Towards the assessment of the limitations on computerized axial tomography

Etude sur les limitations de la tomographie axiale par ordinateur

Experimentelle Erfassung einiger Begrenzungen in der computerisierten axialen Tomographie

  • Originals
  • Published:
Neuroradiology Aims and scope Submit manuscript

Abstract

The results of two computational experiments, performed on digitised X-ray absorption data measured by the EMI scanner, are reported. In the first experiment tomograms reconstructed from the same data, but using different reconstruction algorithms, were compared. In the second, a comparison was made between tomograms reconstructed by one of the programs using different amounts of data as input. For the particular example studied, the results showed that: 1. all algorithms gave equally satisfactory results from a clinical point of view, and that 2. equally usable tomograms could be obtained with fewer measurements. The possible significance of such results in the application of computerised axial tomography in clinical neuroradiology is discussed.

Résumé

Les résultats de deux expériences conduites à l'aide d'un ordinateur utilisant les données digitalisées de l'absorption de rayons X fournies par un “EMI scanner” sont rapportés. Dans la première expérience sont comparés les tomogrammes obtenus d'après les mêmes données mais en utilisant différents algorithmes de reconstruction. Dans la seconde, sont comparés une série de tomogrammes obtenus avec le même programme mais en utilisant différentes projections ou différents points de mesure par projection. Dans le cas clinque étudié, les résultats montrent: 1. que tous les algorithmes de reconstruction utilisés ont donné des resultats similaires et tous sont satisfaisants du point de vue clinique. 2. que les tomogrames des sections obtenues avec moins de points de mesure restent cliniquement valables. La signification de ces résultats pour l'application de la tomographie computérisée axiale en neuroradiologie est discutée.

Zusammenfassung

Bericht über zwei mittels eines Computers durchgeführten Experimente mit Projektionsdaten einer klinischen Untersuchung am EMI-Scanner. Beim ersten Experiment wurden mittels verschiedener Rekonstruktions-Algorhythmen gewonnene Schnittbilder verglichen. Beim zweiten Experiment wurde einer der beschriebenen Rekonstruktions-Algorhythmen dazu verwendet, um die Qualität des rekonstruierten Gehirnquerschnittes in Abhängigkeit von der Anzahl der verwendeten Projektionen sowie der Anzahl Meßwerte pro Projektionen zu zeigen. Das den Experimenten zugrundegelegte klinische Beispiel gestattet folgende Feststellungen: 1. Alle geprüften Rekonstruktions-Algorhythmen führten zu neuroradiologisch gleichermaßen befriedigenden Ergebnissen. 2. Rekonstruktionsbilder von annähernd gleicher klinischer Aussagekraft können auch mit einer reduzierten Menge von Eingabedaten erhalten werden. Die Bedeutung dieser Ergebnisse für die Anwendung der computerisierten axialen Tomographie in der neuroradiologischen Praxis werden diskutiert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Ambrose, J.: Computerized transverse axial scanning (tomography): Part 2 Clinical application. Brit. J. Radiology 46, 1023–1047 (1973)

    Google Scholar 

  2. Führer, K., Rührnschopf, E.P., Schwierz, G.: Computer-Tomographie; Mathematische Verfahren für die Bildrekonstruktion. Presented at the Jahrestagung der Deutschen Gesellschaft für Med. Physik, Essen, 8–10 October (1974)

  3. Hounsfield, G.N.: Patent Specification 1283915. The Patent Office, London (1972)

    Google Scholar 

  4. Hounfield, G.N.: Computerized transverse axial scanning (tomography): Part 1 Description of system. Brit. J. Radiology 46, 1016–1022 (1973)

    Google Scholar 

  5. Klug, A., Crowther, R.A.: Three-dimensional image reconstruction from the viewpoint of information theory. Nature 238, 435–440 (1972)

    Google Scholar 

  6. Ledley, R.S., DiChiro, G., Luessenhop, A.J., Twigg, H.L.: Computerized transaxial X-ray tomography of the human body. Science 186, 207–212 (1974)

    Google Scholar 

  7. Marr, R.B. (editor): ‘Techniques for three-dimensional reconstruction’, Proceedings of an International Workshop held at Brookhaven National Laboratory, Upton, N.Y., 16–19 July (1974)

  8. Mayo, B.J.: Communication at the international workshop on three-dimensional reconstruction techniques held at Brookhaven National Laboratory, Upton, N. Y., 16–19 July (1974)

  9. Müller, H.R., Wüthrich, R., Hünig, R., Elke M., Hochstetter, A.: A graphical reporting system for computerized axial X-ray tomography (EMI scanning). Europ. Neurol. (Basel) 11, 197–207 (1974)

    Google Scholar 

  10. Paxton, R., Ambrose, J.: The EMI-Scanner: a brief review of the first 650 patients. Brit. J. Radiology 47, 330–365 (1974)

    Google Scholar 

  11. Ramachandran, G.N., Lakshminarayanan, A.V.: Threedimensional reconstruction from radiographs and electron micrographs; application of convolutions instead of Fourier transforms. Proc. Nat. Acad. Sci. (Wash.) 68, 2236–2240 (1971)

    Google Scholar 

  12. Shepp, L.A., Logan, B.F.: The Fourier reconstruction of a head section. IEEE Trans. Nuclear Sci. NS-21 (3), 21–43 (1974)

    Google Scholar 

  13. Smith, P.R., Peters, T.M., Bates, R.H.T.: Image reconstruction from finite numbers of projections. J. Phys. A.: Math. Nucl. Gen. 6, 361–382 (1973)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Smith, P.R., Peters, T.M., Müller, H.R. et al. Towards the assessment of the limitations on computerized axial tomography. Neuroradiology 9, 1–8 (1975). https://doi.org/10.1007/BF00330129

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00330129

Keywords

Navigation