First-In-Man Procedural Experience with the Novel EmboTrap® Revascularization Device for the Treatment of Ischemic Stroke—A European Multicenter Series

  • Original Article
  • Published:
Clinical Neuroradiology Aims and scope Submit manuscript

Abstract

Purpose

We describe safety and efficacy aspects of mechanical thrombectomy with the novel stent-type clot retrieval device EmboTrap for revascularization of large-artery occlusions in acute ischemic stroke.

Materials and Methods

A total of 40 patients with acute ischemic stroke due to large-artery occlusion were treated in five European centers. Clinical and procedural data were collected; self-reported angiographic results and neurologic outcome (discharge and 90 days) were assessed in a standardized manner.

Results

The median National Institutes of Health Stroke Scale at admittance was 16; no patient was treated after 12 h. Intravenous thrombolysis (IVT) was received by 65 % of patients. The territorial distribution was as follows: 87.5 % in the anterior circulation: 62.5 % middle cerebral artery and 25 % internal carotid artery (ICA) terminus, comprising 20 % tandem occlusions; and 12.5 % in the posterior circulation, all of which were vertebrobasilar. Revascularization of TICI 2b–3 was achieved in 38 of 40 (95 %) treatments. Additional devices were utilized in 11 of 40 (28 %) cases after a mean of 2.6 passes with EmboTrap prior to switching. The mean number of EmboTrap passes needed was 1.8, with a mean procedural time to reperfusion of 54 min. One patient (2.5 %), who had not received IVT, experienced symptomatic intracranial hemorrhage post-interventionally. Procedural complications were encountered in 2 of 40 cases (5 %); both patients exhibited ICA dissection that was treated conservatively without clinical sequelae. There were no device-related complications. Of 23 available patients, 8 (35 %) had a good outcome after 90 days.

Conclusion

The procedural results from these five centers suggest that EmboTrap is technically safe. Successful recanalization rates can be expected to be within the range of other stent retrievers.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

Abbreviations

BGC:

Balloon guide catheter

ENDOSTROKE:

Endovascular Stroke Treatment Registry

IQR:

Interquartile range

IVT:

Intravenous thrombolysis

NASA:

North American Solitaire Stent Retriever Acute Stroke Registry

sICH:

Symptomatic intracranial hemorrhage

TICI:

Thrombolysis in Cerebral Infarction

References

  1. Castano C, Dorado L, Guerrero C, Millan M, Gomis M, Perez de la Ossa N, Castellanos M, García MR, Domenech S, Dávalos A. Mechanical thrombectomy with the Solitaire AB device in large artery occlusions of the anterior circulation: a pilot study. Stroke. 2010;41(8):1836–40. doi:10.1161/STROKEAHA.110.584904.

    Article  PubMed  Google Scholar 

  2. Castano C, Serena J, Davalos A. Use of the new SolitaireTM AB device for mechanical thrombectomy when Merci Clot Retriever has failed to remove the clot. A case report. Interv Neuroradiol. 2009;15(2):209–14.

    CAS  PubMed  PubMed Central  Google Scholar 

  3. Costalat V, Machi P, Lobotesis K, Maldonado I, Vendrell JF, Riquelme C, Mourand I, Milhaud D, Héroum C, Perrigault PF, Arquizan C, Bonafé A. Rescue, combined, and stand-alone thrombectomy in the management of large vessel occlusion stroke using the solitaire device: a prospective 50-patient single-center study: timing, safety, and efficacy. Stroke. 2011;42(7):1929–35. doi:10.1161/STROKEAHA.110.608976.

    Article  PubMed  Google Scholar 

  4. Koh JS, Lee SJ, Ryu CW, Kim HS. Safety and efficacy of mechanical thrombectomy with solitaire stent retrieval for acute ischemic stroke: a systematic review. Neurointervention. 2012;7(1):1–9. doi:10.5469/neuroint.2012.7.1.1.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Miteff F, Faulder KC, Goh AC, Steinfort BS, Sue C, Harrington TJ. Mechanical thrombectomy with a self-expanding retrievable intracranial stent (Solitaire AB): experience in 26 patients with acute cerebral artery occlusion. AJNR Am J Neuroradiol. 2011;32(6):1078–81. doi:10.3174/ajnr.A2447.

    Article  CAS  PubMed  Google Scholar 

  6. Mpotsaris A, Bussmeyer M, Loehr C, Oelerich M, Buchner H, Weber W. Mechanical thrombectomy in severe acute stroke: preliminary results of the Solitaire stent. J Neurol Neurosurg Psychiatry. 2012;83(1):117–8. doi:10.1136/jnnp.2010.225573.

  7. Park H, Hwang GJ, Jin SC, Jung CK, Bang JS, Han MK, Bae HJ, Choe GY, Oh CW, Kwon OK. A retrieval thrombectomy technique with the Solitaire stent in a large cerebral artery occlusion. Acta Neurochir (Wien). 2011;153(8):1625–31. doi:10.1007/s00701-011-0999-0.

