relative to WBI-2 (figures 4(c), (d)).

3.4. Induced effects for APBI-7

For the APBI-7 plans, similar dose coverage is achieved for all magnetic-field values. Furthermore, no overdose is observed for the three situations (table 2).

Table 2. Dose parameters calculated for APBI-7 at 0 T, 0.35 T, and 1.5 T, respectively, for all breast tumours (n = 11). Prescription was 10 × 3.85 Gy = 38.5 Gy. Values are denoted as: 'mean (standard deviation)'.

  APBI-7
Dose parameter 0 T 0.35 T 1.5 T
PTV      
D95% (%) 97.0 (0.5) 97.0 (0.5) 97.0 (0.7)
D107% (%) 0.0 (0.0) 0.0 (0.0) 0.0 (0.0)
Ips. lung      
MLD (Gy) 2.1 (1.0) 2.0 (1.0) 1.8 (0.8)
V5 Gy (%) 25.1 (12.9) 23.1 (12.6) 20.3 (9.6)
V10 Gy (%) 10.3 (6.1) 9.8 (7.1) 8.0 (4.5)
V20 Gy (%) 2.3 (1.5) 2.3 (1.6) 1.9 (1.3)
Heart      
D2 cc (Gy) 6.9 (2.5) 6.2 (2.0) 5.8 (2.5)
V5 Gy (%) 8.0 (13.3) 6.2 (10.2) 6.0 (10.9)
V10 Gy (%) 0.4 (1.2) 0.0 (0.0) 0.0 (0.0)
Unspecified tissue      
D2cc (Gy) 37.9 (0.5) 38.2 (0.6) 38.4 (0.7)
Contr. breast      
Mean (Gy) 1.1 (0.5) 1.0 (0.4) 1.4 (0.6)
Skin      
D2cc (Gy) 35.5 (4.7) 35.2 (4.9) 35.6 (4.4)
Mean (Gy) 5.2 (2.1) 5.6 (2.4) 5.8 (2.4)

D95% = fraction of the volume receiving at least 95% of the prescribed dose, or 36.6 Gy. D107% = fraction of the volume receiving at least 107% of the prescribed dose, or 41.2 Gy. Ips. lung = ipsilateral lung. MLD = mean lung dose (over both lungs). VXGy = fraction of volume receiving at least X Gy. D2cc = dose received by 2 cc of the structure. Contr. breast = contralateral breast.

The impact of the ERE on the skin dose is less prominent in APBI than in WBI. The mean skin dose in APBI-7 is low when compared to the WBI techniques and is only slightly raised when a magnetic field is present (table 2). This can also be observed in the typical DVH in figure 3(c) and is also reflected in the dose-difference maps of figures 4(e), (f). Since the skin dose is already observed to be low, the V35 Gy for the skin in APBI-7 is low and does not increase considerably (figure 5(c)). This can be clearly seen by the boxes of V35 Gy at 0.35 T and at 1.5 T both largely overlapping with the box at 0 T (p = 0.13 and p < 0.01, respectively). Although the increase at 1.5 T is significant, it does not reflect a clinically relevant difference since the absolute change is very small. The maximum skin dose (D2 cc) is lower when compared to WBI and remains unchanged at non-zero magnetic-field strengths (table 2).

The MLD in APBI-7 is low (∼2 Gy) and remains almost constant for non-zero field strengths. Furthermore, there are no observed significant differences (p > 0.05) in the presence of a magnetic field in the V5 Gy and V10 Gy of the ipsilateral lung, while also the V20 Gy remains unchanged. The V5 Gy, V10 Gy, and D2 cc of the heart in APBI-7 remain constant relative to 0 T at both 0.35 T and 1.5 T (p > 0.05 for all parameter differences). The heart V25 Gy stays zero in APBI-7 (not given in table 2). Relative to WBI, the mean dose of the contralateral breast remains low, whereas the unspecified tissue D2 cc seems to be raised slightly. The maximum raise is 0.5 Gy at 1.5 T compared to 0 T, however, this is statistically insignificant (p = 0.17). Small increases are visible in the dose-difference maps in figures 4(e), (f).

4. Discussion

The results of this treatment-planning study show effects on the dose distribution when treating breast-cancer patients with a 6 MV photon beam in presence of a magnetic field. The induced dosimetric effects on the skin are observed at 0.35 T and 1.5 T. A significant increase in skin dose is observed when conventional treatment by tangential-field WBI is performed. This observation could impair the clinical acceptability of the plans in the MRL. Therefore, a seven-field WBI technique was introduced, in an attempt to reduce the skin dose, as predicted in a previous film-based phantom study by Almberg et al (2011), although the other OARs would receive more dose than with the tangential technique. However, an increase in skin dose is observed there as well. For the APBI technique, the impact on the skin dose is small in comparison with WBI.

Doses to the OARs in WBI are comparable to those from other studies in the 0 T cases (Hong et al 1999, Popescu et al 2006, Johansen et al 2009, Moran et al 2009), except for the V20 Gy of the ipsilateral lung in WBI-7, which is relatively high. Differences in the dose distribution between the WBI-2 and WBI-7 techniques are mainly in the low-dose regions of the OARs as a consequence of the different beam orientations, and in the high-dose region of the ipsilateral lung. Although the MLD is higher in seven-field WBI, the high-dose volume is smaller compared to the tangential set-up, since in the tangential plans the dose is less conformal near the lung. In the seven-field set-up the mean skin dose is lower relative to the tangential-field technique. However, a magnetic-field-induced effect was still observed, thus the plans were clinically not better than for the tangential technique.

