Abstract
Objective
Alternative sacrum fixation with double-plate osteosynthesis in vertical unstable pelvic fractures. The surgical technique allows anatomic reduction and osteosynthesis of the sacrum component.
Indications
All vertical unstable pelvic ring fractures 61-C type according to AO/OTA system with associated displaced sacrum fractures
Contraindications
Hemodynamic unstable patients, fractures with major sacroiliac joint involvement, nondisplaced sacrum fractures.
Surgical technique
In prone position, a midline incision exposing the sacrum fracture is needed. The fracture is distracted and interjacent bone fragments or hematoma can be evacuated. The fracture is anatomically reduced and compressed. Two anatomically bent 3.5-mm locked compression plates (LCP) are placed on the dorsal aspect of the sacrum. In Denis type II and III fractures, the lateral screws can be placed in the lateral aspect of the sacrum. In Denis type I fractures, the lateral screws are placed in the ilium lateral to the sacroiliac joint to improve purchase. Additional short monocortical locking screws in the middle part of the plates increase the vertical stability to the osteosynthesis. Fracture fixation of the anterior ring follows the dorsal procedure.
Postoperative management
The patient is mobilized with partial weight bearing. Clinical and radiological follow-up is planned after 6 and 12 weeks.
Results
A total of 27 patients with vertical unstable pelvic fractures were identified and treated with open reduction and double-plate osteosynthesis. The Iowa pelvic score was rated excellent in 83 % and good in 17 % of cases. One patient developed a nonunion and had to be revised.
Zusammenfassung
Operationsziel
Alternative Fixation einer Sakrumfraktur bei vertikal instabilen Beckenfrakturen mit einer Doppelplattentechnik. Die Operation erlaubt die anatomische Reposition der Sakrumfraktur.
Indikationen
Alle vertikal instabilen Beckenringfrakturen vom Typ 61-C gemäß dem AO/OTA-System mit dislozierter Sakrumfraktur.
Kontraindikationen
Hämodynamisch instabile Patienten. Frakturen, die hauptsächlich transiliosakral verlaufen und undislozierte Sakrumfrakturen.
Operationstechnik
Der Patient wird zuerst in Bauchlage gelagert. Es folgt eine Mittellinieninzision zur submuskulären Darstellung des Sakrums. Anschließend wird die Frakturzone distrahiert und interponierende Knochenfragmente oder Hämatome werden entfernt. Die Fraktur wird reponiert und komprimiert. Zwei vorgebogene 3,5-mm-LCP („locked compression plates“) werden in direktem Knochenkontakt auf das Sakrum appliziert, um die Fraktur zu überbrücken. Bei Denis-Typ-I-Frakturen werden die lateralen Schrauben lateral des Iliosakralgelenks (ISG) ins Ilium gesetzt, um ausreichenden Halt zu gewährleisten. Bei Denis-Typ-II- und -III-Frakturen können die lateralen Schrauben unmittelbar medial des ISG in das laterale Sakrum eingebracht werden. Zusätzliche kurze monokortikale Schrauben in den medialen Plattenlöchern erhöhen die vertikale Stabilität und verhindern Scherbewegungen. Die Stabilisierung des ventralen Beckenrings erfolgt üblicherweise im Anschluss.
Weiterbehandlung
Der Patient wird mit Teilbelastung der betroffenen Seite mobilisiert. Klinische und radiologische Kontrollen sind nach 6 und 12 Wochen geplant.
Ergebnisse
Retrospektiv wurden 27 Patienten mit vertikal instabiler Beckenringfraktur und offen reponierter Sakrumfraktur identifiziert. Der Iowa-Beckenscore wurde in 83 % als sehr gut und in 17 % als gut gewertet. Ein Patient zeigte eine verzögerte Frakturheilung (non-union) und musste revidiert werden.
Similar content being viewed by others
References
Berber O, Amis AA, Day AC (2011) Biomechanical testing of a concept of posterior pelvic reconstruction in rotationally and vertically unstable fractures. J Bone Joint Surg Br 93:237–244
Burgess AR, Eastridge BJ, Young JW et al (1990) Pelvic ring disruptions: effective classification system and treatment protocols. J Trauma 30:848–856
Choy WS, Kim KJ, Lee SK et al (2012) Anterior pelvic plating and sacroiliac joint fixation in unstable pelvic ring injuries. Yonsei Med J 53:422–426
Dalal SA, Burgess AR, Siegel JH et al (1989) Pelvic fracture in multiple trauma: classification by mechanism is key to pattern of organ injury, resuscitative requirements, and outcome. J Trauma 29:981–1000 (discussion 1000–1002)
Denis F, Davis S, Comfort T (1988) Sacral fractures: an important problem. Retrospective analysis of 236 cases. Clin Orthop Relat Res 227:67–81
Griffin DR, Starr AJ, Reinert CM et al (2006) Vertically unstable pelvic fractures fixed with percutaneous iliosacral screws: does posterior injury pattern predict fixation failure? J Orthop Trauma 20:S30–S36 (discussion S36)
Marsh JL, Slongo TF, Agel J et al (2007) Fracture and dislocation classification compendium—2007: Orthopaedic Trauma Association classification, database and outcomes committee. J Orthop Trauma 21:S1–S133
Matta JM, Saucedo T (1989) Internal fixation of pelvic ring fractures. Clin Orthop Relat Res 242:83–97
Nepola JV, Trenhaile SW, Miranda MA et al (1999) Vertical shear injuries: is there a relationship between residual displacement and functional outcome? J Trauma 46:1024–1029 (discussion 1029–1030)
Osterhoff G, Ossendorf C, Wanner GA et al (2011) Percutaneous iliosacral screw fixation in S1 and S2 for posterior pelvic ring injuries: technique and perioperative complications. Arch Orthop Trauma Surg 131:809–813
Pohlemann T, Tscherne H, Baumgartel F et al (1996) Pelvic fractures: epidemiology, therapy and long-term outcome. Overview of the multicenter study of the Pelvis Study Group. Unfallchirurg 99:160–167
Templeman D, Goulet J, Duwelius PJ et al (1996) Internal fixation of displaced fractures of the sacrum. Clin Orthop Relat Res 329:180–185
Templeman DC, Simpson T, Matta JM (2005) Surgical management of pelvic ring injuries. Instr Course Lect 54:395–400
Tile M (1996) Acute pelvic fractures: I. Causation and classification. J Am Acad Orthop Surg 4:143–151
Tile M (1996) Acute pelvic fractures: II. Principles of management. J Am Acad Orthop Surg 4:152–161
Compliance with ethical guidelines
Conflict of interest. Y. P Acklin, G. Marco and C. Sommer state that there are no conflicts of interest. The accompanying manuscript does not include studies on humans or animals.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Acklin, Y., Marco, G. & Sommer, C. Double locking plate fixation of sacral fractures in unstable pelvic ring C-type injuries. Oper Orthop Traumatol 27, 74–79 (2015). https://doi.org/10.1007/s00064-014-0307-2
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00064-014-0307-2