    Article  Google Scholar 

  8. Nogueira RG, Lutsep HL, Gupta R, Jovin TG, Albers GW, Walker GA, Liebeskind DS, Smith WS; TREVO 2 Trialists. Trevo versus Merci retrievers for thrombectomy revascularisation of large vessel occlusions in acute ischaemic stroke (TREVO 2): a randomised trial. Lancet. 2012;380(9849):1231–40. doi:10.1016/S0140-6736(12)61299-9.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Zaidat OO, Castonguay AC, Gupta R, Sun CH, Martin C, Holloway WE, Mueller-Kronast N, English JD, Linfante I, Dabus G, Malisch TW, Marden FA, Bozorgchami H, Xavier A, Rai AT, Froehler MT, Badruddin A, Nguyen TN, Taqi MA, Abraham MG, Janardhan V, Shaltoni H, Novakovic R, Yoo AJ, Abou-Chebl A, Chen PR, Britz GW, Kaushal R, Nanda A, Issa MA, Nogueira RG. North American Solitaire Stent Retriever Acute Stroke registry: post-marketing revascularization and clinical outcome results. J Neurointerv Surg. 2013. doi:10.1136/neurintsurg-2013-010895.

  10. Singer OC, Haring HP, Trenkler J, Nolte CH, Bohner G, Reich A, Wiesmann M, Bussmeyer M, Mpotsaris A, Neumann-Haefelin T, Hohmann C, Niederkorn K, Deutschmann H, Stoll A, Bormann A, Jander S, Turowski B, Brenck J, Schlamann MU, Petzold GC, Urbach H, Liebeskind DS, Berkefeld J. Age dependency of successful recanalization in anterior circulation stroke: the ENDOSTROKE study. Cerebrovasc Dis. 2013;36(5–6):437–45. doi:10.1159/000356213.

    Article  PubMed  Google Scholar 

  11. Higashida RT, Furlan AJ, Roberts H, Tomsick T, Connors B, Barr J, Dillon W, Warach S, Broderick J, Tilley B, Sacks D; Technology Assessment Committee of the American Society of Interventional and Therapeutic Neuroradiology; Technology Assessment Committee of the Society of Interventional Radiology. Trial design and reporting standards for intra-arterial cerebral thrombolysis for acute ischemic stroke. Stroke. 2003;34(8):e109–37. doi:10.1161/01.STR.0000082721.62796.09.

    Article  PubMed  Google Scholar 

  12. Kuntze Soderqvist A, Kaijser M, Soderman M, Holmin S, Wahlgren N, Andersson T. Mechanical thrombectomy in acute ischemic stroke-experience from 6 years of practice. Neuroradiology. 2014;56(6):477–86. doi:10.1007/s00234-014-1353-z.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Singer OC, Haring HP, Trenkler J, Nolte CH, Bohner G, Neumann-Haefelin T, Neumann-Haefelin T, Hofmann E, Reich A, Wiesmann M, Niederkorn K, Deutschmann H, Bussmeyer M, Mpotsaris A, Stoll A, Bormann A, Petzold GC, Urbach H, Jander S, Turowski B, Weimar C, Schlamann M, Gröschel K, Boor S, Berkefeld J; ENDOSTROKE Study Group. Periprocedural aspects in mechanical recanalization for acute stroke: data from the ENDOSTROKE registry. Neuroradiology. 2013;55(9):1143–51. doi:10.1007/s00234-013-1219-9.

    Article  PubMed  Google Scholar 

  14. Cohen JE, Gomori M, Rajz G, Moscovici S, Leker RR, Rosenberg S, Itshayek E. Emergent stent-assisted angioplasty of extracranial internal carotid artery and intracranial stent-based thrombectomy in acute tandem occlusive disease: technical considerations. J Neurointerv Surg. 2012;5(5):440–6. doi:10.1136/neurintsurg-2012-010340.

    Article  PubMed  Google Scholar 

  15. Machi P, Lobotesis K, Maldonado IL, Costalat V, Vendrell JF, Riquelme C, Bonafé A. Endovascular treatment of tandem occlusions of the anterior cerebral circulation with solitaire FR thrombectomy system. Initial experience. Eur J Radiol. 2012;81(11):3479–84. doi:10.1016/j.ejrad.2011.12.032.

    Article  PubMed  Google Scholar 

  16. Mpotsaris A, Bussmeyer M, Buchner H, Weber W. Clinical outcome of neurointerventional emergency treatment of extra- or intracranial tandem occlusions in acute major stroke: antegrade approach with wallstent and solitaire stent retriever. Clin Neuroradiol. 2013;23(3):207–15. doi:10.1007/s00062-013-0197-y.

    Article  CAS  PubMed  Google Scholar 

  17. Soize S, Kadziolka K, Estrade L, Serre I, Barbe C, Pierot L. Outcome after mechanical thrombectomy using a stent retriever under conscious sedation: comparison between tandem and single occlusion of the anterior circulation. J Neuroradiol. 2014;41(2):136–42. doi:10.1016/j.neurad.2013.07.001.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Mpotsaris MD.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kabbasch, C., Mpotsaris, A., Liebig, T. et al. First-In-Man Procedural Experience with the Novel EmboTrap® Revascularization Device for the Treatment of Ischemic Stroke—A European Multicenter Series. Clin Neuroradiol 26, 221–228 (2016). https://doi.org/10.1007/s00062-014-0352-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00062-014-0352-0

Keywords

Navigation