Abstract
Respiratory failure can be difficult to predict. It can develop into a life-threatening condition in just a few minutes, or it can build up more slowly. Thus continuous monitoring of respiratory activity should be mandatory in clinical, high-risk situations, and appropriate monitoring equipment could be life-saving. The review considers non-invasive methods and devices claimed to provide information about respiratory rate or depth, or gas exchange. Methods are categorised into those responding to movement, volume and tissue composition detection; air flow; and blood gas concentration. The merits and limitations of the methods and devices are analysed, considering information content and their ability to minimise the rate of false alarms and false non-alarms. It is concluded that the field of non-invasive respiratory monitoring is still in an exploratory phase, with numerous reports on specific device solutions but less work on evaluation and adaptation to clinical requirements. Convincing evidence of the clinical usefulness of respiratory monitors is still lacking. Devices responding only to respiratory rate, and lacking information about actual gas exchange, will have limited clinical value. Furthermore, enhancement in specificity and sensitivity to avoid false alarms and non-alarms will be necessary to meet clinical requirements. Miniature CO2 sensors are identified as one route towards substantial improvement.
Similar content being viewed by others
References
Allison, R. D., Holmes, E. L., andNyboer, J. (1964): ‘Volumetric dynamics of respiration as measured by electrical impedance plethysmography’,J. Appl. Physiol.,19, pp. 166–173
Ashutosh, K., Gilbert, R., Auchincloss, J. H., Erlebacher, J., andPeppi, D. (1974): ‘Impedance pnemograph and magnetometer methods for monitoring tidal volume’,J. Appl. Physiol.,37, pp. 964–966
Askanazi, J., Silverberg, P. A., Foster, R. J., Hyman, A. I., Milic-Emili, J., andKinney, J. M. (1980): ‘Effects of respiratory apparatus on breathing pattern’,J. Appl. Physiol.,48, pp. 577–580
Barrow, R. E., Vorwald, A. J., andDomeier, E. (1969): ‘Capacitance respirometry’,Arch. Environ. Health,19, pp. 579–585
Barschdorff, D., andZhang, W. (1994): ‘Respiratory rhythm detection with photoplethysmographic methods’,IEEE, pp. 912–913
Bennett, A. D. (2002): ‘Home apnoea monitoring for infants. A discussion of primary care issues’,Adv. Nurse Pract.,10, pp. 48–53
Bierman, M. I. (1995): ‘Respiratory monitoring’, inDantzkner, D. R., MacIntyre, N. R., andBakow, E. D. (Eds): ‘Comprehensive respiratory care’ (WB Saunders Company, Philadelphia, 1995), pp. 243–264
Brown, B. H., Smallwood, R. H., Barber, D. C., Lawford, P. V., andHose, D. R. (1999a): ‘Medical physics and biomedical engineering’ (Institute of Physics Publishing, London, 1999), pp. 556–558
Brown, B. H., Smallwood, R. H., Barber, D. C., Lawford, P. V., andHose, D. R. (1999b): ‘Medical physics and biomedical engineering’ (Institute of Physics Publishing, London, 1999), pp. 567–569
Burdett-Smith, P. (1997): ‘Always check the respiratory rate’,Br. Med. J.,314, p. 1549
Cohen, K. P., Ladd, W. M., Beams, D. M., Sheers, W. S., Radwin, R. G., Tompkins, W. J., andWebster, J. G. (1997): ‘Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction’,IEEE Trans. Biomed. Eng.,44, pp. 555–566
Cop, W. (1988): ‘Methods devices used in ventilatory monitoring’,Encyl. Med. Dev. Instrum.,4, pp. 2870–2877
Coté, C. J., Rolf, N., Liu, L. M. P., Goudsouzian, N. G., Ryan, J. F., Zaslavsky, A., Gore, R., Todres, D., Vassallo, S., Polaner, D., andAlifimoff, J. K. (1991): ‘A single-blind study of combined pulse oximetry and capnography in children’,Anesthesiology,74, pp. 980–987
Cyna, A. M., Kulkarni, V., Tunstall, M. E., Hutchison, J. M. S., andMallard, J. R. (1991): ‘Aura: A new respiratory monitor and apnoea alarm for spontaneously breathing patients’,Br. J. Anaesth.,67, pp. 341–345
Davies, A., Blakeley, A. G. H., andKidd, C. (2001): ‘Human physiology: The respiratory system’ (Churchill Livingstone, Edinburgh, 2001), pp. 647–706
Davis, C., Mazzolini, A., andMurphy, D. (1997): ‘A new fibre optic sensor for respiratory monitoring’,Austral. Phys. Eng. Sci. Med.,20, pp. 214–219
