Abstract
Objective
To determine the reproducibility of laser Doppler flowmetry coupled with intra-dermal saline delivery.
Methods
Delivery of saline was judged visually by two operators (n=100), using a graduated syringe (Becton-Dickinson), by expelling saline onto a weighing boat. Volume was assessed by weight. Skin blood flow following intra-dermal injection of saline was assessed in 18 healthy volunteers; 10 attended twice to assess between-day reproducibility, and 8 attended once to assess between-site reproducibility. Results are expressed as mean value±SEM and 95% confidence interval for mean differences.
Results
There was no difference between operators in mean injection weight, both weights being 10.3±0.1 mg (mean difference 0.08, 95% confidence interval, CI −0.23 to 0.39 mg; n=100, P=0.9). Intra-dermal saline caused a nine-fold increase in blood flow (0.03±0.003 to 0.27±0.02 perfusion units, PU; n=18, P<0.001). This response had a rapid onset, with the maximal effect seen at 4 min and a duration of greater than 30 min. There was no difference in the magnitude of the response between the dominant and non-dominant arms, AUC was 2.9±0.4 and 2.9±0.4, respectively (mean difference −0.05, 95% CI −0.8 to 0.73 PU; n=18, P=0.93). However, there was a trend towards differences between study visits 1 and 2: AUC was 3.2±0.6 and 2.0±0.5, respectively (mean difference 1.2, 95% CI −0.03 to 2.43 PU; n=10, P=0.7). There was no difference in the magnitude of responses between different sites on the forearm (n=64, P=0.6).
Conclusions
These studies demonstrate that the technique of laser Doppler flowmetry coupled with intra-dermal injection is a safe, well-tolerated technique with good reproducibility. A trend towards reduced between-day reproducibility emphasizes the importance of vehicle control sites when investigating the effects of vasoactive compounds. This technique provides a reliable method for the intra-dermal delivery of drugs, despite the direct effect of injection of saline on blood flow.
Similar content being viewed by others
References
Stern MD (1975) In vivo evaluation of microcirculation by coherent light scattering. Nature 254:56–58
Vongsavan N, Matthews B (1993) Some aspects of the use of laser Doppler flow meters for recording tissue blood flow. Exp Physiol 78:1–14
Schabauer AMA, Rookw TW (1994) Cutaneous laser Doppler flowmetry: applications and findings. Mayo Clin Proc 69:564–574
Nilsson GE, Tenland T, Oberg PA (1980) A new instrument for continuous measurement of tissue blood flow by light beating spectroscopy. IEEE Trans Biomed Eng 27:12–19
Hovell CJ, Beasley CR, Mani R, Holgate ST (1987) Laser Doppler flowmetry for determining changes in cutaneous blood flow following intradermal injection of histamine. Clin Allergy 17:469–479
Haynes WG, Cullen MJ, Webb DJ (1991) Increase in skin microcirculatory blood flow after local renin inhibition in man. J Hypertens (Suppl) 9:228–229
Wenzel RR, Noll G, Luscher TF (1994) Endothelin receptor antagonists inhibit endothelin in human skin microcirculation. Hypertension 23:581–586
Wenzel RR, Duthiers N, Noll G, Bucher J, Kaufmann U, Luscher TF (1996) Endothelin and calcium antagonists in the skin microcirculation of patients with coronary artery disease. Circulation 94:316–322
Wenzel RR, Zbinden S, Noll G, Meier B, Luscher TF (1998) Endothelin-1 induces vasodilation in human skin by nociceptor fibres and release of nitric oxide. Br J Clin Pharmacol 45:441–446
Holloway GA Jr (1980) Cutaneous blood flow responses to injection trauma measured by laser Doppler velocimetry. J Invest Dermatol 74:1–4
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i: 307–310
Weber R, Pechere-Bertschi A, Hayoz D, Gerc V, Brouard R, Lahmy JP, Brunner HR, Burnier M (1997) Effects of SR 49059, a new orally active and specific vasopressin V1 receptor antagonist, on vasopressin-induced vasoconstriction in humans. Hypertension 30:1121–1127
Acknowledgements
S.J.L. and J.T.A. were supported by British Heart Foundationships (FS/98040 and FS/2000021).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Leslie, S.J., Affolter, J., Denvir, M.A. et al. Validation of laser Doppler flowmetry coupled with intra-dermal injection for investigating effects of vasoactive agents on the skin microcirculation in man. Eur J Clin Pharmacol 59, 99–102 (2003). https://doi.org/10.1007/s00228-003-0577-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00228-003-0577-3