Simultaneous PET and MR imaging

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation Yiping Shao et al 1997 Phys. Med. Biol. 42 1965 DOI 10.1088/0031-9155/42/10/010

0031-9155/42/10/1965

Abstract

We have developed a prototype PET detector which is compatible with a clinical MRI system to provide simultaneous PET and MR imaging. This single-slice PET system consists of 48 LSO crystals in a 38 mm diameter ring configuration that can be placed inside the receiver coil of the MRI system, coupled to three multi-channel photomultipliers housed outside the main magnetic field via 4 m long and 2 mm diameter optical fibres. The PET system exhibits 2 mm spatial resolution, 41% energy resolution at 511 keV and 20 ns timing resolution. Simultaneous PET and MR phantom images were successfully acquired.