Dodds, D., Purdy, J., andMoulton, D. (1999): ‘The PEP transducer: a new way of measuring respiratory rate in the nonintubated patient’,J. Accid. Emerg. Med.,16, pp. 26–28
Drummond, G. B., Nimmo, A. F., andElton, R. A. (1996): ‘Thoracic impedance used for measuring chest wall movement in postoperative patients’,Br. J. Anaesth.,77, pp. 327–332
Eriksson, I., Berggren, L., andHallgren, S. (1986): ‘CO2 production and breathing pattern during invasive and non-invasive respiratory monitoring’,Acta Anaesthesiologica Scandinavica,30, pp. 438–443
Folke, M., Granstedt, F., Hök, B. andScheer, H. (2002): ‘Comparative provocation test of respiratory monitoring methods’,J. Clin. Monit.,17, pp. 97–103
Franks, C. I., Brown, B. H., andJohnston, D. M. (1976): ‘Contactless respiration monitoring of infants’,Med. Biol. Eng.,May, pp. 306–311
Friesen, R. H., andAlswang, M. (1996): ‘End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: Accuracy and sources of error’,J. Clin. Monit.,12, pp. 155–159
Galland, B. C., Taylor, B. J., andBolton, D. P. (2002): ‘Prone versus supine sleep position: a review of the physiological studies in SIDS research’,J. Paediatr. Child Health,38, pp. 332–338
Gilbert, R., Auchincloss, J. H., Brodsky, J., andBoden, W. (1972): ‘Changes in tidal volume, frequency, and ventilation induced by their measurement’,J. Appl. Physiol.,33, pp. 252–254
Gill, N. P., Wright, B., andReilly, C. S. (1992): ‘Relationship between hypoxaemic and cardiac ischaemic events in the perioperative period’,Br. J. Anaesth.,68, pp. 471–473
Gordh, T., Rawal, N., Ström, S., andHök, B. (1995): ‘Respiratory monitoring during postoperative analgesia’,J. Clin. Monit.,11, pp. 365–372
Gordon, D. H., andThompson, W. L. (1975): ‘A new technique for monitoring spontaneous respiration’,Med. Instrum.,9, pp. 21–22
Granstedt, F., Hök, B., Bjurman, U., Ekström, M., andBacklund, Y. (2001): ‘New CO2 sensor with high resolution and fast response’,IEEE-EMBC, Istanbul, Turkey
Hamilton, L. H., Beard, J. D., Carmean, R. E., andKory, R. C. (1967): ‘An electrical impedance ventilometer to quantitate tidal volume and ventilation’,Med. Res. Eng.,6, pp. 11–16
Hoffman, S., Jedeikin, R., andAtlas, D. (1986): ‘Respiratory monitoring with a new impedance plethysmograph’,Anaesthesia,41, pp. 1139–1142
Hök, B., Wiklund, L., andHenneberg, S. (1993): ‘A new respiratory rate monitor: development and initial clinical experience’,Int. J. Clin. Monit. Comput.,10, pp. 101–107
Johansson, A., andÖberg, P. Å. (1999a): ‘Estimation of respiratory volumes from the photoplethysmographic signal Part I: experimental results’,Med. Biol. Eng. Comput.,37, pp. 42–47
Johansson, A., andÖberg, P. Å. (1999b): ‘Estimation of respiratory volumes from the photoplethysmographic signal Part II: a model study’,Med. Biol. Eng. Comput.,37, pp. 48–53
Jones, J. G., Sapsford, D. J., andWheatley, R. G. (1990): ‘Postoperative hypoxemia: mechanisms and time course’,Anaesthesia,45, pp. 566–573
Jopling, M. W., Manheimer, P. D., andBebout, D. E. (2000): ‘Effects of severe motion on three pulse oximeters designed for use in motion’,Int. J. Intens. Care, pp. 177–179
Kavanagh, B. P., Sandler, A. N., Turner, K. E., Wick, V., andLawson, S. (1992): ‘Use of end-tidal PCO2 and transcutaneous PCO2 as noninvasive measurement of arterial PCO2 in extubated patients recovering from general anesthesia’,J. Clin. Monit.,8, pp. 226–230
Kehlet, H., andRosenberg, J. (1995): ‘Late post-operative hypoxaemia and organ dysfunction’,Eur. J. Anaesth.,12, pp. 31–34
Larsson, C., Davidsson, L., Lundin, P., Gustafsson, G., andVegfors, M. (1999): ‘Respiratory monitoring during MR imaging’,Acta Radiologica,40, pp. 33–36
Larsson, C., andStaun, P. (1999): ‘Evaluation of a new fibre-optic monitor for respiratory rate monitoring’,J. Clin. Monit. Comput.,15, pp. 295–298
Lenz, G., Heipertz, W., andEpple, E. (1991): ‘Capnometry for continuous postoperative monitoring of nonintubated, spontaneously breathing patients’,J. Clin. Monit.,7, pp. 245–248
Levene, M. I., Tudehope, D. I., andThearle, M. J. (2000): ‘Essentials of neonatal medicine. Respiratory physiology, respiratory failure and mechanical ventilation’ (Blackwell Science Ltd, Oxford, 2000), pp. 115–125
Lin, J. C. (1975): ‘Noninvasive microwave measurement of respiration’,Proc. IEEE, p. 1430
