Abstract
Thermal loading is a significant side effect during excimer laser surgery of the cornea. The maximal temperature rise, which is dependent on fluence and repetition rate of the radiation, amounts to 20° C in excised corneas and 7.5° C in the in situ cornea. The temperature decreases exponentially with increasing distance from the incision with a half-value distance of 610 ± 100 μm. Exposure parameters are recommended to avoid temperature rises greater than 11° C.
Similar content being viewed by others
References
Dyer P, Sidhu J (1985) Excimer laser ablation and thermal coupling efficiency to polymer films. J Appl Phys 57:1420–1422
Ito T, Ito A, Hieda K, Kobayashi K (1983) Wavelength dependence of inactivation and membrane damage toSaccharomyces cerevisiae cells by monochromatic synchrotron vacuum — UV radiation (145–190 nm). Radiat Res 96:532–548
Krueger R, Trokel S, Schubert H (1985) Interaction of ultraviolet laser light with the cornea. Invest Ophthalmol 26:1455–1464
Lewis M, Dubin M, Aandahl V (1967) The physical properties of bovine corneal collagen. Exp Eye Res 6:57–69
Marshall J, Trokel S, Rothery S, Schubert H (1985) An ultrastructural study of corneal incisions induced by an excimer laser at 193 nm. Ophthalmology 92:749–758
Seiler T, Wollensak J (1986) In vivo experiments with the excimer laser — technical parameters and healing processes. Ophthalmologica 192:65–70
Sliney D (1985) Laser-tissue interactions. Clin Chest Med 6:203–208
Srinivasan R, Leigh W (1982) Ablative photodecomposition: action of far-ultraviolet (193 nm) laser radiation on poly (-ethylene terephthalate-) films. J Am Chem Soc 104:6784–6789
Srinivasan R, Dyer P, Braren B (1985) Far-UV laser ablation of cornea: photoacoustic studies. IBM Report, Life Sci RC 11620 (no. 52044)
Trokel S, Srinivasan R, Braren B (1983) Excimer laser surgery of the cornea. Am J Ophthalmol 96:710–715
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Bende, T., Seiler, T. & Wollensak, J. Side effects in excimer corneal surgery. Graefe's Arch Clin Exp Ophthalmol 226, 277–280 (1988). https://doi.org/10.1007/BF02181196
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02181196