Abstract
Accuracy and dynamic response of seven different disposable saline-filled transducer-tubing systems were evoluated in two configurations; a pole-moant version (60-inch tubing) and patient-mount version (12-inch tubing). Natural frequencies (Fn) and damping coefficients were measured as were pressure waveform recordings. A 0.03 ml air bubble was inserted into the different systems in close proximity to the transducer and measurements were repeateeL There were dramatic changes in dynamic response with lengthening of tubing to a pole-mount configuration and air bubble additions. Only one transducertubing system had adequate accuracy and dynamic response in both the pole- and patient-mount version. Pressure waveforms in the pole-mount versions and with the air bubble produced systolic overshoots up to 35 mmHg. The performance characteristies dictate that the choice for a disposable transducer-tubing system must be made not only by design, but dynamic response which is directly related to clinical set-up.
Résumé
On a évalué la précision et la réponse dynamique de sept systèmes différents de tubulure jetable pour capteurs remplis de salin en installant les tubes de deux façons differentes: une installation sur tige (tube de 60 po,) et une installation sur les patients (tube de 12 po.). On a mesuré les fréquences naturelles (Fn), les coefficients ďamortissement, ainsi que les enrigistrements de courbes de pression. On a ensuite inséré une bulle ď air de 0.03 ml dans les différents systèmes et les mesures out été répétées. Des changements dramatiques se produisaient dans la réponse dynamique avec ľallongement du tube dans ľinstallation sur tige et avec ľajout ďune bulle ďair. Un seul systime de tubulure pour capteur avail une précision et une réponse dynamique adéquates dans ľinstallation sur tige et dans I installation sur le patient. Ľajout de bulle ďair dans ľinstallation sur tige amenait les courbes de pression à produire des dépassements systoliques allant jusqu’à 35 mmHg. Les caractéristiques de fonctionnement nous indiquent que le choix ď un système ď une tubulure jetable pour capteur ne doit pas se faire seulement selon le modèle (design), mais doit sefaire selon la réponse dynamique qui est directement reliée au montage clinique.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Health Devices Disposable pressure transducers. Emergency Care Research Institute, 5200 Butler Pike, Plymouth Meeting, PA, 19462, U.S.A., vol. 13, No. 11, 1984.
Gardner RM. Direct pressure measurement-dynamic response requirement. Anesthesiology 1981; 54: 227–36.
Shinozaki T, Deane RS, Mazuzan JE. The dynamic response of liquid-filled catheter systems for direct measurements of blood pressure. Anesthesiology 1980; 53: 498–504.
Boutros A, Albert S. Effect of the dynamic response to transducer-tubing system on accuracy of direct blood pressure measurement in patients. Crit Care Med, 1983; 11; 124–7.
Bruner JMP. Handbook of Blood Pressure Monitoring, Littleton, MA., PSG Publishing Co., Inc., 1978; 63–9.
Abrams JH, Olson ML, Marino JA, Cerra FB. Use of a needle valve variable resistor to improve invasive blood pressure monitoring. Crit Care Med 1984; 12: 978–82.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hunziker, P. Accuracy and dynamic response of disposable pressure transducertubing systems. Can J Anaesth 34, 409–414 (1987). https://doi.org/10.1007/BF03010146
Issue Date:
DOI: https://doi.org/10.1007/BF03010146