Lindberg, L.-G., Ugnell, H., andÖberg, P. Å. (1992): ‘Monitoring of respiratory and heart rates using a fibre-optic sensor’,Med. Biol. Eng. Comput.,30, pp. 533–537
Linko, K., andPaloheimo, M. (1989): ‘Monitoring of the inspired and end-tidal oxygen, carbon dioxide, and nitrous oxide concentrations: Clinical applications during anesthesia and recovery’,J. Clin. Monit.,5, pp. 149–156
Loughnan, T. E., Monagle, J., Copland, J. M., Ranjan, P., andChen, M. F. (2000): ‘A comparison of carbon dioxide monitoring and oxygenation between facemask and divided nasal cannula’,Anaesth. Intens. Care,28, pp. 151–154
Nagler, J. (2002): ‘Sudden infant death syndrome’,Curr. Opin. Pediatr.,14, pp. 247–250
Nakajima, K., Tamura, T., andMilke, H. (1996): ‘Monitoring of heart and respiratory rates by photoplethysmography using a digital filtering techniques’,Med. Eng. Phys.,18, pp. 365–372
Nilsson, L., Johansson, A., andKalman, S. (2000): ‘Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique’,J. Clin. Monit. Comput.,16, pp. 309–315
Perez, W., andTobin, M. J. (1985): ‘Separation of factors responsible for change in breathing pattern induced by instrumentation’,J. Appl. Physiol.,59, pp. 1515–1520
Rolfe, P. (1971): ‘A magnetometer respiration for use with premature babies’,Bio-med. Eng.,September, pp. 402–404
Rodate, J. R., andShardonofsky, F. R. (2000): ‘Respiratory system mechanics’ inMurray, J. F., Nadel, J. A., Mason, R. J., andBoushey, H. A. (Eds): ‘Textbook of respiratory medicine’ (WB Saunders Company, Philadelphia, 2000), pp. 91–118
Rosenberg, J., Pedersen, M. H., Ramsing, T., andKehlet, H. (1992): ‘Circadian variation in unexpected postoperative death’,Br. J. Surg.,79, pp. 1300–1302
Roy, J., McNulty, S. E., andTorjman, M. C. (1991): ‘An improved nasal prong apparatus for end-tidal carbon dioxide monitoring in awake, sedated patients’,J. Clin. Monit.,7, pp. 249–252
Sage, J., andGough, W. (1998): ‘A simple inexpensive device for monitoring patient respiration’,Med. Biol. Eng. Comput.,36, pp. 231–232
Santos, L. J., Varon, J., Pic-Aluas, L., andCombs, A. H. (1994): ‘Practical uses of end-tidal carbon dioxide monitoring in the emergency department’,J. Emerg. Med.,12, pp. 633–644
Semmes, B. J., Tobin, M. J., Snyder, J. V., andGrenvik, A. (1985): ‘Subjective and objective measurement of tidal volume in critically ill patients’,Chest,78, pp. 577–579
Severinghaus, J. W. (1987): ‘History, status and future of pulse oximetry’,Adv. Exper. Med. Biol.,220, pp. 3–8
Siivola, J. (1989): ‘New noninvasive piezoelectric transducer for recording of respiration, heart rate and body movements’,Med. Biol. Eng. Comput.,27, pp. 423–424
Sullivan, F. M., andBarlow, S. M. (2001): ‘Review of risk factors for sudden infant death syndrome’,Paediatr. Perinat. Epidemiol.,15, pp. 144–200
Tatara, T., andTsuzaki, K. (1997): ‘An apnea monitor using a rapid-response hygometer’,J. Clin. Monit.,13, pp. 5–9
Tobin, M. J. (1988): ‘Respiratory monitoring in the intensive care unit’,Am. Rev. Respir. Dis.,138, pp. 1625–1642
Vegfors, M., Ugnell, H., Hök, B., Öberg, P. Å., andLennmarken, C. (1993): ‘Experimental evaluation of two new sensors for respiratory rate monitoring’,Physiol. Meas.,14, pp. 171–181
Vegfors, M., Lindberg, L.-G., Pettersson, H., andÖberg, P. Å. (1994): ‘Presentation and evaluation of a new optical sensor for respiratory rate monitoring’,Int. J. Clin. Monit. Comput.,11, pp. 151–156
Waag-Carlson, B., Neelon, V. J., andHsiao, H. (1999): ‘Evaluation of a non-invasive respiratory monitoring system for sleeping subjects’,Physiol. Meas.,20, pp. 53–63
Ward, K. R., andYealy, D. M. (1998): ‘End-tidal carbon dioxide monitoring in emergency medicine, Part 1: basic principles’,Acad. Emerg. Med.,5, pp. 628–636
Wiedemann, H. P., andMcCarthy, K. (1989): ‘Noninvasive monitoring of oxygen and carbon dioxide’,Clin. Chest. Med.,10, pp. 239–254
Wiklund, L., Hök, B., Ståhl, K., andJordeby-Jönsson, A. (1994): ‘Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices’,J. Clin. Anesth.,6, pp. 182–188
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Folke, M., Cernerud, L., Ekström, M. et al. Critical review of non-invasive respiratory monitoring in medical care. Med. Biol. Eng. Comput. 41, 377–383 (2003). https://doi.org/10.1007/BF02348078
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02348078