In APBI, the observed effects on the skin dose are found to be negligible in presence of a magnetic field. Moreover, the absolute dose values on the other OARs are low relative to WBI and comparable to values found in other studies at 0 T (Moon et al 2009, Moran et al 2009), while it should be kept in mind that the fractionation in APBI is different from WBI. Since the OAR doses are not affected adversely under influence of the magnetic field, it can be concluded that the performance of APBI treatments in the MRL is not impaired by the ERE.

Current cosmetic outcome for treatments by WBI after BCS is good to excellent (Taylor et al 1995, Vrieling et al 2000, Munshi et al 2009, Immink et al 2012). For APBI, at least in the studies by Lewin et al (2012) and Shaikh et al (2012), the current cosmetic outcome was shown to be good to excellent. With regard to the skin dose, the presence of the magnetic field is clearly disadvantageous for WBI. Further increases to the skin dose could result in a higher complication rate and severity of complications (Mukesh et al 2012). The actual clinical implications of the observed raised skin dose—especially in the regions receiving 30–40 Gy or more—have to be investigated further, since precise data and calculations on the biological effects of high skin dose are unavailable. However, negative side effects induced by high skin dose are observed (Hopewell 1990, Kurtz 1995, Archambeau et al 1995), implying a potential drawback when considering the performance of WBI treatments in the MRL.

A fluence-based optimization process is used in MRLTP to optimize the dose distribution, which means that the sequencing step for actual dose delivery is not included. Thus, any quality degradation of the plan caused by a sequencer is omitted. The aim of our study, however, was to investigate the induced physical influences on the dose caused by the magnetic field. Although the absolute dose values may differ slightly when including a sequencer, the physical effects on the skin dose in general are already reflected in the dose plans.

The highest dose increases in the skin were observed in the most superficial layer of the skin (figure 4). In our analysis a voxel size of 2 × 2 × 2 mm3 was used. Further investigation of the magnetic-field-induced effects on the dose distribution with a higher-resolution dose grid was performed by Oborn et al (2010). In specific phantom set-ups, layers of tiny voxels (in the order of 10 µm per voxel) were constructed and analysed for a 6 MV photon spectrum and a range of magnetic-field strengths (0 T–3.0 T) using Monte Carlo simulations. It was shown that at non-zero magnetic fields the first 70 µm under the surface receive a larger increase in skin dose than what was to be expected based on lower-resolution calculations. As a consequence, there is a possibility that even in APBI treatments, a dose peak may appear in the top layers of the skin. This should be properly measured and monitored extensively before the clinical implementation of APBI treatments in the MRL. Moreover, it was shown in Monte Carlo dose calculations by Oborn et al (2009) that the contribution of contaminant electrons to both the entry dose and entry-surface dose can cause an increase at 0 T. However, this effect is again limited to about the first 70 µm of the tissue. Contaminant electrons are not incorporated in the MRLTP.

The current study was based on static-image treatment-planning, which means that effects due to intrafractional movements are not accounted for. However, Frazier et al (2004) showed that dose plans for WBI are relatively insensitive to the effects of breast motion during normal breathing. Furthermore, in a study by George et al (2003) it was concluded that there were no significant effects observed in the IMRT delivery when respiratory motion is compensated for in breast RT. Additionally, the MRL is primarily designed for tracking the target volume and performing dynamic planning, thus compensating for all motion.

The added value of MRI for breast RT in the MRL will especially be the visualization of the tumour bed in high contrast to the surrounding anatomy. For WBI alone, the breast boundaries can also be visualized with other modalities such as kV or MV radiographs. However, the purpose of the research was to investigate the dosimetric effects of the magnetic field on RT treatments, including the standard technique of WBI, which involves relatively large field sizes. This can be of special relevance when WBI is to be performed in the MRL in combination with, for example, the delivery of a (stereotactic) boost. In future studies, more novel treatments in the MRL will be investigated as well. A single modality which can perform both standard and more sophisticated RT techniques at once could lead to a very high efficiency in the treatment of the patient.

The use of MRI allows for very precise RT performances in the MRL, from which breast-cancer patients could potentially benefit. However, from the previous discussion we find that the direct application in the MRL of currently standard treatments such as WBI is not straightforward. In both tangential and seven-field WBI, the skin-dose increase caused by the ERE is probable to cause negative side effects, which implies a potential drawback for those treatments. On the other hand, in the APBI set-up considered, negative side effects are much more unlikely, which makes APBI in the MRL feasible. APBI treatments could fully benefit from the capabilities of MRI, such as the high soft-tissue contrast, enabling direct, on-line position verification of the target area itself. This opens new possibilities for developing novel treatment techniques for breast cancer, aiming directly at the primary tumour location (Schmitz et al 2012). Therefore, our future research will focus on MR-guided partial-breast treatments.

5. Conclusion

The treatment of patients in the MRL using a conventional tangential WBI set-up or a seven-field WBI technique, at 0.35 T or 1.5 T, implies an increase in skin dose of the ipsilateral breast. In APBI treatments the skin dose and other OAR doses are relatively low and only minimally affected by the magnetic field. This opens new possibilities for developing MR-guided partial-breast treatment techniques in the MRL.

Please wait… references are loading.
10.1088/0031-9155/58/17/5917