Optimizing functional parameter accuracy for breath-hold DCE-MRI of liver tumours

, , , , , and

Published 17 March 2009 2009 Institute of Physics and Engineering in Medicine
, , Citation Matthew R Orton et al 2009 Phys. Med. Biol. 54 2197 DOI 10.1088/0031-9155/54/7/023

0031-9155/54/7/2197

Abstract

Dynamic contrast enhanced magnetic resonance imaging (DCE-MRI) is a valuable tool for assessing treatment response to novel cancer therapeutics. With appropriate data acquisition, quantitative functional parameter estimates can be obtained by fitting a model to the data. This research focuses on applying a dual-input single-compartment pharmacokinetic model to breath-hold DCE-MRI imaging of the liver. In this paper, the use of two breath-holds, providing greater temporal information, is compared with a single breath-hold approach. Computer simulations are used to assess the accuracy, precision and sensitivity to input function errors obtained for parameters estimated from the two imaging protocols. Data from ten patients were analysed to assess the noise statistics obtained from the two breath-hold protocols. The noise statistics were used with a pharmacokinetic liver model to simulate data, from which the estimation accuracy, precision and sensitivity for the two protocols were assessed. Data from the ten patients were also analysed, and the estimates were compared with literature values. This work demonstrates the feasibility of obtaining functional liver perfusion estimates over a 3D volume using a sequential breath-hold protocol. The simulation results show that the protocol consisting of two images per breath-hold is to be preferred as it requires identical patient co-operation, but provides parameter estimates that have superior accuracy and precision.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/0031-9155/54/7/023